1
|
Racine D, Niemann T, Nemeth B, Manzano LG, Alkadhi H, Viry A, Kubik-Huch RA, Frauenfelder T, Euler A. Dual-Split CT to Simulate Multiple Radiation Doses From a Single Scan-Liver Lesion Detection Compared With Dose-Matched Single-Energy CT. Invest Radiol 2025; 60:131-137. [PMID: 39074298 DOI: 10.1097/rli.0000000000001111] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 07/31/2024]
Abstract
OBJECTIVES The aim of this study was to evaluate the potential use of simulated radiation doses from a dual-split CT scan for dose optimization by comparing their lesion detectability to dose-matched single-energy CT acquisitions at different radiation dose levels using a mathematical model observer. MATERIALS AND METHODS An anthropomorphic abdominal phantom with liver lesions (5-10 mm, both hyperattenuating and hypoattenuating) was imaged using a third-generation dual-source CT in single-energy dual-source mode at 100 kVp and 3 radiation doses (5, 2.5, 1.25 mGy). The tube current was 67% for tube A and 33% for tube B. For each dose, 5 simulated radiation doses (100%, 67%, 55%, 45%, 39%, and 33%) were generated through linear image blending. The phantom was also imaged using traditional single-source single-energy mode at equivalent doses. Each setup was repeated 10 times. Image noise texture was evaluated by the average spatial frequency (f av ) of the noise power spectrum. Liver lesion detection was measured by the area under the receiver operating curve (AUC), using a channelized Hotelling model observer with 10 dense Gaussian channels. RESULTS F av decreased at lower radiation doses and differed between simulated and single-energy images (eg, 0.16 mm -1 vs 0.14 mm -1 for simulated and single-energy images at 1.25 mGy), indicating slightly blotchier noise texture for dual-split CT. For hyperattenuating lesions, the mean AUC ranged between 0.92-0.99, 0.81-0.96, and 0.68-0.89 for single-energy, and between 0.91-0.99, 0.78-0.91, and 0.70-0.85 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. For hypoattenuating lesions, the AUC ranged between 0.90-0.98, 0.75-0.93, and 0.69-0.86 for the single-energy, and between 0.92-0.99, 0.76-0.87, and 0.67-0.81 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. AUC values were similar between both modes at 5 mGy, and slightly lower, albeit not significantly, for the dual-split mode at 2.5 and 1.25 mGy. CONCLUSIONS Lesion detectability was comparable between multiple simulated radiation doses from a dual-split CT scan and dose-matched single-energy CT. Noise texture was slightly blotchier in the simulated images. Simulated doses using dual-split CT can be used to assess the impact of radiation dose reduction on lesion detectability without the need for repeated patient scans.
Collapse
Affiliation(s)
- Damien Racine
- From the Institute of Radiation Physics, University Hospital Lausanne (CHUV), University of Lausanne, Lausanne, Switzerland (D.R., L.G.M., A.V.); Department of Radiology, Kantonsspital Baden, Affiliated Hospital for Research and Teaching of the Faculty of Medicine of the University of Zurich, Baden, Switzerland (T.N., R.A.K.-H., A.E.); Department of Biomedical Imaging and Image-Guided Therapy, Division of Neuroradiology and Musculoskeletal Radiology, Medical University of Vienna, Vienna, Austria (B.N.); and Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland (H.A., T.F.)
| | | | | | | | | | | | | | | | | |
Collapse
|
2
|
Racine D, Mergen V, Viry A, Frauenfelder T, Alkadhi H, Vitzthum V, Euler A. Photon-Counting Detector CT for Liver Lesion Detection-Optimal Virtual Monoenergetic Energy for Different Simulated Patient Sizes and Radiation Doses. Invest Radiol 2024; 59:554-560. [PMID: 38193782 DOI: 10.1097/rli.0000000000001060] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/10/2024]
Abstract
OBJECTIVES The aim of this study was to evaluate the optimal energy level of virtual monoenergetic images (VMIs) from photon-counting detector computed tomography (CT) for the detection of liver lesions as a function of phantom size and radiation dose. MATERIALS AND METHODS An anthropomorphic abdominal phantom with liver parenchyma and lesions was imaged on a dual-source photon-counting detector CT at 120 kVp. Five hypoattenuating lesions with a lesion-to-background contrast difference of -30 HU and -45 HU and 3 hyperattenuating lesions with +30 HU and +90 HU were used. The lesion diameter was 5-10 mm. Rings of fat-equivalent material were added to emulate medium- or large-sized patients. The medium size was imaged at a volume CT dose index of 5, 2.5, and 1.25 mGy and the large size at 5 and 2.5 mGy, respectively. Each setup was imaged 10 times. For each setup, VMIs from 40 to 80 keV at 5 keV increments were reconstructed with quantum iterative reconstruction at a strength level of 4 (QIR-4). Lesion detectability was measured as area under the receiver operating curve (AUC) using a channelized Hotelling model observer with 10 dense differences of Gaussian channels. RESULTS Overall, highest detectability was found at 65 and 70 keV for both hypoattenuating and hyperattenuating lesions in the medium and large phantom independent of radiation dose (AUC range, 0.91-1.0 for the medium and 0.94-0.99 for the large phantom, respectively). The lowest detectability was found at 40 keV irrespective of the radiation dose and phantom size (AUC range, 0.78-0.99). A more pronounced reduction in detectability was apparent at 40-50 keV as compared with 65-75 keV when radiation dose was decreased. At equal radiation dose, detection as a function of VMI energy differed stronger for the large size as compared with the medium-sized phantom (12% vs 6%). CONCLUSIONS Detectability of hypoattenuating and hyperattenuating liver lesions differed between VMI energies for different phantom sizes and radiation doses. Virtual monoenergetic images at 65 and 70 keV yielded highest detectability independent of phantom size and radiation dose.
Collapse
Affiliation(s)
- Damien Racine
- From the Institute of Radiation Physics, University Hospital Lausanne (CHUV), University of Lausanne, Lausanne, Switzerland (D.R., A.V., V.V.); Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland (V.M., T.F., H.A., A.E.); and Department of Radiology, Kantonsspital Baden, Baden, Switzerland (A.E.)
| | | | | | | | | | | | | |
Collapse
|
3
|
Wang TJ, Wang Y, Zhang ZH, Wang M, Wang M, Su T, Xu YH, Ma ZF, Wang J, Chen Y, Jin ZY. Deep learning reconstruction improves the image quality of low-dose temporal bone CT with otitis media and mastoiditis patients. Heliyon 2024; 10:e22810. [PMID: 38148801 PMCID: PMC10750061 DOI: 10.1016/j.heliyon.2023.e22810] [Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/11/2023] [Revised: 11/08/2023] [Accepted: 11/20/2023] [Indexed: 12/28/2023] Open
Abstract
Objective To evaluate the image quality of low-dose temporal bone computed tomography (CT) in otitis media and mastoiditis patients by using deep learning reconstruction (DLR). Materials and methods A total of ninety-seven temporal bones from 53 consecutive adult patients who had suspected otitis media and mastoiditis and underwent temporal bone CT were prospectively enrolled. All patients underwent high resolution CT protocol (group A) and an additional low-dose protocol (group B). In group A, high resolution data were reconstructed by filter back projection (FBP). In group B, low-dose data were reconstructed by DLR mild (B1), DLR standard (B2) and DLR strong (B3). The objective image quality was analyzed by measuring the CT value and image noise on the transverse image and calculating the signal-to-noise ratio (SNR) on incudomallear joint, retroauricular muscle, vestibule and subcutaneous fat. Subjective image quality was analyzed by using a five-point scale to evaluate nine anatomical structures of middle and inner ear. The number of temporal bone lesions which involved in five structures of middle ear were assessed in group A, B1, B2 and B3 images. Results There were no significant differences in the CT values of the four reconstruction methods at four structures (all p > 0.05). The DLR group B1, B2 and B3 had significantly less image noise and a significantly higher SNR than group A at four structures (all p < 0.001). The group B1 had comparable subjective image quality as group A in nine structures (all p > 0.05), however, the group B3 had lower subjective image quality than group A in modiolus, spiral osseous lamina and stapes (all p < 0.001), the group B2 had lower subjective image quality than group A in modiolus and spiral osseous lamina (both p < 0.05). The number of temporal bone lesions which involved in five structures for group A, B1 and B2 images were no significant difference (all p > 0.05), however, the number of temporal bone lesions which involved in mastoid for group B3 images were significantly more than group A (p < 0.05). The radiation dose of high resolution CT protocol and low-dose protocol were 0.55 mSv and 0.11 mSv, respectively. Conclusion Compared with high resolution CT protocol, in the low-dose protocol of temporal bone CT, DLR mild and standard could improve the objective image quality, maintain good subjective image quality and satisfy clinical diagnosis of otitis media and mastoiditis patients.
Collapse
Affiliation(s)
- Tian-Jiao Wang
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Yun Wang
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Zhu-Hua Zhang
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Ming Wang
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Man Wang
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Tong Su
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Ying-Hao Xu
- Canon Medical Systems (China) CO., LTD., Building 205, Yard NO. A10, JiuXianQiao North Road, Beijing, 100015, China
| | - Zhuang-Fei Ma
- Canon Medical Systems (China) CO., LTD., Building 205, Yard NO. A10, JiuXianQiao North Road, Beijing, 100015, China
| | - Jian Wang
- Canon Medical Systems (China) CO., LTD., Building 205, Yard NO. A10, JiuXianQiao North Road, Beijing, 100015, China
| | - Yu Chen
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| | - Zheng-Yu Jin
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuai Fu Yuan, Dong Cheng District, Beijing, 100730, China
| |
Collapse
|
4
|
Göppel M, Anton M, Gala HDLH, Giussani A, Trinkl S, Renger B, Brix G. Dose-efficiency quantification of computed tomography systems using a model-observer. Med Phys 2023; 50:7594-7605. [PMID: 37183490 DOI: 10.1002/mp.16441] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/01/2022] [Revised: 04/01/2023] [Accepted: 04/17/2023] [Indexed: 05/16/2023] Open
Abstract
BACKGROUND Recent advances in computed tomography (CT) technology have considerably improved the quality of CT images and reduced radiation exposure in patients. At present, however, there is no generally accepted figure of merit (FOM) for comparing the dose efficiencies of CT systems. PURPOSE (i) To establish an FOM that characterizes the quality of CT images in relation to the radiation dose by means of a mathematical model observer and (ii) to evaluate the new FOM on different CT systems and image reconstruction algorithms. METHODS Images of a homogeneous phantom with four low-contrast inserts were acquired using three different CT systems at three dose levels and a representative protocol for CT imaging of low-contrast objects in the abdomen. The images were reconstructed using filtered-back projection and iterative algorithms. A channelized hotelling observer with difference-of-Gaussian channels was applied to compute the detectability (d ' $d^{\prime}$ ). This was done for each insert and each of the considered imaging conditions from square regions of interest (ROIs) that were (semi-)automatically centered on the inserts. The estimated detectabilities (d ' $d^{\prime}$ ) were averaged in the first step over the three dose levels (⟨ d ' ⟩ $\langle {d^{\prime}} \rangle $ ), and subsequently over the four contrast inserts (⟨ d ' ⟩ w ${\langle {d^{\prime}} \rangle _{\rm{w}}}$ ). All calculation steps included a dedicated assessment of the related uncertainties following accepted metrological guidelines. RESULTS The determined detectabilities (d ' $d^{\prime}$ ) varied considerably with the contrast and diameter of the four inserts, as well as with the radiation doses and reconstruction algorithms used for image generation (d ' $d^{\prime}\;$ = 1.3-5.5). Thus, the specification of a single detectability as an FOM is not well suited for comprehensively characterizing the dose efficiency of a CT system. A more comprehensive and robust characterization was provided by the averaged detectabilities⟨ d ' ⟩ $\langle {d^{\prime}} \rangle $ and, in particular,⟨ d ' ⟩ w ${\langle {d^{\prime}} \rangle _{\rm{w}}}$ . Our analysis reveals that the model observer analysis is very sensitive to the exact position of the ROIs. CONCLUSIONS The presented automatable software approach yielded with the weighted detectability⟨ d ' ⟩ w ${\langle {d^{\prime}} \rangle _{\rm{w}}}$ an objective FOM to benchmark different CT systems and reconstruction algorithms in a robust and reliable manner. An essential advantage of the proposed model-observer approach is that uncertainties in the FOM can be provided, which is an indispensable prerequisite for type testing.
