1
|
Terzidis E, Nordström F, Götstedt J, Bäck A. Impact of delivery variations on 3D dose distributions for volumetric modulated arc therapy plans of various complexity. Med Phys 2024; 51:8466-8481. [PMID: 39012800 DOI: 10.1002/mp.17310] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/10/2023] [Revised: 06/05/2024] [Accepted: 07/04/2024] [Indexed: 07/18/2024] Open
Abstract
BACKGROUND Delivery variations during radiotherapy can cause discrepancies between planned and delivered dose distribution. These variations could arise from random and systematic offsets in certain machine parameters or systematic offsets related to the calibration process of the treatment unit. PURPOSE The aim of this study was to present a novel simulation-based methodology to evaluate realistic delivery variations in three dimensions (3D). Additionally, we investigated the dosimetric impact of delivery variations for volumetric modulated arc therapy (VMAT) plans for different treatment sites and complexities. METHODS Twelve VMAT plans for different treatment sites (prostate-, head & neck-, lung-, and gynecological cancer) were selected. The clinical plan used for the treatment of each patient was reoptimized to create one plan with reduced complexity (i.e., simple plan) and one of higher complexity (i.e., complex plan). This resulted in a total of 36 plans. Delivery variations were simulated by randomly introducing offsets in multi-leaf collimator position, jaw position, gantry angle and collimator angle simultaneously. Twenty simulations were carried out for each of the 36 plans, yielding 720 simulated deliveries. To explore the impact of individual offsets, additional simulations were conducted for each type of offset separately. A 3D dose calculation was performed for each simulation using the same calculation engine as for the clinical plan. Two standard deviations (2SD) of dose were determined for every voxel for 3D-spatial evaluations. The dose variation in certain DVH metrics, that is, D2% and D98% for the clinical target volume and five different DVH metrics for selected organs at risk, was calculated for the twenty simulated deliveries of each plan. For comparison, the effect of delivery variations was assessed by conducting measurements with the Delta4 phantom. RESULTS The volume of voxels with 2SD above 1% of the prescribed dose was consistently larger for the complex plans in comparison to their corresponding simple and clinical plans. 2SDs larger than 1% were in many cases, found to accumulate outside the planning target volume. For complex plans, regions with 2SDs larger than 1% were detected also inside the high dose region, exhibiting, on average, a size six times larger volume, than those observed in simple plans. Similar results were found for all treatment sites. Variation in the selected DVH metrics for the simulated deliveries was generally largest for the complex plans with few exceptions. When comparing the 2SD distribution of the measurements with the 2SD distribution from the simulations, the spatial information showed deviations outside the PTV in both simulations and measurements. However, the measured values were, on average, 35% higher for the prostate plans and 10% higher for the head & neck plans compared to the simulated values. CONCLUSIONS The presented methodology effectively quantified and localized dose deviations due to delivery offsets. The 3D analysis provided information that was undetectable using the analysis based on DVH metrics. Dosimetric uncertainties due to delivery variations were prominent at the edge of the high-dose region irrespective of treatment site and plan complexity. Dosimetric uncertainties inside the high-dose region was more profound for plans of higher complexity.
Collapse
Affiliation(s)
- Emmanouil Terzidis
- Department of Medical Radiation Sciences, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
- Department of Therapeutic Radiation Physics, Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden
| | - Fredrik Nordström
- Department of Medical Radiation Sciences, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
- Department of Therapeutic Radiation Physics, Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden
| | - Julia Götstedt
- Department of Medical Radiation Sciences, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
- Department of Therapeutic Radiation Physics, Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden
| | - Anna Bäck
- Department of Medical Radiation Sciences, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
- Department of Therapeutic Radiation Physics, Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden
| |
Collapse
|
2
|
Hansen CR, Hussein M, Bernchou U, Zukauskaite R, Thwaites D. Plan quality in radiotherapy treatment planning - Review of the factors and challenges. J Med Imaging Radiat Oncol 2022; 66:267-278. [PMID: 35243775 DOI: 10.1111/1754-9485.13374] [Citation(s) in RCA: 9] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/02/2021] [Accepted: 12/14/2021] [Indexed: 12/25/2022]
Abstract
A high-quality treatment plan aims to best achieve the clinical prescription, balancing high target dose to maximise tumour control against sufficiently low organ-at-risk dose for acceptably low toxicity. Treatment planning (TP) includes multiple steps from simulation/imaging and segmentation to technical plan production and reporting. Consistent quality across this process requires close collaboration and communication between clinical and technical experts, to clearly understand clinical requirements and priorities and also practical uncertainties, limitations and compromises. TP quality depends on many aspects, starting from commissioning and quality management of the treatment planning system (TPS), including its measured input data and detailed understanding of TPS models and limitations. It requires rigorous quality assurance of the whole planning process and it links to plan deliverability, assessable by measurement-based verification. This review highlights some factors influencing plan quality, for consideration for optimal plan construction and hence optimal outcomes for each patient. It also indicates some challenges, sources of difference and current developments. The topics considered include: the evolution of TP techniques; dose prescription issues; tools and methods to evaluate plan quality; and some aspects of practical TP. The understanding of what constitutes a high-quality treatment plan continues to evolve with new techniques, delivery methods and related evidence-based science. This review summarises the current position, noting developments in the concept and the need for further robust tools to help achieve it.
