Wu LL, Wu YS, Chen WY, Zhou W, Tang L, Li B, Liu B. Determination of artemisitene in rat plasma by ultra-performance liquid chromatography/tandem mass spectrometry and its application in pharmacokinetics.
RAPID COMMUNICATIONS IN MASS SPECTROMETRY : RCM 2017;
31:1121-1128. [PMID:
28403574 DOI:
10.1002/rcm.7881]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/03/2017] [Revised: 04/06/2017] [Accepted: 04/09/2017] [Indexed: 06/07/2023]
Abstract
RATIONALE
Artemisitene shows a wide variety of pharmacological activities, such as antioxidant protection in vitro and in vivo. It has been identified as a novel Nrf2 inducer. However, there is no report on an ultra-performance liquid chromatography/tandem mass spectrometry (UPLC/MS/MS) method to quantitate artemisitene in rat plasma and its application to a pharmacokinetic profile study.
METHODS
An ACQUITY UPLC™ BEH Symmetry Shield RP18 column (1.7 μm, 2.1 mm × 100 mm) was used at a flow rate of 0.3 mL·min-1 . Mass detection was performed by electrospray ionization tandem mass spectrometry via multiple reaction monitoring (MRM) in positive mode. Plasma samples were pre-treated by a single-step extraction with 0.1% formic acid aqueous solutions-acetonitrile, and tolbutamide was used as internal standard.
RESULTS
The calibration curve was from 0.98 to 1000 ng∙mL-1 (r2 = 0.995). The extraction recoveries were 61.5-79.4% and 81.7-94.6% for artemisitene and tolbutamide, respectively. The lower limit of quantification (LLOQ) was 0.98 ng∙mL-1 . The absolute bioavailability of artemisitene was 3.7% after intravenous and oral administration in rats.
CONCLUSIONS
The UPLC/MS/MS assay was validated for linearity, accuracy, stability, extraction recovery, matrix effects, and intra-day and inter-day precision. The method, for the first time, achieved some pharmacokinetic parameters and was successfully applied to a pharmacokinetic study Copyright © 2017 John Wiley & Sons, Ltd.