Collapse
Affiliation(s)
- Maximilian Göppel
- Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
| | - Mathias Anton
- Department of Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt, Berlin, Germany
| | - Hugo de Las Heras Gala
- Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
| | - Augusto Giussani
- Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
| | - Sebastian Trinkl
- Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
| | - Bernhard Renger
- Department of Diagnostic and Interventional Radiology, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany
| | - Gunnar Brix
- Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
| |
Collapse
|
5
|
Racine D, Mergen V, Viry A, Eberhard M, Becce F, Rotzinger DC, Alkadhi H, Euler A. Photon-Counting Detector CT With Quantum Iterative Reconstruction: Impact on Liver Lesion Detection and Radiation Dose Reduction. Invest Radiol 2023; 58:245-252. [PMID: 36094810 DOI: 10.1097/rli.0000000000000925] [Citation(s) in RCA: 10] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
OBJECTIVES To assess image noise, diagnostic performance, and potential for radiation dose reduction of photon-counting detector (PCD) computed tomography (CT) with quantum iterative reconstruction (QIR) in the detection of hypoattenuating and hyperattenuating focal liver lesions compared with energy-integrating detector (EID) CT. MATERIALS AND METHODS A medium-sized anthropomorphic abdominal phantom with liver parenchyma and lesions (diameter, 5-10 mm; hypoattenuating and hyperattenuating from -30 HU to +90 HU at 120 kVp) was used. The phantom was imaged on ( a ) a third-generation dual-source EID-CT (SOMATOM Force, Siemens Healthineers) in the dual-energy mode at 100 and 150 kVp with tin filtration and ( b ) a clinical dual-source PCD-CT at 120 kVp (NAEOTOM Alpha, Siemens). Scans were repeated 10 times for each of 3 different radiation doses of 5, 2.5, and 1.25 mGy. Datasets were reconstructed as virtual monoenergetic images (VMIs) at 60 keV for both scanners and as linear-blended images (LBIs) for EID-CT. For PCD-CT, VMIs were reconstructed with different strength levels of QIR (QIR 1-4) and without QIR (QIR-off). For EID-CT, VMIs and LBIs were reconstructed using advanced modeled iterative reconstruction at a strength level of 3. Noise power spectrum was measured to compare image noise magnitude and texture. A channelized Hotelling model observer was used to assess diagnostic accuracy for lesion detection. The potential for radiation dose reduction using PCD-CT was estimated for the QIR strength level with the highest area under the curve compared with EID-CT for each radiation dose. RESULTS Image noise decreased with increasing QIR level at all radiation doses. Using QIR-4, noise reduction was 41%, 45%, and 59% compared with EID-CT VMIs and 12%, 18%, and 33% compared with EID-CT LBIs at 5, 2.5, and 1.25 mGy, respectively. The peak spatial frequency shifted slightly to lower frequencies at higher QIR levels. Lesion detection accuracy increased at higher QIR levels and was higher for PCD-CT compared with EID-CT VMIs. The improvement in detection with PCD-CT was strongest at the lowest radiation dose, with an area under the receiver operating curve of 0.917 for QIR-4 versus 0.677 for EID-CT VMIs for hyperattenuating lesions, and 0.900 for QIR-4 versus 0.726 for EID-CT VMIs for hypoattenuating lesions. Compared with EID-CT LBIs, detection was higher for QIR 1-4 at 2.5 mGy and for QIR 2-4 at 1.25 mGy (eg, 0.900 for QIR-4 compared with 0.854 for EID-CT LBIs at 1.25 mGy). Radiation dose reduction potential of PCD-CT with QIR-4 was 54% at 5 mGy compared with VMIs and 39% at 2.5 mGy compared with LBIs. CONCLUSIONS Compared with EID-CT, PCD-CT with QIR substantially improved focal liver lesion detection, especially at low radiation dose. This enables substantial radiation dose reduction while maintaining diagnostic accuracy.
Collapse
Affiliation(s)
- Damien Racine
- From the Institute of Radiation Physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne
| | - Victor Mergen
- Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich
| | - Anaïs Viry
- From the Institute of Radiation Physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne
| | - Matthias Eberhard
- Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich
| | - Fabio Becce
- Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland
| | - David C Rotzinger
- Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland
| | - Hatem Alkadhi
- Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich
| | - André Euler
- Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich
| |
Collapse
|
6
|
Valeri F, Bartolucci M, Cantoni E, Carpi R, Cisbani E, Cupparo I, Doria S, Gori C, Grigioni M, Lasagni L, Marconi A, Mazzoni LN, Miele V, Pradella S, Risaliti G, Sanguineti V, Sona D, Vannucchi L, Taddeucci A. UNet and MobileNet CNN-based model observers for CT protocol optimization: comparative performance evaluation by means of phantom CT images. J Med Imaging (Bellingham) 2023; 10:S11904. [PMID: 36895439 PMCID: PMC9989681 DOI: 10.1117/1.jmi.10.s1.s11904] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/06/2022] [Accepted: 02/09/2023] [Indexed: 03/09/2023] Open
Abstract
Purpose The aim of this work is the development and characterization of a model observer (MO) based on convolutional neural networks (CNNs), trained to mimic human observers in image evaluation in terms of detection and localization of low-contrast objects in CT scans acquired on a reference phantom. The final goal is automatic image quality evaluation and CT protocol optimization to fulfill the ALARA principle. Approach Preliminary work was carried out to collect localization confidence ratings of human observers for signal presence/absence from a dataset of 30,000 CT images acquired on a PolyMethyl MethAcrylate phantom containing inserts filled with iodinated contrast media at different concentrations. The collected data were used to generate the labels for the training of the artificial neural networks. We developed and compared two CNN architectures based respectively on Unet and MobileNetV2, specifically adapted to achieve the double tasks of classification and localization. The CNN evaluation was performed by computing the area under localization-ROC curve (LAUC) and accuracy metrics on the test dataset. Results The mean of absolute percentage error between the LAUC of the human observer and MO was found to be below 5% for the most significative test data subsets. An elevated inter-rater agreement was achieved in terms of S-statistics and other common statistical indices. Conclusions Very good agreement was measured between the human observer and MO, as well as between the performance of the two algorithms. Therefore, this work is highly supportive of the feasibility of employing CNN-MO combined with a specifically designed phantom for CT protocol optimization programs.
Collapse
Affiliation(s)
- Federico Valeri
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
- Università degli Studi di Firenze, Scuola di Scienze della Salute Umana, Florence, Italy
| | - Maurizio Bartolucci
- Ospedale S. Stefano, Azienda USL Toscana Centro, SOC Radiodiagnostica, Prato, Italy
| | - Elena Cantoni
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
| | - Roberto Carpi
- Ospedale Santa Maria Nuova, Azienda USL Toscana Centro, SOC Radiologia, Florence, Italy
| | - Evaristo Cisbani
- Istituto Superiore di Sanità, Centro Nazionale Tecnologie Innvative in Sanità Pubblica, Rome, Italy
| | - Ilaria Cupparo
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
- Università degli Studi di Firenze, Scuola di Scienze della Salute Umana, Florence, Italy
| | - Sandra Doria
- Istituto di Chimica dei Composti OrganoMetallici, Consiglio Nazionale delle Ricerche, Florence, Italy
- Università degli Studi di Firenze, European Laboratory for Nonlinear Spectroscopy, Florence, Italy
| | - Cesare Gori
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
| | - Mauro Grigioni
- Istituto Superiore di Sanità, Centro Nazionale Tecnologie Innvative in Sanità Pubblica, Rome, Italy
| | - Lorenzo Lasagni
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
- Università degli Studi di Firenze, Scuola di Scienze della Salute Umana, Florence, Italy
| | - Alessandro Marconi
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
| | - Lorenzo Nicola Mazzoni
- Ospedale San Jacopo, Azienda USL Toscana Centro, UO Fisica Sanitaria Prato e Pistoia, Pistoia, Italy
| | - Vittorio Miele
- Azienda Ospedaliero-Universitaria Careggi, SOD Radiodiagnostica di Emergenza-Urgenza, Florence, Italy
| | - Silvia Pradella
- Azienda Ospedaliero-Universitaria Careggi, SOD Radiodiagnostica di Emergenza-Urgenza, Florence, Italy
| | - Guido Risaliti
- Università degli Studi di Firenze, Dipartimento di Fisica e Astronomia, Florence, Italy
| | - Valentina Sanguineti
- Istituto Italiano di Tecnologia, Pattern Analysis & Computer Vision, Genoa, Italy
| | - Diego Sona
- Fondazione Bruno Kessler, Data Science for Health Unit, Trento, Italy
| | - Letizia Vannucchi
- Ospedale S. Jacopo, AUSL Toscana Centro, SOC Radiodiagnostica, Pistoia, Italy
| | - Adriana Taddeucci
- Azienda Ospedaliero-Universitaria Careggi, UO Fisica Sanitaria, Florence, Italy
- Istituto Nazionale di Fisica Nucleare - Sezione di Firenze, Sesto Fiorentino, Italy
| |
Collapse
|
7
|
Funama Y, Shirasaka T, Goto T, Aoki Y, Tanaka K, Yoshida R. Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study. Acta Radiol Open 2022; 11:20584601221109919. [PMID: 35747445 PMCID: PMC9209785 DOI: 10.1177/20584601221109919] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/16/2022] [Accepted: 06/08/2022] [Indexed: 11/24/2022] Open
Abstract
Background Brain CT needs more attention to improve the extremely low image contrast and image texture. Purpose To evaluate the performance of iterative progressive reconstruction with visual modeling (IPV) for the improvement of low-contrast detectability (IPV-LCD) compared with filtered backprojection (FBP) and conventional IPV. Materials and methods Low-contrast and water phantoms were used. Helical scans were conducted with the use of a CT scanner with 64 detectors. The tube voltage was set at 120 kVp; the tube current was adjusted from 60 to 300 mA with a slice thickness of 0.625 mm and from 20 to 150 mA with a slice thickness of 5.0 mm. Images were reconstructed with the FBP, conventional IPV, and IPV-LCD algorithms. The channelized Hotelling observer (CHO) model was applied in conjunction with the use of low-contrast modules in the low-contrast phantom. The noise power spectrum (NPS) and normalized NPS were calculated. Results At the same standard and strong levels, the IPV-LCD method improved low-contrast detectability compared with the conventional IPV, regardless of contrast-rod diameters. The mean CHO values at a slice thickness of 0.625 mm were 1.83, 3.28, 4.40, 4.53, and 5.27 for FBP, IPV STD, IPV-LCD STD, IPV STR, and IPV-LCD STR, respectively. The normalized NPS for the IPV-LCD STD and STR images were slightly shifted to the higher frequency compared with that for the FBP image. Conclusion IPV-LCD images further improve the low-contrast detectability compared with FBP and conventional IPV images while maintaining similar FBP image appearances.
