1
|
Fesmire CC, Williamson RH, Petrella RA, Kaufman JD, Topasna N, Sano MB. Integrated Time Nanosecond Pulse Irreversible Electroporation (INSPIRE): Assessment of Dose, Temperature, and Voltage on Experimental and Clinical Treatment Outcomes. IEEE Trans Biomed Eng 2024; 71:1511-1520. [PMID: 38145519 PMCID: PMC11035095 DOI: 10.1109/tbme.2023.3340718] [Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/27/2023]
Abstract
OBJECTIVE This study sought to investigate a novel strategy using temperature-controlled delivery of nanosecond pulsed electric fields as an alternative to the 50-100 microsecond pulses used for irreversible electroporation. METHODS INSPIRE treatments were carried out at two temperatures in 3D tumor models using doses between 0.001 s and 0.1 s. The resulting treatment zones were quantified using viability staining and lethal electric field intensities were determined numerically. Computational modeling was then used to determine parameters necessary for INSPIRE treatments to achieve equivalent treatment zones to clinical electroporation treatments and evaluate the potential for these treatments to induce deleterious thermal damage. RESULTS Lethal thresholds between 1109 and 709 V/cm were found for nominal 0.01 s treatments with pulses between 350 ns and 2000 ns at physiological temperatures. Further increases in dose resulted in significant decreases in lethal thresholds. Given these experimental results, treatment zones comparable to clinical electroporation are possible by increasing the dose and voltage used with nanosecond duration pulses. Temperature-controlled simulations indicate minimal thermal cell death while achieving equivalent treatment volumes to clinical electroporation. CONCLUSION Nanosecond electrical pulses can achieve comparable outcomes to traditional electroporation provided sufficient electrical doses or voltages are applied. The use of temperature-controlled delivery may minimize thermal damage during treatment. SIGNIFICANCE Intense muscle stimulation and the need for cardiac gating have limited irreversible electroporation. Nanosecond pulses can alleviate these challenges, but traditionally have produced significantly smaller treatment zones. This study suggests that larger ablation volumes may be possible with the INSPIRE approach and that future in vivo studies are warranted.
Collapse
|
2
|
Fesmire CC, Peal B, Ruff J, Moyer E, McParland TJ, Derks K, O’Neil E, Emke C, Johnson B, Ghosh S, Petrella RA, DeWitt MR, Prange T, Fogle C, Sano MB. Investigation of integrated time nanosecond pulse irreversible electroporation against spontaneous equine melanoma. Front Vet Sci 2024; 11:1232650. [PMID: 38352036 PMCID: PMC10861690 DOI: 10.3389/fvets.2024.1232650] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/31/2023] [Accepted: 01/10/2024] [Indexed: 02/16/2024] Open
Abstract
Introduction Integrated time nanosecond pulse irreversible electroporation (INSPIRE) is a novel tumor ablation modality that employs high voltage, alternating polarity waveforms to induce cell death in a well-defined volume while sparing the underlying tissue. This study aimed to demonstrate the in vivo efficacy of INSPIRE against spontaneous melanoma in standing, awake horses. Methods A custom applicator and a pulse generation system were utilized in a pilot study to treat horses presenting with spontaneous melanoma. INSPIRE treatments were administered to 32 tumors across 6 horses and an additional 13 tumors were followed to act as untreated controls. Tumors were tracked over a 43-85 day period following a single INSPIRE treatment. Pulse widths of 500ns and 2000ns with voltages between 1000 V and 2000 V were investigated to determine the effect of these variables on treatment outcomes. Results Treatments administered at the lowest voltage (1000 V) reduced tumor volumes by 11 to 15%. Higher voltage (2000 V) treatments reduced tumor volumes by 84 to 88% and eliminated 33% and 80% of tumors when 500 ns and 2000 ns pulses were administered, respectively. Discussion Promising results were achieved without the use of chemotherapeutics, the use of general anesthesia, or the need for surgical resection in regions which are challenging to keep sterile. This novel therapeutic approach has the potential to expand the role of pulsed electric fields in veterinary patients, especially when general anesthesia is contraindicated, and warrants future studies to demonstrate the efficacy of INSPIRE as a solid tumor treatment.
