Liang J, Fuchs TM, Schäfer R, Kresin VV. Strong permanent magnet gradient deflector for Stern-Gerlach-type experiments on molecular beams.
THE REVIEW OF SCIENTIFIC INSTRUMENTS 2020;
91:053202. [PMID:
32486738 DOI:
10.1063/5.0007602]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/14/2020] [Accepted: 05/04/2020] [Indexed: 06/11/2023]
Abstract
We describe the design, assembly, and testing of a magnet intended to deflect beams of paramagnetic nanoclusters, molecules, and atoms. It is energized by high-grade permanent neodymium magnets. This offers a convenient option in terms of cost, portability, and scalability of the construction while providing field and gradient values (1.1 T, 330 T/m), which are fully comparable with those of commonly used electromagnet deflectors.