Nibe Y, Takahashi T, Hai H, Matsumura T, Takeshita K. Comparative biomechanical analysis of tibial posterior slope in medial open wedge high tibial osteotomy vs. distal tuberosity osteotomy with and without anterior-posterior screw: a study using porcine tibia. SICOT J 2024; 10:41. [PMID: 39431748 PMCID: PMC11492831 DOI: 10.1051/sicotj/2024042] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/11/2024] [Accepted: 09/15/2024] [Indexed: 10/22/2024] Open
Abstract
Purpose While increased posterior tibial slope (PTS) is a concern post-medial open wedge high tibial osteotomy (MOWHTO), the ability of distal tuberosity osteotomy (DTO) to maintain postoperative PTS after cyclic loading remains unverified. This study aims to determine whether PTS alterations significantly differ between DTO and MOWHTO following cyclic loading. METHODS Biomechanical evaluations were conducted on thirty porcine tibias using MOWHTO and DTO, with and without an anterior-posterior (AP) screw. To investigate PTS changes, cyclic testing was carried out for MOWHTO and DTO. Displacement along the mechanical axis during cycles 10th, 100th, 500th, 1000th, 1500th and 2000th, variations in anterior and posterior gaps after 2000 cycles and increased PTS after 2000 cycles, were compared across the three groups. The displacement was evaluated by repeated-measures analysis of variance (ANOVA), and changes in AG and PG and increased PTS were evaluated by one-way ANOVA. The sample size for α and β errors were <0.05 and <0.20, and the effect size was 0.60 for one-way ANOVA and 0.46 for repeated-measures ANOVA. RESULTS There were no significant differences in displacement and anterior gap changes among the groups. A significant difference was observed in the posterior gap changes (P < 0.001) and increased PTS (P = 0.013) among the groups. Post hoc analysis indicated substantial disparities between MOWHTO and DTO without the AP screw (P = 0.035), as well as between MOWHTO and DTO with the AP screw (P = 0.021) concerning the increased PTS. CONCLUSION After cyclic loading, MOWHTO exhibited a notably smaller PTS change than DTO regardless of the presence of an AP screw.
Collapse
Affiliation(s)
- Yoshiya Nibe
- Department of Orthopaedics Surgery, Jichi Medical University 3311-1 Yakushiji Shimotsuke 329-0498 Japan
| | - Tsuneari Takahashi
- Department of Orthopaedics Surgery, Jichi Medical University 3311-1 Yakushiji Shimotsuke 329-0498 Japan
| | - Hironari Hai
- Department of Orthopaedic Surgery, Toyokawa City Hospital 23 Noji Toyokawa 442-0857 Japan
| | - Tomohiro Matsumura
- Department of Emergency and Critical Care Medicine, Jichi Medical University 3311-1 Yakushiji Shimotsuke 329-0498 Japan
| | - Katsushi Takeshita
- Department of Orthopaedics Surgery, Jichi Medical University 3311-1 Yakushiji Shimotsuke 329-0498 Japan
| |
Collapse
|
2
|
Saghaei Z, Salehipour S, Hashemi A. Larger lateral hinges increase the probability of Takeuchi type II and III fractures in high tibial osteotomy. EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY & TRAUMATOLOGY : ORTHOPEDIE TRAUMATOLOGIE 2024; 34:2717-2722. [PMID: 38761202 DOI: 10.1007/s00590-024-03935-5] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 12/14/2023] [Accepted: 03/25/2024] [Indexed: 05/20/2024]
Abstract
PURPOSE Lateral hinge fractures are the main complications in the high tibial osteotomy to treat varus deformities. The aim of present study is to answer the question whether the lateral hinge length (H) has an effect on the type of fracture and required force during the opening in high tibia osteotomy. It was hypothesized in this comparative research that extending the hinge length increased opening force and probability of a type II and type III fractures. METHODS A monoplanar medial open wedge osteotomy with different intact hinge lengths varying from 9 to 32 mm was performed in 20 ostrich bones. A biomechanical experiment using unidirectional tensile testing apparatus was performed to open the wedge, and the required force was increased until a 10 mm opening was reached; then, the presence of fracture in the lateral cortex and its direction were evaluated. Lateral hinge fracture type based on direction was classified as suggested by Takeuchi et al. RESULTS Fracture that grows along the osteotomy line (type I) was observed in 4 samples with the mean hinge length (H) of 11 ± 1.54 mm. For seven bones with Takeuchi fracture type II, with downward crack propagation, the mean H was 16 ± 3.36 mm. For the mean H of 25 ± 6.53 mm, the crack propagated upward to the cutting path, displaying a Takeuchi type III fracture in seven samples. The statistical analysis showed that the fracture type significantly depends on the hinge length (P value < 0.05). Also, the mean opening force significantly increased with hinge lengthening (P value < 0.05). The peak forces at crack initiation were 41.8 ± 21.9, 115.2 ± 41.5, and 167 ± 135.3 N, respectively, for the fracture types I, II, and III samples. CONCLUSION The lateral cortical hinge length was significantly associated with hinge fracture type. The experimental tests indicated that the hinge lengthening increases the risk of type II and III fractures, as classified by Takeuchi.
Collapse
Affiliation(s)
- Zahra Saghaei
- Biomechanical Engineering Group, Department of Biomedical Engineering, Amirkabir University of Technology, Hafez Avenue, Tehran, 15875, Iran
| | - Saeed Salehipour
- Biomechanical Engineering Group, Department of Biomedical Engineering, Amirkabir University of Technology, Hafez Avenue, Tehran, 15875, Iran
| | - Ata Hashemi
- Biomechanical Engineering Group, Department of Biomedical Engineering, Amirkabir University of Technology, Hafez Avenue, Tehran, 15875, Iran.
| |
Collapse
|
3
|
Nibe Y, Takahashi T, Matsumura T, Kubo T, Takeshita K. Effect of plate design on maintenance of anterior and posterior gaps and posterior tibial slope after cyclic loading in medial open-wedge high tibial osteotomy: A biomechanical study using porcine's tibia. J Exp Orthop 2024; 11:e12036. [PMID: 38899050 PMCID: PMC11185942 DOI: 10.1002/jeo2.12036] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/27/2023] [Revised: 03/29/2024] [Accepted: 04/19/2024] [Indexed: 06/21/2024] Open
Abstract
Purpose We aimed to investigate whether a plate adapted to the anatomy of the proximal medial porcine's tibia would provide maintenance of the anterior gap (AG), posterior gap (PG) and posterior tibial slope (PTS). Methods Twenty-seven porcine tibias were biomechanically evaluated by performing MOWHTO and placing TOMOFIX (n = 9), AC plate (n = 9) and TriS (n = 9) anteromedially. Cyclic testing (800 N, 2000 cycles, 0.5 Hz) was performed to investigate the PTS over time for MOWHTO. The particular displacement calculated from the maximum to the minimum point with the load-displacement curve along the mechanical axis during cyclic testing, the final AG and PG changes at the osteotomy site, the increased PTS calculated by subtracting AG from PG after 2000 cycles were compared among the three groups. The displacement was evaluated by repeated-measures analysis of variance (ANOVA), and changes in AG and PG and increased PTS were evaluated by one-way ANOVA. The sample size for α and β errors were <0.05 and <0.20, and the effect size was 0.64 for one-way ANOVA and 0.49 for repeated-measures ANOVA. Results There were no significant differences in displacement among the groups. A significant difference was observed in the AG (p = 0.044) and PG (p = 0.0085) changes. There were no significant differences in increased PTS among the groups. Conclusion When anteromedially placed, the AC plate and TriS resulted in significant maintenance of AG and PG compared with that of TOMOFIX in MOWHTO after cyclic loading. Level of Evidence Level Ⅳ.