Collapse
Affiliation(s)
- Chris C. Fesmire
- Bioelectricity Lab, UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, NC, United States
| | - Bridgette Peal
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Jennifer Ruff
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Elizabeth Moyer
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Thomas J. McParland
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Kobi Derks
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Erin O’Neil
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Carrie Emke
- Clinical Studies Core, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Brianna Johnson
- Clinical Studies Core, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Shatorupa Ghosh
- Bioelectricity Lab, UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, NC, United States
| | - Ross A. Petrella
- Bioelectricity Lab, UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, NC, United States
| | - Matthew R. DeWitt
- Bioelectricity Lab, UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, NC, United States
| | - Timo Prange
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Callie Fogle
- Department of Clinical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| | - Michael B. Sano
- Bioelectricity Lab, UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, NC, United States
- Department of Molecular Biomedical Sciences, NC State College of Veterinary Medicine, Raleigh, NC, United States
| |
Collapse
|
3
|
Real-Time Temperature Rise Estimation during Irreversible Electroporation Treatment through State-Space Modeling. Bioengineering (Basel) 2022; 9:bioengineering9100499. [PMID: 36290467 PMCID: PMC9598795 DOI: 10.3390/bioengineering9100499] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/01/2022] [Revised: 09/14/2022] [Accepted: 09/18/2022] [Indexed: 11/16/2022] Open
Abstract
To evaluate the feasibility of real-time temperature monitoring during an electroporation-based therapy procedure, a data-driven state-space model was developed. Agar phantoms mimicking low conductivity (LC) and high conductivity (HC) tissues were tested under the influences of high (HV) and low (LV) applied voltages. Real-time changes in impedance, measured by Fourier Analysis SpecTroscopy (FAST) along with the known tissue conductivity and applied voltages, were used to train the model. A theoretical finite element model was used for external validation of the model, producing model fits of 95.8, 88.4, 90.7, and 93.7% at 4 mm and 93.2, 58.9, 90.0, and 90.1% at 10 mm for the HV-HC, LV-LC, HV-LC, and LV-HC groups, respectively. The proposed model suggests that real-time temperature monitoring may be achieved with good accuracy through the use of real-time impedance monitoring.
Collapse
|
4
|
Aycock KN, Campelo SN, Davalos RV. A Comparative Modeling Study of Thermal Mitigation Strategies in Irreversible Electroporation Treatments. JOURNAL OF HEAT TRANSFER 2022; 144:031206. [PMID: 35833151 PMCID: PMC8823459 DOI: 10.1115/1.4053199] [Citation(s) in RCA: 5] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 10/05/2021] [Revised: 12/03/2021] [Indexed: 05/09/2023]
Abstract
Irreversible electroporation (IRE), also referred to as nonthermal pulsed field ablation (PFA), is an attractive focal ablation modality for solid tumors and cardiac tissue due to its ability to destroy aberrant cells with limited disruption of the underlying tissue architecture. Despite its nonthermal cell death mechanism, application of electrical energy results in Joule heating that, if ignored, can cause undesired thermal injury. Engineered thermal mitigation (TM) technologies including phase change materials (PCMs) and active cooling (AC) have been reported and tested as a potential means to limit thermal damage. However, several variables affect TM performance including the pulsing paradigm, electrode geometry, PCM composition, and chosen active cooling parameters, meaning direct comparisons between approaches are lacking. In this study, we developed a computational model of conventional bipolar and monopolar probes with solid, PCM-filled, or actively cooled cores to simulate clinical IRE treatments in pancreatic tissue. This approach reveals that probes with integrated PCM cores can be tuned to drastically limit thermal damage compared to existing solid probes. Furthermore, actively cooled probes provide additional control over thermal effects within the probe vicinity and can altogether abrogate thermal damage. In practice, such differences in performance must be weighed against the increased time, expense, and effort required for modified probes compared to existing solid probes.
