Schaefer C, Allers M, Hitzemann M, Nitschke A, Kobelt T, Mörtel M, Schröder S, Ficks A, Zimmermann S. Reliable Detection of Chemical Warfare Agents Using High Kinetic Energy Ion Mobility Spectrometry. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY 2024. [PMID: 39013159 DOI: 10.1021/jasms.4c00240] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 07/18/2024]
Abstract
High Kinetic Energy Ion Mobility Spectrometers (HiKE-IMS) ionize and separate ions at reduced pressures of 10-40 mbar and over a wide range of reduced electric field strengths E/N of up to 120 Td. Their reduced operating pressure is distinct from that of conventional drift tube ion mobility spectrometers that operate at ambient pressure for trace compound detection. High E/N can lead to a field-induced fragmentation pattern that provides more specific structural information about the analytes. In addition, operation at high E/N values adds the field dependence of ion mobility as an additional separation dimension to low-field ion mobility, making interfering compounds less likely to cause a false positive alarm. In this work, we study the chemical warfare agents tabun (GA), sarin (GB), soman (GD), cyclosarin (GF) and sulfur mustard (HD) in a HiKE-IMS at variable E/N in both the reaction and the drift region. The results show that varying E/N can lead to specific fragmentation patterns at high E/N values combined with molecular signals at low E/N. Compared to the operation at a single E/N value in the drift region, the variation of E/N in the drift region also provides the analyte-specific field dependence of ion mobility as additional information. The accumulated data establish a unique fingerprint for each analyte that allows for reliable detection of chemical warfare agents even in the presence of interfering compounds with similar low-field ion mobilities, thus reducing false positives.
Collapse
Affiliation(s)
- Christoph Schaefer
- Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, 30167 Hannover, Germany
| | - Maria Allers
- Bundeswehr Research Institute for Protective Technologies and CBRN Protection, Humboldtstrasse 100, 29633 Munster, Germany
| | - Moritz Hitzemann
- Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, 30167 Hannover, Germany
| | - Alexander Nitschke
- Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, 30167 Hannover, Germany
| | - Tim Kobelt
- Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, 30167 Hannover, Germany
| | - Max Mörtel
- Bundeswehr Research Institute for Protective Technologies and CBRN Protection, Humboldtstrasse 100, 29633 Munster, Germany
| | - Stefanie Schröder
- Bundeswehr Research Institute for Protective Technologies and CBRN Protection, Humboldtstrasse 100, 29633 Munster, Germany
| | - Arne Ficks
- Bundeswehr Research Institute for Protective Technologies and CBRN Protection, Humboldtstrasse 100, 29633 Munster, Germany
| | - Stefan Zimmermann
- Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, Leibniz University Hannover, Appelstr. 9A, 30167 Hannover, Germany
| |
Collapse
|
2
|
Haack A, Schaefer C, Zimmermann S. On the Arrival Time Distribution of Reacting Systems in Ion Mobility Spectrometry. Anal Chem 2024. [PMID: 39009503 DOI: 10.1021/acs.analchem.4c02010] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 07/17/2024]
Abstract
Ion mobility spectrometry (IMS) is a widely used gas-phase separation technique, particularly when coupled with mass spectrometry (MS). Modern IMS instruments often apply elevated reduced field strengths for improved ion separation and ion focusing. These alter the collision dynamics and further drive ion reaction processes that can change the analyte's structure. As a result, the measured arrival time distribution (ATD) can change with the applied reduced field strengths. In this work, we systematically study how the ion collision dynamics and the ion reaction dynamics, as a function of the reduced field strength, can alter the ATD. To this end, we investigate 2,6-di-tert-butylpyridine, methanol, and ethyl acetate using a home-built drift tube IMS coupled to a home-built MS and extensive first-principles Monte Carlo modeling. We show how elevated reduced field strengths can actually lower resolving power through increased ion diffusion and how the field dependency of the ion mobility can introduce uncertainties to collision cross sections (CCS) calculated from the measured mobilities. On top of the collision dynamics, we show how chemical transformation processes that alter the analyte's CCS, e.g., dynamic clustering or fragmentation, can lead to broadened, shifted, or non-Gaussian ATDs and how sensitive these processes are to the applied field strengths. We highlight how first-principles ion dynamics simulations can help to understand and even harness the mentioned effects.