Collapse
Affiliation(s)
- Yoshiya Nibe
- Department of Orthopaedic Surgery, School of MedicineJichi Medical UniversityShimotsukeJapan
| | - Tsuneari Takahashi
- Department of Orthopaedic SurgeryIshibashi General HospitalShimotsukeJapan
| | - Tomohiro Matsumura
- Department of Emergency and Critical Care MedicineJichi Medical UniversityShimotsukeJapan
| | - Tatsuya Kubo
- Department of Orthopaedic Surgery, School of MedicineJichi Medical UniversityShimotsukeJapan
| | - Katsushi Takeshita
- Department of Orthopaedic Surgery, School of MedicineJichi Medical UniversityShimotsukeJapan
| |
Collapse
|
4
|
Wen PY, Sun H, Li J, Fu C, Chen P, Yu J, Chen W, Zhang Y. Is single-stage bilateral medial opening wedge high tibial osteotomy advisable? BMC Musculoskelet Disord 2024; 25:497. [PMID: 38926688 PMCID: PMC11201859 DOI: 10.1186/s12891-024-07501-2] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/20/2023] [Accepted: 05/06/2024] [Indexed: 06/28/2024] Open
Abstract
PURPOSE To validate the safety and clinical results of single-stage bilateral versus unilateral medial opening wedge high tibial osteotomy (HTO). METHODS A propensity-matched cohort study was performed from March 2020 to March 2021 in our medical center. Data were prospectively collected. Including 34 patients who underwent single-stage bilateral medial opening HTO(SSBHTO), and 68 cases in the unilateral group. Propensity-matched ration was 2:1 based on age, sex, and body mass index using R software. Comparisons of the length of hospital stay, operative time, blood loss, postoperative adverse events, 90-day readmission rate, conversion to TKA rate, self-reported VAS and WOMAC scores were made to investigate the safety and clinical results of bilateral HTO. RESULTS The mean length of hospital stay was 7.36 ± 2.23 days for SSBHTO and 7 days (IQR, 3 days; range, 4 to 23 days) for the unilateral group (P = 0.219). The mean operative time was 144 ± 47 min for bilateral HTO and 105(37.5) mins for a unilateral OWHTO (P < 0.001). The mean blood loss was 150(100) ml for SSBHTO and 100(50) ml for unilateral OWHTO (P < 0.001). There were no significant difference of the adverse events and 90-day readmission rate between two groups. No failed HTO or conversion to knee arthroplasty were observed at the end of follow-up. VAS, pain, stiffness, and functional scores of the WOMAC scale were essentially comparable of two groups one year after surgery (P > 0.05). CONCLUSIONS A single-stage bilateral medial opening wedge high tibial osteotomy is advisable for patients with knee osteoarthritis. Patients benefit from avoiding secondary anesthesia, postoperative complications, and substantial cost savings. LEVEL OF EVIDENCE Therapeutic Level III.
Collapse
Affiliation(s)
- Peizhi Yu Wen
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China
| | - Huilian Sun
- The Second Hospital of Hebei Medical University, No. 215 Heping West Road, Shijiazhuang, 050004, People's Republic of China
| | - Jiaqi Li
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China
| | - Chunxu Fu
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China
| | - Pengzhao Chen
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China
| | - Jiahao Yu
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China
| | - Wei Chen
- Trauma Emergency Center, Shijiazhuang, People's Republic of China
- Department of Orthopedic Surgery, NHC Key Laboratory of Intelligent Orthopaedic Equipment, The Third Hospital of Hebei Medical University, No. 139 Ziqiang Road, Qiaoxi District, Shijiazhuang, 050051, People's Republic of China
- Hebei Orthopaedic Clinical Research Center, Shijiazhuang, People's Republic of China
| | - Yingze Zhang
- Orthopaedic Research Institute, Shijiazhuang, Hebei Province, People's Republic of China.
- Trauma Emergency Center, Shijiazhuang, People's Republic of China.
- Department of Orthopedic Surgery, NHC Key Laboratory of Intelligent Orthopaedic Equipment, The Third Hospital of Hebei Medical University, No. 139 Ziqiang Road, Qiaoxi District, Shijiazhuang, 050051, People's Republic of China.
- Hebei Orthopaedic Clinical Research Center, Shijiazhuang, People's Republic of China.
- Engineering Research Center of Orthaepedic, The Third Hospital of Hebei Medical University, 139 Ziqiang Road, Shijiazhuang, 050051, People's Republic of China.
| |
Collapse
|
5
|
Zou Z, Wu Z, Yuan D, Xu S, Yang L, Ye C. Intraoperative radiographic analysis and adjustment of the optimal position of plate in high tibial osteotomy. Acta Radiol 2024; 65:609-615. [PMID: 38204189 DOI: 10.1177/02841851231223661] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/12/2024]
Abstract
BACKGROUND When high tibial osteotomy is performed for genu varus deformity, it is not easy to determine the accurate placement of the plate. PURPOSE To determine a simple way to assess the position of the plate, to provide more effective mechanical support and to reduce the risk of implant rupture and vascular injury. MATERIAL AND METHODS Two human anatomical marks, the patellar ligament and semimembranosus, were connected and divided into four parts to identify points Ⅰ, Ⅱ, and Ⅲ. These points determined the areas for Tomofix placement: anterior, anterolateral, and lateral. Simulated internal fixation placed hole B of Tomofix at points Ⅰ (anterior), Ⅱ (anterolateral), and Ⅲ (lateral). We analyzed the pointing direction of the locking screws in Tomofix holes on MRI to assess potential injury risk to the popliteal neurovascular bundle. RESULTS In the X-ray: holes B and C appeared as the plate in the anterior, only hole C appeared as the plate in the anterolateral, and none of the holes appeared as the plate in the lateral. In the general view of the sawbones, the screw pointed towards the popliteal neurovascular bundle when the plate was in the anterior. CONCLUSION If a small number of holes on the plate is visible under fluoroscopy, then several lateral positions of the plate can be obtained; the direction of the screw tunnel tends to deviate from the popliteal neurovascular bundle with the posterior position of the plate.
Collapse
Affiliation(s)
- Zihao Zou
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| | - Zhanyu Wu
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| | - Daizhu Yuan
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| | - Shunen Xu
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| | - Long Yang
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| | - Chuan Ye
- Department of Orthopaedics, The Affiliated Hospital of Guizhou Medical University, Guiyang, PR China
- Center for Tissue Engineering and Stem Cell Research, Guizhou Medical University, Guiyang, PR China
- National-Local Joint Engineering Laboratory of Cell Engineering and Biomedicine, Guiyang, PR China
| |
Collapse
|
6
|
Ando J, Takahashi T, Matsumura T, Nibe Y, Takeshita K. Biomechanical comparisons of plate placement for medial tibial plateau fractures (Schatzker type IV): A biomechanical study using porcine tibias. Injury 2024; 55:111158. [PMID: 38579154 DOI: 10.1016/j.injury.2023.111158] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/03/2023] [Revised: 10/22/2023] [Accepted: 10/23/2023] [Indexed: 04/07/2024]
Abstract
INTRODUCTION Biomechanical studies on medial tibial plateau fractures (MTPFs) (Schatzker classification type IV) are currently few, while studies using locking plates (LPs) placed on medial proximal tibias are unavailable. Hence, we compared the biomechanical properties of plate osteosynthesis at the medial and anteromedial placements using large- and small-fragment LPs in porcine bones. MATERIALS AND METHODS MTPFs were internally fixed using LPs on 40 porcine tibias. Specimens were equally divided into four groups: medial placement using a large-fragment LP (LPs for the medial (LM) group), anteromedial placement using a large-fragment LP (LAM group), medial placement using a small-fragment LP (SM group), and anteromedial placement using a small-fragment LP (SAM group). The translation patterns of the constructs in each group were examined by cycling loading test (displacement and translation along the mechanical axis at 10-100, 100-500, 500-1000, 1000-1500, and 1500-2000 cycles). Then, articular gaps and step-off changes after 2000 cycles were compared among the four groups. RESULTS One-way analysis of variance (ANOVA) revealed no significant differences in displacement and translation during cyclic loading. One-way ANOVA followed by post hoc analysis revealed that the anterior gap was lower in LPs for the medial (LM) than in SM (P = 0.029) and SAM (P = 0.0026). The central gap was also lower in LM than in SM (P = 0.042) and SAM (P < 0.001), and it was lower in LAM than in SAM (P = 0.047). Likewise, the posterior gap was lower in LM than in LAM (P = 0.025) and SAM (P < 0.001). Furthermore, the central step-off of SAM was higher than that of LM, LAM, and SM (P < 0.001, P = 0.0014, and P = 0.0077, respectively). The posterior step-off was lower in LM than in SAM and LAM (P = 0.037 and P < 0.001), and it was also lower in SM than in SAM (P = 0.0082). CONCLUSION Medial LP placement for MTPFs in porcine bones resulted in significantly lower posterior step-offs after cyclic loading than anteromedial placement, and large-fragment LPs for MTPFs caused significantly lower fracture gaps in the central articular after cyclic loading than small-fragment LPs.
