Braun P, Delgado R, Drago M, Fanti D, Fleury H, Hofmann J, Izopet J, Kühn S, Lombardi A, Mancon A, Marcos MA, Mileto D, Sauné K, O'Shea S, Pérez-Rivilla A, Ramble J, Trimoulet P, Vila J, Whittaker D, Artus A, Rhodes D. A European multicientre study on the comparison of HIV-1 viral loads between VERIS HIV-1 Assay and Roche COBAS® TAQMAN® HIV-1 test, Abbott RealTime HIV-1 Assay, and Siemens VERSANT HIV-1 Assay. J Clin Virol 2017; 92:75-82. [PMID: 28599228 DOI: 10.1016/j.jcv.2017.05.003] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/01/2017] [Revised: 05/03/2017] [Accepted: 05/06/2017] [Indexed: 11/30/2022]
Abstract
BACKGROUND Viral load monitoring is essential for patients under treatment for HIV. Beckman Coulter has developed the VERIS HIV-1 Assay for use on the novel, automated DxN VERIS Molecular Diagnostics System.¥ OBJECTIVES: Evaluation of the clinical performance of the new quantitative VERIS HIV-1 Assay at multiple EU laboratories. STUDY DESIGN Method comparison with the VERIS HIV-1 Assay was performed with 415 specimens at 5 sites tested with COBAS® AmpliPrep/COBAS® TaqMan® HIV-1 Test, v2.0, 169 specimens at 3 sites tested with RealTime HIV-1 Assay, and 202 specimens from 2 sites tested with VERSANT HIV-1 Assay. Patient monitoring sample results from 4 sites were also compared. RESULTS Bland-Altman analysis showed the average bias between VERIS HIV-1 Assay and COBAS HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay to be 0.28, 0.39, and 0.61 log10 cp/mL, respectively. Bias at low end levels below 1000cp/mL showed predicted bias to be <0.3 log10 cp/mL for VERIS HIV-1 Assay versus COBAS HIV-1 Test and RealTime HIV-1 Assay, and <0.5 log10cp/mL versus VERSANT HIV-1 Assay. Analysis on 174 specimens tested with the 0.175mL volume VERIS HIV-1 Assay and COBAS HIV-1 Test showed average bias of 0.39 log10cp/mL. Patient monitoring results using VERIS HIV-1 Assay demonstrated similar viral load trends over time to all comparators. CONCLUSIONS The VERIS HIV-1 Assay for use on the DxN VERIS System demonstrated comparable clinical performance to COBAS® HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay.
Collapse
Affiliation(s)
- Patrick Braun
- Laboratory Dr. Knechten, Medical Center for HIV and Hepatitis, Aachen, Germany
| | - Rafael Delgado
- Microbiology Department, Hospital Universitario 12 de Octubre and Insituto de Investigation, Hospital 12 de Octubre (i+12) Madrid, Spain
| | - Monica Drago
- Clinical Chemistry and Microbiology Laboratories, Niguarda Ca' Granda Hospital, Milan, Italy
| | - Diana Fanti
- Clinical Chemistry and Microbiology Laboratories, Niguarda Ca' Granda Hospital, Milan, Italy
| | - Hervé Fleury
- Virology Department, Hôpital Pellegrin, CHU, Bordeaux, France
| | - Jörg Hofmann
- Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany
| | - Jacques Izopet
- Department of Virology, Federative Institute of Biology, CHU, Toulouse, France
| | - Sebastian Kühn
- Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany
| | - Alessandra Lombardi
- Clinical Microbiology, Virology and Bioemergence Diagnosis, L. Sacco University Hospital, Milan, Italy
| | - Alessandro Mancon
- Clinical Microbiology, Virology and Bioemergence Diagnosis, L. Sacco University Hospital, Milan, Italy
| | - Mª Angeles Marcos
- Department of Clinical Microbiology, Hospital Clinic, School of Medicine, University of Barcelona, Centre for International Health Research (CRESIB), Barcelona, Spain
| | - Davide Mileto
- Virology Department, Labor Berlin - Charité Vivantes GmbH, Berlin, Germany
| | - Karine Sauné
- Department of Virology, Federative Institute of Biology, CHU, Toulouse, France
| | - Siobhan O'Shea
- Viapath Analytics, Infection Sciences, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom
| | - Alfredo Pérez-Rivilla
- Microbiology Department, Hospital Universitario 12 de Octubre and Insituto de Investigation, Hospital 12 de Octubre (i+12) Madrid, Spain
| | - John Ramble
- Viapath Analytics, Infection Sciences, Guy's and St. Thomas' NHS Foundation Trust, London, United Kingdom
| | | | - Jordi Vila
- Department of Clinical Microbiology, Hospital Clinic, School of Medicine, University of Barcelona, Centre for International Health Research (CRESIB), Barcelona, Spain
| | - Duncan Whittaker
- Laboratory Medicine Building, North Lane, Northern General Hospital, Sheffield, United Kingdom
| | - Alain Artus
- Beckman Coulter, Immunotech, 130 Ave Lattre de Tassigny, Marseille 13009, France
| | - Daniel Rhodes
- Beckman Coulter, Immunotech, 130 Ave Lattre de Tassigny, Marseille 13009, France.