Collapse
Affiliation(s)
- Christian Rønn Hansen
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark.,Department of Clinical Research, University of Southern Denmark, Odense, Denmark.,Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia.,Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark
| | - Mohammad Hussein
- Metrology for Medical Physics Centre, National Physical Laboratory, Teddington, UK
| | - Uffe Bernchou
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark.,Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Ruta Zukauskaite
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark.,Department of Oncology, Odense University Hospital, Odense, Denmark
| | - David Thwaites
- Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia
| |
Collapse
|
3
|
Hu J, Gu S, Wang N, Cui F, Zhang S, Yin C, Cai Y, Gou C, Zou L, Wu Z. Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy. Technol Cancer Res Treat 2022; 21:15330338221114499. [PMID: 36112945 PMCID: PMC9478705 DOI: 10.1177/15330338221114499] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022] Open
Abstract
Purpose: To compare the sensitivity of ArcCHECK (AC), portal
dosimetry (PD), and an in-house logfile-based system (LF) to multileaf
collimators (MLC) aperture errors and the ability to identify these errors.
Methods and Materials: For 12 retrospective original head and
neck volumetric modulated arc therapy (VMAT) plans, MLC aperture errors
of ± 0.4mm, ± 1.2mm, ± 2mm, and ± 3mm were introduced for each plan, resulting
in 96 plans with errors. AC, PD, and LF were used for the gamma evaluation at
3%/3mm, 3%/2mm, and 2%/2mm criteria. Gradient analysis was used to evaluate the
sensitivity to MLC aperture errors. The area under the curve (AUC) obtained from
the receiver operating characteristic (ROC) curve was used to evaluate the
ability to identify MLC aperture errors and dose errors, and the optimal cut-off
value to identify the error was obtained. Results: The gamma pass
rate (%GP) of LF had the smallest descent gradient as the MLC error increases in
any case. The descent gradient of PD was larger than AC, except for the case at
the 2%/2mm criteria. For the 3%/3mm criteria, the MLC aperture errors that can
be perfectly identified by AC, PD, and LF were ± 3mm, ± 2mm, and ± 1.2mm,
respectively, and the average percent dose error (%DEs) of dose metrics in
targets that can be perfectly identified were 4% to 5%, 3% to 4%, and 2% to 3%,
respectively. For the 3%/2mm criteria, the errors that AC, PD, and LF can
perfectly identify were the same as the 3%/3mm criteria. For the 2%/2mm
criteria, AC can perfectly identify the MLC error of ± 2mm and the %DE of 3% to
4%. PD and LF can identify the MLC error of ± 1.2mm and the %DE of 2% to 3%.
Conclusion: Different patient-specific quality assurance (PSQA)
systems have different sensitivity and recognition abilities to MLC aperture
errors. Institutions should formulate their own customized %GP limits based on
their PSQA process through ROC or other methods.
