Ganter MT, Hofer CK, Zollinger A, Spahr T, Pasch T, Zalunardo MP. Accuracy and performance of a modified continuous intravascular blood gas monitoring device during thoracoscopic surgery.
J Cardiothorac Vasc Anesth 2004;
18:587-91. [PMID:
15578469 DOI:
10.1053/j.jvca.2004.07.017]
[Citation(s) in RCA: 8] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
OBJECTIVE
The only commercially available continuous intravascular blood gas monitoring system for adults, the Paratrend (Diametrics Medical Inc, High Wycombe, UK), was modified by the manufacturer to the Paratrend 7+ (PT7+) in 1999. The aim of this study was to evaluate the modified probe over a wide range of blood gas and pH values during thoracoscopic surgery in a similar setup as done with the previous model.
DESIGN
Prospective methods comparison study.
SETTING
University hospital.
PARTICIPANTS
Twenty-three patients.
INTERVENTIONS
Elective thoracoscopic surgery.
MEASUREMENTS AND MAIN RESULTS
One hundred thirteen PT7+ readings and their corresponding arterial blood gas and pH measurements (ABGA) were evaluated. The ranges for ABGA values were 50 to 474 mmHg for PO(2), 29 to 58 mmHg for PCO(2), and 7.28 to 7.49 for pH. Bland and Altman analysis revealed a bias +/- 2 standard deviation of -20 +/- 86 mmHg for PO(2), 3 +/- 9 mmHg for PCO(2), and -0.01 +/- 0.06 for pH. No specific complications attributable to the probe were observed.
CONCLUSION
In patients undergoing thoracoscopic surgery with rapidly changing blood gas parameters, the PT7+ device is a valuable trend indicator and hence may be helpful for clinical decision making. However, the underestimation of PO(2) values by 20 mmHg on average and the wide limits of agreement documented in this study must be regarded as limiting factors.