Collapse
Affiliation(s)
- Kenneth N. Aycock
- Bioelectromechanical Systems Lab, Virginia Tech—Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech Department of Biomedical Engineering and Mechanics, 320 Kelly Hall, 325 Stanger Street, Blacksburg, VA 24061
- Corresponding author. e-mail:
| | - Sabrina N. Campelo
- Bioelectromechanical Systems Lab, Virginia Tech—Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech Department of Biomedical Engineering and Mechanics, 320 Kelly Hall, 325 Stanger Street, Blacksburg, VA 24061
- e-mail:
| | - Rafael V. Davalos
- Bioelectromechanical Systems Lab, Virginia Tech—Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech Department of Biomedical Engineering and Mechanics, 320 Kelly Hall, 325 Stanger Street, Blacksburg, VA 24061
- e-mail:
| |
Collapse
|
5
|
Zhang B, Liu F, Fang Z, Ding L, Moser MAJ, Zhang W. An in vivo study of a custom-made high-frequency irreversible electroporation generator on different tissues for clinically relevant ablation zones. Int J Hyperthermia 2021; 38:593-603. [PMID: 33853496 DOI: 10.1080/02656736.2021.1912417] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/21/2022] Open
Abstract
PURPOSE To examine the ablation zone, muscle contractions, and temperature increases in both rabbit liver and kidney models in vivo for a custom-made high-frequency irreversible electroporation (H-FIRE) generator. MATERIALS AND METHODS A total of 18 New Zealand white rabbits were used to investigate five H-FIRE protocols (n = 3 for each protocol) and an IRE protocol (n = 3) for the performance of the designed H-FIRE device in both liver and kidney tissues. The ablation zone was determined by using histological analysis 72 h after treatment. The extent of muscle contractions and temperature change during the application of pulse energy were measured by a commercial accelerometer attached to animals and fiber optic temperature probe inserted into organs with IRE electrodes, respectively. RESULTS All H-FIRE protocols were able to generate visible ablation zones without muscle contractions, for both liver and kidney tissues. The area of ablation zone generated in H-FIRE pulse protocols (e.g., 0.3-1 μs, 2000 V, and 90-195 bursts) appears similar to that of IRE protocol (100 μs, 1000 V, and 90 pulses) in both liver and kidney tissues. No significant temperature increase was noticed except for the protocol with the highest pulse energy (e.g., 1 μs, 2000 V, and 180 bursts). CONCLUSION Our work serves to complement the current H-FIRE pulse waveforms, which can be optimized to significantly improve the quality of ablation zone in terms of precision for liver and kidney tumors in clinical setting.
Collapse
Affiliation(s)
- Bing Zhang
- Energy-based Tumor Ablation Laboratory, School of Mechatronic Engineering and Automation, Shanghai University, Shanghai, China
| | - Fanning Liu
- Energy-based Tumor Ablation Laboratory, School of Mechatronic Engineering and Automation, Shanghai University, Shanghai, China
| | - Zheng Fang
- Energy-based Tumor Ablation Laboratory, School of Mechatronic Engineering and Automation, Shanghai University, Shanghai, China
| | - Lujia Ding
- Energy-based Tumor Ablation Laboratory, School of Mechatronic Engineering and Automation, Shanghai University, Shanghai, China
| | - Michael A J Moser
- Department of Surgery, University of Saskatchewan, Saskatoon, Canada
| | - Wenjun Zhang
- Division of Biomedical Engineering, University of Saskatchewan, Saskatoon, Canada
| |
Collapse
|
6
|
Comparison between high-frequency irreversible electroporation and irreversible electroporation ablation of small swine liver: follow-up of DCE-MRI and pathological observations. Chin Med J (Engl) 2021; 134:2081-2090. [PMID: 34172620 PMCID: PMC8439989 DOI: 10.1097/cm9.0000000000001663] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/21/2022] Open
Abstract
Background: High-frequency irreversible electroporation (H-FIRE) is a novel, next-generation nanoknife technology with the advantage of relieving irreversible electroporation (IRE)-induced muscle contractions. However, the difference between IRE and H-FIRE with distinct ablation parameters was not clearly defined. This study aimed to compare the efficacy of the two treatments in vivo. Methods: Ten Bama miniature swine were divided into two group: five in the 1-day group and five in the 7-day group. The efficacy of IRE and H-FIRE ablation was compared by volume transfer constant (Krans), rate constant (Kep) and extravascular extracellular volume fraction (Ve) value of dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI), size of the ablation zone, and histologic analysis. Each animal underwent the IRE and H-FIRE. Temperatures of the electrodes were measured during ablation. DCE-MRI images were obtained 1, 4, and 7 days after ablation in the 7-day group. All animals in the two groups were euthanized 1 day or 7 days after ablation, and subsequently, IRE and H-FIRE treated liver tissues were collected for histological examination. Student's t test or Mann-Whitney U test was applied for comparing any two groups. One-way analysis of variance (ANOVA) test and Welch's ANOVA test followed by Holm-Sidak's multiple comparisons test, one-way ANOVA with repeated measures followed by Bonferroni test, or Kruskal-Wallis H test followed by Dunn's multiple comparison test was used for multiple group comparisons and post hoc analyses. Pearson correlation coefficient test was conducted to analyze the relationship between two variables. Results: Higher Ve was seen in IRE zone than in H-FIRE zone (0.14 ± 0.02 vs. 0.08 ± 0.05, t = 2.408, P = 0.043) on day 4, but no significant difference was seen in Ktrans or Kep between IRE and H-FIRE zones at all time points (all P > 0.05). For IRE zone, the greatest Ktrans was seen on day 7, which was significantly higher than that on day 1 (P = 0.033). The ablation zone size of H-FIRE was significantly larger than IRE 1 day (4.74 ± 0.88 cm2vs. 3.20 ± 0.77 cm2, t = 3.241, P = 0.009) and 4 days (2.22 ± 0.83 cm2vs. 1.30 ± 0.50 cm2, t = 2.343, P = 0.041) after treatment. Apoptotic index (0.05 ± 0.02 vs. 0.73 ± 0.06 vs. 0.68 ± 0.07, F = 241.300, P < 0.001) and heat shock protein 70 (HSP70) (0.03 ± 0.01 vs. 0.46 ± 0.09 vs. and 0.42 ± 0.07, F = 64.490, P < 0.001) were significantly different between the untreated, IRE and H-FIRE zones, but no significant difference was seen in apoptotic index or HSP70 between IRE and H-FIRE zone (both P > 0.05). Electrode temperature variations were not significantly different between the two zones (18.00 ± 3.77°C vs. 16.20 ± 7.45°C, t = 0.682, P = 0.504). The Ktrans value (r = 0.940, P = 0.017) and the Kep value (r = 0.895, P = 0.040) of the H-FIRE zone were positively correlated with the number of hepatocytes in the ablation zone. Conclusions: H-FIRE showed a comparable ablation effect to IRE. DCE-MRI has the potential to monitor the changes of H-FIRE ablation zone.
Collapse
|
7
|
Granata V, Grassi R, Fusco R, Belli A, Palaia R, Carrafiello G, Miele V, Grassi R, Petrillo A, Izzo F. Local ablation of pancreatic tumors: State of the art and future perspectives. World J Gastroenterol 2021; 27:3413-3428. [PMID: 34163121 PMCID: PMC8218359 DOI: 10.3748/wjg.v27.i23.3413] [Citation(s) in RCA: 27] [Impact Index Per Article: 9.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/07/2021] [Revised: 03/28/2021] [Accepted: 05/08/2021] [Indexed: 02/06/2023] Open
Abstract
BACKGROUND Currently, the technologies most commonly used to treat locally advanced pancreatic cancer are radiofrequency ablation (RFA), microwave ablation, and irreversible (IRE) or reversible electroporation combined with low doses of chemotherapeutic drugs.
AIM To report an overview and updates on ablative techniques in pancreatic cancer.
METHODS Several electronic databases were searched. The search covered the years from January 2000 to January 2021. Moreover, the reference lists of the found papers were analysed for papers not indexed in the electronic databases. All titles and abstracts were analysed.