Collapse
Affiliation(s)
- Yoshinori Funama
- Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan
| | - Takashi Shirasaka
- Graduate School of Health Sciences, Kumamoto University, Kumamoto, Japan
- Division of Radiology, Department of Medical Technology, Kyushu University, Fukuoka, Japan
| | - Taiga Goto
- Rad Diagnostic R&D Division, Radiation Diagnostic Systems Division, FUJIFILM Healthcare Corporation, Tokyo, Japan
| | - Yuko Aoki
- Rad Diagnostic R&D Division, Radiation Diagnostic Systems Division, FUJIFILM Healthcare Corporation, Tokyo, Japan
| | - Kana Tanaka
- Rad Diagnostic R&D Division, Radiation Diagnostic Systems Division, FUJIFILM Healthcare Corporation, Tokyo, Japan
| | - Ryo Yoshida
- Rad Diagnostic R&D Division, Radiation Diagnostic Systems Division, FUJIFILM Healthcare Corporation, Tokyo, Japan
| |
Collapse
|
8
|
Saeed SU, Fu Y, Stavrinides V, Baum ZMC, Yang Q, Rusu M, Fan RE, Sonn GA, Noble JA, Barratt DC, Hu Y. Image quality assessment for machine learning tasks using meta-reinforcement learning. Med Image Anal 2022; 78:102427. [PMID: 35344824 DOI: 10.1016/j.media.2022.102427] [Citation(s) in RCA: 18] [Impact Index Per Article: 6.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/09/2021] [Revised: 01/24/2022] [Accepted: 03/18/2022] [Indexed: 11/23/2022]
Abstract
In this paper, we consider image quality assessment (IQA) as a measure of how images are amenable with respect to a given downstream task, or task amenability. When the task is performed using machine learning algorithms, such as a neural-network-based task predictor for image classification or segmentation, the performance of the task predictor provides an objective estimate of task amenability. In this work, we use an IQA controller to predict the task amenability which, itself being parameterised by neural networks, can be trained simultaneously with the task predictor. We further develop a meta-reinforcement learning framework to improve the adaptability for both IQA controllers and task predictors, such that they can be fine-tuned efficiently on new datasets or meta-tasks. We demonstrate the efficacy of the proposed task-specific, adaptable IQA approach, using two clinical applications for ultrasound-guided prostate intervention and pneumonia detection on X-ray images.
Collapse
Affiliation(s)
- Shaheer U Saeed
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK.
| | - Yunguan Fu
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK; InstaDeep, London, UK
| | - Vasilis Stavrinides
- Division of Surgery & Interventional Science, University College London, London, UK; Department of Urology, University College Hospital NHS Foundation Trust, London, UK
| | - Zachary M C Baum
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK
| | - Qianye Yang
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK
| | - Mirabela Rusu
- Department of Radiology, Stanford University, Stanford, California, USA
| | - Richard E Fan
- Department of Urology, Stanford University, Stanford, California, USA
| | - Geoffrey A Sonn
- Department of Radiology, Stanford University, Stanford, California, USA; Department of Urology, Stanford University, Stanford, California, USA
| | - J Alison Noble
- Department of Engineering Science, University of Oxford, Oxford, UK
| | - Dean C Barratt
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK
| | - Yipeng Hu
- Centre for Medical Image Computing, Wellcome/EPSRC Centre for Interventional & Surgical Sciences, and Department of Medical Physics & Biomedical Engineering, University College London, London, UK; Department of Engineering Science, University of Oxford, Oxford, UK
| |
Collapse
|
9
|
Mikayama R, Shirasaka T, Kojima T, Sakai Y, Yabuuchi H, Kondo M, Kato T. Deep-learning reconstruction for ultra-low-dose lung CT: Volumetric measurement accuracy and reproducibility of artificial ground-glass nodules in a phantom study. Br J Radiol 2022; 95:20210915. [PMID: 34908478 PMCID: PMC8822562 DOI: 10.1259/bjr.20210915] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/03/2023] Open
Abstract
OBJECTIVES The lung nodule volume determined by CT is used for nodule diagnoses and monitoring tumor responses to therapy. Increased image noise on low-dose CT degrades the measurement accuracy of the lung nodule volume. We compared the volumetric accuracy among deep-learning reconstruction (DLR), model-based iterative reconstruction (MBIR), and hybrid iterative reconstruction (HIR) at an ultra-low-dose setting. METHODS Artificial ground-glass nodules (6 mm and 10 mm diameters, -660 HU) placed at the lung-apex and the middle-lung field in chest phantom were scanned by 320-row CT with the ultra-low-dose setting of 6.3 mAs. Each scan data set was reconstructed by DLR, MBIR, and HIR. The volumes of nodules were measured semi-automatically, and the absolute percent volumetric error (APEvol) was calculated. The APEvol provided by each reconstruction were compared by the Tukey-Kramer method. Inter- and intraobserver variabilities were evaluated by a Bland-Altman analysis with limits of agreements. RESULTS DLR provided a lower APEvol compared to MBIR and HIR. The APEvol of DLR (1.36%) was significantly lower than those of the HIR (8.01%, p = 0.0022) and MBIR (7.30%, p = 0.0053) on a 10-mm-diameter middle-lung nodule. DLR showed narrower limits of agreement compared to MBIR and HIR in the inter- and intraobserver agreement of the volumetric measurement. CONCLUSIONS DLR showed higher accuracy compared to MBIR and HIR for the volumetric measurement of artificial ground-glass nodules by ultra-low-dose CT. ADVANCES IN KNOWLEDGE DLR with ultra-low-dose setting allows a reduction of dose exposure, maintaining accuracy for the volumetry of lung nodule, especially in patients which deserve a long-term follow-up.
Collapse
Affiliation(s)
- Ryoji Mikayama
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
| | - Takashi Shirasaka
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
| | | | - Yuki Sakai
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
| | - Hidetake Yabuuchi
- Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Fukuoka, Japan
| | - Masatoshi Kondo
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
| | - Toyoyuki Kato
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
| |
Collapse
|
10
|
Bornet PA, Villani N, Gillet R, Germain E, Lombard C, Blum A, Gondim Teixeira PA. Clinical acceptance of deep learning reconstruction for abdominal CT imaging: objective and subjective image quality and low-contrast detectability assessment. Eur Radiol 2022; 32:3161-3172. [PMID: 34989850 DOI: 10.1007/s00330-021-08410-x] [Citation(s) in RCA: 9] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/26/2021] [Revised: 09/25/2021] [Accepted: 10/13/2021] [Indexed: 01/29/2023]
Abstract
OBJECTIVE To evaluate the image quality and clinical acceptance of a deep learning reconstruction (DLR) algorithm compared to traditional iterative reconstruction (IR) algorithms. METHODS CT acquisitions were performed with two phantoms and a total of nine dose levels. Images were reconstructed with two types of IR algorithms, DLR and filtered-back projection. Spatial resolution, image texture, mean noise value, and objective and subjective low-contrast detectability were compared. Ten senior radiologists evaluated the clinical acceptance of these algorithms by scoring ten CT exams reconstructed with the DLR and IR algorithms evaluated. RESULTS Compared to MBIR, DLR yielded a lower noise and a higher low-contrast detectability index at low doses (CTDIvol ≤ 2.2 and ≤ 4.5 mGy, respectively). Spatial resolution and detectability at higher doses were better with MBIR. Compared to HIR, DLR yielded a higher spatial resolution, a lower noise, and a higher detectability index. Despite these differences in algorithm performance, significant differences in subjective low-contrast performance were not found (p ≥ 0.005). DLR texture was finer than that of MBIR and closer to that of HIR. Radiologists preferred DLR images for all criteria assessed (p < 0.0001), whereas MBIR was rated worse than HIR (p < 0.0001) in all criteria evaluated, except for noise (p = 0.044). DLR reconstruction time was 12 times faster than that of MBIR. CONCLUSION DLR yielded a gain in objective detection and noise at lower dose levels with the best clinical acceptance among the evaluated reconstruction algorithms. KEY POINTS • DLR yielded improved objective low-contrast detection and noise at lower dose levels. • Despite the differences in objective detectability among the algorithms evaluated, there were no differences in subjective detectability. • DLR presented significantly higher clinical acceptability scores compared to MBIR and HIR.
Collapse
Affiliation(s)
- Pierre-Antoine Bornet
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France.
| | - Nicolas Villani
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| | - Romain Gillet
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| | - Edouard Germain
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| | - Charles Lombard
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| | - Alain Blum
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| | - Pedro Augusto Gondim Teixeira
- Guilloz Imaging Department, University of Lorraine, Central Hospital, University Hospital Center of Nancy, Nancy, France
| |
Collapse
|
11
|
Lubis LE, Basith RA, Hariyati I, Ryangga D, Mart T, Bosmans H, Soejoko DS. Novel phantom for performance evaluation of contrast-enhanced 3D rotational angiography. Phys Med 2021; 90:91-98. [PMID: 34571289 DOI: 10.1016/j.ejmp.2021.09.002] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/22/2021] [Revised: 08/22/2021] [Accepted: 09/03/2021] [Indexed: 11/19/2022] Open
Abstract
PURPOSE This technical note presents an in-house phantom with a specially designed contrast-object module constructed to address the need for three-dimensional rotational angiography (3DRA) testing. METHODS The initial part of the study was a brief evaluation on the commercially available phantom used for 3DRA and computed tomography angiography (CTA) to confirm the need for a special phantom for 3D angiography. Once confirmed, an in-house phantom was constructed. The novel phantom was tested to evaluate the basic image performance metrics, i.e., unsharpness (MTF) and noise characterization (NPS), as well as to show its capability for vessel contrast visibility study. RESULTS The low contrast objects in the commercially available tools dedicated for CT is found to yield significantly lower signal difference to noise ratio (SDNR) when used for 3DRA, therefore deemed inadequate for 3DRA contrast evaluation. The constructed in-house phantom demonstrates a capability to serve for basic imaging performance check (MTF, NPS, and low contrast evaluation) for 3DRA and CTA. With higher and potentially adjustable visibility of contrast objects as artificial vessels, the in-house phantom also makes more clinically relevant tests, e.g., human- or model observer study and task-based optimization, possible. CONCLUSION The novel phantom with special contrast object module shows higher visibility in 3DRA compared to the currently available commercial phantom and, therefore, is recommended for use in 3D angiography.
Collapse
Affiliation(s)
- L E Lubis
- Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok 16424, Indonesia
| | - R A Basith
- Radiology Department, R. Syamsuddin S.H. Regional General Hospital, Sukabumi 43113, Indonesia
| | - I Hariyati
- Radiology Department, Gading Pluit Hospital, Jakarta 14250, Indonesia
| | - D Ryangga
- Radiotherapy Department, Pasar Minggu Regional General Hospital, Jakarta 12550, Indonesia
| | - T Mart
- Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok 16424, Indonesia.
| | - H Bosmans
- Medical Physics and Quality Assessment, Catholic University of Leuven, Leuven 3000, Belgium
| | - D S Soejoko
- Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok 16424, Indonesia
| |
Collapse
|
12
|
Duan J, Mou X. Image quality guided iterative reconstruction for low-dose CT based on CT image statistics. Phys Med Biol 2021; 66. [PMID: 34352735 DOI: 10.1088/1361-6560/ac1b1b] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/26/2021] [Accepted: 08/05/2021] [Indexed: 11/12/2022]
Abstract
Iterative reconstruction framework shows predominance in low dose and incomplete data situation. In the iterative reconstruction framework, there are two components, i.e., fidelity term aims to maintain the structure details of the reconstructed object, and the regularization term uses prior information to suppress the artifacts such as noise. A regularization parameter balances them, aiming to find a good trade-off between noise and resolution. Currently, the regularization parameters are selected as a rule of thumb or some prior knowledge assumption is required, which limits practical uses. Furthermore, the computation cost of regularization parameter selection is also heavy. In this paper, we address this problem by introducing CT image quality assessment (IQA) into the iterative reconstruction framework. Several steps are involved during the study. First, we analyze the CT image statistics using the dual dictionary method. Regularities are observed and concluded, revealing the relationship among the regularization parameter, iterations, and CT image quality. Second, with derivation and simplification of DDL procedure, a CT IQA metric named SODVAC is designed. The SODVAC locates the optimal regularization parameter that results in the reconstructed image with distinct structures and with no noise or little noise. Third, we introduce SODVAC into the iterative reconstruction framework and then propose a general image-quality-guided iterative reconstruction (QIR) framework and give a specific framework example (sQIR) by introducing SODVAC into the iterative reconstruction framework. sQIR simultaneously optimizes the reconstructed image and the regularization parameter during the iterations. Results confirm the effectiveness of the proposed method. No prior information needed and low computation cost are the advantages of our method compared with existing state-of-theart L-curve and ZIP selection strategies.
Collapse
Affiliation(s)
- Jiayu Duan
- Institute of Image Processing and Pattern Recognition School of Electronics and Information Engineering, Xi'an Jiaotong University, Xi'an, Shannxi, CHINA
| | - Xuanqin Mou
- Institute of Image Processing and Pattern Recognition School of Electronics and Information Engineering, Xi'an Jiaotong University, Xi'an, Shaanxi 710049, Xi'an, 710049, CHINA
| |
Collapse
|
13
|
Othman N, Simon AC, Montagu T, Berteloot L, Grévent D, Habib Geryes B, Benkreira M, Bigand E, Capdeville S, Desrousseaux J, Farman B, Garnier E, Gempp S, Nigoul JM, Nomikossoff N, Vincent M. Toward a comparison and an optimization of CT protocols using new metrics of dose and image quality part I: prediction of human observers using a model observer for detection and discrimination tasks in low-dose CT images in various scanning conditions. Phys Med Biol 2021; 66. [PMID: 33887706 DOI: 10.1088/1361-6560/abfad8] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/03/2020] [Accepted: 04/22/2021] [Indexed: 11/11/2022]
Abstract
In the context of reducing the patient dose coming from CT scanner examinations without penalizing the diagnosis, the assessment of both patient dose and image quality (IQ) with relevant metrics is crucial. The present study represents the first stage in a larger work, aiming to compare and optimize CT protocols using dose and IQ new metrics. We proposed here to evaluate the capacity of the Non-PreWhitening matched filter with an eye (NPWE) model observer to be a robust and accurate estimation of IQ. We focused our work on two types of clinical tasks: a low contrast detection task and a discrimination task. We designed a torso-shaped phantom, including Plastic Water®slabs with cylindrical inserts of different diameters, sections and compositions. We led a human observer study with 13 human observers on images acquired in multiple irradiation and reconstruction scanning conditions (voltage, pitch, slice thickness, noise level of the reconstruction algorithm, energy level in dual-energy mode and dose), to evaluate the behavior of the model observer compared to the human responses faced to changing conditions. The model observer presented the same trends as the human observers with generally better results. We rescaled the NPWE model on the human responses by scanning conditions (kVp, pitch, slice thickness) to obtain the best agreement between both observer types, estimated using the Bland-Altman method. The impact of some scanning parameters was estimated using the correct answer rate given by the rescaled NPWE model, for both tasks and each insert size. In particular, the comparison between the dual-energy mode at 74 keV and the single-energy mode at 120 kVp showed that, if the 120 kVp voltage provided better results for the smallest insert at the lower doses for both tasks, their responses were equivalent in many cases.
Collapse
Affiliation(s)
- Nadia Othman
- Université Paris-Saclay, CEA, List, F-91120 Palaiseau, France
| | | | - Thierry Montagu
- Université Paris-Saclay, CEA, List, F-91120 Palaiseau, France
| | - Laureline Berteloot
- Necker-Enfants Malades University Hospital, Paediatric Radiology Department, Paris, France
| | - David Grévent
- Necker-Enfants Malades University Hospital, Paediatric Radiology Department, Paris, France
| | - Bouchra Habib Geryes
- Necker-Enfants Malades University Hospital, Paediatric Radiology Department, Paris, France
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
14
|
Narita K, Nakamura Y, Higaki T, Akagi M, Honda Y, Awai K. Deep learning reconstruction of drip-infusion cholangiography acquired with ultra-high-resolution computed tomography. Abdom Radiol (NY) 2020; 45:2698-2704. [PMID: 32248261 DOI: 10.1007/s00261-020-02508-4] [Citation(s) in RCA: 25] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/24/2022]
Abstract
PURPOSE Deep learning reconstruction (DLR) introduces deep convolutional neural networks into the reconstruction flow. We examined the clinical applicability of drip-infusion cholangiography (DIC) acquired on an ultra-high-resolution CT (U-HRCT) scanner reconstructed with DLR in comparison to hybrid and model-based iterative reconstruction (hybrid-IR, MBIR). METHODS This retrospective, single-institution study included 30 patients seen between January 2018 and November 2019. A radiologist recorded the standard deviation of attenuation in the paraspinal muscle as the image noise and calculated the contrast-to-noise ratio (CNR) in the common bile duct. The overall visual image quality of the bile duct on thick-slab maximum intensity projections was assessed by two other radiologists and graded on a 5-point confidence scale ranging from 1 (not delineated) to 5 (clearly delineated). The difference among hybrid-IR, MBIR, and DLR images was compared. RESULTS The image noise was significantly lower on DLR than hybrid-IR and MBIR images and the CNR and the overall visual image quality of the bile duct were significantly higher on DLR than on hybrid-IR and MBIR images (all: p < 0.001). CONCLUSION DLR resulted in significant quantitative and qualitative improvement of DIC acquired with U-HRCT.
Collapse
Affiliation(s)
- Keigo Narita
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan
| | - Yuko Nakamura
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan.
| | - Toru Higaki
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan
| | - Motonori Akagi
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan
| | - Yukiko Honda
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan
| | - Kazuo Awai
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, 734-8551, Japan
| |
Collapse
|
15
|
Task-based characterization of a deep learning image reconstruction and comparison with filtered back-projection and a partial model-based iterative reconstruction in abdominal CT: A phantom study. Phys Med 2020; 76:28-37. [DOI: 10.1016/j.ejmp.2020.06.004] [Citation(s) in RCA: 39] [Impact Index Per Article: 7.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/02/2020] [Revised: 04/28/2020] [Accepted: 06/02/2020] [Indexed: 12/12/2022] Open
|
16
|
Funama Y, Takahashi H, Goto T, Aoki Y, Yoshida R, Kumagai Y, Awai K. Improving Low-contrast Detectability and Noise Texture Pattern for Computed Tomography Using Iterative Reconstruction Accelerated with Machine Learning Method: A Phantom Study. Acad Radiol 2020; 27:929-936. [PMID: 31918961 DOI: 10.1016/j.acra.2019.09.007] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/17/2019] [Revised: 09/01/2019] [Accepted: 09/11/2019] [Indexed: 11/15/2022]
Abstract
RATIONALE AND OBJECTIVES To evaluate the performance of iterative reconstruction (IR) and filtered back projection (FBP) images in terms of low-contrast detectability at different radiation doses, IR levels, and slice thickness using the mathematical model observer with a focus on low-contrast detectability. MATERIALS AND METHODS The CCT189 MITA CT IQ Low-Contrast Phantom was used and helical scans were performed using a 64-detector CT scanner. Tube voltage was set at 120 kVp and tube current was adjusted from 45 to 600 mA. Images were reconstructed at slice thicknesses of 0.625 and 5.0 mm with FBP and five types of iterative progressive reconstruction with visual modeling (IPV) algorithms. The noise power spectrum (NPS) and normalized NPS were calculated. To evaluate low-contrast detectability, the model observer with the channelized Hotelling observer model was applied using low-contrast modules in the phantom. RESULTS The NPS and normalized NPS for IPV images had similar curves as that for FBP images. At a slice thickness of 0.625 mm and equivalent radiation dose level, the mean improvement of low-contrast detectability for IPV images was 1.19-2.15-fold greater than FBP images with corresponding noise reduction levels. At equivalent noise levels of 5.0-8.0 HU, low-contrast detectability of the IPVstd2 to IPVstr2 images as almost the same or better than that of the FBP images. However, the detectability of the IPVstr4 image was lower than that of the FBP image (p = 0.02). CONCLUSION Low-contrast detectability of the IPV images was improved with a similar normalized NPS as with FBP images. Furthermore, a radiation reduction of >50% was achieved for the IPV images, while maintaining similar low-contrast detectability.
