Silveira BMF, Pereira MCB, Cardoso DR, Ribeiro-Samora GA, Martins HR, Parreira VF. New method for evaluating maximal respiratory pressures: Concurrent validity, test-retest, and inter-rater reliability.
Braz J Phys Ther 2021;
25:741-748. [PMID:
34119441 DOI:
10.1016/j.bjpt.2021.04.012]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/30/2020] [Revised: 02/12/2021] [Accepted: 04/21/2021] [Indexed: 11/17/2022] Open
Abstract
BACKGROUND
Maximal respiratory pressures (MRP) obtained at functional residual capacity (FRC) may reflect the real respiratory muscle pressure.
OBJECTIVES
To evaluate concurrent validity, test-retest, and inter-rater reliability of MRP performed with a new instrument in healthy individuals, and to compare values obtained at different volumes in healthy individuals and individuals with COPD.
METHODS
MRP of 100 healthy individuals were obtained using the TrueForce and the MicroRPM® at residual volume (RV) and total lung capacity (TLC) to evaluate concurrent validity. MRP were obtained at FRC using the TrueForce to evaluate reliability. Comparisons of inspiratory pressure values (FRC compared to RV) and expiratory pressure values (FRC compared to TLC) were performed with 100 healthy individuals and 15 individuals with COPD.
RESULTS
The intraclass correlation coefficient (ICC) was 0.77 and 0.86 for concurrent validity for inspiratory and expiratory pressures, respectively. Test-retest reliability showed an ICC of 0.87 for inspiratory pressure, and 0.78 for expiratory pressure; inter-rater reliability showed an ICC of 0.91 for inspiratory pressure, and 0.84 for expiratory pressure. Measurements performed at RV and TLC were higher when compared to FRC [mean difference (95%CI)= -8.30 (-11.82, -4.78) cmH2O; -37.29 (-42.63, -31.96) cmH2O] in healthy individuals, and -11.09 (-15.83, -6.35) cmH2O; -57.14 (-71.05, -43.05) cmH2O in COPD, for inspiratory and expiratory pressures, respectively.
CONCLUSION
MRP performed with the TrueForce presented good concurrent validity, good test-retest reliability, excellent inter-rater reliability for inspiratory pressure and good inter-rater reliability for expiratory pressure. MRP were lower when obtained at FRC for healthy individuals and with COPD.