Zielinska D, Poels I, Pietraszkiewicz M, Radecki J, Geise HJ, Nagels LJ. Potentiometric detection of organic acids in liquid chromatography using polymeric liquid membrane electrodes incorporating macrocyclic hexaamines.
J Chromatogr A 2001;
915:25-33. [PMID:
11358255 DOI:
10.1016/s0021-9673(01)00644-6]
[Citation(s) in RCA: 30] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
Potentiometric detection employing coated-wire electrodes was applied to the determination of organic acids in liquid chromatography (LC). Poly(vinyl chloride)-based liquid membranes, incorporating lipophilic macrocyclic hexaamines as neutral ionophores were used as electrode coatings. The selectivity and sensitivity of the macrocycle-based electrodes were found to be superior to an electrode based on a lipophilic anion exchanger (a quaternary ammonium salt). Sensitive detection was obtained for the di- and tricarboxylic acids tartaric, malonic, malic, citric, fumaric, succinic, pyruvic, 2-oxoglutaric and maleic acids after separation in reversed-phase LC. Detection limits (signal/4sigmanoise=3) of 6 pmol for malonic acid and 2 pmol for maleic acid were attained. The detection was explained using a molecular recognition model. The hexaamine-based potentiometric electrodes had a 1-s response time at 1 ml min(-1) flow-rates. They were stable for at least 4 months, with an intra-electrode variation of 3.2% (n=5).
Collapse