Hurley L, Alashban Y, Albeshan S, England A, McEntee MF. The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study.
Diagn Interv Radiol 2023;
29:555-560. [PMID:
37129301 PMCID:
PMC10679606 DOI:
10.4274/dir.2023.232126]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/20/2023] [Accepted: 04/03/2023] [Indexed: 05/03/2023]
Abstract
PURPOSE
The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations.
METHODS
Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student's t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding.
RESULTS
A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (P = 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant.
CONCLUSION
This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols.