Han Q, Huo Y, Yang L, Yang X, He Y, Wu J. Determination of Trace Nickel in Water Samples by Graphite Furnace Atomic Absorption Spectrometry after Mixed Micelle-Mediated Cloud Point Extraction.
Molecules 2018;
23:molecules23102597. [PMID:
30309038 PMCID:
PMC6222332 DOI:
10.3390/molecules23102597]
[Citation(s) in RCA: 16] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/03/2018] [Revised: 09/26/2018] [Accepted: 09/27/2018] [Indexed: 11/16/2022] Open
Abstract
A simple and sensitive cloud point extraction method for the preconcentration of ultra-trace amounts of nickel as a prior step to its determination by graphite furnace atomic absorption spectrometry was proposed. It is based on the reaction of nickel with 2-(5-bromo-2-pyridylazo)-5-dimethylaminoaniline (5-Br-PADMA) in HAc–NaAc buffer media and mixed micelle-mediated extraction of the complex using the anionic surfactant sodium dodecyl sulfate sodium (SDS) and non-ionic surfactant (1,1,3,3-Tetramethylbutyl)phenyl-polyethylene (Triton X-114). The optimal reaction and extraction conditions such as pH, concentration of 5-Br-PADMA, SDS and Triton X-114, equilibrium temperature, incubation, and centrifuge time were evaluated and optimized. Under the optimal conditions, the calibration graph was linear over the range 0.1–5.5 ng/mL of nickel with a correlation coefficient of 0.9942. The detection limit obtained was 0.031 ng/mL, and the relative standard deviation was 2.1% for nickel (c = 2 ng/mL, n = 6). The proposed method was successfully applied to the determination of nickel in water samples.
Collapse