Collapse
Affiliation(s)
- Jiro Ando
- Department of Orthopedics, School of Medicine, Jichi Medical University, Shimotsuke, Japan
| | - Tsuneari Takahashi
- Department of Orthopedic Surgery, Ishibashi General Hospital, Shimotsuke, 1-15-4, Shimotsuke 329-0502, Japan.
| | - Tomohiro Matsumura
- Department of Emergency and Critical Care Medicine, Jichi Medical University, Shimotsuke, Japan
| | - Yoshiya Nibe
- Department of Orthopedics, School of Medicine, Jichi Medical University, Shimotsuke, Japan
| | - Katsushi Takeshita
- Department of Orthopedics, School of Medicine, Jichi Medical University, Shimotsuke, Japan
| |
Collapse
|
7
|
Palmer J, Getgood A, Lobenhoffer P, Nakamura R, Monk P. Medial opening wedge high tibial osteotomy for the treatment of medial unicompartmental knee osteoarthritis: A state-of-the-art review. J ISAKOS 2024; 9:39-52. [PMID: 37839705 DOI: 10.1016/j.jisako.2023.10.004] [Citation(s) in RCA: 4] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/31/2023] [Revised: 07/13/2023] [Accepted: 10/08/2023] [Indexed: 10/17/2023]
Abstract
Medial unicompartmental knee osteoarthritis is a common condition that is frequently associated with significant pain and dysfunction. Medial opening wedge high tibial osteotomy (MOWHTO) offers a unique opportunity to preserve the knee joint and potentially alter the course of the degenerative process. Recent advances in this field of surgery have enabled surgeons to perform a MOWHTO in a safe, reliable and reproducible manner. This state-of-the-art review highlights the most important advances in the field of MOWHTO. Key concepts related to patient selection, pre-operative planning, surgical accuracy and patient outcome are considered. The importance of an individualized approach is emphasized and its influence on the future direction of the procedure is discussed.
Collapse
Affiliation(s)
| | - Alan Getgood
- Fowler Kennedy Sport Medicine Clinic, University of Western Ontario, London, Ontario, Canada, N6A 3K7
| | | | - Ryuichi Nakamura
- Joint Preservation and Sports Orthopaedic Center, Harue Hospital, Sakai, 919-0476, Japan
| | - Paul Monk
- Unisports Orthopaedics, Auckland, 1072, New Zealand; Department of Trauma and Orthopaedics, Auckland City Hospital, Auckland, 1023, New Zealand.
| |
Collapse
|
8
|
Nibe Y, Takahashi T, Kubo T, Matsumura T, Takeshita K. Effect of plate position on tibial displacement and posterior tibial slope after cyclic loading in medial open wedge high tibial osteotomy: A biomechanical study using porcine tibia. Clin Biomech (Bristol, Avon) 2023; 109:106076. [PMID: 37634465 DOI: 10.1016/j.clinbiomech.2023.106076] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/19/2022] [Revised: 08/15/2023] [Accepted: 08/21/2023] [Indexed: 08/29/2023]
Abstract
BACKGROUND Medial open wedge high tibial osteotomy is currently the most common osteotomy, but its complication is an increased posterior tibial slope over time. However, no study has clarified whether the plate position of medial open wedge high tibial osteotomy could reduce the increase in posterior tibial slope after cyclic loading. METHODS Fourteen porcine bones were biomechanically evaluated by performing medial open wedge high tibial osteotomy and placing TOMOFIX medially and anteromedially. Cyclic testing was performed to investigate the posterior tibial slope over time for medial open wedge high tibial osteotomy with medial or anteromedial plate. The displacement along the mechanical axis during cyclic testing from 10 to 100th, 100-500th, 500-1000th, 1000-1500th, and 1500-2000th cycles, and changes in anterior and posterior gaps after 2000 cycles were compared between plate position. FINDINGS There were no significant differences in displacement. A significant difference was found in posterior gap changes (-0.20 ± 0.84 mm in group of using medial plate, 1.07 ± 0.82 mm in group of using anteromedial plate) (P = 0.014), but none was found in anterior gap changes. INTERPRETATION Medial plate placement in medial open wedge high tibial osteotomy resulted in significantly less increased posterior tibial slope than anteromedial plate placement after cyclic loading.
Collapse
Affiliation(s)
- Yoshiya Nibe
- Department of Orthopaedis, School of Medicine, Jichi Medical University, Shimotsuke, Japan
| | - Tsuneari Takahashi
- Department of Orthopaedic Surgery, Ishibashi General Hospital, Shimotsuke, Japan.
| | - Tatsuya Kubo
- Department of Orthopaedic Surgery, Shin-Oyama City Hospital, Oyama, Japan
| | - Tomohiro Matsumura
- Department of Emergency and Critical Care Medicine, Jichi Medical University, Shimotsuke, Japan
| | - Katsushi Takeshita
- Department of Orthopaedis, School of Medicine, Jichi Medical University, Shimotsuke, Japan
| |
Collapse
|
9
|
Mochizuki H, Yoshioka T, Kikuchi N, Yamazaki M. Bilateral Knee Osteoarthritis Treated With Medial Open-Wedge High Tibial Osteotomy Using Two Types of β-Tricalcium Phosphate With Differing Placements in Each Knee: A Report of Two Cases. Cureus 2023; 15:e45427. [PMID: 37859897 PMCID: PMC10581886 DOI: 10.7759/cureus.45427] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 09/17/2023] [Indexed: 10/21/2023] Open
Abstract
In medial open-wedge high tibial osteotomy (MOWHTO) for knee osteoarthritis, synthetic bone is commonly used as a replacement material for the opening gap. Unidirectional porous β-tricalcium phosphate (UDPTCP) and spherical porous β-tricalcium phosphate (SPTCP) have been widely used in this regard. In general, the two prostheses are placed parallel to the osteotomy opening gap. In this report, we discuss two cases involving a 63-year-old woman and a 51-year-old man who underwent MOWHTO for bilateral knee osteoarthritis. Both patients had experienced bilateral knee pain. In both patients, UDPTCP was placed anteriorly and SPTCP was placed posteriorly in one knee, with the placement reversed in the other knee. The remodeling of each type of β-TCP was evaluated using CT immediately after the surgery and one year postoperatively. The postoperative corrective loss and clinical outcomes were also evaluated. Remodeling with β-TCP was found to be faster with UDPTCP than with SPTCP, even though the anteroposterior placement differed laterally in each patient. Furthermore, there was no correction loss, and the clinical outcomes were comparable, regardless of the placement of β-TCP.