RESULTS We found 30 studies (14 studies for RFA, 3 for microwave therapy, 10 for IRE, and 3 for electrochemotherapy), comprising 1047 patients, which were analysed further. Two randomized trials were found for IRE. Percutaneous and laparotomy approaches were performed. In the assessed patients, the median maximal diameter of the lesions was in the range of 2.8 to 4.5 cm. All series included patients unfit for surgical treatment, but Martin et al assessed a subgroup of patients with borderline resectable tumours who underwent resection with margin attenuation with IRE. Most studies administered chemotherapy prior to ablative therapies. However, several studies suggest that the key determinant of improved survival is attributable to ablative treatment alone. Nevertheless, the authors suggested chemotherapy before local therapies for several reasons. This strategy may not only downstage a subgroup of patients to curative-intent surgery but also support to recognize patients with biologically unfavourable tumours who would likely not benefit from ablation treatments. Ablation therapies seem safe based on the 1047 patients assessed in this review. The mortality rate ranged from 1.8% to 2%. However, despite the low mortality, the reported rates of severe post procedural complications ranged from 0%-42%. Most reported complications have been self-limiting and manageable. Median overall survival varied between 6.0 and 33 mo. Regarding the technical success rate, assessed papers reported an estimated rate in the range of 85% to 100%. However, the authors reported early recurrence after treatment. A distinct consideration should be made on whether local treatments induce an immune response in the ablated area. Preclinical and clinical studies have shown that RFA is a promising mechanism for inducing antigen-presenting cell infiltration and enhancing the systemic antitumour T-cell immune response and tumour regression.
CONCLUSION In the management of patients with pancreatic cancer, the possibility of a multimodal approach should be considered, and conceptually, the combination of RFA with immunotherapy represents a novel angle of attack against this tumour.
Collapse
Affiliation(s)
- Vincenza Granata
- Department of Radiology, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| | - Roberta Grassi
- Department of Radiology, Università degli Studi della Campania Luigi Vanvitelli, Naples 80127, Italy
| | - Roberta Fusco
- Department of Radiology, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| | - Andrea Belli
- Department of Surgery, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| | - Raffaele Palaia
- Department of Surgery, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| | | | - Vittorio Miele
- Italian Society of Medical and Interventional Radiology SIRM, SIRM Foundation Milan 20122, Italy
- Department of Emergency Radiology, San Camillo Hospital, Firenze 50139, Italy
| | - Roberto Grassi
- Department of Radiology, Università degli Studi della Campania Luigi Vanvitelli, Naples 80127, Italy
- Italian Society of Medical and Interventional Radiology SIRM, SIRM Foundation Milan 20122, Italy
| | - Antonella Petrillo
- Department of Radiology, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| | - Francesco Izzo
- Department of Surgery, Istituto Nazionale Tumori –IRCCS- Fondazione G. Pascale, Naples 80131, Italy
| |
Collapse
|
8
|
Aycock KN, Zhao Y, Lorenzo MF, Davalos RV. A Theoretical Argument for Extended Interpulse Delays in Therapeutic High-Frequency Irreversible Electroporation Treatments. IEEE Trans Biomed Eng 2021; 68:1999-2010. [PMID: 33400646 PMCID: PMC8291206 DOI: 10.1109/tbme.2021.3049221] [Citation(s) in RCA: 16] [Impact Index Per Article: 5.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/12/2022]
Abstract
High-frequency irreversible electroporation (H-FIRE) is a tissue ablation modality employing bursts of electrical pulses in a positive phase-interphase delay (d1)-negative phase-interpulse delay (d2) pattern. Despite accumulating evidence suggesting the significance of these delays, their effects on therapeutic outcomes from clinically-relevant H-FIRE waveforms have not been studied extensively. OBJECTIVE We sought to determine whether modifications to the delays within H-FIRE bursts could yield a more desirable clinical outcome in terms of ablation volume versus extent of tissue excitation. METHODS We used a modified spatially extended nonlinear node (SENN) nerve fiber model to evaluate excitation thresholds for H-FIRE bursts with varying delays. We then calculated non-thermal tissue ablation, thermal damage, and excitation in a clinically relevant numerical model. RESULTS Excitation thresholds were maximized by shortening d1, and extension of d2 up to 1,000 μs increased excitation thresholds by at least 60% versus symmetric bursts. In the ablation model, long interpulse delays lowered the effective frequency of burst waveforms, modulating field redistribution and reducing heat production. Finally, we demonstrate mathematically that variable delays allow for increased voltages and larger ablations with similar extents of excitation as symmetric waveforms. CONCLUSION Interphase and interpulse delays play a significant role in outcomes resulting from H-FIRE treatment. SIGNIFICANCE Waveforms with short interphase delays (d1) and extended interpulse delays (d2) may improve therapeutic efficacy of H-FIRE as it emerges as a clinical tissue ablation modality.
