Olko P, Bilski P. MICRODOSIMETRIC UNDERSTANDING OF DOSE RESPONSE AND RELATIVE EFFICIENCY OF THERMOLUMINESCENCE DETECTORS.
RADIATION PROTECTION DOSIMETRY 2020;
192:165-177. [PMID:
33418564 PMCID:
PMC7840113 DOI:
10.1093/rpd/ncaa211]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 10/20/2020] [Revised: 10/20/2020] [Accepted: 11/23/2020] [Indexed: 06/12/2023]
Abstract
LiF:Mg,Ti detectors show relative efficiency η for heavy charged particles significantly lower than one. It was for a long time not recognised that η varies also for electron energies and, as a consequence for photons. For LiF:Mg,Cu,P detectors measured photon energy response was named 'anomalous' because it differed significantly from the ratio of photon absorption coefficients. The decrease of η was explained as a microdosimetric effect due to local saturation of trapping centres around the electron track. For TLD-100 it was noticed by Horowitz that the measured photon energy response disagrees with the ratio of absorption coefficient by about 10%. It was demonstrated that a fraction of the TL signal in LiF:Mg,Ti is generated in the supralinear dose-response range, due to the high local doses generated by photon-induced tracks. Prediction of TL efficiency is particularly important in space dosimetry and in dosimetry of therapeutic beams like protons or carbon ions.