Collapse
Affiliation(s)
- Hiromi Mochizuki
- Department of Orthopaedic Surgery, Tsukuba Central Hospital, Ushiku, JPN
| | - Tomokazu Yoshioka
- Division of Regenerative Medicine for Musculoskeletal System, Institute of Medicine, University of Tsukuba, Tsukuba, JPN
| | - Naoya Kikuchi
- Department of Orthopaedic Surgery, Institute of Medicine, University of Tsukuba, Tsukuba, JPN
| | - Masashi Yamazaki
- Department of Orthopaedic Surgery, Institute of Medicine, University of Tsukuba, Tsukuba, JPN
| |
Collapse
|
10
|
Nakamura R, Komatsu N, Fujita K, Kuroda K, Takahashi M, Okano A, Katsuki Y. Anteromedial plating without filling the gap in open wedge high tibial osteotomy may increase the risk of screw breakage, which can be reduced by medial plating and bone-substitute insertion. J Orthop Sci 2023; 28:416-425. [PMID: 35074292 DOI: 10.1016/j.jos.2021.12.017] [Citation(s) in RCA: 4] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/28/2021] [Revised: 11/12/2021] [Accepted: 12/22/2021] [Indexed: 11/30/2022]
Abstract
BACKGROUND Open wedge high tibial osteotomy (OWHTO) brings new complications such as screw breakages with or without correction loss and time-dependent increased posterior tibial slopes (PTS) due to the opening gap. For preventive purposes, we changed our OWHTO procedure from anteromedial plating without filling the gap (non-grafted group, n = 40, 2009-2012) to medial plating with bone-substitute insertion (grafted group, n = 45, 2012-2015). The objectives of this study were to evaluate the complication patterns and the effects of the modifications. METHODS Patients undergoing OWHTO with TomoFix plates were included in this retrospective study. Demographics, clinical outcomes (flexion range and Japanese Orthopaedic Association score) and radiological outcomes (femorotibial angle) were assessed in both groups pre-operatively and 2-years postoperatively. The plate installation angle [PIA] and screw insertion depth [SID] were measured on computed tomographic slices at 6 months. PIA/SID was defined as the angle between the tibial anteroposterior axis and plate-width axis/the distance between the proximal screw tip and the proximal tibiofibular joint. The non-grafted group was further divided into complication and non-complication subgroups. Screw breakages were assessed during plate removal (1.5-2.5 years postoperatively). RESULTS There were no differences in baseline characteristics or radiological/clinical outcomes between the non-grafted and grafted groups. There were 0 and 11 complications in the grafted and non-grafted groups, respectively. Complications included 7 screw breakages, 4 correction losses, and 5 time-dependent increased PTS with some overlaps. PIA and SID were significantly lower (p < 0.001) and higher (p < 0.001), respectively, in the grafted group and significantly lower (p = 0.018) and higher (p = 0.040), respectively, in the non-complication subgroup within the non-grafted group. The cutoff value of PIA for complications was calculated at 48.0°. CONCLUSIONS Medial plating OWHTO (PIA<48°) using bone-substitute with deeper screw insertion reinforces the opening gap for better angular stability compared with anteromedial plating without bone-substitute.
Collapse
Affiliation(s)
- Ryuichi Nakamura
- Joint Preservation and Sports Orthopaedic Center, Harue Hospital, Sakai, Japan; Department of Orthopaedic Surgery, Yawata Medical Center, Komatsu, Japan.
| | - Nahoko Komatsu
- Department of Dermatology, Yawata Medical Center, Komatsu, Japan
| | - Kenji Fujita
- Department of Orthopaedic Surgery, Yawata Medical Center, Komatsu, Japan
| | - Kazunari Kuroda
- Department of Orthopaedic Surgery, Yawata Medical Center, Komatsu, Japan
| | - Masaki Takahashi
- Department of Orthopaedic Surgery, Yawata Medical Center, Komatsu, Japan
| | - Akira Okano
- Joint Preservation and Sports Orthopaedic Center, Harue Hospital, Sakai, Japan
| | - Yasuo Katsuki
- Department of Orthopaedic Surgery, Yawata Medical Center, Komatsu, Japan
| |
Collapse
|
11
|
Ryu DJ, Park SJ, Lee DH, Kwon KB, Choi GH, Kim IS, Wang JH. Does the anteromedial plate position affect proximal screw length and worsen the clinical outcomes in medial opening wedge high tibial osteotomy? BMC Musculoskelet Disord 2023; 24:14. [PMID: 36611141 PMCID: PMC9824977 DOI: 10.1186/s12891-022-06080-4] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/10/2022] [Accepted: 12/12/2022] [Indexed: 01/08/2023] Open
Abstract
BACKGROUND During medial opening wedge high tibial osteotomy (MOWHTO), sometimes the plate tends to be positioned anteromedially. The plate position can affect the length of the proximal screw, which significantly affects stability after osteotomy. Therefore, research on the correlation among plate position, screw length, and clinical outcomes is needed. METHODS This retrospective review examines 196 knees in 175 patients who underwent MOWHTO from May 2012 to December 2018, for symptomatic medial compartment osteoarthritis with a varus alignment of > 5°. We evaluated the anteroposterior plate position, length of proximal screw, and postoperative computed tomography (CT). We reviewed patients' clinical outcome scores, presence of lateral hinge fracture, neurovascular complications, and infection. The correlation among proximal plate position, proximal screw length, and clinical outcomes was evaluated using Pearson's correlation analysis. A subgroup analysis by screw angle (> 48 ° or < 48 °) was also performed using chi-square test and Student t-test. RESULTS The mean proximal plate position was 16.28% (range, 5.17-44.74) of the proximal tibia's anterior-to-posterior distance ratio, and the proximal screw length averaged 63.8 mm (range, 44-80 mm). Proximal posteromedial plate position and proximal screw length were significantly correlated (r2 = 0.667, P < .001), as were screw angle and length (r2 = 0.746, P < .001). Medial plating (< 48°) can use a longer proximal screw; nevertheless, no significant difference occurred in clinical outcomes between the two groups. Also, no differences occurred in complication rate, including hinge fracture. CONCLUSION With more medially positioned plating during MOWHTO, we can use longer proximal screws. However, there was no significant difference in clinical outcomes and the incidence of lateral hinge fractures regardless of plate position and screw length.