Collapse
Affiliation(s)
- Alexander Haack
- Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, 30167 Hannover, Germany
| | - Christoph Schaefer
- Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, 30167 Hannover, Germany
| | - Stefan Zimmermann
- Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, 30167 Hannover, Germany
| |
Collapse
|
3
|
Krzyściak W, Bystrowska B, Karcz P, Chrzan R, Bryll A, Turek A, Mazur P, Śmierciak N, Szwajca M, Donicz P, Furman K, Pilato F, Kozicz T, Popiela T, Pilecki M. Association of Blood Metabolomics Biomarkers with Brain Metabolites and Patient-Reported Outcomes as a New Approach in Individualized Diagnosis of Schizophrenia. Int J Mol Sci 2024; 25:2294. [PMID: 38396971 PMCID: PMC10888632 DOI: 10.3390/ijms25042294] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/10/2024] [Revised: 02/06/2024] [Accepted: 02/10/2024] [Indexed: 02/25/2024] Open
Abstract
Given its polygenic nature, there is a need for a personalized approach to schizophrenia. The aim of the study was to select laboratory biomarkers from blood, brain imaging, and clinical assessment, with an emphasis on patients' self-report questionnaires. Metabolomics studies of serum samples from 51 patients and 45 healthy volunteers, based on the liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS/MS), led to the identification of 3 biochemical indicators (cortisol, glutamate, lactate) of schizophrenia. These metabolites were sequentially correlated with laboratory tests results, imaging results, and clinical assessment outcomes, including patient self-report outcomes. The hierarchical cluster analysis on the principal components (HCPC) was performed to identify the most homogeneous clinical groups. Significant correlations were noted between blood lactates and 11 clinical and 10 neuroimaging parameters. The increase in lactate and cortisol were significantly associated with a decrease in immunological parameters, especially with the level of reactive lymphocytes. The strongest correlations with the level of blood lactate and cortisol were demonstrated by brain glutamate, N-acetylaspartate and the concentrations of glutamate and glutamine, creatine and phosphocreatine in the prefrontal cortex. Metabolomics studies and the search for associations with brain parameters and self-reported outcomes may provide new diagnostic evidence to specific schizophrenia phenotypes.
Collapse
Affiliation(s)
- Wirginia Krzyściak
- Department of Medical Diagnostics, Jagiellonian University Medical College, Faculty of Pharmacy, 30-688 Krakow, Poland;
| | - Beata Bystrowska
- Department of Biochemical Toxicology, Jagiellonian University Medical College, Faculty of Pharmacy, 30-688 Krakow, Poland;
| | - Paulina Karcz
- Department of Electroradiology, Jagiellonian University Medical College, Faculty of Health Sciences, 31-126 Krakow, Poland;
| | - Robert Chrzan
- Department of Radiology, Jagiellonian University Medical College, Faculty of Medicine, 31-503 Krakow, Poland; (R.C.); (A.B.); (T.P.)
| | - Amira Bryll
- Department of Radiology, Jagiellonian University Medical College, Faculty of Medicine, 31-503 Krakow, Poland; (R.C.); (A.B.); (T.P.)
| | - Aleksander Turek
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| | - Paulina Mazur
- Department of Medical Diagnostics, Jagiellonian University Medical College, Faculty of Pharmacy, 30-688 Krakow, Poland;
| | - Natalia Śmierciak
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| | - Marta Szwajca
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| | - Paulina Donicz
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| | - Katarzyna Furman
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| | - Fabio Pilato
- Neurology, Neurophysiology and Neurobiology Unit, Department of Medicine, Università Campus Bio-Medico di Roma, 00128 Rome, Italy;
| | - Tamas Kozicz
- Department of Clinical Genomics, Center for Individualized Medicine, Mayo Clinic, Rochester, MN 55905, USA;
| | - Tadeusz Popiela
- Department of Radiology, Jagiellonian University Medical College, Faculty of Medicine, 31-503 Krakow, Poland; (R.C.); (A.B.); (T.P.)
| | - Maciej Pilecki
- Department of Child and Adolescent Psychiatry and Psychotherapy, Faculty of Medicine, Jagiellonian University Medical College, 31-501 Krakow, Poland; (A.T.); (N.Ś.); (M.S.); (P.D.); (K.F.); (M.P.)
