Lefèvre J, Lafitte S, Reant P, Perron JM, Roudaut R. Optimization of patent foramen ovale detection by contrast transthoracic echocardiography using second harmonic imaging.
Arch Cardiovasc Dis 2008;
101:213-9. [PMID:
18654095 DOI:
10.1016/s1875-2136(08)73695-7]
[Citation(s) in RCA: 13] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/14/2022]
Abstract
BACKGROUND
Patent foramen ovale is an anomaly responsible for paradoxical embolizations and cerebral ischemic events. Aims. - We want to show second harmonic transthoracic echography sensitized by contrast agent perfusion is as well as transesophageal echography to patent foramen ovale detection.
METHODS
Onene hundred twenty one patients referred for transesophageal echocardiography for patent foramen ovale detection, underwent additive second harmonic transthoracic echocardiography with one of three randomized contrast agents: a mixture A of dextrose and air, mixture B of dextrose and air and blood, or mixture C of hydroxyethylamidon. The severity of atrial shunting was evaluated on recordings by semi-quantitative scoring. Intensity of contrast was also assessed by objective quantitative videodensitometry.
RESULTS
No difference was observed between the two techniques, nor between mixture A, B and C in terms of PFO detection during each exam. However, quantitative contrast analysis showed higher intensity with mixtures B and C with mixture A during transthoracic echography.
CONCLUSIONS
When performed with a contrast agent, second harmonic transthoracic echography and transoesophageal echography are comparable when it comes to patent foramen ovale detection. Although the composition of the contrast agent does not appear to affect the rate of this detection, contrast quality in the right atrium during transthoracic exam is better with mixtures B and C than with mixture A.
Collapse