Maquer G, Favre P. [Enriching in vivo clinical trials with in silico models for orthopedic implants]. Med Sci (Paris) 2022; 38:38-44. [PMID: 35060885 DOI: 10.1051/medsci/2021243] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/14/2022] Open
Abstract
Clinical trials are used by the medical device industry to confirm products safety, performance, and clinical benefits. Traditional clinical studies typically follow a limited number of volunteers, which prevents capturing the full breath of patient demographics and implant use. New tools are required to overcome the limitations of traditional trials while fulfilling increasingly demanding regulatory requirements. Computer simulations have the potential to enrich traditional clinical trials with so called in silico clinical trials (ISCT) by providing data on a much broader spectrum of patients, clinical conditions and implant configurations. The historical use of simulation in the orthopedic device industry is described here to explain how it is now technically possible to model virtual populations. We also discuss the multiple benefits of such a translational research approach for the patients, healthcare systems, and manufacturers, but also the challenges to overcome. A more detailed version is available in English [1].
Collapse
|
2
|
A Perspective on Biotribology in Arthroplasty: From In Vitro toward the Accurate In Silico Wear Prediction. APPLIED SCIENCES-BASEL 2020. [DOI: 10.3390/app10186312] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/13/2022]
Abstract
Nowadays hip arthroplasty is recognized as one of the most successful orthopedic surgical procedures, even if it involves challenges to overcome, such that lately, younger and more active patients are in need of total arthroplasty. Wear is still one of the main issues affecting joint prostheses endurance, and often causes loosening accompanied by implant failures. Actual in vitro wear tests executed by mechanical simulators have a long duration, are very expensive, and do not take into account all the possible daily activities of the patients; thus, the challenge to obtain a complete in silico tribological and dynamical model of (bio) tribo-systems could give the possibility to overcome the actual testing procedures and could contribute as a tool for a more accurate tribological design of human prostheses. This prospective paper is intended to underline actual research trends toward the challenge of having accurate numerical algorithms to be used both in preclinical testing and in the optimizations of the prostheses design. With this aim we depicted the possible in silico approach in artificial joints’ wear assessment over time, accounting for contact mechanics, numerical stress–strain analysis, musculoskeletal multibody, and synovial lubrication modelling (boundary/mixed, hydrodynamic, and elastohydrodynamic).
Collapse
|
3
|
Tatani I, Megas P, Panagopoulos A, Diamantakos I, Nanopoulos P, Pantelakis S. Comparative analysis of the biomechanical behavior of two different design metaphyseal-fitting short stems using digital image correlation. Biomed Eng Online 2020; 19:65. [PMID: 32814586 PMCID: PMC7437017 DOI: 10.1186/s12938-020-00806-y] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/21/2020] [Accepted: 08/04/2020] [Indexed: 01/26/2023] Open
Abstract
BACKGROUND The progressive evolution in hip replacement research is directed to follow the principles of bone and soft tissue sparing surgery. Regarding hip implants, a renewed interest has been raised towards short uncemented femoral implants. A heterogeneous group of short stems have been designed with the aim to approximate initial, post-implantation bone strain to the preoperative levels in order to minimize the effects of stress shielding. This study aims to investigate the biomechanical properties of two distinctly designed femoral implants, the TRI-LOCK Bone Preservation Stem, a shortened conventional stem and the Minima S Femoral Stem, an even shorter and anatomically shaped stem, based on experiments and numerical simulations. Furthermore, finite element models of implant-bone constructs should be evaluated for their validity against mechanical tests wherever it is possible. In this work, the validation was performed via a direct comparison of the FE calculated strain fields with their experimental equivalents obtained using the digital image correlation technique. RESULTS Design differences between Trilock BPS and Minima S femoral stems conditioned different strain pattern distributions. A distally shifting load distribution pattern as a result of implant insertion and also an obvious decrease of strain in the medial proximal aspect of the femur was noted for both stems. Strain changes induced after the implantation of the Trilock BPS stem at the lateral surface were greater compared to the non-implanted femur response, as opposed to those exhibited by the Minima S stem. Linear correlation analyses revealed a reasonable agreement between the numerical and experimental data in the majority of cases. CONCLUSION The study findings support the use of DIC technique as a preclinical evaluation tool of the biomechanical behavior induced by different implants and also identify its potential for experimental FE model validation. Furthermore, a proximal stress-shielding effect was noted after the implantation of both short-stem designs. Design-specific variations in short stems were sufficient to produce dissimilar biomechanical behaviors, although their clinical implication must be investigated through comparative clinical studies.