Collapse
Affiliation(s)
- Yoshinori Funama
- Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto 862-0976, Japan.
| | | | - Taiga Goto
- Hitachi Ltd. Healthcare Business Unit, Kashiwa, Chiba, Japan
| | - Yuko Aoki
- Hitachi Ltd. Healthcare Business Unit, Kashiwa, Chiba, Japan
| | - Ryo Yoshida
- Hitachi Ltd. Healthcare Business Unit, Kashiwa, Chiba, Japan
| | - Yukio Kumagai
- Hitachi Ltd. Healthcare Business Unit, Kashiwa, Chiba, Japan
| | - Kazuo Awai
- Department of Diagnostic Radiology, Graduate School of Biomedical Sciences, Hiroshima University, Hiroshima, Japan
| |
Collapse
|
17
|
Rampado O, Depaoli A, Marchisio F, Gatti M, Racine D, Ruggeri V, Ruggirello I, Darvizeh F, Fonio P, Ropolo R. Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study. Radiol Med 2020; 126:55-62. [PMID: 32495272 DOI: 10.1007/s11547-020-01228-5] [Citation(s) in RCA: 12] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/20/2020] [Accepted: 05/12/2020] [Indexed: 01/19/2023]
Abstract
PURPOSE The purpose of this study was to verify the maintenance of low-contrast detectability at different CT dose reduction levels, in patients of different sizes, as a consequence of the application of iterative reconstruction at different strengths combined with tube current modulation. METHODS Anthropomorphic abdominal phantoms of two sizes (small and large) were imaged at a fixed noise with iterative algorithm ASIR-V percentages in the range between 0 and 70% and corresponding dose reductions in the range of 0-83%. A total of 1400 images with and without liver low-contrast simulated lesions were evaluated by five radiologists, using the receiver operating characteristics (ROC) paradigm and evaluating the area under the ROC curve (AUC). The human observer results were then compared with AUC obtained with a channelized Hotelling observer (CHO). CNR values were also calculated. RESULTS For the small phantom, the AUC values lie between 0.90 and 0.93 for human evaluations of images acquired without iterative reconstruction, with 30% ASIR-V and with 50% ASIR-V. The AUC decreased significantly to 0.81 (p = 0.0001) at 70% ASIR-V. The CHO results were in coherence with human observer scores. Also, similar results were observed for the large size phantom. CNR values were stable for the different ASIR-V percentages. CONCLUSIONS The iterative algorithm maintained the low-contrast detectability up to a dose reduction of about 70%, following application of a 50% ASIR-V combined with automatic tube current modulation, regardless of the phantom size. At further dose reductions using greater iterative percentages, a significant decrease in detectability was observed.
Collapse
Affiliation(s)
- Osvaldo Rampado
- Medical Physics Unit, S.C. Fisica Sanitaria, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy.
| | - Alessandro Depaoli
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Filippo Marchisio
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Marco Gatti
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Damien Racine
- Institute of Radiation Physics, Lausanne University Hospital, Rue du Grand-Pré 1, 1007, Lausanne, Switzerland
| | - Valeria Ruggeri
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Irene Ruggirello
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Fatemeh Darvizeh
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Paolo Fonio
- University Radiodiagnostic Unit, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| | - Roberto Ropolo
- Medical Physics Unit, S.C. Fisica Sanitaria, A.O.U. Città della Salute e della Scienza di Torino, Corso Bramante 88, 10126, Turin, Italy
| |
Collapse
|
18
|
Possibility of Deep Learning in Medical Imaging Focusing Improvement of Computed Tomography Image Quality. J Comput Assist Tomogr 2020; 44:161-167. [PMID: 31789682 DOI: 10.1097/rct.0000000000000928] [Citation(s) in RCA: 34] [Impact Index Per Article: 6.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/16/2022]
Abstract
Deep learning (DL), part of a broader family of machine learning methods, is based on learning data representations rather than task-specific algorithms. Deep learning can be used to improve the image quality of clinical scans with image noise reduction. We review the ability of DL to reduce the image noise, present the advantages and disadvantages of computed tomography image reconstruction, and examine the potential value of new DL-based computed tomography image reconstruction.
Collapse
|
19
|
Rubert N, Southard R, Hamman SM, Robison R. Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study. Pediatr Radiol 2020; 50:345-356. [PMID: 31705156 DOI: 10.1007/s00247-019-04561-8] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/26/2019] [Revised: 09/23/2019] [Accepted: 10/16/2019] [Indexed: 12/20/2022]
Abstract
BACKGROUND Iterative reconstruction is offered by all vendors to achieve similar or better CT image quality at lower doses than images reconstructed with filtered back-projection. OBJECTIVE The purpose of this study was to investigate the dose-reduction potential for pediatric abdominal CT imaging when using either a commercially available hybrid or a commercially available model-based iterative reconstruction algorithm from a single manufacturer. MATERIALS AND METHODS A phantom containing four low-contrast inserts and a uniform background with total attenuation equivalent to the abdomen of an average 8-year-old child was imaged on a CT scanner (IQon; Philips Healthcare, Cleveland, OH). We reconstructed images using both hybrid iterative reconstruction (iDose4) and model-based iterative reconstruction (Iterative Model Reconstruction). The four low-contrast inserts had circular cross-section with diameters of 3 mm, 5 mm, 7 mm and 10 mm and contrasts of 14 Hounsfield units (HU), 7 HU, 5 HU and 3 HU, respectively. Helical scans with identical kilovoltage (kV), pitch, rotation time, and collimation were repeated on the phantom at volume CT dose index (CTDIvol) of 2.0 milligrays (mGy), 3.0 mGy, 4.5 mGy and 6.0 mGy. We measured the contrast-to-noise ratio (CNR) in each rod across scans. Additionally, we collected sub-images containing each rod and sub-images containing the background and used them in two-alternative forced choice observer experiments with four observers (two radiologists and two physicists). We calculated the dose-reduction potential of both iterative reconstruction algorithms relative to a scan performed at 6 mGy and reconstructed with filtered back-projection. RESULTS We calculated dose-reduction potential by either matching average equal observer performance in the two-alternative forced choice experiments or matching CNR. When matching CNR, the dose-reduction potential was 34% to 45% for hybrid iterative reconstruction and 89% to 95% for model-based iterative reconstruction. When matching average observer performance, the dose-reduction potential was 9% to 30% for hybrid iterative reconstruction and 57% to 74% for model-based iterative reconstruction. The range in dose-reduction potential depended on the rod size and contrast level. CONCLUSION Observer performance in this phantom study indicates that the dose-reduction potential indicated by an observer study is less than that of CNR; extrapolating the results to clinical studies suggests that the dose-reduction potential would also be less.
Collapse
Affiliation(s)
- Nicholas Rubert
- Department of Radiology, Phoenix Children's Hospital, 1919 E. Thomas Road, Phoenix, AZ, 85016, USA.
| | - Richard Southard
- Department of Radiology, Phoenix Children's Hospital, 1919 E. Thomas Road, Phoenix, AZ, 85016, USA.,College of Medicine, University of Arizona, Phoenix, AZ, USA
| | - Susan M Hamman
- Section of Pediatric Radiology, C. S. Mott Children's Hospital, Department of Radiology, University of Michigan Health System, Ann Arbor, MI, USA
| | - Ryan Robison
- Department of Radiology, Phoenix Children's Hospital, 1919 E. Thomas Road, Phoenix, AZ, 85016, USA.,College of Nursing and Health Innovation, Arizona State University, Phoenix, AZ, USA
| |
Collapse
|
20
|
Nakamura Y, Higaki T, Tatsugami F, Zhou J, Yu Z, Akino N, Ito Y, Iida M, Awai K. Deep Learning-based CT Image Reconstruction: Initial Evaluation Targeting Hypovascular Hepatic Metastases. Radiol Artif Intell 2019; 1:e180011. [PMID: 33937803 DOI: 10.1148/ryai.2019180011] [Citation(s) in RCA: 53] [Impact Index Per Article: 8.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/22/2018] [Revised: 06/03/2019] [Accepted: 07/05/2019] [Indexed: 02/07/2023]
Abstract
Purpose To evaluate the effect of a deep learning-based reconstruction (DLR) method on the conspicuity of hypovascular hepatic metastases on abdominal CT images. Materials and Methods This retrospective study with institutional review board approval included 58 patients with hypovascular hepatic metastases. A radiologist recorded the standard deviation of attenuation in the paraspinal muscle as the image noise and the contrast-to-noise ratio (CNR). CNR was calculated as region of interest ([ROI]L - ROIT)/N, where ROIL is the mean liver parenchyma attenuation, ROIT, the mean tumor attenuation, and N, the noise. Two other radiologists graded the conspicuity of the liver lesion on a five-point scale where 1 is unidentifiable and 5 is detected without diagnostic compromise. Only the smallest liver lesion in each patient, classified as smaller or larger than 10 mm, was evaluated. The difference between hybrid iterative reconstruction (IR) and DLR images was determined by using a two-sided Wilcoxon signed-rank test. Results The image noise was significantly lower, and the CNR was significantly higher on DLR images than hybrid IR images (median image noise: 19.2 vs 12.8 HU, P < .001; median CNR: tumors < 10 mm: 1.9 vs 2.5; tumors > 10 mm: 1.7 vs 2.2, both P < .001). The scores for liver lesions were significantly higher for DLR images than hybrid IR images (P < .01 for both in tumors smaller or larger than 10 mm). Conclusion DLR improved the quality of abdominal CT images for the evaluation of hypovascular hepatic metastases.© RSNA, 2019Supplemental material is available for this article.
Collapse
Affiliation(s)
- Yuko Nakamura
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Toru Higaki
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Fuminari Tatsugami
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Jian Zhou
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Zhou Yu
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Naruomi Akino
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Yuya Ito
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Makoto Iida
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| | - Kazuo Awai
- Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan 734-8551 (Y.N., T.H., F.T., M.I., K.A.); Canon Medical Research USA, Vernon Hills, Ill (J.Z., Z.Y.); and Canon Medical Systems, Tochigi, Japan (N.A., Y.I.)
| |
Collapse
|
21
|
Balta C, Bouwman RW, Broeders MJM, Karssemeijer N, Veldkamp WJH, Sechopoulos I, van Engen RE. Optimization of the difference-of-Gaussian channel sets for the channelized Hotelling observer. J Med Imaging (Bellingham) 2019; 6:035501. [PMID: 31572746 PMCID: PMC6763759 DOI: 10.1117/1.jmi.6.3.035501] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/29/2019] [Accepted: 08/30/2019] [Indexed: 10/15/2023] Open
Abstract
The channelized-Hotelling observer (CHO) was investigated as a surrogate of human observers in task-based image quality assessment. The CHO with difference-of-Gaussian (DoG) channels has shown potential for the prediction of human detection performance in digital mammography (DM) images. However, the DoG channels employ parameters that describe the shape of each channel. The selection of these parameters influences the performance of the DoG CHO and needs further investigation. The detection performance of the DoG CHO was calculated and correlated with the detection performance of three humans who evaluated DM images in 2-alternative forced-choice experiments. A set of DM images of an anthropomorphic breast phantom with and without calcification-like signals was acquired at four different dose levels. For each dose level, 200 square regions-of-interest (ROIs) with and without signal were extracted. Signal detectability was assessed on ROI basis using the CHO with various DoG channel parameters and it was compared to that of the human observers. It was found that varying these DoG parameter values affects the correlation (r 2 ) of the CHO with human observers for the detection task investigated. In conclusion, it appears that the the optimal DoG channel sets that maximize the prediction ability of the CHO might be dependent on the type of background and signal of ROIs investigated.
