Witkowska A, Petre EN, Moussa AM, Santos E, Sarkar D, Lis E, Cornelis FH. Feasibility and Safety of Percutaneous CT-Guided Bone Biopsies in Patients with Cancer Using a Patient-Mounted Robotic System: A Retrospective Analysis of 40 Consecutive Biopsies.
J Vasc Interv Radiol 2023;
34:2174-2179. [PMID:
37673400 PMCID:
PMC11260433 DOI:
10.1016/j.jvir.2023.08.040]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/05/2023] [Revised: 08/23/2023] [Accepted: 08/26/2023] [Indexed: 09/08/2023] Open
Abstract
This retrospective study evaluated the feasibility and safety of percutaneous computed tomography (CT)-guided bone biopsies in patients with cancer using a patient-mounted robotic system with steering capabilities. The study included 39 patients (17 women, 22 men; median age, 65.5 years; interquartile range [IQR], 54.8-71.0 years). Forty biopsies were performed in the pelvis, spine, ribs, shoulder, femur, and sternum. The technical success rate was 100%, and the median trajectory length was 55.9 mm (IQR, 47.1-73.6 mm). Intermediate checkpoints were used in 8 biopsies. Median time from the first to final scan was 21 minutes (IQR, 17-37 minutes). The overall procedure time was 30 minutes (IQR, 24-36 minutes). The median dose length product and effective dose were 536.6 mGy⋅cm (IQR, 396.2-837.7 mGy∗cm) and 7.1 mSv (IQR, 4.7-10.8 mSv), respectively. No adverse events occurred. The diagnostic yield for cancer was 72.5%. Percutaneous robotic-assisted bone biopsies demonstrated high technical success, adequate diagnostic yield, and favorable safety profile.