Schröder S, Mäurer S, Eppig T, Seitz B, Rubly K, Langenbucher A. Comparison of Corneal Tomography: Repeatability, Precision, Misalignment, Mean Elevation, and Mean Pachymetry.
Curr Eye Res 2018;
43:709-716. [PMID:
29482368 DOI:
10.1080/02713683.2018.1441873]
[Citation(s) in RCA: 19] [Impact Index Per Article: 3.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/17/2022]
Abstract
PURPOSE
To compare corneal tomography and its statistical uncertainty for measurements obtained by three clinically used corneal tomographers: A Scheimpflug camera (Pentacam HR), a swept source optical coherence tomography system (CASIA SS-1000), and Placido ring imaging (TMS-5).
MATERIAL AND METHODS
Repeated measurements with all three devices on 34 normal eyes were used to estimate the repeatability, precision, and mean values of corneal elevation and pachymetry within 8 mm diameter. The repeatability (standard deviation) was calculated for each data point of the corneal elevation data-maps of anterior and posterior cornea as well as for the pachymetry data-maps. Uncertainty on the position of the eye at each measurement might contribute to the differences between elevation data-maps. To take this into account, we defined the precision as the standard deviation for the elevation data-maps of anterior and posterior cornea after correction of misalignment-effects (rotation, translation). The mean elevation and pachymetry data-maps were fitted with Zernike polynomials for interdevice-comparison.
RESULTS
Pentacam HR offered the best repeatability and precision for the anterior corneal elevation (<3 and <1.6 μm, respectively). CASIA SS-1000 offered good repeatability and precision with high resolution for posterior corneal elevation, and the best repeatability for pachymetry (<3 μm). TMS-5 measured anterior elevation with similar repeatability to CASIA SS-1000 (<6 μm). The data-maps of the three tomographers could not be used interchangeably. The largest differences were observed for pachymetry and posterior corneal elevation data-maps.
CONCLUSIONS
Misalignment limited the repeatability of TMS-5 and Pentacam HR, but had little influence on the repeatability of CASIA SS-1000.