Ispierto L, Muñoz J, Cladellas JM, Cuadras P, Capellades J, Latorre P, Dávalos A, Vancamp T, Álvarez R. Post-Operative Localization of Deep Brain Stimulation Electrodes in the Subthalamus Using Transcranial Sonography.
Neuromodulation 2017;
21:574-581. [PMID:
29178240 DOI:
10.1111/ner.12733]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/20/2017] [Revised: 05/19/2017] [Accepted: 06/16/2017] [Indexed: 11/28/2022]
Abstract
OBJECTIVES
The correct positioning of deep brain stimulation electrodes determines the success of surgery. In this study, we attempt to validate transcranial sonography (TCS) as a method for early postoperative confirmation of electrode location in the subthalamic nucleus (STN).
MATERIALS AND METHODS
Nineteen patients diagnosed with Parkinson's disease were enrolled in the study. Postoperative TCS was applied to measure the distance between the implanted electrodes and the third ventricle in the axial plane. Whether the electrodes were positioned within or outside the substantia nigra (SN) was evaluated through measurements in the coronal plane. The obtained metrics through TCS were compared with those from postoperative computed tomography (CT) and magnetic resonance imaging (MRI).
RESULTS
A statistically significant correlation between distances from electrode to third ventricle by TCS and CT/MRI (r = 0.75, p < 0.01) was observed. Distances from third ventricle to electrodes tips were different when sonographically they showed to be inside or outside the SN (p < 0.01). A cut-off value of 8.85mm in these distances was the most sensitive (100%) and specific (90.5%) to predict if electrodes were positioned inside the SN (CI 95% 0.81-10.30, p = 0.001).
CONCLUSIONS
Transcranial sonography is a useful technique to reliably identify targeted positioning of deep brain stimulation electrodes in or out of the SN.
Collapse