Gelves-Meza J, Lang RM, Valderrama-Achury MD, Zamorano JL, Vargas-Acevedo C, Medina HM, Salazar G. Tricuspid Regurgitation Related to Cardiac Implantable Electronic Devices: An Integrative Review.
J Am Soc Echocardiogr 2022;
35:1107-1122. [PMID:
35964911 DOI:
10.1016/j.echo.2022.08.004]
[Citation(s) in RCA: 15] [Impact Index Per Article: 7.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/22/2019] [Revised: 08/01/2022] [Accepted: 08/02/2022] [Indexed: 10/15/2022]
Abstract
The use of cardiac implantable electronic devices, including permanent pacemakers, implantable cardiac defibrillators and cardiac resynchronization therapy, has dramatically increased in recent years. The interaction between the device lead and tricuspid valve leaflets is a potential cause of tricuspid regurgitation which in turn has an impact on morbidity and mortality. Echocardiography is necessary for grading of tricuspid regurgitation severity. The use of three-dimensional imaging helps determine whether the device lead is interfering with normal leaflet coaptation. Early identification of lead-related tricuspid regurgitation is critical to select the optimal treatment, which may include lead extraction or even tricuspid valve repair/replacement in severe cases. This review aims to provide a thorough assessment of the evidence about lead-associated tricuspid regurgitation, the benefits of using 3D echocardiography with some technical considerations, and finally, propose a treatment algorithm.
Collapse