Paepen J, Altzitzoglou T, Van Ammel R, Sibbens G, Pommé S. Half-life measurement of 124Sb.
Appl Radiat Isot 2010;
68:1555-60; discussion 1560. [PMID:
20044262 DOI:
10.1016/j.apradiso.2009.11.050]
[Citation(s) in RCA: 8] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/20/2022]
Abstract
The half-life of (124)Sb was determined experimentally by following the decay of a source from a radiopure solution with a Centronic IG12 ionisation chamber. Thousands of measurements were performed over a period of 358 days, i.e. about six half-life periods. However, the data analysis was restricted to the first 221 days, in order to limit the dominant uncertainty component associated with the hypothetical possibility of a systematic error on background subtraction. The resulting value for the (124)Sb half-life, 60.212 (11) days, is found to be in very good agreement with published values, but carries a lower uncertainty. Major uncertainty contributions pertain to possible systematic errors in background correction, long-term changes in source-detector geometry and medium- and long-term instability of the instrument. Additional measurements were performed with a high-purity germanium detector to confirm the above value.
Collapse