Aykanat F, Kose O, Guneri B, Celik HK, Cakar A, Tasatan E, Ulmeanu ME. Comparison of four different screw configurations for the fixation of Fulkerson osteotomy: a finite element analysis. J Orthop Traumatol 2023; 24:30. [PMID: 37358664 DOI: 10.1186/s10195-023-00714-6] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/23/2022] [Accepted: 06/04/2023] [Indexed: 06/27/2023] Open
Abstract
BACKGROUND Conventionally, two 4.5 mm cortical screws inserted toward the posterior tibial cortex are usually advocated for the fixation of Fulkerson osteotomy. This finite element analysis aimed to compare the biomechanical behavior of four different screw configurations to fix the Fulkerson osteotomy. MATERIALS AND METHODS Fulkerson osteotomy was modeled using computerized tomography (CT) data of a patient with patellofemoral instability and fixed with four different screw configurations using two 4.5 mm cortical screws in the axial plane. The configurations were as follows: (1) two screws perpendicular to the osteotomy plane, (2) two screws perpendicular to the posterior cortex of the tibia, (3) the upper screw perpendicular to the osteotomy plane, but the lower screw is perpendicular to the posterior cortex of the tibia, and (4) the reverse position of the screw configuration in the third scenario. Gap formation, sliding, displacement, frictional stress, and deformation of the components were calculated and reported. RESULTS The osteotomy fragment moved superiorly after loading the models with 1654 N patellar tendon traction force. Since the proximal cut is sloped (bevel-cut osteotomy), the osteotomy fragment slid and rested on the upper tibial surface. Afterward, the upper surface of the osteotomy fragment acted as a fulcrum, and the distal part of the fragment began to separate from the tibia while the screws resisted the displacement. The resultant total displacement was 0.319 mm, 0.307 mm, 0.333 mm, and 0.245 mm from the first scenario to the fourth scenario, respectively. The minimum displacement was detected in the fourth scenario (upper screw perpendicular to the osteotomy plane and lower screw perpendicular to the posterior tibial cortex). Maximum frictional stress and maximum pressure between components on both surfaces were highest in the first scenario (both screws perpendicular to the osteotomy plane). CONCLUSIONS A divergent screw configuration in which the upper screw is inserted perpendicular to the osteotomy plane and the lower screw is inserted perpendicular to the posterior tibial cortex might be a better option for the fixation of Fulkerson osteotomy. Level of evidence Level V, mechanism-based reasoning.
Collapse
Affiliation(s)
- Faruk Aykanat
- Vocational School of Health Services, SANKO University, Gaziantep, Turkey
| | - Ozkan Kose
- Department of Orthopedics and Traumatology, Antalya Training and Research Hospital, Varlık mah., Kazım Karabekir cd., Muratpasa, 07100, Antalya, Turkey.
| | - Bulent Guneri
- Department of Orthopedics and Traumatology, Adana City Education and Research Hospital, Adana, Turkey
| | - H Kursat Celik
- Agricultural Faculty, Department of Agricultural Machinery and Technology Engineering, Akdeniz University, Antalya, Turkey
| | - Albert Cakar
- Department of Orthopedics and Traumatology, Istanbul Training and Research Hospital, Istanbul, Turkey
| | - Ersin Tasatan
- Department of Orthopedics and Traumatology, Prof. Dr. Cemil Tascioglu City Hospital, Istanbul, Turkey
| | - Mihaela-Elena Ulmeanu
- Department of Manufacturing, Polytechnic University of Bucharest, Bucharest, Romania
| |
Collapse
|
2
|
Chang CW, Chen CH, Li CT, Chen YN, Yang TH, Chang CJ, Chang CH. Role of an additional third screw in the fixation of transverse patellar fracture with two parallel cannulated screw and anterior wire. BMC Musculoskelet Disord 2020; 21:752. [PMID: 33189156 PMCID: PMC7666768 DOI: 10.1186/s12891-020-03744-x] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/30/2020] [Accepted: 10/26/2020] [Indexed: 01/08/2023] Open
Abstract
BACKGROUND Two parallel cannulated screws along with an anterior wire to construct a tension band is a popular approach in transverse patellar fractures. However, the optimal screw proximity, either deep or superficial screw placements, remains controversial. Hence, a new concept of the addition of a third screw to form a triangular configuration along with the original two parallel screws was proposed in this study. Therefore, the biomechanical effect of the additional third screw on the stability of the fractured patella was investigated with finite element (FE) simulation. METHODS An FE knee model including the distal femur, proximal tibia, and fractured patella (type AT/OTA 34-C) was developed in this study. Four different screw configurations, including two parallel cannulated screws with superficial (5-mm proximity) and deep (10-mm proximity) placements and two parallel superficial screws plus a third deep screw, and two parallel deep screws plus a third superficial screw, with or without the anterior wire, were considered for the simulation. RESULTS Results indicated that the addition of a third screw increased stability by reducing the dorsal gap opening when two parallel screws were deeply placed, particularly on the fractured patella without an anterior wire. However, the third screw was of little value when two parallel screws were superficially placed. In the existence of two deep parallel screws and the anterior wire, the third screw reduced the gap opening by 23.5% (from 1.15 mm to 0.88 mm) and 53.6% (from 1.21 mm to 0.61 mm) in knee flexion 45° and full extension, respectively. Furthermore, in the absence of the anterior wire, the third screw reduced the gap opening by 73.5% (from 2 mm to 0.53 mm) and 72.2% (from 1.33 mm to 0.37 mm) in knee flexion 45° and full extension, respectively. CONCLUSION Based on the results, a third cannulated screw superficially placed (5-mm proximity) is recommended to increase stability and maintain contact of the fractured patella, fixed with two parallel cannulated screws deeply placed (10-mm proximity), particularly when an anterior wire was not used. Furthermore, the third screw deeply placed is not recommended in a fractured patella with two parallel superficial screws.
Collapse
Affiliation(s)
- Chih-Wei Chang
- Department of Orthopedics, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan City, Taiwan.,Department of Orthopedics, College of Medicine, National Cheng Kung University, Tainan City, Taiwan
| | - Chih-Hsien Chen
- Department of Orthopaedic Surgery, Tainan Municipal Hospital (Managed by Show Chwan Medical Care Corporation), Tainan City, Taiwan
| | - Chun-Ting Li
- Institute of Geriatric Welfare Technology & Science, Mackay Medical College, New Taipei City, Taiwan
| | - Yen-Nien Chen
- Department of Physical Therapy, Asia University, 500, Lioufeng Rd, Wufeng, Taichung City, 41354, Taiwan.
| | - Tai-Hua Yang
- Department of Orthopedics, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan City, Taiwan. .,Department of Orthopedics, College of Medicine, National Cheng Kung University, Tainan City, Taiwan. .,Department of BioMedical Engineering, National Cheng Kung University, No.1, University Road, Tainan City, 701, Taiwan. .,Skeleton Materials and Bio-compatibility Core Lab, Research Center of Clinical Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan City, Taiwan. .,Medical Device Innovation Center, National Cheng Kung University, Tainan City, Taiwan.
| | - Chia-Jung Chang
- Department of BioMedical Engineering, National Cheng Kung University, No.1, University Road, Tainan City, 701, Taiwan
| | - Chih-Han Chang
- Department of BioMedical Engineering, National Cheng Kung University, No.1, University Road, Tainan City, 701, Taiwan.,Medical Device Innovation Center, National Cheng Kung University, Tainan City, Taiwan
| |
Collapse
|