Cox JA, Dabek-Zlotorzynska E. High-performance liquid chromatography of sulfur-containing amino acids and related compounds with amperometric detection at a modified electrode.
J Chromatogr A 1991;
543:226-32. [PMID:
1885683 DOI:
10.1016/s0021-9673(01)95772-3]
[Citation(s) in RCA: 22] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/29/2022]
Abstract
Organic disulfides generally are not oxidized at bare electrodes under conditions that are suited to routine amperometric detection, and thiols are typically oxidized in a manner that leads to partial blockage of the surface. Modification of a carbon electrode with a film of Ru(III,IV) oxide stabilized with cyanocross-links permits the amperometric detection of cystine, cysteine, glutathione, methionine, and glutathione disulfide under conditions compatible with their chromatographic separation on a strong cation-exchange column. Detection limits of 0.2-0.6 microM and linear dynamic ranges of at least 1-50 microM were obtained. The electrode was stable for at least 11 days with a pH 1 citrate, phosphate mobile phase.