Collapse
Affiliation(s)
- Christiana Balta
- Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Nijmegen, The Netherlands
| | | | - Mireille J. M. Broeders
- Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Department for Health Evidence, Nijmegen, The Netherlands
| | - Nico Karssemeijer
- Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Nijmegen, The Netherlands
| | | | - Ioannis Sechopoulos
- Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Nijmegen, The Netherlands
| | | |
Collapse
|
22
|
Brat H, Zanca F, Montandon S, Racine D, Rizk B, Meicher E, Fournier D. Local clinical diagnostic reference levels for chest and abdomen CT examinations in adults as a function of body mass index and clinical indication: a prospective multicenter study. Eur Radiol 2019; 29:6794-6804. [PMID: 31144074 DOI: 10.1007/s00330-019-06257-x] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/10/2019] [Revised: 04/16/2019] [Accepted: 04/26/2019] [Indexed: 11/30/2022]
Abstract
OBJECTIVES To compare institutional dose levels based on clinical indication and BMI class to anatomy-based national DRLs (NDRLs) in chest and abdomen CT examinations and to assess local clinical diagnostic reference levels (LCDRLs). METHODS From February 2017 to June 2018, after protocol optimization according to clinical indication and body mass index (BMI) class (< 25; ≥ 25), 5310 abdomen and 1058 chest CT series were collected from 5 CT scanners in a Swiss multicenter group. Clinical indication-based institutional dose levels were compared to the Swiss anatomy-based NDRLs. Statistical significance was assessed (p < 0.05). LCDRLs were calculated as the third quartile of the median dose values for each CT scanner. RESULTS For chest examinations, dose metrics based on clinical indication were always below P75 NDRL for CTDIvol (range 3.9-6.4 vs. 7.0 mGy) and DLP (164.0-211.2 vs. 250 mGycm) in all BMI classes except for DLP in BMI ≥ 25 (248.8-255.4 vs. 250.0 mGycm). For abdomen examinations, they were significantly lower or not different than P50 NDRLs for all BMI classes (3.8-9.0 vs. 10.0 mGy and 192.9-446.8 vs. 470mGycm). The estimated LCDRLs show a drop in CTDIvol (21% for chest and 32% for abdomen, on average) with respect to current DRLs. When considering BMI stratification, the largest LCDRL difference within the same clinical indication is for renal tumor (4.6 mGy for BMI < 25 vs. 10.0 mGy for BMI ≥ 25; - 117%). CONCLUSION The results suggest the necessity of estimating clinical indication-based DRLs, especially for abdomen examinations. Stratifying per BMI class allows further optimization of the CT doses. KEY POINTS • Our data show that clinical indication-based DRLs might be more appropriate than anatomy-based DRLs and might help in reducing large variations in dose levels for the same type of examinations. • Stratifying the data per patient-size subgroups (non-overweight, overweight) allows a better optimization of CT doses and therefore the possibility to set LCDRLs based on BMI class. • Institutions who are fostering continuous dose optimization and LDRLs should consider defining protocols based on clinical indication and BMI group, to achieve ALARA.
Collapse
Affiliation(s)
- Hugues Brat
- Institut de Radiologie de Sion, Groupe 3R, Sion, Switzerland.
| | - Federica Zanca
- GE Healthcare, Buc, France.,Palindromo Consulting, Leuven, Belgium
| | | | - Damien Racine
- Institute of Radiation Physics (IRA), Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland
| | - Benoit Rizk
- Institut de Radiologie de Sion, Groupe 3R, Sion, Switzerland
| | - Eric Meicher
- Institut de Radiologie de Sion, Groupe 3R, Sion, Switzerland
| | | |
Collapse
|
23
|
Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images. J Med Imaging (Bellingham) 2019; 6:025501. [DOI: 10.1117/1.jmi.6.2.025501] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/07/2018] [Accepted: 04/15/2019] [Indexed: 11/14/2022] Open
|
24
|
Dilger SKN, Yu L, Chen B, Favazza CP, Carter RE, Fletcher JG, McCollough CH, Leng S. Localization of liver lesions in abdominal CT imaging: I. Correlation of human observer performance between anatomical and uniform backgrounds. Phys Med Biol 2019; 64:105011. [PMID: 30995611 DOI: 10.1088/1361-6560/ab1a45] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The purpose of this study was to determine the correlation between human observer performance for localization of small low contrast lesions within uniform water background versus an anatomical liver background, under the conditions of varying dose, lesion size, and reconstruction algorithm. Liver lesions (5 mm, 7 mm, and 9 mm, contrast: -21 HU) were digitally inserted into CT projection data of ten normal patients in vessel-free liver regions. Noise was inserted into the projection data to create three image sets: full dose and simulated half and quarter doses. Images were reconstructed with a standard filtered back projection (FBP) and an iterative reconstruction (IR) algorithm. Lesion and noise insertion procedures were repeated with water phantom data. Two-dimensional regions of interest (66 lesion-present, 34 lesion-absent) were selected, randomized, and independently reviewed by three medical physicists to identify the most likely location of the lesion and provide a confidence score. Locations and confidence scores were assessed using the area under the localization receiver operating characteristic curve (AzLROC). We examined the correlation between human performance for the liver and uniform water backgrounds as dose, lesion size, and reconstruction algorithm varied. As lesion size or dose increased, reader localization performance improved. For full dose IR images, the AzLROC for 5, 7, and 9 mm lesions were 0.53, 0.91, and 0.97 (liver) and 0.51, 0.96, and 0.99 (uniform water), respectively. Similar trends were seen with other parameters. Performance values for liver and uniform backgrounds were highly correlated for both reconstruction algorithms, with a Spearman correlation of ρ = 0.97, and an average difference in AzLROC of 0.05 ± 0.04. For the task of localizing low contrast liver lesions, human observer performance was highly correlated between anatomical and uniform backgrounds, suggesting that lesion localization studies emulating a clinical test of liver lesion detection can be performed using a uniform background.
Collapse
Affiliation(s)
- Samantha K N Dilger
- Department of Radiology, Mayo Clinic, Rochester, MN, United States of America
| | | | | | | | | | | | | | | |
Collapse
|
25
|
Akagi M, Nakamura Y, Higaki T, Narita K, Honda Y, Zhou J, Yu Z, Akino N, Awai K. Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol 2019; 29:6163-6171. [PMID: 30976831 DOI: 10.1007/s00330-019-06170-3] [Citation(s) in RCA: 237] [Impact Index Per Article: 39.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/06/2019] [Revised: 02/22/2019] [Accepted: 03/14/2019] [Indexed: 12/12/2022]
Abstract
OBJECTIVES Deep learning reconstruction (DLR) is a new reconstruction method; it introduces deep convolutional neural networks into the reconstruction flow. This study was conducted in order to examine the clinical applicability of abdominal ultra-high-resolution CT (U-HRCT) exams reconstructed with a new DLR in comparison to hybrid and model-based iterative reconstruction (hybrid-IR, MBIR). METHODS Our retrospective study included 46 patients seen between December 2017 and April 2018. A radiologist recorded the standard deviation of attenuation in the paraspinal muscle as the image noise and calculated the contrast-to-noise ratio (CNR) for the aorta, portal vein, and liver. The overall image quality was assessed by two other radiologists and graded on a 5-point confidence scale ranging from 1 (unacceptable) to 5 (excellent). The difference between CT images subjected to hybrid-IR, MBIR, and DLR was compared. RESULTS The image noise was significantly lower and the CNR was significantly higher on DLR than hybrid-IR and MBIR images (p < 0.01). DLR images received the highest and MBIR images the lowest scores for overall image quality. CONCLUSIONS DLR improved the quality of abdominal U-HRCT images. KEY POINTS • The potential degradation due to increased noise may prevent implementation of ultra-high-resolution CT in the abdomen. • Image noise and overall image quality for hepatic ultra-high-resolution CT images improved with deep learning reconstruction as compared to hybrid- and model-based iterative reconstruction.
Collapse
Affiliation(s)
- Motonori Akagi
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan
| | - Yuko Nakamura
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan.
| | - Toru Higaki
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan
| | - Keigo Narita
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan
| | - Yukiko Honda
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan
| | - Jian Zhou
- Canon Medical Research USA, Inc., Vernon Hills, IL, USA
| | - Zhou Yu
- Canon Medical Research USA, Inc., Vernon Hills, IL, USA
| | | | - Kazuo Awai
- Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima, Japan
| |
Collapse
|
26
|
Goto M, Tominaga C, Taura M, Azumi H, Sato K, Homma N, Mori I. A method to measure slice sensitivity profiles of CT images under low-contrast and high-noise conditions. Phys Med 2019; 60:100-110. [PMID: 31000069 DOI: 10.1016/j.ejmp.2019.03.010] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/27/2018] [Revised: 03/03/2019] [Accepted: 03/11/2019] [Indexed: 11/17/2022] Open
Abstract
Noise reduction features of iterative reconstruction (IR) methods in computed tomography might accompany the sacrifice of the longitudinal resolution, or slice sensitivity profile (SSP), at low contrast-to-noise ratio (CNR) conditions. To assess the benefit of IR methods correctly, the difference of SSP between IR methods and filtered-backprojection (FBP) must be taken into account. Therefore, SSP measurement under low-CNR conditions is necessary. Although edge methods are predominantly used, their performance under low-CNR conditions appears to be not fully established. We developed a method that is compatible with extremely low-CNR conditions. Thin plastic disk-shaped sheets embedded in acrylic resin were used as low-contrast test objects. The lowest peak contrast used was approximately 17 [HU]. We assessed the performance of our method by using FBP images. We identified a source of measurement instability aside from noise: the measured thin-slice SSP is dependent on the orbital phase of helical scan, presumably because of cone-beam artifacts. This impediment to high accuracy is manageable using phase-controlled scans. We confirmed that table position repeatability is much better than the value of the specifications, and therefore the ensemble-averaged images of multiple scans can be used for SSP measurement. Accurate measurement of SSP under extremely low-CNR conditions is possible, even when the test object is visually indiscernible from the noisy background. Low-contrast SSP behavior is elucidated for IR methods (AIDR-3D, FIRST, and AiSR-V) by using this measurement method.
Collapse
Affiliation(s)
- Mitsunori Goto
- Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan; Miyagi Cancer Center Natori, 981-1293, Japan.
| | - Chiaki Tominaga
- Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan; Chiba University Hospital, Chiba 260-8677, Japan
| | - Masaaki Taura
- Tohoku Medical and Pharmaceutical University Hospital, Sendai 983-8512, Japan
| | | | - Kazuhiro Sato
- Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan
| | - Noriyasu Homma
- Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan
| | - Issei Mori
- Graduate School of Medicine, Tohoku University, Sendai 980-8575, Japan
| |
Collapse
|
27
|
Petrov D, Marshall NW, Young KC, Bosmans H. Systematic approach to a channelized Hotelling model observer implementation for a physical phantom containing mass-like lesions: Application to digital breast tomosynthesis. Phys Med 2019; 58:8-20. [PMID: 30824154 DOI: 10.1016/j.ejmp.2018.12.033] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/21/2018] [Revised: 12/05/2018] [Accepted: 12/25/2018] [Indexed: 10/27/2022] Open
Abstract
PURPOSE to develop a channelized model observer (CHO) that matches human reader (HR) scoring of a physical phantom containing breast simulating structure and mass lesion-like targets for use in quality control of digital breast tomosynthesis (DBT) imaging systems. METHODS A total of 108 DBT scans of the phantom was acquired using a Siemens Inspiration DBT system. The detectability of mass-like targets was evaluated by human readers using a 4-alternative forced choice (4-AFC) method. The percentage correct (PC) values were then used as the benchmark for CHO tuning, again using a 4-AFC method. Three different channel functions were considered: Gabor, Laguerre-Gauss and Difference of Gaussian. With regard to the observer template, various methods for generating the expected signal were studied along with the influence of the number of training images used to form the covariance matrix for the observer template. Impact of bias in the training process on the observer template was evaluated next, as well as HR and CHO reproducibility. RESULTS HR performance was most closely matched by 8 Gabor channels with tuned phase, orientation and frequency, using an observer template generated from training image data. Just 24 DBT image stacks were required to give robust CHO performance with 0% bias, although a bias of up to 33% in the training images also gave acceptable performance. CHO and HR reproducibility were similar (on average 3.2 PC versus 3.4 PC). CONCLUSIONS The CHO algorithm developed matches human reader performance and is therefore a promising candidate for automated readout of phantom studies.