Collapse
Affiliation(s)
- Jinyou Hu
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China.,Cancer Center, 89669Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, Chengdu, China
| | - Shaoxian Gu
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Ningyu Wang
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Fengjie Cui
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Shengyuan Zhang
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Chuou Yin
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Yunzhu Cai
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Chengjun Gou
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| | - Lian Zou
- Cancer Center, 89669Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, Chengdu, China
| | - Zhangwen Wu
- Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, 12530Sichuan University, Chengdu, China
| |
Collapse
|
4
|
Gray A, Bawazeer O, Arumugam S, Vial P, Descallar J, Thwaites D, Holloway L. Evaluation of the ability of three commercially available dosimeters to detect systematic delivery errors in step-and-shoot IMRT plans. Rep Pract Oncol Radiother 2021; 26:793-803. [PMID: 34760314 DOI: 10.5603/rpor.a2021.0093] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/20/2021] [Accepted: 07/03/2021] [Indexed: 11/25/2022] Open
Abstract
Background There is limited data on error detectability for step-and-shoot intensity modulated radiotherapy (sIMRT) plans, despite significant work on dynamic methods. However, sIMRT treatments have an ongoing role in clinical practice. This study aimed to evaluate variations in the sensitivity of three patient-specific quality assurance (QA) devices to systematic delivery errors in sIMRT plans. Materials and methods Four clinical sIMRT plans (prostate and head and neck) were edited to introduce errors in: Multi-Leaf Collimator (MLC) position (increasing field size, leaf pairs offset (1-3 mm) in opposite directions; and field shift, all leaves offset (1-3 mm) in one direction); collimator rotation (1-3 degrees) and gantry rotation (0.5-2 degrees). The total dose for each plan was measured using an ArcCHECK diode array. Each field, excluding those with gantry offsets, was also measured using an Electronic Portal Imager and a MatriXX Evolution 2D ionisation chamber array. 132 plans (858 fields) were delivered, producing 572 measured dose distributions. Measured doses were compared to calculated doses for the no-error plan using Gamma analysis with 3%/3 mm, 3%/2 mm, and 2%/2 mm criteria (1716 analyses). Results Generally, pass rates decreased with increasing errors and/or stricter gamma criteria. Pass rate variations with detector and plan type were also observed. For a 3%/3 mm gamma criteria, none of the devices could reliably detect 1 mm MLC position errors or 1 degree collimator rotation errors. Conclusions This work has highlighted the need to adapt QA based on treatment plan type and the need for detector specific assessment criteria to detect clinically significant errors.
Collapse
Affiliation(s)
- Alison Gray
- Liverpool and Macarthur Cancer Therapy Centres, South Western Sydney Local Health District, Sydney, NSW, Australia.,Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.,South Western Sydney Clinical School, School of Medicine, University of New South Wales, Sydney, NSW, Australia
| | - Omemh Bawazeer
- Physics Department, Umm Al-Qura University, Mecca, Saudi Arabia
| | - Sankar Arumugam
- Liverpool and Macarthur Cancer Therapy Centres, South Western Sydney Local Health District, Sydney, NSW, Australia.,Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.,South Western Sydney Clinical School, School of Medicine, University of New South Wales, Sydney, NSW, Australia
| | - Philip Vial
- Liverpool and Macarthur Cancer Therapy Centres, South Western Sydney Local Health District, Sydney, NSW, Australia.,Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.,South Western Sydney Clinical School, School of Medicine, University of New South Wales, Sydney, NSW, Australia.,Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia
| | - Joseph Descallar
- Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.,South Western Sydney Clinical School, School of Medicine, University of New South Wales, Sydney, NSW, Australia
| | - David Thwaites
- Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia
| | - Lois Holloway
- Liverpool and Macarthur Cancer Therapy Centres, South Western Sydney Local Health District, Sydney, NSW, Australia.,Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.,South Western Sydney Clinical School, School of Medicine, University of New South Wales, Sydney, NSW, Australia.,Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia.,Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia
| |
Collapse
|
5
|
Alharthi T, George A, Arumugam S, Holloway L, Thwaites D, Vial P. An investigation of the IQM signal variation and error detection sensitivity for patient specific pre-treatment QA. Phys Med 2021; 86:6-18. [PMID: 34049118 DOI: 10.1016/j.ejmp.2021.05.005] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/05/2021] [Revised: 04/11/2021] [Accepted: 05/03/2021] [Indexed: 11/19/2022] Open