| |
Collapse
|
4
|
Schaefer C, Lippmann M, Beukers M, Beijer N, van de Kamp B, Knotter J, Zimmermann S. Detection of Triacetone Triperoxide by High Kinetic Energy Ion Mobility Spectrometry. Anal Chem 2023; 95:17099-17107. [PMID: 37946366 PMCID: PMC10666079 DOI: 10.1021/acs.analchem.3c04101] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/12/2023] [Revised: 10/24/2023] [Accepted: 10/26/2023] [Indexed: 11/12/2023]
Abstract
High Kinetic Energy Ion Mobility Spectrometry (HiKE-IMS) is a versatile technique for the detection of gaseous target molecules that is particularly useful in complex chemical environments, while the instrumental effort is low. Operating HiKE-IMS at reduced pressures from 10 to 60 mbar results in fewer ion-neutral collisions than at ambient pressure, reducing chemical cross-sensitivities and eliminating the need for a preceding separation dimension, e.g., by gas chromatography. In addition, HiKE-IMS allows operation over a wide range of reduced electric field strengths E/N up to 120 Td, allowing separation of ions by low-field ion mobility and exploiting the field dependence of ion mobility, potentially allowing separation of ion species at high E/N despite similar low-field ion mobilities. Given these advantages, HiKE-IMS can be a useful tool for trace gas analysis such as triacetone triperoxide (TATP) detection. In this study, we employed HiKE-IMS to detect TATP. We explore the ionization of TATP and the field-dependent ion mobilities, providing a database of the ion mobilities depending on E/N. Confirming the literature results, ionization of TATP by proton transfer with H3O+ in HiKE-IMS generates fragments, but using NH4+ as the primary reactant ion leads to the TATP·NH4+ adduct. This adduct fragments at high E/N, which could provide additional information for reliable detection of TATP. Thus, operating HiKE-IMS at variable E/N in the drift region generates a unique fingerprint of TATP made of all ion species related to TATP and their ion mobilities depending on E/N, potentially reducing the rate of false positives.
Collapse
Affiliation(s)
- Christoph Schaefer
- Institute
of Electrical Engineering and Measurement Technology, Department of
Sensors and Measurement Technology, Leibniz
University Hannover, Appelstr. 9A, Hannover 30167, Germany
| | - Martin Lippmann
- Institute
of Electrical Engineering and Measurement Technology, Department of
Sensors and Measurement Technology, Leibniz
University Hannover, Appelstr. 9A, Hannover 30167, Germany
| | - Michiel Beukers
- Research
Group Technologies for Criminal Investigations, Saxion University of Applied Sciences, M.H Tromplaan 28, Enschede 7513AB, The Netherlands
- Knowledge
Centre of Digitalization, Intelligence, and Technology, Police Academy of The Netherlands, Arnhemseweg 348, Apeldoorn 7334AC, The Netherlands
| | - Niels Beijer
- Research
Group Technologies for Criminal Investigations, Saxion University of Applied Sciences, M.H Tromplaan 28, Enschede 7513AB, The Netherlands
- Knowledge
Centre of Digitalization, Intelligence, and Technology, Police Academy of The Netherlands, Arnhemseweg 348, Apeldoorn 7334AC, The Netherlands
| | - Ben van de Kamp
- Research
Group Technologies for Criminal Investigations, Saxion University of Applied Sciences, M.H Tromplaan 28, Enschede 7513AB, The Netherlands
- Knowledge
Centre of Digitalization, Intelligence, and Technology, Police Academy of The Netherlands, Arnhemseweg 348, Apeldoorn 7334AC, The Netherlands
| | - Jaap Knotter
- Research
Group Technologies for Criminal Investigations, Saxion University of Applied Sciences, M.H Tromplaan 28, Enschede 7513AB, The Netherlands
- Knowledge
Centre of Digitalization, Intelligence, and Technology, Police Academy of The Netherlands, Arnhemseweg 348, Apeldoorn 7334AC, The Netherlands
| | - Stefan Zimmermann
- Institute
of Electrical Engineering and Measurement Technology, Department of
Sensors and Measurement Technology, Leibniz
University Hannover, Appelstr. 9A, Hannover 30167, Germany
| |
Collapse
|
5
|
Haack A, Schaefer C, Zimmermann S, Hopkins WS. Validation of Field-Dependent Ion-Solvent Cluster Modeling via Direct Measurement of Cluster Size Distributions. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY 2023; 34:1035-1046. [PMID: 37116175 DOI: 10.1021/jasms.3c00012] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/08/2023]
Abstract