Collapse
Affiliation(s)
- I Tatani
- Orthopaedic Department, University Hospital of Patras, Papanikolaou 1, Rio-Patra, 26504, Patras, Greece.
| | - P Megas
- Orthopaedic Department, University Hospital of Patras, Papanikolaou 1, Rio-Patra, 26504, Patras, Greece
| | - A Panagopoulos
- Orthopaedic Department, University Hospital of Patras, Papanikolaou 1, Rio-Patra, 26504, Patras, Greece
| | - I Diamantakos
- Laboratory of Technology and Strength of Materials, Department of Mechanical Engineering and Aeronautics, University of Patras, Patras, Greece
| | - Ph Nanopoulos
- Department of Computer Engineering & Informatics, University of Patras, Patras, Greece
| | - Sp Pantelakis
- Laboratory of Technology and Strength of Materials, Department of Mechanical Engineering and Aeronautics, University of Patras, Patras, Greece
| |
Collapse
|
4
|
Pappalardo F, Russo G, Tshinanu FM, Viceconti M. In silico clinical trials: concepts and early adoptions. Brief Bioinform 2020; 20:1699-1708. [PMID: 29868882 DOI: 10.1093/bib/bby043] [Citation(s) in RCA: 119] [Impact Index Per Article: 29.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/01/2018] [Revised: 04/18/2018] [Indexed: 02/07/2023] Open
Abstract
Innovations in information and communication technology infuse all branches of science, including life sciences. Nevertheless, healthcare is historically slow in adopting technological innovation, compared with other industrial sectors. In recent years, new approaches in modelling and simulation have started to provide important insights in biomedicine, opening the way for their potential use in the reduction, refinement and partial substitution of both animal and human experimentation. In light of this evidence, the European Parliament and the United States Congress made similar recommendations to their respective regulators to allow wider use of modelling and simulation within the regulatory process. In the context of in silico medicine, the term 'in silico clinical trials' refers to the development of patient-specific models to form virtual cohorts for testing the safety and/or efficacy of new drugs and of new medical devices. Moreover, it could be envisaged that a virtual set of patients could complement a clinical trial (reducing the number of enrolled patients and improving statistical significance), and/or advise clinical decisions. This article will review the current state of in silico clinical trials and outline directions for a full-scale adoption of patient-specific modelling and simulation in the regulatory evaluation of biomedical products. In particular, we will focus on the development of vaccine therapies, which represents, in our opinion, an ideal target for this innovative approach.
Collapse
Affiliation(s)
| | - Giulia Russo
- Department of Biomedical and Biotechnological Sciences, University of Catania, Catania 95123, Italy
| | - Flora Musuamba Tshinanu
- Federal Agency for Medicines and Health Products, Brussels, Belgium and INSERM U1248, Université de Limoges, Limoges, France
| | - Marco Viceconti
- Department of Mechanical Engineering, University of Sheffield, Sheffield, UK and INSIGNEO Institute for In Silico Medicine, University of Sheffield, Sheffield, UK
| |
Collapse
|
5
|
Cachão JH, Soares dos Santos MP, Bernardo R, Ramos A, Bader R, Ferreira JAF, Torres Marques A, Simões JAO. Altering the Course of Technologies to Monitor Loosening States of Endoprosthetic Implants. SENSORS 2019; 20:s20010104. [PMID: 31878028 PMCID: PMC6982938 DOI: 10.3390/s20010104] [Citation(s) in RCA: 12] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 10/02/2019] [Revised: 11/07/2019] [Accepted: 11/10/2019] [Indexed: 02/02/2023]
Abstract
Musculoskeletal disorders are becoming an ever-growing societal burden and, as a result, millions of bone replacements surgeries are performed per year worldwide. Despite total joint replacements being recognized among the most successful surgeries of the last century, implant failure rates exceeding 10% are still reported. These numbers highlight the necessity of technologies to provide an accurate monitoring of the bone–implant interface state. This study provides a detailed review of the most relevant methodologies and technologies already proposed to monitor the loosening states of endoprosthetic implants, as well as their performance and experimental validation. A total of forty-two papers describing both intracorporeal and extracorporeal technologies for cemented or cementless fixation were thoroughly analyzed. Thirty-eight technologies were identified, which are categorized into five methodologies: vibrometric, acoustic, bioelectric impedance, magnetic induction, and strain. Research efforts were mainly focused on vibrometric and acoustic technologies. Differently, approaches based on bioelectric impedance, magnetic induction and strain have been less explored. Although most technologies are noninvasive and are able to monitor different loosening stages of endoprosthetic implants, they are not able to provide effective monitoring during daily living of patients.
