Sedaghat MR, Ostadi-Moghadam H, Jabbarvand M, Askarizadeh F, Momeni-Moghaddam H, Narooie-Noori F. Corneal hysteresis and corneal resistance factor in pellucid marginal degeneration.
J Curr Ophthalmol 2017;
30:42-47. [PMID:
29564407 PMCID:
PMC5859559 DOI:
10.1016/j.joco.2017.08.002]
[Citation(s) in RCA: 16] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/13/2017] [Revised: 07/02/2017] [Accepted: 08/01/2017] [Indexed: 01/09/2023] Open
Abstract
Purpose
To evaluate and compare corneal hysteresis (CH) and corneal resistance factor (CRF) in pellucid marginal degeneration (PMD), keratoconus (KCN), and normal eyes using the Ocular Response Analyzer (ORA).
Methods
In this retrospective study, corneal biomechanical parameters were measured in patients with PMD (n = 102) and KCN (n = 202) and normal subjects (n = 208) using the ORA. Data, including full patient history as well as the results of refraction, slit-lamp biomicroscopy, Pentacam HR (Oculus), and ORA (Reichert; Buffalo, New York, USA), were collected from medical records. Also, the data of only one eye per individual were selected for the analysis. The inclusion criteria for PMD and KCN groups were a reliable diagnosis of these ectatic disorders based on the clinical and corneal tomographic findings. CH, CRF, CH–CRF, intraocular pressure (IOP) measurements were assessed for each subject. Data were analyzed with SPSS and MedCalc using the ANOVA, Pearson Correlation, and receiver operating characteristic (ROC) curve analysis.
Results
The mean CH was 8.91 mmHg ± 1.05 [standard deviation (SD)], 8.43 ± 0.78, and 10.89 ± 1.08 in the PMD, KCN, and normal group, respectively. Also, the mean CRF was 8.21 ± 1.35, 7.19 ± 1.11, and 10.69 ± 1.41 in the PMD, KCN, and normal group, respectively. ANOVA showed differences in the mean CH, CRF, and CH–CRF between three groups (P < 0.001). Also, ROC curve analysis showed the cut-off points ≤9.5, ≤9.5, and >1.3 mmHg for CH, CRF, and CH–CRF in the PMD group, respectively. For biomechanical parameters in PMD eyes, CRF had the highest sensitivity (75.49%) while the greatest area under the ROC curve (AUC) was seen for CH (0.903). Moreover, central corneal thickness (CCT) showed no correlation with CH (P = 0.30, r = −0.104) or CRF (P = 0.75, r = 0.033) in the PMD group.
Conclusions
This study presented the values of corneal biomechanics for PMD using the ORA. The results of the ORA were markedly different between PMD, KCN, and normal eyes.