Green MV, Seidel J, Choyke PL, Jagoda EM. A simple device to convert a small-animal PET scanner into a multi-sample tissue and injection syringe counter.
Nucl Med Biol 2017;
53:9-13. [PMID:
28692851 PMCID:
PMC5610080 DOI:
10.1016/j.nucmedbio.2017.05.012]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/24/2017] [Revised: 05/26/2017] [Accepted: 05/30/2017] [Indexed: 10/19/2022]
Abstract
INTRODUCTION
We describe a simple fixture that can be added to the imaging bed of a small-animal PET scanner that allows for automated counting of multiple organ or tissue samples from mouse-sized animals and counting of injection syringes prior to administration of the radiotracer. The combination of imaging and counting capabilities in the same machine offers advantages in certain experimental settings.
METHODS
A polyethylene block of plastic, sculpted to mate with the animal imaging bed of a small-animal PET scanner, is machined to receive twelve 5-ml containers, each capable of holding an entire organ from a mouse-sized animal. In addition, a triangular cross-section slot is machined down the centerline of the block to secure injection syringes from 1-ml to 3-ml in size. The sample holder is scanned in PET whole-body mode to image all samples or in one bed position to image a filled injection syringe. Total radioactivity in each sample or syringe is determined from the reconstructed images of these objects using volume re-projection of the coronal images and a single region-of-interest for each. We tested the accuracy of this method by comparing PET estimates of sample and syringe activity with well counter and dose calibrator estimates of these same activities.
RESULTS
PET and well counting of the same samples gave near identical results (in MBq, R2=0.99, slope=0.99, intercept=0.00-MBq). PET syringe and dose calibrator measurements of syringe activity in MBq were also similar (R2=0.99, slope=0.99, intercept=- 0.22-MBq).
CONCLUSION
A small-animal PET scanner can be easily converted into a multi-sample and syringe counting device by the addition of a sample block constructed for that purpose. This capability, combined with live animal imaging, can improve efficiency and flexibility in certain experimental settings.