Gustavo Cuña E, Schulz H, Tuzzi E, Biagi L, Bosco P, García-Fontes M, Mattos J, Tosetti M, Engelmann J, Scheffler K, Hagberg GE. Simulated and experimental phantom data for multi-center quality assurance of quantitative susceptibility maps at 3 T, 7 T and 9.4 T.
Phys Med 2023;
110:102590. [PMID:
37116389 DOI:
10.1016/j.ejmp.2023.102590]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/28/2022] [Revised: 04/07/2023] [Accepted: 04/12/2023] [Indexed: 04/30/2023] Open
Abstract
PURPOSE
To develop methods for quality assurance of quantitative susceptibility mapping (QSM) using MRI at different magnetic field strengths, and scanners, using different MR-sequence protocols, and post-processing pipelines.
METHODS
We built a custom phantom based on iron in two forms: homogeneous susceptibility ('free iron') and with fine-scaled variations in susceptibility ('clustered iron') at different iron concentrations. The phantom was measured at 3.0 T (two scanners), 7.0 T and 9.4 T using multi-echo, gradient echo acquisition sequences. A digital phantom analogue to the iron-phantom, tailored to obtain similar results as in experimentation was developed, with similar geometry and susceptibility values. Morphology enabled dipole inversion was applied to the phase images to obtain QSM for experimental and simulated data using the MEDI + 0 approach for background regularization.
RESULTS
Across all scanners, QSM-values showed a linear increase with iron concentrations. The QSM-relaxivity was 0.231 ± 0.047 ppm/mM for free and 0.054 ± 0.013 ppm/mM for clustered iron, with adjusted determination coefficients (DoC) ≥ 0.87. Similarly, the simulations yielded linear increases (DoC ≥ 0.99). In both the experimental and digital phantoms, the estimated molar susceptibility was lower with clustered iron, because clustering led to highly localized field effects.
CONCLUSION
Our iron phantom can be used to evaluate the capability of QSM to detect local variations in susceptibility across different field strengths, when using different MR-sequence protocols. The devised simulation method captures the effect of iron clustering in QSM as seen experimentally and could be used in the future to optimize QSM processing pipelines and achieve higher accuracy for local field effects, as also seen in Alzheimer's beta-amyloid plaques.