Collapse
Affiliation(s)
- João Henrique Cachão
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
| | - Marco P. Soares dos Santos
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
- Center for Mechanical Technology & Automation (TEMA), University of Aveiro, 3810-193 Aveiro, Portugal
- Associated Laboratory for Energy, Transports and Aeronautics (LAETA), 4150-179 Porto, Portugal
- Correspondence:
| | - Rodrigo Bernardo
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
| | - António Ramos
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
- Center for Mechanical Technology & Automation (TEMA), University of Aveiro, 3810-193 Aveiro, Portugal
| | - Rainer Bader
- Department of Orthopedics, University Medicine Rostock, 18057 Rostock, Germany
| | - Jorge A. F. Ferreira
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
- Center for Mechanical Technology & Automation (TEMA), University of Aveiro, 3810-193 Aveiro, Portugal
| | - António Torres Marques
- Associated Laboratory for Energy, Transports and Aeronautics (LAETA), 4150-179 Porto, Portugal
- Mechanical Engineering Department, University of Porto, 4200-465 Porto, Portugal
| | - José A. O. Simões
- Department of Mechanical Engineering, University of Aveiro, 3810-193 Aveiro, Portugal
- Center for Mechanical Technology & Automation (TEMA), University of Aveiro, 3810-193 Aveiro, Portugal
| |
Collapse
|
6
|
Tatani I, Panagopoulos A, Diamantakos I, Sakellaropoulos G, Pantelakis S, Megas P. Comparison of two metaphyseal-fitting (short) femoral stems in primary total hip arthroplasty: study protocol for a prospective randomized clinical trial with additional biomechanical testing and finite element analysis. Trials 2019; 20:359. [PMID: 31208433 PMCID: PMC6580512 DOI: 10.1186/s13063-019-3445-x] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/18/2018] [Accepted: 05/13/2019] [Indexed: 02/07/2023] Open
Abstract
BACKGROUND Total hip replacement has recently followed a progressive evolution towards principles of bone- and soft-tissue-sparing surgery. Regarding femoral implants, different stem designs have been developed as an alternative to conventional stems, and there is a renewed interest towards short versions of uncemented femoral implants. Based on both experimental testing and finite element modeling, the proposed study has been designed to compare the biomechanical properties and clinical performance of the newly introduced short-stem Minima S, for which clinical data are lacking with an older generation stem, the Trilock Bone Preservation Stem with an established performance record in short to midterm follow-up. METHODS/DESIGN In the experimental study, the transmission of forces as measured by cortical surface-strain distribution in the proximal femur will be evaluated using digital image correlation (DIC), first on the non-implanted femur and then on the implanted stems. Finite element parametric models of the bone, the stem and their interface will be also developed. Finite element predictions of surface strains in implanted composite femurs, after being validated against biomechanical testing measurements, will be used to assist the comparison of the stems by deriving important data on the developed stress and strain fields, which cannot be measured through biomechanical testing. Finally, a prospective randomized comparative clinical study between these two stems will be also conducted to determine (1) their clinical performance up to 2 years' follow-up using clinical scores and gait analysis (2) stem fixation and remodeling using a detailed radiographic analysis and (3) incidence and types of complications. DISCUSSION Our study would be the first that compares not only the clinical and radiological outcome but also the biomechanical properties of two differently designed femoral implants that are theoretically classified in the same main category of cervico-metaphyseal-diaphyseal short stems. We can hypothesize that even these subtle variations in geometric design between these two stems may create different loading characteristics and thus dissimilar biomechanical behaviors, which in turn could have an influence to their clinical performance. TRIAL REGISTRATION International Standard Randomized Controlled Trial Number, ID: ISRCTN10096716 . Retrospectively registered on May 8 2018.
Collapse
Affiliation(s)
- I Tatani
- Orthopaedic Department, University Hospital of Patras, Patras, Greece
| | - A Panagopoulos
- Orthopaedic Department, University Hospital of Patras, Patras, Greece.
