Lozares S, Font JA, Gandía A, Campos A, Flamarique S, Ibáñez R, Villa D, Alba V, Jiménez S, Hernández M, Casamayor C, Vicente I, Hernando E, Rubio P. In vivo dosimetry in low-voltage IORT breast treatments with XR-RV3 radiochromic film.
Phys Med 2021;
81:173-181. [PMID:
33465753 DOI:
10.1016/j.ejmp.2020.12.011]
[Citation(s) in RCA: 6] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/21/2020] [Revised: 12/12/2020] [Accepted: 12/16/2020] [Indexed: 12/30/2022] Open
Abstract
PURPOSE
The objectives of the study were to establish a procedure for in vivo film-based dosimetry for intraoperative radiotherapy (IORT), evaluate the typical doses delivered to organs at risk, and verify the dose prescription.
MATERIALS AND METHODS
In vivo dose measurements were studied using XR-RV3 radiochromic films in 30 patients with breast cancer undergoing IORT using the Axxent® device (Xoft Inc.). The stability of the radiochromic films in the energy ranges used was verified by taking measurements at different depths. The stability of the scanner response was tested, and 5 different calibration curves were constructed for different beam qualities. Six pieces of film were placed in each of the 30 patients. All the pieces were correctly sterilized and checked to ensure that the process did not affect the outcome. All calibration and dose measurements were analyzed using the Radiochromic.com software application.
RESULTS
The doses were measured for 30 patients. The doses in contact with the applicator (prescription zone) were 19.8 ± 0.9 Gy. In the skin areas, the doses were as follows: 1-2 cm from the applicator, 1.86 ± 0.77 Gy; 2-5 cm, 0.73 ± 0.14 Gy; and greater than 5 cm, 0.28 ± 0.17 Gy. The dose delivered to the pectoral muscle (tungsten shielding disc) was 0.51 ± 0.27 Gy.
CONCLUSIONS
The study demonstrated the viability of XR-RV3 films for in vivo dose measurement in the dose and energy ranges applied in a complex procedure, such as breast IORT. The doses in organs at risk were far below the tolerances for cases such as those studied.
Collapse