Reinders FC, de Ridder M, Stijnman PR, Doornaert PA, Raaijmakers CP, Philippens ME. Detectability and intra-fraction motion of individual elective lymph nodes in head and neck cancer patients on the Magnetic Resonance Image guided linear accelerator.
Phys Imaging Radiat Oncol 2024;
29:100532. [PMID:
38317852 PMCID:
PMC10839763 DOI:
10.1016/j.phro.2024.100532]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/02/2023] [Revised: 12/22/2023] [Accepted: 12/30/2023] [Indexed: 02/07/2024] Open
Abstract
Background and purpose
Individual elective lymph node irradiation instead of elective neck irradiation is a new concept for head-and-neck cancer (HNC) patients developed for the Magnetic Resonance Image guided linear accelerator (MR-linac). To prepare this, the detectability, volume changes and intra-fraction motion of elective lymph nodes on the MR-linac was assessed.
Materials and methods
A total of 15 HNC patients underwent diagnostic pre-treatment MRI. Additionally, two MR-linac scans were obtained with a 10-minute time difference in the first week of radiation treatment. Elective lymph node contours inside lymph node levels (Ib-V) were segmented on the pre-treatment MRI and the MR-linac scans and compared on number and maximal transversal diameter. Intra-fraction motion of elective lymph nodes on the MR-linac was estimated using Center of Mass (COM) distances and incremental isotropic expansion of lymph node segmentations.
Results
Of all 679 detected lymph nodes on the pre-treatment MRI, eight lymph nodes were not detectable on the first MR-linac scan and 16 new lymph nodes were detected. Lymph node diameters between the pre-treatment MRI scan and the MR-linac scan varied from -0.19 to + 0.13 mm. COM distances varied from 1.2 to 1.7 mm and lymph node contours had to be expanded with 3 mm.
Conclusions
Nearly all elective lymph nodes were detectable on the 1.5T MR-linac scan with no major changes in target volumes compared to the pre-treatment MRI. Simulated intra-fraction motion during the MR-linac scans was smaller than the 5-mm margin that will be used in the first elective lymph node radiation treatment.