Farkash G, Liberman G, Martinho RP, Frydman L. Improving MRI's slice selectivity in the presence of strong, metal-derived inhomogeneities.
Magn Reson Imaging 2020;
69:71-80. [PMID:
32179096 DOI:
10.1016/j.mri.2020.03.003]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/13/2019] [Revised: 03/08/2020] [Accepted: 03/10/2020] [Indexed: 11/16/2022]
Abstract
PURPOSE
To develop schemes that deliver faithful 2D slices near field heterogeneities of the kind arising from non-ferromagnetic metal implants, with reduced artifacts and shorter scan times.
METHODS
An excitation scheme relying on cross-term spatio-temporal encoding (xSPEN) was used as basis for developing the new inhomogeneity-insensitive, slice-selective pulse scheme. The resulting Fully refOCUSED cross-term SPatiotemporal ENcoding (FOCUSED-xSPEN) approach involved four adiabatic sweeps. The method was evaluated in silico, in vitro and in vivo using mice models, and compared against a number of existing and of novel alternatives based on both conventional and swept RF pulses, including an analogous method based on LASER's selectivity spatial selectivity.
RESULTS
Calculations and experiments confirmed that multi-sweep derivatives of xSPEN and LASER can deliver localized excitation profiles, centered at the intended positions and endowed with enhanced immunity to B0 and B1 distortions. This, however, is achieved at the expense of higher SAR than non-swept counterparts. Furthermore, single-shot FOCUSED-xSPEN and LASER profiles covered limited off-resonance ranges. This could be extended to bands covering arbitrary off-resonance values with uniform slice widths, by looping the experiments over a number of scans possessing suitable transmission and reception offsets.
CONCLUSIONS
A series of novel approaches were introduced to select slices near metals, delivering robustness against Bo and B1+ field inhomogeneities.