Kämmerer PW, Kumar VV, Brüllmann D, Götz H, Kann PH, Al-Nawas B, Klein MO. Evaluation of ultrasound transmission velocity and 3-dimensional radiology in different bone types for dental implantology: a comparative ex vivo study.
Oral Surg Oral Med Oral Pathol Oral Radiol 2012;
116:e77-84. [PMID:
22762916 DOI:
10.1016/j.oooo.2011.11.028]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/26/2011] [Revised: 10/31/2011] [Accepted: 11/07/2011] [Indexed: 11/29/2022]
Abstract
OBJECTIVE
To evaluate ultrasound transmission velocity (UTV) for assessment of mechanical bone quality, an ex vivo comparison of different bone types measured with UTV, 2-dimensional (2D) histomorphometry and with 3-dimensional (3D) radiology (cone-beam computerized tomography [CBCT], computerized microtomography [μCT]) was conducted.
STUDY DESIGN
Clinical cortical, cancellous, and mixed bone (each n = 6) was measured via UTV (m/s), CBCT (white pixel/black pixel ratio [WP/BP]), μCT (bone volume/total volume [μBV/TV]), and histomorphometry (bone volume/total volume [hBV/TV]). UTV values were correlated with 2D-histomorphometry and 3D-radiologic results.
RESULTS
For the cortical, cancellous, and mixed bone samples, respectively, UTV values were 1,945.17, 1,266.9,and 1472.2 m/s, WP/BP quotients were 0.96, 0.15, and 0.33, μBV/TV quotients were 0.94, 0.2, and 0.47, and hBV/TV quotients were 0.96, 0.24 and 0.39. Significant correlations between UTV and the other methods were seen (P < .0001).
CONCLUSIONS
Similar to the other methods, UTV is able to discriminate between different bone types ex vivo.
Collapse