| | - I Diamantakos
- Laboratory of Technology and Strength of Materials, Department of Mechanical Engineering and Aeronautics, University of Patras, Patras, Greece
| | - G Sakellaropoulos
- Department of Medical Physics, School of Medicine, University of Patras, Patras, Greece
| | - Sp Pantelakis
- Laboratory of Technology and Strength of Materials, Department of Mechanical Engineering and Aeronautics, University of Patras, Patras, Greece
| | - P Megas
- Orthopaedic Department, University Hospital of Patras, Patras, Greece
| |
Collapse
|
7
|
O'Rourke D, Bottema M, Taylor M. Sampling strategies for approximating patient variability in population-based finite element studies of total hip replacement. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING 2019; 35:e3168. [PMID: 30394696 DOI: 10.1002/cnm.3168] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/06/2018] [Revised: 10/26/2018] [Accepted: 10/27/2018] [Indexed: 06/08/2023]
Abstract
Total hip replacements must be robust to patient variability for long-term success in the population. The challenge during the design process is evaluating an implant in a diverse population but the computational cost of simulating a population of subject-specific finite element (FE) models is not practical. We examined five strategies to generate representative subsets of subjects from a cohort of 103 implanted hip joint FE models to approximate the variability in output metrics. Comparing with the median and distribution of the 95th percentile composite peak micromotion (CPM) and polar gap in the full cohort (CPM median: 136 μm, interquartile range [IQR]: 74-230 μm) (Polar Gap median: 467 μm, IQR: 434-548 μm), the Anatomic Sampling strategy (12 subjects) achieved the best balance of computational cost and approximation of the output metrics (CPM median: 169 μm, IQR: 78-236 μm) (Polar Gap median: 469 μm, IQR: 448-537 μm). Convex hull sampling (41 subjects) more closely captured the output metrics (CPM median: 99 μm, IQR: 70-191 μm) (Polar Gap median: 456 μm, IQR: 418-533 μm) but required over three times the number of subjects. Volume reduction of the convex hull captured the extremes of variability with subsets of 5 to 20 subjects, while the largest minimum-distance strategy captured the variability toward the middle of the cohort. These strategies can estimate the level of variability in FE model output metrics with a low computational cost when evaluating implants during the design process.
Collapse
Affiliation(s)
- Dermot O'Rourke
- Medical Device Research Institute, Flinders University, 1284 South Road, SA, 5042, Australia
| | - Murk Bottema
- Medical Device Research Institute, Flinders University, 1284 South Road, SA, 5042, Australia
| | - Mark Taylor
- Medical Device Research Institute, Flinders University, 1284 South Road, SA, 5042, Australia
| |
Collapse
|
8
|
Affatato S, Merola M. Does the Hirsch Index Improve Research Quality in the Field of Biomaterials? A New Perspective in the Biomedical Research Field. MATERIALS 2018; 11:ma11101967. [PMID: 30322125 PMCID: PMC6213325 DOI: 10.3390/ma11101967] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 09/21/2018] [Revised: 10/08/2018] [Accepted: 10/11/2018] [Indexed: 11/17/2022]
Abstract
Orthopaedic implants offer valuable solutions to many pathologies of bones and joints. The research in this field is driven by the aim of realizing durable and biocompatible devices; therefore, great effort is spent on material analysis and characterization. As a demonstration of the importance assumed by tribology in material devices, wear and friction are two of the main topics of investigation for joint prostheses. Research is led and supported by public institutions, whether universities or research centers, based on the laboratories’ outputs. Performance criteria assessing an author’s impact on research contribute somewhat to author inflation per publication. The need to measure the research activity of an institution is an essential goal and this leads to the development of indicators capable of giving a rating to the publication that disseminates them. The main purpose of this work was to observe the variation of the Hirsch Index (h-index) when the position of the authors is considered. To this end, we conducted an analysis evaluating the h-index by excluding the intermediate positions. We found that the higher the h value, the larger the divergence between this value and the corrected one. The correction relies on excluding publications for which the author does not have a relevant position. We propose considering the authorship order in a publication in order to obtain more information on the impact that authors have on their research field. We suggest giving the users of researcher registers (e.g., Scopus, Google Scholar) the possibility to exclude from the h-index evaluation the objects of research where the scientist has a marginal position.
Collapse
Affiliation(s)
- Saverio Affatato
- Laboratorio di Tecnologia Medica, IRCCS-Istituto Ortopedico Rizzoli, Via di Barbiano, 1/10 40136 Bologna, Italy.
| | - Massimiliano Merola
- Laboratorio di Tecnologia Medica, IRCCS-Istituto Ortopedico Rizzoli, Via di Barbiano, 1/10 40136 Bologna, Italy.