Abstract
PURPOSE To evaluate the Integral Quality Monitor (IQM) as a clinical dosimetry device for detecting photon beam delivery errors in clinically relevant conditions. MATERIALS AND METHODS The IQM's ability to detect delivery errors introduced into clinical VMAT plans for two different treatment sites was assessed. This included measuring 103 nasopharynx VMAT plans and 78 lung SBRT VMAT plans with introduced errors in gantry angle (1-5°) and in MLC-defined field size and field shift (1-5 mm). The IQM sensitivity was compared to ArcCheck detector performance. Signal dependence on field position for on-axis and asymmetrically offset square field sizes from 1 × 1 cm2 to 30 × 30 cm2 was also investigated. RESULTS The IQM detected almost all introduced clinically-significant MLC field size errors, but not some small gantry angle errors or most MLC field shift errors. The IQM sensitivity was comparable to the ArcCheck for lung SBRT, but worse for the nasopharynx plans. Differences between IQM calculated/predicted and measured signals were within ± 2% for all on-axis square fields, but up to 60% for the smallest asymmetrically offset fields at large offsets. CONCLUSION The IQM performance was consistent and reproducible. It showed highest sensitivity to the field size errors for these plans, but did not detect some clinically-significant introduced gantry angle errors or most MLC field shift errors. The IQM calculation model is still being developed, which should improve small offset-field performance. Care is required in IQM use for plan verification or online monitoring, especially for small fields that are off-axis in the detector gradient direction.
Collapse
Affiliation(s)
- Thahabah Alharthi
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia; School of Medicine, Taif University, Taif, Saudi Arabia; Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia.
| | - Armia George
- Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia.
| | - Sankar Arumugam
- Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; South Western Sydney Clinical School, University of New South Wales, Sydney, NSW, Australia.
| | - Lois Holloway
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia; Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; South Western Sydney Clinical School, University of New South Wales, Sydney, NSW, Australia; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
| | - David Thwaites
- Institute of Medical Physics, School of Physics, The University of Sydney, Sydney, New South Wales, Australia.
| | - Phil Vial
- Liverpool and Macarthur Cancer Therapy Centers, Liverpool, NSW, Australia; Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
| |
Collapse
|
6
|
Lechner W, Primeßnig A, Nenoff L, Wesolowska P, Izewska J, Georg D. The influence of errors in small field dosimetry on the dosimetric accuracy of treatment plans. Acta Oncol 2020; 59:511-517. [PMID: 31694438 DOI: 10.1080/0284186x.2019.1685127] [Citation(s) in RCA: 12] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/25/2022]
Abstract
Background: Dosimetric effects of inaccuracies of output factors (OFs) implemented in treatment planning systems (TPSs) were investigated.Materials and methods: Modified beam models (MBM) for which the OFs of small fields (down to 1 × 1 cm2) were increased by up to 12% compared to the original beam models (OBM) were created for two TPSs. These beam models were used to recalculate treatment plans of different complexity. Treatment plans using stereotactic 3D-conformal (s3D-CRT) for brain metastasis as well as VMAT plans for head and neck and prostate cancer patients were generated. Dose distributions calculated with the MBM and the OBM were compared to measured dose distributions acquired using film dosimetry and a 2D-detector-array. For the s3D-CRT plans the calculated and measured dose at the isocenter was evaluated. For VMAT, gamma pass rates (GPRs) were calculated using global gamma index with 3%/3 mm, 2%/3 mm, 1%/3 mm and 2%/2 mm with a 20% threshold. Contribution of small fields to the total fluence was expressed as the ratio (F) of fluence trough leaf openings smaller than 2 cm to the total fluence.Results: Using film dosimetry for the s3D-CRT plans, the average of the ratio of calculated dose to measured dose at the isocenter was 1.01 and 1.06 for the OBM and MBM model, respectively. A significantly lower GPR of the MBM compared to the OBM was only found for the localized prostate cases (F = 12.4%) measured with the 2D-detector-array and an acceptance criterion of 1%/3 mm.Conclusion: The effects of uncertainties in small field OFs implemented in TPSs are most pronounced for s3D-CRT cases and can be clearly identified using patient specific quality assurance. For VMAT these effects mainly remain undetected using standard patient specific quality assurance. Using tighter acceptance criteria combined with an analysis of the fluence generated by small fields can help identifying inaccuracies of OFs implemented in TPSs.
