Karim MKA, Rahim NA, Matsubara K, Hashim S, Mhareb MHA, Musa Y. The effectiveness of bismuth breast shielding with protocol optimization in CT Thorax examination.
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY 2019;
27:139-147. [PMID:
30584178 DOI:
10.3233/xst-180397]
[Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/09/2023]
Abstract
BACKGROUND
Numerous techniques had been proposed to reduce radiation exposure in computed tomography (CT) including the use of radiation shielding.
OBJECTIVE
This study aims to evaluate efficacy of using a bismuth breast shield and optimized scanning parameter to reduce breast absorbed doses from CT thorax examination.
METHODS
Five protocols comprising the standard CT thorax clinical protocol (CP1) and four modified protocols (CP2 to CP5) were applied in anthropomorphic phantom scans. The phantom was configured as a female by placing a breast component on the chest. The breast component was divided into four quadrants, where 2 thermoluminescence dosimeters (TLD-100) were inserted into each quadrant to measure the absorbed dose. The bismuth shield was placed over the breast component during CP4 and CP5 scans.
RESULTS
The pattern of absorbed doses in each breast and quadrant were approximately the same for all protocols, where the 4th quadrant > 3rd quadrant > 2nd quadrant > 1st quadrant. The mean absorbed dose value in CP3 was reduced to almost 34% of CP1's mean absorbed dose. It was reduced even lower to 15% of CP1's mean absorbed dose when the breast shield was used in CP5.
CONCLUSION
This study showed that CT radiation exposure on the breast could be reduced by using a bismuth shield and low tube potential protocol without compromising the image quality.