| |
Collapse
|
9
|
Affatato S, Ruggiero A, Jaber SA, Merola M, Bracco P. Wear Behaviours and Oxidation Effects on Different UHMWPE Acetabular Cups Using a Hip Joint Simulator. MATERIALS 2018; 11:ma11030433. [PMID: 29547536 PMCID: PMC5873012 DOI: 10.3390/ma11030433] [Citation(s) in RCA: 16] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 02/22/2018] [Revised: 03/07/2018] [Accepted: 03/14/2018] [Indexed: 11/16/2022]
Abstract
Given the long-term problem of polyethylene wear, medical interest in the new improved cross-linked polyethylene (XLPE), with or without the adding of vitamin E, has risen. The main aim of this study is to gain further insights into the mutual effects of radiation cross-linking and addition of vitamin E on the wear performance of ultra-high-molecular-weight polyethylene (UHMWPE). We tested four different batches of polyethylene (namely, a standard one, a vitamin E-stabilized, and two cross-linked) in a hip joint simulator for five million cycles where bovine calf serum was used as lubricant. The acetabular cups were then analyzed using a confocal profilometer to characterize the surface topography. Moreover; the cups were analyzed by using Fourier Transformed Infrared Spectroscopy and Differential Scanning Calorimetry in order to assess the chemical characteristics of the pristine materials. Comparing the different cups’ configuration, mass loss was found to be higher for standard polyethylene than for the other combinations. Mass loss negatively correlated to the cross-link density of the polyethylenes. None of the tested formulations showed evidence of oxidative degradation. We found no correlation between roughness parameters and wear. Furthermore, we found significantly differences in the wear behavior of all the acetabular cups. XLPEs exhibited lower weight loss, which has potential for reduced wear and decreased osteolysis. However, surface topography revealed smoother surfaces of the standard and vitamin E stabilized polyethylene than on the cross-linked samples. This observation suggests incipient crack generations on the rough and scratched surfaces of the cross-linked polyethylene liners.
Collapse
Affiliation(s)
- Saverio Affatato
- Medical Technology Laboratory, IRCCS-Rizzoli Orthopaedic Institute, Via di Barbiano, 1/10, 40136 Bologna, Italy.
| | - Alessandro Ruggiero
- Department of Industrial Engineering, University of Salerno, 84084 Fisciano, Italy.
| | - Sami Abdel Jaber
- Medical Technology Laboratory, IRCCS-Rizzoli Orthopaedic Institute, Via di Barbiano, 1/10, 40136 Bologna, Italy.
| | - Massimiliano Merola
- Department of Industrial Engineering, University of Salerno, 84084 Fisciano, Italy.
| | - Pierangiola Bracco
- Chemistry Department and Nanostructured Interfaces and Surfaces (NIS) Centre, University of Turin, Via Giuria 7, 10125 Turin, Italy.
| |
Collapse
|
10
|
Trommer RM, Maru MM. Importance of preclinical evaluation of wear in hip implant designs using simulator machines. Rev Bras Ortop 2017; 52:251-259. [PMID: 28702381 PMCID: PMC5496994 DOI: 10.1016/j.rboe.2016.07.004] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/20/2016] [Accepted: 07/05/2016] [Indexed: 11/24/2022] Open
Abstract
Total hip arthroplasty (THA) is a surgical procedure that involves the replacement of the damaged joint of the hip by an artificial device. Despite the recognized clinical success of hip implants, wear of the articulating surfaces remains as one of the critical issues influencing performance. Common material combinations used in hip designs comprise metal-on-polymer (MoP), ceramic-on-polymer (CoP), metal-on-metal (MoM), and ceramic-on-ceramic (CoC). However, when the design of the hip implant is concerned besides the materials used, several parameters can influence its wear performance. In this scenario, where the safety and efficacy for the patient are the main issues, it is fundamental to evaluate and predict the wear rate of the hip implant design before its use in THA. This is one of the issues that should be taken into account in the preclinical evaluation step of the product, in which simulated laboratory tests are necessary. However, it is fundamental that the applied motions and loads can reproduce the wear mechanisms physiologically observed in the patient. To replicate the in vivo angular displacements and loadings, special machines known as joint simulators are employed. This article focuses on the main characteristics related to the wear simulation of hip implants using mechanical simulators, giving information to surgeons, researchers, regulatory bodies, etc., about the importance of preclinical wear evaluation. A critical analysis is performed on the differences in the principles of operation of simulators and their effects on the final results, and about future trends in wear simulation.