Collapse
Affiliation(s)
- Kenneth N. Aycock
- Department of Biomedical Engineering and Mechanics, Bioelectromechanical Systems Laboratory at Virginia Tech, Blacksburg, VA 24061 USA
| | - Yajun Zhao
- Department of Biomedical Engineering and Mechanics, Bioelectromechanical Systems Laboratory at Virginia Tech, Blacksburg, VA 24061 USA
| | - Melvin F. Lorenzo
- Department of Biomedical Engineering and Mechanics, Bioelectromechanical Systems Laboratory at Virginia Tech, Blacksburg, VA 24061 USA
| | - Rafael V. Davalos
- Department of Biomedical Engineering and Mechanics, Bioelectromechanical Systems Laboratory at Virginia Tech, Blacksburg, VA 24061 USA
| |
Collapse
|
9
|
High-Frequency and High-Voltage Asymmetric Bipolar Pulse Generator for Electroporation Based Technologies and Therapies. ELECTRONICS 2021. [DOI: 10.3390/electronics10101203] [Citation(s) in RCA: 7] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
Abstract
Currently, in high-frequency electroporation, much progress has been made but limited to research groups with custom-made laboratory prototype electroporators. According to the review of electroporators and economic evaluations, there is still an area of pulse parameters that needs to be investigated. The development of an asymmetric bipolar pulse generator with a maximum voltage of 4 kV and minimum duration time of a few hundred nanoseconds, would enable in vivo evaluation of biological effects of high-frequency electroporation pulses. Herein, from a series of most commonly used drivers and optical isolations in high-voltage pulse generators the one with optimal characteristics was used. In addition, the circuit topology of the developed device is described in detail. The developed device is able to generate 4 kV pulses, with theoretical 131 A maximal current and 200 ns minimal pulse duration, the maximal pulse repetition rate is 2 MHz and the burst maximal repetition rate is 1 MHz. The device was tested in vivo. The effectiveness of electrochemotherapy of high-frequency electroporation pulses is compared to “classical” electrochemotherapy pulses. In vivo electrochemotherapy with high-frequency electroporation pulses was at least as effective as with “classical” well-established electric pulses, resulting in 86% and 50% complete responses, respectively. In contrast to previous reports, however, muscle contractions were comparable between the two protocols.
Collapse
|
10
|
Sano MB, DeWitt MR. Thermochromic Tissue Phantoms for Evaluating Temperature Distribution in Simulated Clinical Applications of Pulsed Electric Field Therapies. Bioelectricity 2020; 2:362-371. [PMID: 34476365 PMCID: PMC8370349 DOI: 10.1089/bioe.2020.0023] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022] Open
Abstract
Background: Irreversible electroporation (IRE) induces cell death through nonthermal mechanisms, however, in extreme cases, the treatments can induce deleterious thermal transients. This study utilizes a thermochromic tissue phantom to enable visualization of regions exposed to temperatures above 60°C. Materials and Methods: Poly(vinyl alcohol) hydrogels supplemented with thermochromic ink were characterized and processed to match the electrical properties of liver tissue. Three thousand volt high-frequency IRE protocols were administered with delivery rates of 100 and 200 μs/s. The effect of supplemental internal applicator cooling was then characterized. Results: Baseline treatments resulted thermal areas of 0.73 cm2, which decreased to 0.05 cm2 with electrode cooling. Increased delivery rates (200 μs/s) resulted in thermal areas of 1.5 and 0.6 cm2 without and with cooling, respectively. Conclusions: Thermochromic tissue phantoms enable rapid characterization of thermal effects associated with pulsed electric field treatments. Active cooling of applicators can significantly reduce the quantity of tissue exposed to deleterious temperatures.
