Chen N, Miao GX, Peng LQ, Li YH, Gu J, He Y, Chen T, Fu XY, Xing ZX. Bow-and-arrow sign on point-of-care ultrasound for diagnosis of pacemaker lead-induced heart perforation: A case report and literature review.
World J Clin Cases 2023;
11:1615-1626. [PMID:
36926400 PMCID:
PMC10011982 DOI:
10.12998/wjcc.v11.i7.1615]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/24/2022] [Revised: 12/17/2022] [Accepted: 02/10/2023] [Indexed: 03/02/2023] Open
Abstract
BACKGROUND
Pacemaker lead-induced heart perforation is a rare but life-threatening complication of pacemaker implantation, and timely diagnosis remains a challenge for clinicians. Here, we report a case of pacemaker lead-induced cardiac perforation rapidly diagnosed by a “bow-and-arrow” sign on point-of-care ultrasound (POCUS).
CASE SUMMARY
A 74-year-old Chinese woman who had undergone permanent pacemaker implantation 26 d before suddenly developed severe dyspnea, chest pain, and hypotension. The patient had received emergency laparotomy for an incarcerated groin hernia and was transferred to the intensive care unit 6 d before. Computed tomography was not available due to unstable hemodynamic status, so POCUS was performed at the bedside and revealed severe pericardial effusion and cardiac tamponade. Subsequent pericardiocentesis yielded a large volume of bloody pericardial fluid. Further POCUS by an ultrasonographist revealed a unique “bow-and-arrow” sign indicating right ventricular (RV) apex perforation by the pacemaker lead, which facilitated the rapid diagnosis of lead perforation. Given the persistent drainage of pericardial bleeding, urgent off-pump open chest surgery was performed to repair the perforation. However, the patient died of shock and multiple organ dysfunction syndrome within 24 h post-surgery. In addition, we also performed a literature review on the sonographic features of RV apex perforation by lead.
CONCLUSION
POCUS enables the early diagnosis of pacemaker lead perforation at the bedside. A step-wise ultrasonographic approach and the “bow-and-arrow” sign on POCUS are helpful for rapid diagnosis of lead perforation.