Ion mobility spectrometry is widely used in analytical chemistry, either as a stand-alone technique or coupled to mass spectrometry. Ions in the gas phase tend to form loosely bound clusters with surrounding solvent vapors, artificially increasing the collisional cross section and the mass of the ion. This, in turn, affects ion mobility and influences separation. Further, ion-solvent clusters play an important role in most ionization mechanisms occurring in the gas phase. Consequently, a deeper understanding of ion-solvent cluster association and dissociation processes is desirable to guide experimental design and interpretation. A few computational models exist, which aim to describe the amount of clustering as a function of the reduced electric field strength, bath gas pressure and temperature, and the chemical species probed. It is especially challenging to model ion mobility under high reduced electrical field strengths due to the nonthermal conditions created by the field. In this work, we aim to validate a recently proposed first-principles model by comparing its predictions with direct measurements of cluster size distributions over a range of 20-120 Td as observed using a High Kinetic Energy Ion Mobility Spectrometer coupled to a mass spectrometer (HiKE-IMS-MS). By studying H+(H2O)n, [MeOH + H + n(H2O)]+, [ACE + H + n(H2O)]+, and [PhNH2 + H + n(H2O)]+ as test systems, we find very good agreement between model and experiment, supporting the validity of the computational workflow. Further, the detailed information gained from the modeling yields important insights into the cluster dynamics within the HiKE-IMS, allowing for better interpretation of the measured ion mobility spectra.
Collapse
Affiliation(s)
- Alexander Haack
- Department of Chemistry, University of Waterloo, 200 University Avenue W, Waterloo, Ontario N2L 3G1, Canada
| | - Christoph Schaefer
- Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, 30167 Hannover, Germany
| | - Stefan Zimmermann
- Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz University Hannover, 30167 Hannover, Germany
| | - W Scott Hopkins
- Department of Chemistry, University of Waterloo, 200 University Avenue W, Waterloo, Ontario N2L 3G1, Canada
- Watermine Innovation, Waterloo, Ontario N0B 2T0, Canada
- Centre for Eye and Vision Research, Hong Kong Science Park, New Territories, 999077, Hong Kong
| |
Collapse
|
6
|
Naylor CN, Schaefer C, Kirk AT, Zimmermann S. The origin of isomerization of aniline revealed by high kinetic energy ion mobility spectrometry (HiKE-IMS). Phys Chem Chem Phys 2023; 25:1139-1152. [PMID: 36515135 DOI: 10.1039/d2cp01994a] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022]
Abstract
Although aniline is a relatively simple small molecule, the origin of its two peaks observed in ion mobility spectrometry (IMS) has remained under debate for at least 30 years. First hypothesized as a difference in protonation site (amine vs. benzene ring), each ion mobility peak differs by one Dalton when coupled with mass spectrometry where the faster mobility peak is the molecular ion peak, and the slower mobility peak is protonated. To complicate the deconvolution of structures, some previous literature shows the peaks as unresolved and thus proposes these species exist in equilibrium. In this work, we show that when measured with high kinetic energy ion mobility spectrometry (HiKE-IMS), the two peaks observed in spectra of both aniline and all n-fluoroanilines are fully separated (chromatographic resolution from 2-7, Rp > 110) and therefore not in equilibrium. The HiKE-IMS is capable of changing ionization conditions independently of drift region conditions, and our results agree with previous literature showing that ionization source settings (including possible fragmentation at this stage) are the only influence determining the speciation of the two aniline peaks. Finally, when the drift and reactant gas are changed to nitrogen, a third peak appears at high E/N for 2-fluoroaniline and 4-fluoroaniline for the first time in reported literature. As observed by HiKE-IMS-MS, the new third peak is also protonated showing that the para-protonated aniline and resulting fragment ion, molecular ion aniline, can be fully separated in the mobility domain for the first time. The appearance of the third peak is only possible due to the increased separation of the other two peaks within the HiKE-IMS.
Collapse
Affiliation(s)
- Cameron N Naylor
- Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, 30167 Hannover, Germany.
| | - Christoph Schaefer
- Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, 30167 Hannover, Germany.
| | - Ansgar T Kirk
- Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, 30167 Hannover, Germany.
| | - Stefan Zimmermann
- Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Department of Sensors and Measurement Technology, 30167 Hannover, Germany.
| |
Collapse
|