Collapse
Affiliation(s)
- Dong Jin Ryu
- Department of Orthopaedic Surgery, Inha University Hospital, Inha University school of Medicine, Incheon, South Korea
| | - Sang Jun Park
- Department of Orthopaedic Surgery, Cheongju Micro Hospital, Cheongju-si, Chung-cheong bukdo South Korea
| | - Dae Hee Lee
- grid.264381.a0000 0001 2181 989XDepartment of Orthopaedic Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, 06351 South Korea
| | - Kyeu-Back Kwon
- Department of Orthopaedic Surgery, Samsung Bone Hospital, Osan-si, Gyeonggi-do South Korea
| | - Geun Hong Choi
- Department of Orthopaedic Surgery, Samsung Maditop hospital, Sungnam-si, Gyeonggi-do South Korea
| | - Il Su Kim
- grid.264381.a0000 0001 2181 989XDepartment of Orthopaedic Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, 06351 South Korea
| | - Joon Ho Wang
- grid.264381.a0000 0001 2181 989XDepartment of Orthopaedic Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, 06351 South Korea ,grid.264381.a0000 0001 2181 989XDepartment of Health Sciences and Technology, SAIHST, Sungkyunkwan University, Seoul, South Korea ,grid.264381.a0000 0001 2181 989XDepartment of Medical Device Management and Research, SAIHST, Sungkyunkwan University, Seoul, South Korea
| |
Collapse
|
12
|
Forgivingness of an Anteromedially Positioned Small Locked Plate for High Tibial Osteotomy in Case of Overcorrection and Lateral Hinge Fracture. LIFE (BASEL, SWITZERLAND) 2022; 12:life12081265. [PMID: 36013443 PMCID: PMC9409794 DOI: 10.3390/life12081265] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 07/21/2022] [Revised: 08/16/2022] [Accepted: 08/17/2022] [Indexed: 11/17/2022]
Abstract
High tibial osteotomy (HTO) represents a sensible treatment option for patients with moderate unicondylar osteoarthritis of the knee and extraarticular malalignment. The possibility of a continuously variable correction setting and a surgical approach low in complications has meant that the medial opening osteotomy has prevailed over the past decades. The objective of the present study was to determine whether anteromedially positioned small plates are nevertheless forgiving under biomechanically unfavourable conditions (overcorrection and lateral hinge fracture). In this study, a simulated HTO was performed on composite tibiae with a 10-mm wedge and fixed-angle anteromedial osteosynthesis with a small implant. Force was applied axially in a neutral mechanical axis, a slight and a marked overcorrection into valgus, with and without a lateral hinge fracture in each case. At the same time, a physiological gait with a dual-peak force profile and a peak load of 2.4 kN was simulated. Interfragmentary motion and rigidity were determined. The rigidity of the osteosynthesis increased over the cycles investigated. A slight overcorrection into valgus led to the lowest interfragmentary motion, compared with pronounced valgisation and neutral alignment. A lateral hinge fracture led to a significant decrease in rigidity and increase in interfragmentary motion. However, in no case was the limit of 1 mm interfragmentary motion critical for osteotomy healing exceeded. The degree of correction of the leg axis, and the presence of a lateral hinge fracture, have an influence on rigidity and interfragmentary motion. From a mechanically neutral axis ranging up to pronounced overcorrection, the implant investigated offers sufficient stability to allow healing of the osteotomy, even if a lateral hinge fracture is present.
Collapse
|
13
|
Comparison of Anatomical Conformity between TomoFix Anatomical Plate and TomoFix Conventional Plate in Open-Wedge High Tibial Osteotomy. MEDICINA (KAUNAS, LITHUANIA) 2022; 58:medicina58081045. [PMID: 36013511 PMCID: PMC9413536 DOI: 10.3390/medicina58081045] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 06/24/2022] [Revised: 07/20/2022] [Accepted: 08/01/2022] [Indexed: 11/18/2022]
Abstract
Background and Objectives: The TomoFix anatomical plate was developed to improve plate position, proximal screw direction, and post-correction tibial contouring. The purpose of this study was to compare postoperative configurations between the TomoFix anatomical plate and the TomoFix conventional plate. It was hypothesized that the new modified plate provides a better fixative coaptation than the conventional plate. Materials and Methods: A total of 116 cases (112 patients) were enrolled in this study from March 2015 to February 2021. Among them, 63 patients underwent surgery using the TomoFix conventional plate, and 53 underwent surgery using the TomoFix anatomical plate. The radiographic outcomes, including the hip−knee−ankle (HKA) angle, medial proximal tibial angle (MPTA), tibial slope, plate angle, proximal screw angles, and plate-to-cortex distance at #1 hole (just below the osteotomy site) were compared between the two groups. Results: Patients with the TomoFix anatomical plate showed similar results in terms of the pre- and postoperative HKA angle, MPTA, and tibial slope. The TomoFix anatomical group showed a significantly greater plate angle (39.2° ± 8.1° vs. 31.7° ± 7.0°, p < 0.001) and less screw angles, indicating that the TomoFix anatomical plates allowed a more posterior plate position than the conventional plate. The plate-to-cortex distance was significantly less in the TomoFix anatomical group than in the TomoFix conventional group (p < 0.001). Conclusion: The TomoFix anatomical plate showed a more posteromedial plating position, better proximal screw direction to the lateral hinge, and improved post-correction tibial contour compared to the TomoFix conventional plate.
Collapse
|
14
|
Combined use of beta-tricalcium phosphate with different porosities can accelerate bone remodelling in open-wedge high tibial osteotomy. Asia Pac J Sports Med Arthrosc Rehabil Technol 2022; 29:30-34. [PMID: 35847191 PMCID: PMC9262697 DOI: 10.1016/j.asmart.2022.05.004] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/10/2022] [Revised: 04/22/2022] [Accepted: 05/24/2022] [Indexed: 11/04/2022] Open
Abstract
Background/Objective Beta-tricalcium phosphate (β-TCP) is often used as a gap filler in open-wedge high tibial osteotomy (OWHTO). The aim of the present study was to investigate the effects of using β-TCP with different porosities on bone remodelling after OWHTO.Methods: We evaluated 29 knees in 26 patients that underwent OWHTO using β-TCP with porosities of 60% and 75% (combined group). A further 30 knees in 28 patients that underwent OWHTO using β-TCP with 60% porosity alone were allocated as a control group. In the combined group, a β-TCP block with 75% porosity was inserted into the gap at the cancellous bone site and a β-TCP block with 60% porosity was inserted into the medial cortical bone side. In the control group, a β-TCP block with 60% porosity was inserted into the osteotomy gap. The bone remodelling phases of the inserted β-TCP blocks were evaluated on standard anteroposterior radiographs using the modified van Hemert classification at 3 and 6 months post-operatively. Results The rate of satisfactory bone remodelling at the cancellous bone sites was 86.2% (25/29) in the combined group and 0% (0/30) in the control group at 3 months post-operatively (p<0.05), progressing to 96.6% (28/29) in the combined group and 20% (6/30) in the control group at 6 months post-operatively (p<0.05). Conclusion The present study demonstrated that combined use of β-TCP with high and low porosities can significantly enhance bone formation. The combined use of artificial bones with different porosities is useful for early bone remodelling in OWHTO.
Collapse
|
15
|
Ramos Marques N, Morais B, Barreira M, Nóbrega J, Ferrão A, Torrinha Jorge J. Anterior Slope Correction-Flexion Osteotomy in Traumatic Genu Recurvatum. Arthrosc Tech 2022; 11:e889-e893. [PMID: 35646563 PMCID: PMC9134481 DOI: 10.1016/j.eats.2022.01.007] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 02/03/2023] Open
Abstract
A decreased posterior tibial slope has been associated with an increased risk of posterior cruciate ligament failure, anterior knee pain, and premature knee osteoarthritis. Trauma is a common cause of osseous genu recurvatum. Surgical management is recommended to correct the tibial slope and prevent knee pain and osteoarthritis progression. This article discusses our preferred treatment using a proximal tibial opening-wedge osteotomy for surgical management of genu recurvatum secondary to significant anterior tibial slope.