Collapse
Affiliation(s)
- Rafael Mello Trommer
- Instituto Nacional de Metrologia, Qualidade e Tecnologia (Inmetro), Divisão de Metrologia de Materiais, Duque de Caxias, RJ, Brazil
| | - Márcia Marie Maru
- Instituto Nacional de Metrologia, Qualidade e Tecnologia (Inmetro), Divisão de Metrologia de Materiais, Duque de Caxias, RJ, Brazil
| |
Collapse
|
11
|
Importância da avaliação pré‐clínica do desgaste em projetos de implantes de quadril usando máquinas simuladoras. Rev Bras Ortop 2017. [DOI: 10.1016/j.rbo.2016.07.009] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/21/2022] Open
|
12
|
Mukherjee K, Gupta S. The effects of musculoskeletal loading regimes on numerical evaluations of acetabular component. Proc Inst Mech Eng H 2016; 230:918-29. [DOI: 10.1177/0954411916661368] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/09/2015] [Accepted: 07/05/2016] [Indexed: 11/17/2022]
Abstract
The importance of clinical studies notwithstanding, the failure assessment of implant–bone structure has alternatively been carried out using finite element analysis. However, the accuracy of the finite element predicted results is dependent on the applied loading and boundary conditions. Nevertheless, most finite element–based evaluations on acetabular component used a few selective load cases instead of the eight load cases representing the entire gait cycle. These in silico evaluations often suffer from limitations regarding the use of simplified musculoskeletal loading regimes. This study attempts to analyse the influence of three different loading regimes representing a gait cycle, on numerical evaluations of acetabular component. Patient-specific computer tomography scan-based models of intact and resurfaced pelvises were used. One such loading regime consisted of the second load case that corresponded to peak hip joint reaction force. Whereas the other loading regime consisted of the second and fifth load cases, which corresponded to peak hip joint reaction force and peak muscle forces, respectively. The third loading regime included all the eight load cases. Considerable deviations in peri-acetabular strains, standard error ranging between 115 and 400 µε, were observed for different loading regimes. The predicted bone strains were lower when selective loading regimes were used. Despite minor quantitative variations in bone density changes (less than 0.15 g cm−3), the final bone density pattern after bone remodelling was found to be similar for all the loading regimes. Underestimations in implant–bone micromotions (40–50 µm) were observed for selective loading regimes after bone remodelling. However, at immediate post-operative condition, such underestimations were found to be less (less than 5 µm). The predicted results highlight the importance of inclusion of eight load cases representing the gait cycle for in silico evaluations of resurfaced pelvis.
Collapse
Affiliation(s)
- Kaushik Mukherjee
- Department of Mechanical Engineering, Indian Institute of Technology Kharagpur, Kharagpur, India
| | - Sanjay Gupta
- Department of Mechanical Engineering, Indian Institute of Technology Kharagpur, Kharagpur, India
| |
Collapse
|
13
|
Four decades of finite element analysis of orthopaedic devices: where are we now and what are the opportunities? J Biomech 2014; 48:767-78. [PMID: 25560273 DOI: 10.1016/j.jbiomech.2014.12.019] [Citation(s) in RCA: 95] [Impact Index Per Article: 9.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 11/26/2014] [Indexed: 11/23/2022]
Abstract
Finite element has been used for more than four decades to study and evaluate the mechanical behaviour total joint replacements. In Huiskes seminal paper "Failed innovation in total hip replacement: diagnosis and proposals for a cure", finite element modelling was one of the potential cures to avoid poorly performing designs reaching the market place. The size and sophistication of models has increased significantly since that paper and a range of techniques are available from predicting the initial mechanical environment through to advanced adaptive simulations including bone adaptation, tissue differentiation, damage accumulation and wear. However, are we any closer to FE becoming an effective screening tool for new devices? This review contains a critical analysis of currently available finite element modelling techniques including (i) development of the basic model, the application of appropriate material properties, loading and boundary conditions, (ii) describing the initial mechanical environment of the bone-implant system, (iii) capturing the time dependent behaviour in adaptive simulations, (iv) the design and implementation of computer based experiments and (v) determining suitable performance metrics. The development of the underlying tools and techniques appears to have plateaued and further advances appear to be limited either by a lack of data to populate the models or the need to better understand the fundamentals of the mechanical and biological processes. There has been progress in the design of computer based experiments. Historically, FE has been used in a similar way to in vitro tests, by running only a limited set of analyses, typically of a single bone segment or joint under idealised conditions. The power of finite element is the ability to run multiple simulations and explore the performance of a device under a variety of conditions. There has been increasing usage of design of experiments, probabilistic techniques and more recently population based modelling to account for patient and surgical variability. In order to have effective screening methods, we need to continue to develop these approaches to examine the behaviour and performance of total joint replacements and benchmark them for devices with known clinical performance. Finite element will increasingly be used in the design, development and pre-clinical testing of total joint replacements. However, simulations must include holistic, closely corroborated, multi-domain analyses which account for real world variability.
