Kosaka T, Takatsu J, Inoue T, Iijima K, Suzuki M, Murakami N, Shikama N. Dosimetric evaluation in Helical TomoTherapy for lung SBRT using Monte Carlo-based independent dose verification software.
J Appl Clin Med Phys 2024;
25:e14305. [PMID:
38368607 PMCID:
PMC11087163 DOI:
10.1002/acm2.14305]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/02/2023] [Revised: 01/24/2024] [Accepted: 02/05/2024] [Indexed: 02/20/2024] Open
Abstract
PURPOSE
To elucidate the dosimetric errors caused by a model-based algorithm in lung stereotactic body radiation therapy (SBRT) with Helical TomoTherapy (HT) using Monte Carlo (MC)-based dose verification software.
METHODS
For 38 plans of lung SBRT, the dose calculation accuracy of a treatment planning system (TPS) of HT was compared with the results of DoseCHECK, the commercial MC-based independent verification software. The following indices were extracted to evaluate the correlation of dosimetric errors: (1) target volume, (2) average computed tomography (CT) value of the planning target volume (PTV) margin, and (3) average CT value of surrounding 2-mm area of the PTV (PTV ring). Receiver operating characteristic (ROC) analyses determined the threshold for 5% of differences in PTV D95%. Then, the 38 plans were classified into two groups using the cutoff values of ROC analysis for these three indices. Dosimetric differences between groups were statistically compared using the Mann-Whitney U test.
RESULTS
TPS of HT overestimated by more than 5% in the PTV D95% in 16 of 38 plans. The PTV ring showed the strongest correlation with dosimetric differences. The cutoff value for the target volume, the PTV margin, and the PTV ring was 14.7 cc, -754 HU, and -708 HU, respectively. The area under the curve (AUC) for the target volume, the PTV margin, and the PTV ring were 0.835, 0.878, and 0.932, respectively. Dosimetric errors more than 5% were observed when the PTV volume was less than 15 cc or when the CT value around the target was less than -700 HU.
CONCLUSION
The TPS of HT might overestimate the PTV dose by more than 5% if any the three indices in this study were below threshold. Therefore, independent verification with an MC-based algorithm should be strongly recommended for lung SBRT in HT.
Collapse