Collapse
Affiliation(s)
- Nuno Ramos Marques
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| | - Bruno Morais
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| | - Mariana Barreira
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| | - João Nóbrega
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| | - Ana Ferrão
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| | - João Torrinha Jorge
- Deparment of Orthopedic Surgery, Hospital Curry Cabral, Centro Hospitalar Universitário Lisboa Central, Lisboa, Portugal
| |
Collapse
|
16
|
Nejima S, Kumagai K, Yamada S, Sotozawa M, Kumagai D, Yamane H, Inaba Y. Risk of interference between the tibial tunnel and locking screws in medial meniscus posterior root repair and open wedge high tibial osteotomy. J Exp Orthop 2022; 9:25. [PMID: 35292866 PMCID: PMC8924329 DOI: 10.1186/s40634-022-00464-0] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/07/2021] [Accepted: 03/04/2022] [Indexed: 11/10/2022] Open
Abstract
PURPOSE To evaluate whether the frequency of interference between locking screws for the plate fixation and tibial tunnels differs depending on the tibial tunnel positions in a surgical simulation of the transtibial pull-out repair of medial meniscus posterior root tears (MMPRTs) in patients undergoing biplanar open wedge high tibial osteotomy (OWHTO). METHODS Sixty-five patients (75 knees) who underwent OWHTO with TomoFix small plate (Depuy Synthes, PA, USA) for medial knee osteoarthritis with varus malalignment were enrolled in this study. Surgical simulation of transtibial pull-out repair of MMPRTs was performed using postoperative computed tomography images. The tibial tunnel was created in the anatomical attachment area of the medial meniscus posterior root. Another aperture of the tibial tunnel was created on the anteromedial (AM) tibial cortex, the posteromedial (PM) tibial cortex, and the anterolateral (AL) tibial cortex in the proximal tibial fragment. The frequency of interference between the tibial tunnel and A-D locking screws was compared in the 3 tibial tunnel positions. In each tibial tunnel position, the locking plate position with and without interference between the tibial tunnel and at least one locking screw was compared. RESULTS For screw A, the frequency of interference with the tibial tunnel in the AL position was higher than that in the AM (P = 0.048) and PM positions (P < 0.001). For screws B and C, the frequency of interference with the tibial tunnel in the AM position was higher than that in the PM (P < 0.001, P = 0.007) and AL positions (P < 0.001, P = 0.001), respectively. For screw D, there was no difference in the frequency of interference with the tibial tunnel among the three positions. The frequency of interference between the tibial tunnel and at least one screw in the AM position was 100%, which was higher than that in the PM (P < 0.001) and AL positions (P < 0.001). In the PM position, the locking plate was placed more posteriorly in the group where the locking screw interfered with the tibial tunnel. In the AL position, the locking plate was placed more parallel to the medial/lateral axis of the tibial plateau in the interference group. CONCLUSION Making the tibial tunnel in the AM position should be avoided because interference with locking screws was inevitable. When the tibial tunnel is created in the PM position, interference between the tibial tunnel and screw C should be paid attention. Anterior placement of the locking plate could be useful to prevent interference between locking screws and the tibial tunnel in the PM position. In addition, when the tibial tunnel is created in the AL position, interference between the tibial tunnel and especially screw A among screws A-C should be paid attention. Placing the locking plate in an anteromedial direction could be useful to prevent interference between locking screws and the tibial tunnel in the AL position. LEVEL OF EVIDENCE IV.
Collapse
Affiliation(s)
- Shuntaro Nejima
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| | - Ken Kumagai
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan.
| | - Shunsuke Yamada
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| | - Masaichi Sotozawa
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| | - Dan Kumagai
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| | - Hironori Yamane
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| | - Yutaka Inaba
- Department of Orthopaedic Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama, 236-0004, Japan
| |
Collapse
|
17
|
Kim JH, Jung WH, Jeon SS, Kim JH. Combination of Cylindrical Autologous Bone Grafting Technique With a Metallic Block Insertion in Open-Wedge High Tibial Osteotomy. Arthrosc Tech 2021; 10:e367-e373. [PMID: 33680768 PMCID: PMC7917032 DOI: 10.1016/j.eats.2020.10.012] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/09/2020] [Accepted: 10/09/2020] [Indexed: 02/03/2023] Open
Abstract
Open-wedge high tibial osteotomy (OW-HTO) is an effective surgical intervention for medial-compartment knee osteoarthritis. However, the osteotomized gap might be a disadvantage in OW-HTO because it can cause problems such as delayed bone union or loss of correction. These issues can be minimized by using autologous bone graft in the osteotomized gap, which is known to be the fastest and most clinically satisfactory gap filler. The primary mechanical stability of the osteotomy site in OW-HTO is essential for early weight bearing after surgery. Therefore, we introduce the combination of a cylindrical autologous bone grafting technique and a metallic block insertion for faster bone union and better primary stability of the site in OW-HTO. We expect that the described procedure will enable early postoperative weight bearing and, thereby, allow an early return to normal function.
Collapse
Affiliation(s)
- Jong Hyun Kim
- Samsung Orthopaedic Clinic, Chungju, Republic of Korea
| | - Woon Hwa Jung
- Department of Orthopaedic Surgery, Murup Hospital, Masan, Republic of Korea
| | - Seung Soo Jeon
- Department of Orthopaedic Surgery, Koggiry Hue Hospital, Gwangju, Republic of Korea
| | - Jae Hyoung Kim
- Department of Orthopaedic Surgery, Woori Hospital, Suwon, Republic of Korea,Address correspondence to Jong Hyun Kim, M.D., Ph.D., Samsung Orthopaedic Clinic, 33, Jungang-ro, Chungju-si, Chungcheongbuk-do 27406, Republic of Korea.
| |
Collapse
|
18
|
Solomin LN, Chugaev DV, Filippova AV, Kulesh PN. High Tibial Osteotomy for Genu Varum in Adults: Do Proprietary Implants Limit the Quality of Correction? Strategies Trauma Limb Reconstr 2020; 15:13-22. [PMID: 33363636 PMCID: PMC7744664 DOI: 10.5005/jp-journals-10080-1449] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/23/2022] Open
Abstract
The surgical technique of proximal tibial osteotomy for genu varum in adults has evolved from a procedure using closing wedges of estimated sizes with staple fixation in the 1960s to using standard trauma internal fixation implants and, more recently, to gradual correction with software-guided hexapod external fixators. In the last two decades, implant manufacturers have also produced anatomical implants specific for such corrective osteotomies. This study evaluates the limits of using such proprietary implants for proximal tibial osteotomy in genu varum. Materials and methods Scanograms (teleradiograms) of lower limbs of a patient were used to derive skiagrams (two-dimensional bony outlines of the extremities). From these, two-dimensional and three-dimensional models of varus deformities of the tibia with different values of mechanical medial proximal tibial angle (mMPTA, from 85° to 40°) were created. An analysis of the created deformity was carried out and a simulation for surgical correction was performed using an open wedge high tibial osteotomy with fixation using a proprietary (Tomofix, Synthes) implant. In addition, a 3D simulation technique was used to check the accuracy of the results obtained from the 2D simulation. Results In cases of mMPTA ≥80° with localisation of the apex of varus deformity at the level of the knee joint line, the standard technique used with the proprietary medial tibial plate produces good results.In cases of mMPTA ≤70°, fixation of the osteotomised fragments by the proprietary medial plate is poor owing to the anatomical contours of the implant. In these cases, a different type of osteosynthesis is needed.In cases of mMPTA ≤70°, the distance between the lower edge of the bone plate and the medial surface of the tibia after a proximal tibial osteotomy exceeds 11 mm and will result in unacceptable soft tissue tension around the implant.Mechanical axis deviation to the Fujisawa point produces mMPTA values outside the reference range of normal values. Conclusion An osteotomy of the proximal tibia using a prescribed technique linked to a proprietary implant achieves good results only if performed within a certain range of deformity values. Pronounced varus deformities require a fundamentally different approach. This study reveals that surgeons undertaking corrective proximal tibial osteotomies for genu varum need to perform a comprehensive analysis of the deformity to allow for appropriate selection of patients. This will enable a consideration of the size and other characteristics of the deformity that will reduce the technical complications that may arise if the correction was performed using the recommended technique linked to a proprietary implant. How to cite this article Solomin LN, Chugaev DV, Filippova AV, e t a l. High Tibial Osteotomy for Genu Varum in Adults: Do Proprietary Implants Limit the Quality of Correction? Strategies Trauma Limb Reconstr 2020;15(1):13-22.
