Abstract
BACKGROUND
Validated UPLC method and TLC densitometric method were prescribed for determination of antihypertensive components. Objectives: To establish and validate rapid and accurate Ultra performance liquid chromatographic (UPLC) and TLC densitometric methods for determination of Xipamide and Triamterene in pure and dosage forms.
METHODS
The first method; UPLC method, depended on using Agilent Zorbax Eclipse Plus C8 (50 mm × 2.1 mm, 1.8 μm), as the column, mobile phase composed of (acetonitrile-water) (70 + 30, v/v) adjusted by acetic acid to obtain (pH 3), 0.2 mL/min flow rate and UV detection at 231.4 nm. The second method was a thin layer chromatography (TLC) densitometric method, separation was achieved by using toluene-methanol-ethyl chloride-acetic acid (7 + 2 + 1 + 0.2, v/v/v) as the mobile phase, pre coated silica gel plates as the stationary phase and UV detection at 300.0 nm.
RESULTS
The obtained results were validated and statistically compared with official and reported methods. The obtained results showed high accuracy and reproducible results with excellent mean recoveries for both drugs.
CONCLUSIONS
The UPLC method showed shorter retention time for both Xipamide (0.88 min) and Triamterene (0.63 min), lower detection limit less than 0.055 µg/mL for both drugs with high selectivity, decreased injection volume (1 µL) and lower flow rate other than any HPLC method. Both proposed methods were sensitive, selective, and effectively applied to pure and dosage forms (Epitens®).
HIGHLIGHTS
Unprecedented sensitive, rapid, and reproducible UPLC and TLC methods were developed for selective determination of mixture of Xipamide and Triamterene with LOD less than 0.076 µg/mL for both drugs.