Rocha KN, da Silva JAF, de Jesus DP. Capillary electrophoresis with capacitively coupled contactless conductivity detection (C
4 D) for rapid and simple determination of lactate in sweat.
Electrophoresis 2024;
45:392-399. [PMID:
38072648 DOI:
10.1002/elps.202300179]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/15/2023] [Revised: 11/28/2023] [Accepted: 11/30/2023] [Indexed: 03/20/2024]
Abstract
An analytical method based on capillary electrophoresis (CE) using capacitively coupled contactless conductivity detection (C4 D) was developed and validated for fast, straightforward, and reliable determination of lactate in artificial and human sweat samples. The background electrolyte was composed of equimolar concentrations (10 mmol/L) of 2-(N-morpholino)ethanesulfonic acid and histidine, with 0.2 mmol/L of cetyltrimethylammonium bromide as electroosmotic flow inverter. The limit of detection and quantification were 3.1 and 10.3 µmol/L, respectively. Recoveries in the 97 to 118% range were obtained using sweat samples spiked with lactate at three concentration levels, indicating an acceptable accuracy. The intraday and interday precisions were 1.49 and 7.08%, respectively. The proposed CE-C4 D method can be a starting point for monitoring lactate concentrations in sweat samples for diagnostics, physiological studies, and sports performance assessment applications.
Collapse