Collapse
Affiliation(s)
- Wolfgang Lechner
- Department of Radiation Oncology, Division Medical Physics, Medical University Vienna, Vienna, Austria
- Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Vienna, Austria
| | - Alexander Primeßnig
- Department of Radiation Oncology, Division Medical Physics, Medical University Vienna, Vienna, Austria
- Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Vienna, Austria
| | - Lena Nenoff
- Department of Radiation Oncology, Division Medical Physics, Medical University Vienna, Vienna, Austria
- Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Vienna, Austria
| | - Paulina Wesolowska
- International Atomic Energy Agency, Vienna International Centre, Vienna, Austria
| | - Joanna Izewska
- International Atomic Energy Agency, Vienna International Centre, Vienna, Austria
| | - Dietmar Georg
- Department of Radiation Oncology, Division Medical Physics, Medical University Vienna, Vienna, Austria
- Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Vienna, Austria
| |
Collapse
|
7
|
Bertelsen AS, Schytte T, Møller PK, Mahmood F, Riis HL, Gottlieb KL, Agergaard SN, Dysager L, Hansen O, Gornitzka J, Veldhuizen E, ODwyer DB, Christiansen RL, Nielsen M, Jensen HR, Brink C, Bernchou U. First clinical experiences with a high field 1.5 T MR linac. Acta Oncol 2019; 58:1352-1357. [PMID: 31241387 DOI: 10.1080/0284186x.2019.1627417] [Citation(s) in RCA: 74] [Impact Index Per Article: 12.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/26/2022]
Abstract
Purpose: A 1.5 T MR Linac (MRL) has recently become available. MRL treatment workflows (WF) include online plan adaptation based on daily MR images (MRI). This study reports initial clinical experiences after five months of use in terms of patient compliance, cases, WF timings, and dosimetric accuracy. Method and materials: Two different WF were used dependent on the clinical situation of the day; Adapt To Position WF (ATP) where the reference plan position is adjusted rigidly to match the position of the targets and the OARs, and Adapt To Shape WF (ATS), where a new plan is created to match the anatomy of the day, using deformable image registration. Both WFs included three 3D MRI scans for plan adaptation, verification before beam on, and validation during IMRT delivery. Patient compliance and WF timings were recorded. Accuracy in dose delivery was assessed using a cylindrical diode phantom. Results: Nineteen patients have completed their treatment receiving a total of 176 fractions. Cases vary from prostate treatments (60Gy/20F) to SBRT treatments of lymph nodes (45 Gy/3F) and castration by ovarian irradiation (15 Gy/3F). The median session time (patient in to patient out) for 127 ATPs was 26 (21-78) min, four fractions lasted more than 45 min due to additional plan adaptation. For the 49 ATSs a median time of 12 (1-24) min was used for contouring resulting in a total median session time of 42 (29-91) min. Three SBRT fractions lasted more than an hour. The time on the MRL couch was well tolerated by the patients. The median gamma pass rate (2 mm,2% global max) for the adapted plans was 99.2 (93.4-100)%, showing good agreement between planned and delivered dose. Conclusion: MRL treatments, including daily MRIs, plan adaptation, and accurate dose delivery, are possible within a clinically acceptable timeframe and well tolerated by the patients.
Collapse
Affiliation(s)
- Anders S. Bertelsen
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Tine Schytte
- Department of Oncology, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Pia K. Møller
- Department of Oncology, Odense University Hospital, Odense, Denmark
- Research Unit of Oncology, Odense University Hospital, Odense, Denmark
| | - Faisal Mahmood
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Hans L. Riis
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
- Department of Oncology, Odense University Hospital, Odense, Denmark
| | - Karina L. Gottlieb
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Søren N. Agergaard
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Lars Dysager
- Department of Oncology, Odense University Hospital, Odense, Denmark
| | - Olfred Hansen
- Department of Oncology, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Janne Gornitzka
- Department of Oncology, Odense University Hospital, Odense, Denmark
| | | | - Dean B. ODwyer
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Rasmus L. Christiansen
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Morten Nielsen
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Henrik R. Jensen
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
| | - Carsten Brink
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| | - Uffe Bernchou
- Laboratory of Radiation Physics, Odense University Hospital, Odense, Denmark
- Department of Clinical Research, University of Southern Denmark, Odense, Denmark
| |
Collapse
|