D P, D K, P M, Bhalla NK, R P, M C, Puri A. A Study to Analyze Impact of Treatment Couch and Immobilization Devices on Surface Dose for Megavoltage Photon Beams.
Asian Pac J Cancer Prev 2023;
24:4133-4138. [PMID:
38156848 PMCID:
PMC10909093 DOI:
10.31557/apjcp.2023.24.12.4133]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/23/2023] [Accepted: 12/17/2023] [Indexed: 01/03/2024] Open
Abstract
PURPOSE/OBJECTIVE
The purpose of this study is to investigate the effect of treatment couch and immobilization devices on surface dose for megavoltage photon beams.
MATERIAL/METHODS
Percentage surface dose (PSD) measurement was carried out in Elekta Synergy™ Linear accelerator using PTW Markus® Parallel plate ionization chamber of volume 0.05cm3 with water equivalent RW3 Slab phantom (PTW, Germany). The measurement depth was considered at 0.07mm. The reference PSD was measured at 0° gantry angle with 10×10cm2, 20×20cm2 and 30×30cm2 field sizes and 100cm SSD for 4MV, 6MV and 15MV photon beams. For comparison, PSD measurement was carried out at 180° gantry angle inclusion of treatment couch (TC), All in One positioning system (AIO - PS) and Vac lok Cushions (VLC).
RESULTS
Beam angle at 0°, for field sizes 10×10cm2, 20×20cm2 and 30×30cm2, the PSD was observed as 30.9%, 40.5%, 48.7% for 4MV; 23.7%, 33.8%, 42.2% for 6MV; and 17.0%, 29.6%, 38.6% for 15MV respectively. Beam angle at 180° with TC, an increase in PSD by maximum of 65.0% for 4MV, 64.9% for 6MV and 55.9% for 15MV as compared to 0° angle. The PSD increased when beam angle was 180° with TC and AIO - PS were 65.0% for 4MV, 67.4% for 6MV, and 60.9% for 15MV than 0° angle. Similarly, increased PSD for beam angle at 180° with TC and VLC were 66.8% for 4MV, 66.8% for 6MV and 61.3% for 15MV as compared to 0° angle.
CONCLUSION
For all three-photon energies, at 180° gantry angle, the PSD increased significantly in case of TC, VLC, and AIO - PS for all the field sizes as compared to gantry angle at 0°. It is necessary to consider TC, AIO - PS and VLC during dose calculation to ensure accuracy of patient treatment delivery.