Collapse
Affiliation(s)
- Dimitar Petrov
- Dept. of Medical Physics and Quality Assessment, KU Leuven, Leuven, Belgium.
| | - Nicholas W Marshall
- Dept. of Medical Physics and Quality Assessment, KU Leuven, Leuven, Belgium; Dept. of Radiology, UZ Leuven, Belgium
| | | | - Hilde Bosmans
- Dept. of Medical Physics and Quality Assessment, KU Leuven, Leuven, Belgium; Dept. of Radiology, UZ Leuven, Belgium
| |
Collapse
|
28
|
Balta C, Bouwman RW, Sechopoulos I, Broeders MJM, Karssemeijer N, van Engen RE, Veldkamp WJH. Can a channelized Hotelling observer assess image quality in acquired mammographic images of an anthropomorphic breast phantom including image processing? Med Phys 2018; 46:714-725. [PMID: 30561108 DOI: 10.1002/mp.13342] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/06/2018] [Revised: 11/20/2018] [Accepted: 11/30/2018] [Indexed: 12/28/2022] Open
Abstract
PURPOSE To study the feasibility of a channelized Hotelling observer (CHO) to predict human observer performance in detecting calcification-like signals in mammography images of an anthropomorphic breast phantom, as part of a quality control (QC) framework. METHODS A prototype anthropomorphic breast phantom with inserted gold disks of 0.25 mm diameter was imaged with two different digital mammography x-ray systems at four different dose levels. Regions of interest (ROIs) were extracted from the acquired processed and unprocessed images, signal-present and signal-absent. The ROIs were evaluated by a CHO using four different formulations of the difference of Gaussian (DoG) channel sets. Three human observers scored the ROIs in a two-alternative forced-choice experiment. We compared the human and the CHO performance on the simple task to detect calcification-like disks in ROIs with and without postprocessing. The proportion of correct responses of the human reader (PCH ) and the CHO (PCCHO ) was calculated and the correlation between the two was analyzed using a mixed-effect regression model. To address the signal location uncertainty, the impact of shifting the DoG channel sets in all directions up to two pixels was evaluated. Correlation results including the goodness of fit (r2 ) of PCH and PCCHO for all different parameters were evaluated. RESULTS Subanalysis by system yielded strong correlations between PCH and PCCHO , with r2 between PCH and PCCHO was found to be between 0.926 and 0.958 for the unshifted and between 0.759 and 0.938 for the shifted channel sets, respectively. However, the linear fit suggested a slight system dependence. PCCHO with shifted channel sets increased CHO performance but the correlation with humans was decreased. These correlations were not considerably affected by of the DoG channel set used. CONCLUSIONS There is potential for the CHO to be used in QC for the evaluation of detectability of calcification-like signals. The CHO can predict the PC of humans in images of calcification-like signals of two different systems. However, a global model to be used for all systems requires further investigation.
Collapse
Affiliation(s)
- C Balta
- Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, Wijchenseweg 101, 6538 SW, Nijmegen, The Netherlands.,Department of Radiology and Nuclear Medicine, Radboud University Medical Center, PO Box 9101, 6500 HB, Nijmegen, The Netherlands
| | - R W Bouwman
- Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, Wijchenseweg 101, 6538 SW, Nijmegen, The Netherlands
| | - I Sechopoulos
- Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, Wijchenseweg 101, 6538 SW, Nijmegen, The Netherlands.,Department of Radiology and Nuclear Medicine, Radboud University Medical Center, PO Box 9101, 6500 HB, Nijmegen, The Netherlands
| | - M J M Broeders
- Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, Wijchenseweg 101, 6538 SW, Nijmegen, The Netherlands.,Department for Health Evidence, Radboud University Medical Center, PO Box 9101, 6500 HB, Nijmegen, The Netherlands
| | - N Karssemeijer
- Department of Radiology and Nuclear Medicine, Radboud University Medical Center, PO Box 9101, 6500 HB, Nijmegen, The Netherlands
| | - R E van Engen
- Dutch Expert Centre for Screening (LRCB), Radboud University Medical Center, Wijchenseweg 101, 6538 SW, Nijmegen, The Netherlands
| | - W J H Veldkamp
- Department of Radiology, Leiden University Medical Centre, Albinusdreef 2, 2333 ZA, Leiden, The Netherlands
| |
Collapse
|
29
|
Balta C, Bouwman RW, Veldkamp WJH, Broeders MJM, Sechopoulos I, van Engen RE. Signal template generation from acquired images for model observer-based image quality analysis in mammography. J Med Imaging (Bellingham) 2018; 5:035503. [PMID: 30840714 PMCID: PMC6129177 DOI: 10.1117/1.jmi.5.3.035503] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/07/2018] [Accepted: 08/13/2018] [Indexed: 09/29/2023] Open
Abstract
Mammography images undergo vendor-specific processing, which may be nonlinear, before radiologist interpretation. Therefore, to test the entire imaging chain, the effect of image processing should be included in the assessment of image quality, which is not current practice. For this purpose, model observers (MOs), in combination with anthropomorphic breast phantoms, are proposed to evaluate image quality in mammography. In this study, the nonprewhitening MO with eye filter and the channelized Hotelling observer were investigated. The goal of this study was to optimize the efficiency of the procedure to obtain the expected signal template from acquired images for the detection of a 0.25-mm diameter disk. Two approaches were followed: using acquired images with homogeneous backgrounds (approach 1) and images from an anthropomorphic breast phantom (approach 2). For quality control purposes, a straightforward procedure using a single exposure of a single disk was found adequate for both approaches. However, only approach 2 can yield templates from processed images since, due to its nonlinearity, image postprocessing cannot be evaluated using images of homogeneous phantoms. Based on the results of the current study, a phantom should be designed, which can be used for the objective assessment of image quality.
Collapse
Affiliation(s)
- Christiana Balta
- Radboud University Medical Center, Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Nijmegen, The Netherlands
| | - Ramona W. Bouwman
- Radboud University Medical Center, Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
| | | | - Mireille J. M. Broeders
- Radboud University Medical Center, Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Radboud Institute for Health Sciences (RIHS), Nijmegen, The Netherlands
| | - Ioannis Sechopoulos
- Radboud University Medical Center, Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
- Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Nijmegen, The Netherlands
| | - Ruben E. van Engen
- Radboud University Medical Center, Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands
| |
Collapse
|
30
|
Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study. Eur Radiol 2018; 28:5203-5210. [PMID: 29858638 PMCID: PMC6223860 DOI: 10.1007/s00330-018-5518-8] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/09/2018] [Revised: 04/19/2018] [Accepted: 04/26/2018] [Indexed: 11/23/2022]
Abstract
Objective We investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question. Methods An anthropomorphic abdominal phantom with two optional rings was scanned on 68 CT systems from 62 centres using the local clinical acquisition parameters of the portal venous phase for the detection of focal liver lesions. Low-contrast detectability (LCD) was assessed objectively with channelised Hotelling observer (CHO) using the receiver operating characteristic (ROC) paradigm. For each lesion size, the area under the ROC curve (AUC) was calculated and considered as a figure of merit. The volume computed tomography dose index (CTDIvol) was used to indicate radiation dose exposure. Results The median CTDIvol used was 5.8 mGy, 10.5 mGy and 16.3 mGy for the small, medium and large phantoms, respectively. The median AUC obtained from clinical CT protocols was 0.96, 0.90 and 0.83 for the small, medium and large phantoms, respectively. Conclusions Our study used a model observer to highlight the difference in image quality levels when dealing with the same clinical question. This difference was important and increased with growing phantom size, which generated large variations in patient exposure. In the end, a standardisation initiative may be launched to ensure comparable diagnostic information for well-defined clinical questions. The image quality requirements, related to the clinical question to be answered, should be the starting point of patient dose optimisation. Key Points • Model observers enable to assess image quality objectively based on clinical tasks. • Objective image quality assessment should always include several patient sizes. • Clinical diagnostic image quality should be the starting point for patient dose optimisation. • Dose optimisation by applying DRLs only is insufficient for ensuring clinical requirements.
Collapse
|
31
|
Ba A, Abbey CK, Baek J, Han M, Bouwman RW, Balta C, Brankov J, Massanes F, Gifford HC, Hernandez-Giron I, Veldkamp WJH, Petrov D, Marshall N, Samuelson FW, Zeng R, Solomon JB, Samei E, Timberg P, Förnvik H, Reiser I, Yu L, Gong H, Bochud FO. Inter-laboratory comparison of channelized hotelling observer computation. Med Phys 2018; 45:3019-3030. [PMID: 29704868 DOI: 10.1002/mp.12940] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/04/2017] [Revised: 04/11/2018] [Accepted: 04/15/2018] [Indexed: 01/14/2023] Open
Abstract
PURPOSE The task-based assessment of image quality using model observers is increasingly used for the assessment of different imaging modalities. However, the performance computation of model observers needs standardization as well as a well-established trust in its implementation methodology and uncertainty estimation. The purpose of this work was to determine the degree of equivalence of the channelized Hotelling observer performance and uncertainty estimation using an intercomparison exercise. MATERIALS AND METHODS Image samples to estimate model observer performance for detection tasks were generated from two-dimensional CT image slices of a uniform water phantom. A common set of images was sent to participating laboratories to perform and document the following tasks: (a) estimate the detectability index of a well-defined CHO and its uncertainty in three conditions involving different sized targets all at the same dose, and (b) apply this CHO to an image set where ground truth was unknown to participants (lower image dose). In addition, and on an optional basis, we asked the participating laboratories to (c) estimate the performance of real human observers from a psychophysical experiment of their choice. Each of the 13 participating laboratories was confidentially assigned a participant number and image sets could be downloaded through a secure server. Results were distributed with each participant recognizable by its number and then each laboratory was able to modify their results with justification as model observer calculation are not yet a routine and potentially error prone. RESULTS Detectability index increased with signal size for all participants and was very consistent for 6 mm sized target while showing higher variability for 8 and 10 mm sized target. There was one order of magnitude between the lowest and the largest uncertainty estimation. CONCLUSIONS This intercomparison helped define the state of the art of model observer performance computation and with thirteen participants, reflects openness and trust within the medical imaging community. The performance of a CHO with explicitly defined channels and a relatively large number of test images was consistently estimated by all participants. In contrast, the paper demonstrates that there is no agreement on estimating the variance of detectability in the training and testing setting.