Collapse
Affiliation(s)
- L N Solomin
- National Medical Research Center of Traumatology and Orthopedics, Saint Petersburg State University, Saint Petersburg, Russia, Russian Federation
| | - D V Chugaev
- National Medical Research Center of Traumatology and Orthopedics, Saint Petersburg State University, Saint Petersburg, Russia, Russian Federation
| | - A V Filippova
- Department Saint Petersburg State University, Saint Petersburg, Russia, Russian Federation
| | - P N Kulesh
- St. George City Hospital, Saint Petersburg, Russia, Russian Federation
| |
Collapse
|
19
|
Ogawa H, Matsumoto K, Sengoku M, Yoshioka H, Yamamoto K, Shimokawa T, Ohnishi K, Akiyama H. Clinical course and outcomes of simultaneous-versus staged-bilateral medial opening wedge high tibial osteotomy. ASIA-PACIFIC JOURNAL OF SPORT MEDICINE ARTHROSCOPY REHABILITATION AND TECHNOLOGY 2020; 23:13-17. [PMID: 33344173 PMCID: PMC7725662 DOI: 10.1016/j.asmart.2020.11.003] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 07/22/2020] [Revised: 11/12/2020] [Accepted: 11/26/2020] [Indexed: 11/08/2022]
Abstract
Background Difference in the clinical course and outcomes between simultaneous- and staged-bilateral medial opening wedge high tibial osteotomies (OWHTOs) over time was unknown. The study hypothesis was that patients who underwent simultaneous-bilateral OWHTO (SMBO) have a more rapid improvement in knee function than those who underwent staged-bilateral OWHTO (STBO) due to difference in the change of lower limb alignment between SMBO and STBO. Methods The records of 56 knees in 28 patients who underwent either SMBO (n = 28) or STBO (n = 28) were retrospectively analysed. The time course data of weight-bearing line percentage (%WBL), joint line convergence angle (JLCA), and Knee Society Score were compared between the two procedures. Results Hospitalisation for SMBO was longer than that for STBO by 1 week. No significant difference was observed in %WBL between the two procedures. The JLCA was significantly lower with SMBO than with the first-stage surgery of STBO (P < 0.05), but it became equivalent in both groups at the last follow-up. The knee scores in both SMBO and the first-stage surgery of STBO significantly improved in approximately 1 year. The function scores in the first-stage surgery of STBO did not significantly improve until the completion of the second-stage surgery whereas those in SMBO significantly improved 1 year after surgery and become stable. The function score 1 year after surgery was significantly higher in SMBO than in the first-stage surgery of STBO (p < 0.001). Conclusions Although both SMBO and STBO achieved the desired therapeutic results, SMBO led to earlier functional improvement and decreased JLCA compared with STBO.
Collapse
Affiliation(s)
- Hiroyasu Ogawa
- Department of Orthopaedic Surgery, Ogaki Tokushukai Hospital. Hayashi-machi 6-85-1, Ogaki, 503-0015, Japan.,Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine. Yanagido 1-1, Gifu, 501-1194, Japan
| | - Kazu Matsumoto
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine. Yanagido 1-1, Gifu, 501-1194, Japan
| | - Masaya Sengoku
- Department of Orthopaedic Surgery, Ogaki Tokushukai Hospital. Hayashi-machi 6-85-1, Ogaki, 503-0015, Japan
| | - Hiroki Yoshioka
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine. Yanagido 1-1, Gifu, 501-1194, Japan
| | - Kyosuke Yamamoto
- Department of Orthopaedic Surgery, Ogaki Tokushukai Hospital. Hayashi-machi 6-85-1, Ogaki, 503-0015, Japan
| | - Tetsuya Shimokawa
- Department of Orthopaedic Surgery, Ogaki Tokushukai Hospital. Hayashi-machi 6-85-1, Ogaki, 503-0015, Japan
| | - Kazuichiro Ohnishi
- Department of Orthopaedic Surgery, Ogaki Tokushukai Hospital. Hayashi-machi 6-85-1, Ogaki, 503-0015, Japan
| | - Haruhiko Akiyama
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine. Yanagido 1-1, Gifu, 501-1194, Japan
| |
Collapse
|
20
|
Comparative analysis of fixation configurations and their effect on outcome after medial open-wedge high tibial osteotomy. J Orthop Sci 2020; 25:627-634. [PMID: 31320145 DOI: 10.1016/j.jos.2019.06.018] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/31/2019] [Revised: 06/18/2019] [Accepted: 06/27/2019] [Indexed: 11/22/2022]
Abstract
PURPOSE The purposes of this study were to analyze and compare 1) the postoperative fixation configurations, and 2) radiological and clinical outcomes between the new conceptual fixation device (LCfit) and the conventional locking plates. METHODS Retrospective comparative analysis of 304 cases who underwent open-wedge high tibial osteotomy with four different locking plate systems including the LCfit system was conducted. In analyzing the fixation configuration of each locking plate system, the plate position, the gap between plate and bone, screw angle, and screw length were evaluated from postoperative CT scans. The clinical outcomes and radiological parameters of four patient groups corresponding to the four locking plate systems were also compared. RESULTS In the analysis of the fixation configurations, the LCfit system showed more posteromedial plate position, decreased gap between plate and bone, reduced screw angle, and longer screw length compared to the other three locking plates. However, there were no significant differences in clinical outcomes and radiological parameters such as hip-knee-ankle angle, weight-bearing line position, medial proximal tibial angle, and posterior slope between the four groups. CONCLUSION The newly designed fixation system (LCfit) for open-wedge high tibial osteotomy showed fixation configurations that have theoretical mechanical advantages over conventional anatomical locking plates. However, there were no significant differences in clinical outcomes and radiological parameters between LCfit and the other locking plate systems. LEVEL OF EVIDENCE Level III, retrospective cohort study.
Collapse
|
21
|
Goshima K, Sawaguchi T, Shigemoto K, Iwai S, Fujita K, Kataoka T. Plate removal without loss of correction after open-wedge high tibial osteotomy is possible when posterior cortex bone union reaches osteotomy gap center even in incompletely filled gaps. Knee Surg Sports Traumatol Arthrosc 2020; 28:1827-1834. [PMID: 31273409 DOI: 10.1007/s00167-019-05615-y] [Citation(s) in RCA: 10] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/06/2019] [Accepted: 07/01/2019] [Indexed: 11/27/2022]
Abstract
PURPOSE The study aimed to evaluate the optimal timing for plate removal after open-wedge high tibial osteotomy (OWHTO) without loss of correction and to investigate risk factors for loss of correction after plate removal. The study presents the hypothesis that plate removal without loss of correction was possible when gap filling reached zone 2 (25-50%) on anteroposterior radiographs. METHODS Ninety-one patients (101 knees) who underwent OWHTO using the TomoFix® plate were enrolled. Plate removal was performed at an average 16.4 ± 5.4 months after OWHTO. Clinical evaluation included plate-related symptoms, the Japanese Orthopedic Association Knee Score (JOA score), and Oxford Knee Score (OKS). Radiological outcomes, including the hip-knee-ankle angle (HKA), weight-bearing line ratio (WBLR), medial proximal tibial angle (MPTA), and posterior tibial slope (PTS), were evaluated preoperatively, at plate removal and at 1 year after plate removal. Computed tomography (CT) was performed at plate removal to evaluate the flange bone union, progression rates of gap filling, and posterior cortex bone union. In addition, the risk factors for loss of correction after plate removal were evaluated. RESULTS At plate removal, 63 (62.4%) knees had plate-related symptoms (mild, 56 knees; moderate, 7 knees; severe, 0). After plate removal, the JOA score did not change, whereas OKS further improved; six knees developed loss of correction. On CT evaluation at plate removal, the flange bone union was achieved in all cases; the progression rates of gap filling and posterior cortex bone union were 47.0% ± 16.6% and 62.8% ± 16.5%, respectively. A posterior cortex union rate of < 43.3% was the only predictor for loss of correction after plate removal (odds ratio: 1.38, P < 0.01). CONCLUSIONS Plate removal without loss of correction after OWHTO was possible when bone union of the posterior cortex reached the center of the osteotomy gap even in incompletely filled gaps. LEVEL OF EVIDENCE Therapeutic case series, Level IV.