Collapse
Affiliation(s)
- Michael B. Sano
- UNC/NCSU Joint Department of Biomedical Engineering, Raleigh, North Carolina, USA
| | | |
Collapse
|
11
|
Petrella RA, Fesmire CC, Kaufman JD, Topasna N, Sano MB. Algorithmically Controlled Electroporation: A Technique for Closed Loop Temperature Regulated Pulsed Electric Field Cancer Ablation. IEEE Trans Biomed Eng 2020; 67:2176-2186. [PMID: 32673194 DOI: 10.1109/tbme.2019.2956537] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
OBJECTIVE To evaluate the effect of a closed-loop temperature based feedback algorithm on ablative outcomes for pulsed electric field treatments. METHODS A 3D tumor model of glioblastoma was used to assess the impact of 2 μs duration bipolar waveforms on viability following exposure to open and closed-loop protocols. Closed-loop treatments evaluated transient temperature increases of 5, 10, 15, or 22 °C above baseline. RESULTS The temperature controlled ablation diameters were conditionally different than the open-loop treatments and closed-loop treatments generally produced smaller ablations. Closed-loop control enabled the investigation of treatments with steady state 42 °C hyperthermic conditions which were not feasible without active feedback. Baseline closed-loop treatments at 20 °C resulted in ablations measuring 9.9 ± 0.3 mm in diameter while 37 °C treatments were 20% larger (p < 0.0001) measuring 11.8 ± 0.3 mm indicating that this protocol induces a thermally mediated biological response. CONCLUSION A closed-loop control algorithm which modulated the delay between successive pulse waveforms to achieve stable target temperatures was demonstrated. Algorithmic control enabled the evaluation of specific treatment parameters at physiological temperatures not possible with open-loop systems due to excessive Joule heating. SIGNIFICANCE Irreversible electroporation is generally considered to be a non-thermal ablation modality and temperature monitoring is not part of the standard clinical practice. The results of this study indicate ablative outcomes due to exposure to pulses on the order of one microsecond may be thermally mediated and dependent on local tissue temperatures. The results of this study set the foundation for experiments in vivo utilizing temperature control algorithms.
Collapse
|
12
|
Electro-Thermal Therapy Algorithms and Active Internal Electrode Cooling Reduce Thermal Injury in High Frequency Pulsed Electric Field Cancer Therapies. Ann Biomed Eng 2020; 49:191-202. [PMID: 32415482 DOI: 10.1007/s10439-020-02524-x] [Citation(s) in RCA: 12] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2019] [Accepted: 04/27/2020] [Indexed: 10/24/2022]
Abstract
Thermal tissue injury is an unintended consequence in current irreversible electroporation treatments due to the induction of Joule heating during the delivery of high voltage pulsed electric fields. In this study active temperature control measures including internal electrode cooling and dynamic energy delivery were investigated as a process for mitigating thermal injury during treatment. Ex vivo liver was used to examine the extent of thermal injury induced by 5000 V treatments with delivery rates up to five times faster than current clinical practice. Active internal cooling of the electrode resulted in a 36% decrease in peak temperature vs. non-cooled control treatments. A temperature based feedback algorithm (electro-thermal therapy) was demonstrated as capable of maintaining steady state tissue temperatures between 30 and 80 °C with and without internal electrode cooling. Thermal injury volumes of 2.6 cm3 were observed for protocols with 60 °C temperature set points and electrode cooling. This volume reduced to 1.5 and 0.1 cm3 for equivalent treatments with 50 °C and 40 °C set points. Finally, it was demonstrated that the addition of internal electrode cooling and active temperature control algorithms reduced ETT treatment times by 84% (from 343 to 54 s) vs. non-cooled temperature control strategies with equivalent thermal injury volumes.
