Sheng K, Cao M, Godley A, Lin MH, Henze L, Hammond L, Delombaerde L, Hierholz K, Kouptsidis J. Quantification of Dosimetry Improvement With or Without Patient Surface Guidance.
Adv Radiat Oncol 2024;
9:101570. [PMID:
39188998 PMCID:
PMC11345286 DOI:
10.1016/j.adro.2024.101570]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/04/2023] [Accepted: 06/21/2024] [Indexed: 08/28/2024] Open
Abstract
Purpose
Noncoplanar beams and arcs are routinely used to improve dosimetry for intracranial cases, but their application for extracranial cases has been hampered by the risk of collision. This has led to conservative beam selection whose impact on plan dosimetry has not been previously studied.
Methods and Materials
A full-body 3-dimensional patient surface was acquired using optical cameras for a single lung patient at the time of computed tomography simulation. Eight stereotactic body radiation therapy (SBRT) plans were created for the patient, with varying degrees of noncoplanarity and deliverability. The plans included volumetric modulated arc therapy and intensity modulated radiation therapy (IMRT) plans ranging from simple, coplanar arcs to multiple noncoplanar arcs and IMRT beams. A total of 70 fields were created across the 8 plans, of which 21 fields were undeliverable with a 5-cm buffer. Organs-at-risk (OARs) metrics including R50, Dmax 2 cm from the PTV, lung V20, and chest wall V30 were evaluated. Five expert SBRT dosimetrists from 5 institutions evaluated field deliverability, with or without the guidance of the clearance map.
Results
In the dosimetry evaluation, a clear trend in increasing dosimetric compactness and OAR sparing is observed with increasing plan noncoplanarity. R50, Dmax 2 cm, lung V20, and chest wall V30 decreased 41%, 39%, 43%, and 57%, respectively, from plan 1 (2 coplanar partial arcs) to plan 8 (19 noncoplanar IMRT beams). In the observer tests, the expert dosimetrists' ability to accurately discern beam deliverability because of collision significantly increases with the clearance map. The errors in predicting colliding fields were eliminated using the whole-body surface and clearance map, and the user was able to select fields based on plan quality and patient comfort instead of being overly conservative.
Conclusion
The study shows that incorporating a personalized, whole-body clearance map in the treatment planning workflow can facilitate the adoption of noncoplanar beams or arcs that benefit the SBRT plan dosimetry.