| |
Collapse
|
2
|
European Multicenter Study on Analytical Performance of Veris HIV-1 Assay. J Clin Microbiol 2017; 55:2055-2063. [PMID: 28424254 DOI: 10.1128/jcm.00078-17] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/24/2017] [Accepted: 04/12/2017] [Indexed: 11/20/2022] Open
Abstract
The analytical performance of the Veris HIV-1 assay for use on the new, fully automated Beckman Coulter DxN Veris molecular diagnostics system was evaluated at 10 European virology laboratories. The precision, analytical sensitivity, performance with negative samples, linearity, and performance with HIV-1 groups/subtypes were evaluated. The precision for the 1-ml assay showed a standard deviation (SD) of 0.14 log10 copies/ml or less and a coefficient of variation (CV) of ≤6.1% for each level tested. The 0.175-ml assay showed an SD of 0.17 log10 copies/ml or less and a CV of ≤5.2% for each level tested. The analytical sensitivities determined by probit analysis were 19.3 copies/ml for the 1-ml assay and 126 copies/ml for the 0.175-ml assay. The performance with 1,357 negative samples demonstrated 99.2% with not detected results. Linearity using patient samples was shown from 1.54 to 6.93 log10 copies/ml. The assay performed well, detecting and showing linearity with all HIV-1 genotypes tested. The Veris HIV-1 assay demonstrated analytical performance comparable to that of currently marketed HIV-1 assays. (DxN Veris products are Conformité Européenne [CE]-marked in vitro diagnostic products. The DxN Veris product line has not been submitted to the U.S. FDA and is not available in the U.S. market. The DxN Veris molecular diagnostics system is also known as the Veris MDx molecular diagnostics system and the Veris MDx system.).
Collapse
|
3
|
Evaluation of the Aptima HIV-1 Quant Dx Assay Using Plasma and Dried Blood Spots. J Clin Microbiol 2016; 54:2597-601. [PMID: 27535684 DOI: 10.1128/jcm.01569-16] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/20/2016] [Accepted: 08/08/2016] [Indexed: 11/20/2022] Open
Abstract
HIV-1 RNA quantitation in plasma, or virus load testing, is the primary method by which the response to antiretroviral therapy is monitored. Here we describe evaluation of the Aptima HIV-1 Quant Dx assay (Aptima) performed on the automated Panther system. The clinical performance of Aptima was compared to that of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test v2.0 (CAP/CTM) using 162 EDTA plasma samples collected from patients undergoing HIV-1 monitoring. Overall agreement was 84.0% (136/162), with a kappa statistic of 0.723 (standard error, 0.047; 95% confidence interval [CI], 0.630 to 0.815), indicating substantial agreement. Using the 86 clinical samples quantifiable by both methods, Passing-Bablok regression revealed a regression line of Y = (1.069 × X) - 0.346 (95% CI of the slope [1.003 to 1.139] and intercept [-0.666 to -0.074]), and Bland-Altman analysis demonstrated a mean difference (Aptima-CAP/CTM) of -0.075 log10 copies/ml (95% limits of agreement of -0.624 to 0.475), consistent with negative bias. Comparison of Aptima results for paired dried blood spot (DBS) and plasma specimens archived from participants in the Peninsula AIDS Research Cohort Study (PARC) demonstrated an overall agreement of 94.7% (90/95) when 1,000 copies/ml was used as the threshold. In conclusion, the Aptima HIV-1 Quant Dx assay provides a suitable alternative for HIV-1 monitoring in plasma and DBS.