Collapse
|
14
|
Hua X, Wang L, Al-Hajjar M, Jin Z, Wilcox RK, Fisher J. Experimental validation of finite element modelling of a modular metal-on-polyethylene total hip replacement. Proc Inst Mech Eng H 2014; 228:682-92. [PMID: 24963036 DOI: 10.1177/0954411914541830] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
Finite element models are becoming increasingly useful tools to conduct parametric analysis, design optimisation and pre-clinical testing for hip joint replacements. However, the verification of the finite element model is critically important. The purposes of this study were to develop a three-dimensional anatomic finite element model for a modular metal-on-polyethylene total hip replacement for predicting its contact mechanics and to conduct experimental validation for a simple finite element model which was simplified from the anatomic finite element model. An anatomic modular metal-on-polyethylene total hip replacement model (anatomic model) was first developed and then simplified with reasonable accuracy to a simple modular total hip replacement model (simplified model) for validation. The contact areas on the articulating surface of three polyethylene liners of modular metal-on-polyethylene total hip replacement bearings with different clearances were measured experimentally in the Leeds ProSim hip joint simulator under a series of loading conditions and different cup inclination angles. The contact areas predicted from the simplified model were then compared with that measured experimentally under the same conditions. The results showed that the simplification made for the anatomic model did not change the predictions of contact mechanics of the modular metal-on-polyethylene total hip replacement substantially (less than 12% for contact stresses and contact areas). Good agreements of contact areas between the finite element predictions from the simplified model and experimental measurements were obtained, with maximum difference of 14% across all conditions considered. This indicated that the simplification and assumptions made in the anatomic model were reasonable and the finite element predictions from the simplified model were valid.
Collapse
Affiliation(s)
- Xijin Hua
- Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK
| | - Ling Wang
- State Key Laboratory for Manufacturing Systems Engineering, Xi'an Jiaotong University, Xi'an, Shanxi, China
| | - Mazen Al-Hajjar
- Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK
| | - Zhongmin Jin
- Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK State Key Laboratory for Manufacturing Systems Engineering, Xi'an Jiaotong University, Xi'an, Shanxi, China
| | - Ruth K Wilcox
- Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK
| | - John Fisher
- Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK
| |
Collapse
|
15
|
Clarke S, Phillips A, Bull A. Evaluating a suitable level of model complexity for finite element analysis of the intact acetabulum. Comput Methods Biomech Biomed Engin 2013; 16:717-24. [DOI: 10.1080/10255842.2011.633906] [Citation(s) in RCA: 22] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/14/2022]
|
16
|
Wik TS. Experimental evaluation of new concepts in hip arthroplasty. Acta Orthop 2012; 83:1-26. [PMID: 22489909 DOI: 10.3109/17453674.2012.678804] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
In this thesis we evaluated two different hip arthroplasty concepts trough in vitro studies and numerical analyses. The cortical strains in the femoral neck area were increased by 10 to 15 % after insertion of a resurfacing femoral component compared to values of the intact femur, shown in an in vitro study on human cadaver femurs. There is an increased risk of femoral neck fracture after hip resurfacing arthroplasty. An increase of 10 to 15 % in femoral neck strains is limited, and cannot alone explain these fractures. Together with patient specific and surgical factors, however, increased strain can contribute to increased risk of fracture. An in vitro study showed that increasing the neck length in combination with retroversion or reduced neck shaft angle on a standard cementless femoral stem does not compromise the stability of the stem. The strain pattern in the proximal femur increased significantly at several measuring sites when the version and length of neck were altered. However, the changes were probably too small to have clinical relevance. In a validation study we have shown that a subject specific finite element analysis is able to perform reasonable predictions of strains and stress shielding after insertion of a femoral stem in human cadaver femurs. The usage of finite element models can be a valuable supplement to in vitro tests of femoral strain pattern around hip arthroplasty. Finally, a patient case shows that bone resorption around an implant caused by stress shielding can in extreme cases lead to periprosthetic fracture.
Collapse
Affiliation(s)
- Tina Strømdal Wik
- Orthopaedic Research Centre, St Olav Hospital, Trondheim University Hospital, Postbox 3250 Sluppen, NO-7006 Trondheim, Norway.
| |
Collapse
|
17
|
Martelli S, Taddei F, Schileo E, Cristofolini L, Rushton N, Viceconti M. Biomechanical robustness of a new proximal epiphyseal hip replacement to patient variability and surgical uncertainties: A FE study. Med Eng Phys 2012; 34:161-71. [DOI: 10.1016/j.medengphy.2011.07.006] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/29/2010] [Revised: 06/08/2011] [Accepted: 07/11/2011] [Indexed: 11/27/2022]
|
18
|
Martelli S, Taddei F, Cristofolini L, Schileo E, Rushton N, Viceconti M. A new hip epiphyseal prosthesis: Design revision driven by a validated numerical procedure. Med Eng Phys 2011; 33:1203-11. [DOI: 10.1016/j.medengphy.2011.05.010] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/29/2010] [Revised: 05/13/2011] [Accepted: 05/16/2011] [Indexed: 11/30/2022]
|
19
|
Relationships between viscoelastic properties of lumbar intervertebral disc and degeneration grade assessed by MRI. J Mech Behav Biomed Mater 2011; 4:593-9. [DOI: 10.1016/j.jmbbm.2011.01.007] [Citation(s) in RCA: 28] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/05/2010] [Revised: 01/14/2011] [Accepted: 01/20/2011] [Indexed: 12/19/2022]
|
20
|
Martelli S, Taddei F, Cristofolini L, Gill HS, Viceconti M. Extensive Risk Analysis of Mechanical Failure for an Epiphyseal Hip Prothesis: A Combined Numerical—Experimental Approach. Proc Inst Mech Eng H 2010; 225:126-40. [DOI: 10.1243/09544119jeim728] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Abstract
There has been recent renewed interest in proximal femur epiphyseal replacement as an alternative to conventional total hip replacement. In many branches of engineering, risk analysis has proved to be an efficient tool for avoiding premature failures of innovative devices. An extensive risk analysis procedure has been developed for epiphyseal hip prostheses and the predictions of this method have been compared to the known clinical outcomes of a well-established contemporary design, namely hip resurfacing devices. Clinical scenarios leading to revision (i.e. loosening, neck fracture and failure of the prosthetic component) were associated with potential failure modes (i.e. overload, fatigue, wear, fibrotic tissue differentiation and bone remodelling). Driving parameters of the corresponding failure mode were identified together with their safe thresholds. For each failure mode, a failure criterion was identified and studied under the most relevant physiological loading conditions. All failure modes were investigated with the most suitable investigation tool, either numerical or experimental. Results showed a low risk for each failure scenario either in the immediate postoperative period or in the long term. These findings are in agreement with those reported by the majority of clinical studies for correctly implanted devices. Although further work is needed to confirm the predictions of this method, it was concluded that the proposed risk analysis procedure has the potential to increase the efficacy of preclinical validation protocols for new epiphyseal replacement devices.
Collapse
Affiliation(s)
- S Martelli
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Bologna, Italy
| | - F Taddei
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Bologna, Italy
| | - L Cristofolini
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Bologna, Italy
- Engineering Faculty, University of Bologna, Italy
| | - H S Gill
- Nuffield Department Of Orthopaedics, Rheumatology and Musculoskeletal Sciences, University of Oxford, UK
| | - M Viceconti
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Bologna, Italy
| |
Collapse
|
21
|
Taddei F, Martelli S, Gill HS, Cristofolini L, Viceconti M. Finite Element Modeling of Resurfacing Hip Prosthesis: Estimation of Accuracy Through Experimental Validation. J Biomech Eng 2010; 132:021002. [DOI: 10.1115/1.4000065] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
Metal-on-metal hip resurfacing is becoming increasingly popular, and a number of new devices have been recently introduced that, in the short term, appear to have satisfactory outcome but many questions are still open on the biomechanics of the resurfaced femur. This could be investigated by means of finite element analysis, but, in order to be effective in discerning potential critical conditions, the accuracy of the models’ predictions should be assessed. The major goal of this study was to validate, through a combined experimental-numerical study, a finite element modeling procedure for the simulation of resurfaced femurs. In addition, a preliminary biomechanical analysis of the changes induced in the femoral neck biomechanics by the presence of the device was performed, under a physiologic range of hip joint reaction directions. For this purpose, in vitro tests and a finite element model based on the same specimen were developed using a cadaver femur. The study focused on the Conserve Plus, one of the most common contemporary resurfacing designs. Five loading configurations were identified to correspond to the extremes of physiological directions for the hip joint. The agreement between experimental measurements and numerical predictions was good both in the prediction of the femoral strains (R2>0.9), and in the prosthesis micromotions (error<20 μm), giving confidence in the model predictions. The preliminary biomechanical analysis indicated that the strains in the femoral neck are moderately affected by the presence of the prosthesis, apart from localized strain increments that can be considerable, always predicted near the stem. Low micromotions and contact pressure were predicted, suggesting a good stability of the prosthesis. The model accuracy was good in the prediction of the femoral strains and moderately good in the prediction of the bone-prosthesis micromovements. Although the investigated loading conditions were not completely physiological, the preliminary biomechanical analysis showed relatively small changes for the proximal femur after implantation. This validated model can support realistic simulations to examine physiological load configurations and the effects of variations in prosthesis design and implantation technique.
Collapse
Affiliation(s)
- Fulvia Taddei
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136 Bologna, Italy
| | - Saulo Martelli
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136 Bologna, Italy; Facoltà di Ingegneria, Università degli Studi di Bologna, Viale Risorgimento 2, 40136 Bologna, Italy
| | - Harinderjit Singh Gill
- Nuffield Department of Orthopaedic Surgery, University of Oxford, Nuffield Orthopaedic Centre, Oxford, UK
| | - Luca Cristofolini
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136 Bologna, Italy; Facoltà di Ingegneria, Università degli Studi di Bologna, Viale Risorgimento 2, 40136 Bologna, Italy
| | - Marco Viceconti
- Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136 Bologna, Italy
| |
Collapse
|