Collapse
|
13
|
Sano MB, Petrella RA, Kaufman JD, Fesmire CC, Xing L, Gerber D, Fogle CA. Electro-thermal therapy: Microsecond duration pulsed electric field tissue ablation with dynamic temperature control algorithms. Comput Biol Med 2020; 121:103807. [PMID: 32568680 DOI: 10.1016/j.compbiomed.2020.103807] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/19/2019] [Revised: 04/27/2020] [Accepted: 05/02/2020] [Indexed: 12/13/2022]
Abstract
Electro-thermal therapy (ETT) is a new cancer treatment modality which combines the use of high voltage pulsed electric fields, dynamic energy delivery rates, and closed loop thermal control algorithms to rapidly and reproducibly create focal ablations. This study examines the ablative potential and profile of pulsed electric field treatments delivered in conjunction with precise temperature control algorithms. An ex vivo perfused liver model was utilized to demonstrate the capability of 5000 V 2 μs duration bipolar electrical pulses and dynamic temperature control algorithms to produce ablations. Using a three applicator array, 4 cm ablation zones were created in under 27 min. In this configuration, the algorithms were able to rapidly achieve and maintain temperatures of 80 °C at the tissue-electrode interface. A simplified single applicator and grounding pad approach was used to correlate the measured ablation zones to electric field isocontours in order to determine lethal electric field thresholds of 708 V/cm and 867 V/cm for 45 °C and 60 °C treatments, respectively. These results establish ETT as a viable method for hepatic tumor treatment with ablation profiles equivalent to other energy based techniques. The single applicator and multi-applicator approaches demonstrated may enable the treatment of complex tumor geometries. The flexibility of ETT temperature control yields a malleable intervention which gives clinicians robust control over the ablation modality, treatment time, and safety profile.
Collapse
Affiliation(s)
- Michael B Sano
- UNC/ NCSU Joint Department of Biomedical Engineering, Raleigh, NC, USA.
| | - Ross A Petrella
- UNC/ NCSU Joint Department of Biomedical Engineering, Raleigh, NC, USA
| | - Jacob D Kaufman
- UNC/ NCSU Joint Department of Biomedical Engineering, Raleigh, NC, USA
| | | | - Lei Xing
- Stanford University School of Medicine, Division of Radiation Physics, Stanford, CA, USA
| | - David Gerber
- Division of Abdominal Transplantation, Department of Surgery, University of North Carolina School of Medicine, Chapel Hill, NC, USA
| | - Callie A Fogle
- Department of Clinical Sciences and Population Health & Pathobiology, North Carolina State University College of Veterinary Medicine, USA
| |
Collapse
|
14
|
DeWitt MR, Latouche EL, Kaufman JD, Fesmire CC, Swet JH, Kirks RC, Baker EH, Vrochides D, Iannitti DA, McKillop IH, Davalos RV, Sano MB. Simplified Non-Thermal Tissue Ablation With a Single Insertion Device Enabled by Bipolar High-Frequency Pulses. IEEE Trans Biomed Eng 2019; 67:2043-2051. [PMID: 31751216 DOI: 10.1109/tbme.2019.2954122] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/02/2023]
Abstract
OBJECTIVE To demonstrate the feasibility of a single electrode and grounding pad approach for delivering high frequency irreversible electroporation treatments (H-FIRE) in in-vivo hepatic tissue. METHODS Ablations were created in porcine liver under surgical anesthesia by adminstereing high frequency bursts of 0.5-5.0 μs pulses with amplitudes between 1.1-1.7 kV in the absence of cardiac synchronization or intraoperative paralytics. Finite element simulations were used to determine the electric field strength associated with the ablation margins (ELethal) and predict the ablations feasible with next generation electronics. RESULTS All animals survived the procedures for the protocol duration without adverse events. ELethal of 2550, 1650, and 875 V/cm were found for treatments consisting of 100x bursts containing 0.5 μs pulses and 25, 50, and 75 μs of energized-time per burst, respectively. Treatments with 1 μs pulses consisting of 100 bursts with 100 μs energized-time per burst resulted in ELethal of 650 V/cm. CONCLUSION A single electrode and grounding pad approach was successfully used to create ablations in hepatic tissue. This technique has the potential to reduce challenges associated with placing multiple electrodes in anatomically challenging environments. SIGNIFICANCE H-FIRE is an in situ tumor ablation approach in which electrodes are placed within or around a targeted region to deliver high voltage electrical pulses. Electric fields generated around the electrodes induce irrecoverable cell membrane damage leading to predictable cell death in the relative absence of thermal damage. The sparing of architectural integrity means H-FIRE offers potential advantages compared to thermal ablation modalities for ablating tumors near critical structures.
Collapse
|