Acierno MJ, Seaton D, Mitchell MA, da Cunha A. Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats.
J Am Vet Med Assoc 2010;
237:402-6. [PMID:
20707750 DOI:
10.2460/javma.237.4.402]
[Citation(s) in RCA: 37] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
OBJECTIVE
To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats.
DESIGN
Evaluation study.
ANIMALS
21 cats undergoing routine spaying or neutering.
PROCEDURES
A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units.
RESULTS
There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was -14.9 mm Hg (95% limits of agreement [LOA], -52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, -26.0 to 34.8 mm Hg), and -1.3 mm Hg (95% LOA, -26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was -10.3 mm Hg (95% LOA, -52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, -32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, -32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was -13.4 mm Hg (95% LOA, -51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, -25.5 to 41.6 mm Hg), and -3.6 mm Hg (95% LOA, -31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively.
CONCLUSIONS AND CLINICAL RELEVANCE
Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.
Collapse