Collapse
|
4
|
Tung YC, Ke LY, Lu PL, Lin KH, Lee SC, Lin YY, Chou LC, Tsai WC. Comparison of the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test v1.0 with v2.0 in HIV-1 viral load quantification. Kaohsiung J Med Sci 2015; 31:188-93. [PMID: 25835274 DOI: 10.1016/j.kjms.2014.12.010] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/29/2014] [Revised: 12/30/2014] [Accepted: 12/16/2014] [Indexed: 11/17/2022] Open
Abstract
Roche modified the COBAS AmpliPrep/COBAS TaqMan human immunodeficiency virus type 1 (HIV-1) test version 1.0 (CAP/CTM v1.0), resulting in the COBAS AmpliPrep/COBAS TaqMan HIV-1 test version 2.0 (CAP/CTM v2.0). The aim of this study was to evaluate the performance of the CAP/CTM v2.0 and to compare this performance with that of the CAP/CTM v1.0. The study was conducted in a small local study group in Kaohsiung Medical University Hospital, Kaohsiung, Taiwan. A total of 86 plasma samples from HIV-1-seropositive patients were tested using the two assays. The correlation and concordance of results between the two assays were calculated. The CAP/CTM v2.0 generated higher values than did the CAP/CTM v1.0, and five samples (5.8%) yielded a difference of > 1 log10 copies/mL. In addition, our data show that CAP/CTM v1.0 and CAP/CTM v2.0 yielded relatively consistent values for 23 samples with low viral loads (< 200 copies/mL). Furthermore, when viral loads were in a medium range (2-5 log10 copies/mL), the results of the two assays were more compatible. This study shows a good correlation between CAP/CTM v1.0 and v2.0 in HIV-1 viral load measurement. Further attention must be paid to those cases in which measured viral loads present larger differences between the two assays.
Collapse
Affiliation(s)
- Yi-Ching Tung
- Department of Public Health and Environmental Medicine, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan; Graduate Institute, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan
| | - Liang-Yin Ke
- Department of Laboratory Medicine Kaohsiung Medical University Hospital, Kaohsiung, Taiwan
| | - Po-Liang Lu
- Department of Laboratory Medicine Kaohsiung Medical University Hospital, Kaohsiung, Taiwan; Department of Internal Medicine, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan
| | - Kuei-Hsiang Lin
- Laboratory Diagnosis of Medicine, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan
| | - Su-Chen Lee
- Laboratory Diagnosis of Medicine, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan
| | - Yi-Ying Lin
- Department of Laboratory Medicine Kaohsiung Medical University Hospital, Kaohsiung, Taiwan
| | - Li-Chiu Chou
- Department of Laboratory Medicine Kaohsiung Medical University Hospital, Kaohsiung, Taiwan
| | - Wen-Chan Tsai
- Department of Laboratory Medicine Kaohsiung Medical University Hospital, Kaohsiung, Taiwan; Department of Internal Medicine, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan.
| |
Collapse
|
5
|
Henrich TJ, Wood BR, Kuritzkes DR. Increased risk of virologic rebound in patients on antiviral therapy with a detectable HIV load <48 copies/mL. PLoS One 2012; 7:e50065. [PMID: 23166820 PMCID: PMC3499490 DOI: 10.1371/journal.pone.0050065] [Citation(s) in RCA: 40] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/02/2012] [Accepted: 10/18/2012] [Indexed: 11/18/2022] Open
Abstract
We investigated the independent effects of HIV-1 "target not detected" measurements versus those that were detectable but below the limit of quantification by Taqman RT-PCR assay on subsequent viral rebound as there are conflicting data regarding the clinical implications of arbitrary or isolated low-level viremia. Cox proportional hazard regression modeling was used to investigate the independent effects of the first HIV-1 load measurement after introduction of the Taqman RT-PCR assay (time-point 0 [T0]), pre-T0 viral loads, CD4 T cell count, race/ethnicity, gender, age and NNRTI use on risk of a confirmed VL >50, >200, >400 and >1000 copies/mL at 22 months follow-up in analyses of all patients and propensity-matched baseline cohorts. 778 patients had a viral load that was either not detected by RT-PCR (N = 596) or detectable, but below the limit of quantification (N = 182) at T0. Detectable viremia, lower T0 CD4 count, decreased age, and having detectable or unknown VL within a year prior to T0 were each associated with viral rebound to >50, >200 and >400 copies/mL. Overall failure rates were low and <5.5% of all patients had confirmed VL >1000 copies/mL. A majority of patients with rebound >200 copies/mL subsequently re-suppressed (28 of 53). A detectable VL <48 copies/mL was independently and significantly associated with subsequent viral rebound, and is cause for clinical concern.
Collapse
Affiliation(s)
- Timothy J Henrich
- Division of Infectious Diseases, Brigham and Women's Hospital and Harvard Medical School, Boston, Massachusetts, United States of America.
| | | | | |
Collapse
|