Collapse
Affiliation(s)
- Kenichi Goshima
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan.
| | - Takeshi Sawaguchi
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan
| | - Kenji Shigemoto
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan
| | - Shintaro Iwai
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan
| | - Kenji Fujita
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan
| | - Tomoyuki Kataoka
- Department of Orthopedic Surgery and Joint Reconstructive Surgery, Toyama Municipal Hospital, 2-1 Imaizumi Hokubu-machi, Toyama, 939-8511, Japan
| |
Collapse
|
22
|
Ogawa H, Matsumoto K, Akiyama H. Effects of lateral opening wedge and medial closing wedge distal femoral osteotomies on axial load stability. Knee 2020; 27:760-766. [PMID: 32563434 DOI: 10.1016/j.knee.2020.04.019] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/15/2020] [Revised: 02/27/2020] [Accepted: 04/21/2020] [Indexed: 02/02/2023]
Abstract
BACKGROUND The stability of the current distal femoral osteotomy is insufficient for early weight bearing and bone healing. The purpose of this study was to test the stability of medial closing wedge- (MCW-) and lateral opening wedge-distal femoral osteotomies (LOW-DFO), which have improved with the recent advances in technology. METHODS We used composite bone models with MCW-DFO, improved with biplanar osteotomy technique and a Tomofix™ MDF plate, or LOW-DFO improved with biplanar osteotomy technique, a Tomofix™ LDF plate, and a β-TCP bone substitute, BONISH® graft. A cyclic axial loading test was employed to evaluate the strain on a plate and hinge point. The breaking axial load was measured. RESULTS Tensile strain on the plate produced by cyclic axial loading was significantly lower for LOW-DFO (2.0 ± 0.8 MPa) than for MCW-DFO (3.9 ± 1.6 MPa, P < .05). Compressive strain on the hinge point produced by cyclic axial loading was lower for LOW-DFO (6.6 ± 2.9 MPa) than for MCW-DFO (7.7 ± 4.6 MPa,). The maximum breaking axial load was significantly higher for LOW-DFO (5511 ± 945 N) than for MCW-DFO (4303 ± 518 N, P < .05). CONCLUSIONS LOW-DFO improved with recent advanced technology was superior to MCW-DFO improved with advanced technology in both cyclic axial loading test and breaking axial load test. This suggests that LOW-DFO facilitates earlier weight bearing and bone healing than does MCW-DFO.
Collapse
Affiliation(s)
- Hiroyasu Ogawa
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine, Gifu, Japan; Department of Advanced Joint Reconstructive Surgery, Gifu University Graduate School of Medicine, Gifu, Japan
| | - Kazu Matsumoto
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine, Gifu, Japan.
| | - Haruhiko Akiyama
- Department of Orthopaedic Surgery, Gifu University Graduate School of Medicine, Gifu, Japan
| |
Collapse
|
23
|
Itou J, Itoh M, Maruki C, Tajimi T, So T, Kuwashima U, Okazaki K. Deep peroneal nerve has a potential risk of injury during open-wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 2020; 28:1372-1379. [PMID: 30834477 DOI: 10.1007/s00167-019-05445-y] [Citation(s) in RCA: 10] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/26/2018] [Accepted: 02/25/2019] [Indexed: 02/02/2023]
Abstract
PURPOSE To assess the risk of injury to the neurovascular bundle on the interosseous membrane of the leg during drilling for distal screw insertion in open-wedge high tibial osteotomy (OWHTO), and to investigate the possible influence of the method of plate placement on the risk. METHODS This retrospective study involved, 55 patients (32 with a TomoFix plate, 23 with a TriS plate) who underwent postoperative CT scanning of the knee following OWHTO between 2009 and 2018. The angle and position of the locking plate, and the direction of screw insertion relative to the interosseous membrane were analysed. RESULTS All distal screws had a risk of neurovascular injury. In particular, 25 screws at the #4 hole (45%) had an extended insertion trajectory that intersected with the interosseous membrane. The angle of the proximal part of the TomoFix plate was a significant risk factor. In contrast, methods of TriS plate placement showed no statistically significant differences. CONCLUSIONS Extended insertion trajectories of distal screws were likely to intersect with the interosseous membrane with the neurovascular bundle potentially on its surface; thus, drilling for bicortical fixation posed a risk of neurovascular injury. The risk increased as the TomoFix plate was placed more medially, suggesting that bicortical drilling must be performed with the utmost attention when the plate is placed at the medial position. Given the particularly high risk at the #3 and #4 screw holes, monocortical fixation of a few distal screws is recommended as long as good stability is ensured.
Collapse
Affiliation(s)
- Junya Itou
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Masafumi Itoh
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Chiyomi Maruki
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Takahiro Tajimi
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Takaaki So
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Umito Kuwashima
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan
| | - Ken Okazaki
- Department of Orthopaedic Surgery, Tokyo Women's Medical University, 8-1 Kawada-cho, Shinjuku-ku, Tokyo, 162-8666, Japan.
| |
Collapse
|
24
|
Faster union rate and better clinical outcomes using autologous bone graft after medial opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 2020; 28:1380-1387. [PMID: 30903222 DOI: 10.1007/s00167-019-05463-w] [Citation(s) in RCA: 20] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/01/2018] [Accepted: 03/04/2019] [Indexed: 12/31/2022]
Abstract
PURPOSE To compare the radiological bone union rate after medial opening wedge high tibial osteotomy (MOW-HTO) and stabilization using a TomoFix™ plate (Synthes, Oberdorf, Switzerland) in three patient groups. METHODS Retrospective analysis of 137 knees that underwent MOW-HTO between January 2014 and January 2017 was using a TomoFix™ plate. Osteotomy gaps were filled with β-tricalcium phosphate (β-TCP) (group A), left unfilled (group B), and subject to autologous bone graft and β-TCP (group C). Radiological bone union using simple radiography was determined by a modified version of the Brosset et al. osteotomy filling index. Clinical outcomes were evaluated using the International Knee Documentation Committee (IKDC) and Lysholm score. Statistical analyses using the Chi-square and ANOVA tests were performed between the groups. RESULTS The mean time for radiological bone union was 8.3 ± 3.1 months in group A, 7.2 ± 3.2 in group B and 3.4 ± 1.5 in group C (p = 0.001). There was statistically significant faster bone union in Group C. If the opening distance was more than 10 mm, group A united in 8.6 ± 3.6 months, group B in 8.8 ± 3.4, and group C in 3.5 ± 1.7 (p = 0.001). IKDC and Lysholm knee scores improved significantly (p = 0.004 for IKDC and 0.001 for Lysholm knee scores) in group C when compared to groups A and B at sixth month follow-up. At final follow-up, there was no difference in IKDC and Lysholm knee scores. Less delayed union occurred in group C. CONCLUSIONS MOW-HTO with autologous bone graft and β-TCP had the fastest radiological bone union and best clinical scores at 6 month follow-up. LEVEL OF EVIDENCE Retrospective comparative study, Level III.
Collapse
|