Rowan NJ, Kremer T, McDonnell G. A review of Spaulding's classification system for effective cleaning, disinfection and sterilization of reusable medical devices: Viewed through a modern-day lens that will inform and enable future sustainability.
THE SCIENCE OF THE TOTAL ENVIRONMENT 2023;
878:162976. [PMID:
36963674 DOI:
10.1016/j.scitotenv.2023.162976]
[Citation(s) in RCA: 8] [Impact Index Per Article: 8.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 12/10/2022] [Revised: 02/23/2023] [Accepted: 03/17/2023] [Indexed: 05/13/2023]
Abstract
Despite advances in medicine and innovations in many underpinning fields including disease prevention and control, the Spaulding classification system, originally proposed in 1957, remains widely used for defining the disinfection and sterilization of contaminated re-usable medical devices and surgical instruments. Screening PubMed and Scopus databases using a PRISMA guiding framework generated 272 relevant publications that were used in this review. Findings revealed that there is a need to evolve how medical devices are designed, and processed by cleaning, disinfection (and/or sterilization) to mitigate patient risks, including acquiring an infection. This Spaulding Classification remains in use as it is logical, easily applied and understood by users (microbiologists, epidemiologists, manufacturers, industry) and by regulators. However, substantial changes have occurred over the past 65 years that challenge interpretation and application of this system that includes inter alia emergence of new pathogens (viruses, mycobacteria, protozoa, fungi), a greater understanding of innate and adaptive microbial tolerance to disinfection, toxicity risks, increased number of vulnerable patients and associated patient procedures, and greater complexity in design and use of medical devices. Common cited examples include endoscopes that enable non- or minimal invasive procedures but are highly sophisticated with various types of materials (polymers, electronic components etc), long narrow channels, right angle and heat-sensitive components and various accessories (e.g., values) that can be contaminated with high levels of microbial bioburden and patient tissues after use. Contaminated flexible duodenoscopes have been a source of several significant infection outbreaks, where at least 9 reported cases were caused by multidrug resistant organisms [MDROs] with no obvious breach in processing detected. Despite this, there is evidence of the lack of attention to cleaning and maintenance of these devices and associated equipment. Over the last few decades there is increasing genomic evidence of innate and adaptive resistance to chemical disinfectant methods along with adaptive tolerance to environmental stresses. To reduce these risks, it has been proposed to elevate classification of higher-risk flexible endoscopes (such as duodenoscopes) from semi-critical [contact with mucous membrane and intact skin] to critical use [contact with sterile tissue and blood] that entails a transition to using low-temperature sterilization modalities instead of routinely using high-level disinfection; thus, increasing the margin of safety for endoscope processing. This timely review addresses important issues surrounding use of the Spaulding classification system to meet modern-day needs. It specifically addresses the need for automated, robust cleaning and drying methods combined with using real-time monitoring of device processing. There is a need to understand entire end-to-end processing of devices instead of adopting silo approaches that in the future will be informed by artificial intelligence and deep-learning/machine learning. For example, combinational solutions that address the formation of complex biofilms that harbour pathogenic and opportunistic microorganisms on the surfaces of processed devices. Emerging trends are addressed including future sustainability for the medical devices sector that can be enabled via a new Quintuple Helix Hub approach that combines academia, industry, healthcare, regulators, and society to unlock real world solutions.
Collapse