Collapse
Affiliation(s)
- Alexandre Ba
- Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland
| | - Craig K Abbey
- Department of Psychological and Brain Sciences, University of California Santa Barbara, Santa Barbara, CA, 93106, USA
| | - Jongduk Baek
- School of Integrated Technology, Yonsei University, 406-840, Incheon, Korea
| | - Minah Han
- School of Integrated Technology, Yonsei University, 406-840, Incheon, Korea
| | - Ramona W Bouwman
- Dutch Expert Centre for Screening, Radboud University Nijmegen Medical Centre (LRCB), P.O. Box 6873, 6503 GJ, Nijmegen, The Netherlands
| | - Christiana Balta
- Dutch Expert Centre for Screening, Radboud University Nijmegen Medical Centre (LRCB), P.O. Box 6873, 6503 GJ, Nijmegen, The Netherlands
| | - Jovan Brankov
- Department of Electrical and Computer Engineering, Illinois Institute of Technology, 3301 South Dearborn Street, Chicago, IL, 60616, USA
| | - Francesc Massanes
- Department of Electrical and Computer Engineering, Illinois Institute of Technology, 3301 South Dearborn Street, Chicago, IL, 60616, USA
| | - Howard C Gifford
- Department of Biomedical Engineering, University of Houston, Houston, TX, 77204, USA
| | - Irene Hernandez-Giron
- Radiology Department, Leiden University Medical Center (LUMC), Albinusdreef 2, 2333ZA, Leiden, The Netherlands
| | - Wouter J H Veldkamp
- Radiology Department, Leiden University Medical Center (LUMC), Albinusdreef 2, 2333ZA, Leiden, The Netherlands
| | - Dimitar Petrov
- Department of Medical Physics and Quality Assessment, KU Leuven, Leuven, Belgium
| | - Nicholas Marshall
- Department of Medical Physics and Quality Assessment, KU Leuven, Leuven, Belgium.,Department of Radiology, UZ Leuven, Leuven, Belgium
| | - Frank W Samuelson
- Division of Imaging, Diagnostics and Software Reliability, US Food and Drug Administration, 10903 New Hampshire Ave Building 62, Room 3102, Silver Spring, MD, 20903-1058, USA
| | - Rongping Zeng
- Division of Imaging, Diagnostics and Software Reliability, US Food and Drug Administration, 10903 New Hampshire Ave Building 62, Room 3102, Silver Spring, MD, 20903-1058, USA
| | - Justin B Solomon
- Carl E. Ravin Advanced Imaging Laboratories, Departments of Radiology, Electrical and Computer Engineering, Biomedical Engineering, and Physics, Clinical Imaging Physics Group, Medical Physics Graduate Program, Duke University, Durham, NC, 27705, USA
| | - Ehsan Samei
- Carl E. Ravin Advanced Imaging Laboratories, Departments of Radiology, Electrical and Computer Engineering, Biomedical Engineering, and Physics, Clinical Imaging Physics Group, Medical Physics Graduate Program, Duke University, Durham, NC, 27705, USA
| | - Pontus Timberg
- Department of Medical Radiation Physics, Translational Medicine Malmö, Lund University, Malmö, Sweden
| | - Hannie Förnvik
- Department of Medical Radiation Physics, Translational Medicine Malmö, Lund University, Malmö, Sweden
| | - Ingrid Reiser
- Department of Radiology, University of Chicago, 5841 S Maryland Ave, MC 2026, Chicago, IL, 60637, USA
| | - Lifeng Yu
- Department of Radiology, Mayo Clinic, Rochester, MN, USA
| | - Hao Gong
- Department of Radiology, Mayo Clinic, Rochester, MN, USA
| | - François O Bochud
- Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland
| |
Collapse
|
32
|
Karbaschi Z, Gifford HC. Assessing CT acquisition parameters with visual-search model observers. J Med Imaging (Bellingham) 2018; 5:025501. [PMID: 29662920 DOI: 10.1117/1.jmi.5.2.025501] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/22/2017] [Accepted: 03/12/2018] [Indexed: 11/14/2022] Open
Abstract
A principal difference between the channelized Hotelling (CH) and visual-search (VS) model observers is how they respond to noise texture in images. We compared the two observers in lesion-detection studies to evaluate linear and angular sampling parameters for CT. Simulated lung images were generated from a single two-dimensional mathematical torso phantom containing circular lesions of fixed radius and relative contrast. Projection datasets were produced for two detector pixel sizes and from 15 to 128 projections at 15 and 65 M counts per set. Filtered backprojection reconstructions were obtained with dimensions of [Formula: see text] and [Formula: see text]. A localization receiver operating characteristic study was conducted with two human observers, three single-feature VS observers, and a feature-adaptive VS observer. The effects of the sampling parameters on performance were similar for all of these observers. The CH observer, applied in location-known studies with and without background variability, was not affected by the variations in angular sampling. The two-stage VS framework was an effective modification of the CH observer for assessing the effects of noise texture on human-observer performance in this study.
Collapse
Affiliation(s)
- Zohreh Karbaschi
- University of Houston, Department of Biomedical Engineering, Houston, Texas, United States
| | - Howard C Gifford
- University of Houston, Department of Biomedical Engineering, Houston, Texas, United States
| |
Collapse
|
33
|
Solomon J, Ba A, Bochud F, Samei E. Comparison of low-contrast detectability between two CT reconstruction algorithms using voxel-based 3D printed textured phantoms. Med Phys 2017; 43:6497. [PMID: 27908164 DOI: 10.1118/1.4967478] [Citation(s) in RCA: 51] [Impact Index Per Article: 6.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE To use novel voxel-based 3D printed textured phantoms in order to compare low-contrast detectability between two reconstruction algorithms, FBP (filtered-backprojection) and SAFIRE (sinogram affirmed iterative reconstruction) and determine what impact background texture (i.e., anatomical noise) has on estimating the dose reduction potential of SAFIRE. METHODS Liver volumes were segmented from 23 abdominal CT cases. The volumes were characterized in terms of texture features from gray-level co-occurrence and run-length matrices. Using a 3D clustered lumpy background (CLB) model, a fitting technique based on a genetic optimization algorithm was used to find CLB textures that were reflective of the liver textures, accounting for CT system factors of spatial blurring and noise. With the modeled background texture as a guide, four cylindrical phantoms (Textures A-C and uniform, 165 mm in diameter, and 30 mm height) were designed, each containing 20 low-contrast spherical signals (6 mm diameter at nominal contrast levels of ∼3.2, 5.2, 7.2, 10, and 14 HU with four repeats per signal). The phantoms were voxelized and input into a commercial multimaterial 3D printer (Object Connex 350), with custom software for voxel-based printing (using principles of digital dithering). Images of the textured phantoms and a corresponding uniform phantom were acquired at six radiation dose levels (SOMATOM Flash, Siemens Healthcare) and observer model detection performance (detectability index of a multislice channelized Hotelling observer) was estimated for each condition (5 contrasts × 6 doses × 2 reconstructions × 4 backgrounds = 240 total conditions). A multivariate generalized regression analysis was performed (linear terms, no interactions, random error term, log link function) to assess whether dose, reconstruction algorithm, signal contrast, and background type have statistically significant effects on detectability. Also, fitted curves of detectability (averaged across contrast levels) as a function of dose were constructed for each reconstruction algorithm and background texture. FBP and SAFIRE were compared for each background type to determine the improvement in detectability at a given dose, and the reduced dose at which SAFIRE had equivalent performance compared to FBP at 100% dose. RESULTS Detectability increased with increasing radiation dose (P = 2.7 × 10-59) and contrast level (P = 2.2 × 10-86) and was higher in the uniform phantom compared to the textured phantoms (P = 6.9 × 10-51). Overall, SAFIRE had higher d' compared to FBP (P = 0.02). The estimated dose reduction potential of SAFIRE was found to be 8%, 10%, 27%, and 8% for Texture-A, Texture-B, Texture-C and uniform phantoms. CONCLUSIONS In all background types, detectability was higher with SAFIRE compared to FBP. However, the relative improvement observed from SAFIRE was highly dependent on the complexity of the background texture. Iterative algorithms such as SAFIRE should be assessed in the most realistic context possible.
Collapse
Affiliation(s)
- Justin Solomon
- Department of Radiology, Carl E. Ravin Advanced Imaging Laboratories, Clinical Imaging Physics Group, Duke University Medical Center, Durham, North Carolina 27705
| | - Alexandre Ba
- Institute of Radiation Physics, Lausanne University Hospital, Lausanne 1007, Switzerland
| | - François Bochud
- Institute of Radiation Physics, Lausanne University Hospital, Lausanne 1007, Switzerland
| | - Ehsan Samei
- Department of Radiology, Carl E. Ravin Advanced Imaging Laboratories, Clinical Imaging Physics Group, Duke University Medical Center, Durham, North Carolina 27705 and Departments of Biomedical Engineering, Physics, and Electrical and Computer Engineering, Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705
| |
Collapse
|
34
|
Bujila R, Fransson A, Poludniowski G. Practical approaches to approximating MTF and NPS in CT with an example application to task-based observer studies. Phys Med 2016; 33:16-25. [PMID: 28003136 DOI: 10.1016/j.ejmp.2016.10.016] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/31/2016] [Revised: 09/07/2016] [Accepted: 10/19/2016] [Indexed: 01/20/2023] Open
Abstract
PURPOSE To investigate two methods of approximating the Modulation Transfer Function (MTF) and Noise Power Spectrum (NPS) in computed tomography (CT) for a range of scan parameters, from limited image acquisitions. METHODS The two methods consist of 1) using a linear systems approach to approximate the NPS for different filtered backprojection (FBP) kernels with a filter function derived from the kernel ratio of determined MTFs and 2) using an empirical fitted model to approximate the MTF and NPS. In both cases a scaling function accounts for variations in mAs and kV. The two methods of approximating the MTF/NPS are further investigated by comparing image quality figure of merits (FOM) d' and AUC calculated using approximations of the MTF/NPS and MTF/NPS that have been determined for different mAs/kV levels and reconstruction kernels. RESULTS The greatest RMSE for NPS approximated for a range of mAs/kVp/convolution kernels using both methods and compared to determined NPS was 0.05 of the peak value. The RMSE for FOM with the kernel ratio method were at most 0.1 for d' and 0.01 for the AUC. Using the empirical model method, the RMSE for FOM were at most 0.02 for d' and 0.001 for the AUC. CONCLUSIONS The two methods proposed in this paper can provide a convenient way of approximating the MTF and NPS for use in, among other things, mathematical observer studies. Both methods require a relatively small number of direct determinations of NPS from scan acquisitions to model the NPS/MTF for arbitrary mAs and kV.
Collapse
Affiliation(s)
- Robert Bujila
- Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-17176 Stockholm, Sweden; Department of Physics, Royal Institute of Technology, SE-106 91 Stockholm, Sweden.
| | - Annette Fransson
- Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-17176 Stockholm, Sweden
| | - Gavin Poludniowski
- Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-17176 Stockholm, Sweden
| |
Collapse
|
35
|
Noferini L, Taddeucci A, Bartolini M, Bruschi A, Menchi I. CT image quality assessment by a Channelized Hotelling Observer (CHO): Application to protocol optimization. Phys Med 2016; 32:1717-1723. [DOI: 10.1016/j.ejmp.2016.11.002] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/16/2016] [Revised: 10/28/2016] [Accepted: 11/06/2016] [Indexed: 10/20/2022] Open
|
36
|
|