1
|
Asadollahzadeh N, Razavi S, Zare MH. A survey on mean glandular dose in mammography examination and the factors affecting it in Shahid Sadoughi Hospital, Yazd, Iran. RADIATION PROTECTION DOSIMETRY 2024; 200:809-821. [PMID: 38811346 DOI: 10.1093/rpd/ncae121] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/26/2023] [Revised: 02/12/2024] [Accepted: 05/02/2024] [Indexed: 05/31/2024]
Abstract
BACKGROUND mammography is the gold standard in the early detection of breast cancer. Due to the increase in the rate of women suffering from this malignancy all over the world, this imaging modality has been widely used. Considering the side effects caused by ionizing radiation to measure the carcinogenic risk of mammography X-rays, mean glandular dose (MGD) is the best parameter to evaluate the dose received by patients undergoing mammography. The aims of this study were to measure MGD in mammography in mammographic craniocaudal (CC) and mediolateral oblique (MLO) projections and investigate the relationship of MGD with compressed breast thickness (CBT), body mass index, age of the patient, and device exposure conditions. MATERIALS AND METHODS this cross-sectional analytical study was conducted on women aged 30-70 referring to the mammography unit of Shahid Sadoughi Hospital in Yazd city from May to August 2022. The TLD-GR 200 (thermoluminescence dosimeter) was placed on the breast of the patients for CC and MLO projections, and then the MGD was obtained by multiplying the entrance surface dose and the normalized glandular dose. Analysis of data (such as demographic information of patients, CBT, kVp, and mAs) was done by SPSS 23 software. The normality of the data was checked using Shapiro-Wilk tests. It was found that except for age, other variables did not have a normal distribution; therefore, equivalent parametric and nonparametric tests were used. In this regard, Spearman's correlation was used to assess the correlation between variables. P-value < .05 was considered as level of significance. RESULTS The mean ± standard deviation (SD) age of patients was 47.3 ± 7.1 years. The median ± IQR (the interquartile range (IQR) is a measure of statistical dispersion, which is the spread of the data. It is defined as the difference between the 75th and 25th percentiles of the data.) (mean ± SD) value of MGD per woman was 1.2 ± 0.4 mGy (1.3 ± 0.3 mGy). The median ± IQR MGD in the MLO and CC projections was 1.6 ± 0.6 mGy and 0.9 ± 0.4 mGy, respectively. Significant relationships (P-value ≤ .001) were observed between MGD with CBT (R = 0.62) and age (R = -0.85). CONCLUSION The results showed that the mammography unit at Shahid Sadoughi Hospital in Yazd is functioning properly. The calculated median ± IQR MGD per woman referring to this unit (1.2 ± 0.4 mGy) was clearly below the dose limit recommended by American College of Radiology and International Commission on Radiological Protection (3 mGy). Moreover, among the factors affecting MGD, the highest correlation was seen between MGD and age (R = -0.85).
Collapse
Affiliation(s)
- Najmeh Asadollahzadeh
- Yazd Shahid Sadoughi University of Medical Sciences and Health Services, Yazd, 8915173149, Iran
| | - SeidKazem Razavi
- Medical Physics Department, Yazd Shahid Sadoughi University of Medical Sciences and Health Services, Yazd, 8915164481, Iran
| | - Mohammad Hossein Zare
- Medical Physics Department, Yazd Shahid Sadoughi University of Medical Sciences and Health Services, Yazd, 8915164481, Iran
| |
Collapse
|
2
|
Marshall NW, Bosmans H. Performance evaluation of digital breast tomosynthesis systems: physical methods and experimental data. Phys Med Biol 2022; 67. [DOI: 10.1088/1361-6560/ac9a35] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/22/2022] [Accepted: 10/13/2022] [Indexed: 11/17/2022]
Abstract
Abstract
Digital breast tomosynthesis (DBT) has become a well-established breast imaging technique, whose performance has been investigated in many clinical studies, including a number of prospective clinical trials. Results from these studies generally point to non-inferiority in terms of microcalcification detection and superior mass-lesion detection for DBT imaging compared to digital mammography (DM). This modality has become an essential tool in the clinic for assessment and ad-hoc screening but is not yet implemented in most breast screening programmes at a state or national level. While evidence on the clinical utility of DBT has been accumulating, there has also been progress in the development of methods for technical performance assessment and quality control of these imaging systems. DBT is a relatively complicated ‘pseudo-3D’ modality whose technical assessment poses a number of difficulties. This paper reviews methods for the technical performance assessment of DBT devices, starting at the component level in part one and leading up to discussion of system evaluation with physical test objects in part two. We provide some historical and basic theoretical perspective, often starting from methods developed for DM imaging. Data from a multi-vendor comparison are also included, acquired under the medical physics quality control protocol developed by EUREF and currently being consolidated by a European Federation of Organisations for Medical Physics working group. These data and associated methods can serve as a reference for the development of reference data and provide some context for clinical studies.
Collapse
|
3
|
Evaluation of Breast Galactography Using Digital Breast Tomosynthesis: A Clinical Exploratory Study. Diagnostics (Basel) 2021; 11:diagnostics11112060. [PMID: 34829407 PMCID: PMC8622426 DOI: 10.3390/diagnostics11112060] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/28/2021] [Revised: 11/04/2021] [Accepted: 11/05/2021] [Indexed: 11/28/2022] Open
Abstract
Objectives: To compare the application value of digital breast tomosynthesis (DBT) and full-field digital mammography (FFDM) in breast galactography. Materials and Methods: A total of 128 patients with pathological nipple discharge (PND) were selected to undergo galactography. DBT and FFDM were performed for each patient after injecting the contrast agent; the radiation dose of DBT and FFDM was calculated, and the image quality was evaluated in consensus by two senior breast radiologists. Histopathologic data were found in 49 of the 128 patients. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for both FFDM- and DBT-galactography were calculated using histopathologic results as a reference standard. Data were presented as percentages along with their 95% confidence intervals (CI). Results: The average age of the 128 patients was 46.53 years. The average glandular dose (AGD) of DBT-galactography was slightly higher than that of FFDM-galactography (p < 0.001). DBT-galactography was 30.7% higher than FFDM-galactography in CC view, while DBT-galactography increased by 21.7% compared with FFDM-galactography in ML view. Regarding catheter anatomic distortion, structure detail, and overall image quality groups, DBT scores were higher than FFDM scores, and the differences were significant for all measures (p < 0.05). In 49 patients with pathological nipple discharge, we found that the DBT-galactography had higher sensitivity, specificity, PPV, and NPV (93.3%, 75%, 97.7%, and 50%, respectively) than FFDM-galactography (91.1%, 50%, 95.3%, and 33.3%, respectively). Conclusions: Compared to FFDM-galactography, within the acceptable radiation dose range, DBT-galactography increases the sensitivity and specificity of lesion detection by improving the image quality, providing more confidence for the diagnosis of clinical ductal lesions.
Collapse
|
4
|
Aşlar E, Şahiner E, Polymeris GS, Meriç N. Feasibility of determining Entrance Surface Dose (ESD) and mean glandular dose (MGD) using OSL signal from BeO dosimeters in mammography. Radiat Phys Chem Oxf Engl 1993 2020. [DOI: 10.1016/j.radphyschem.2020.109151] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/23/2022]
|
5
|
Shinohara S, Araki F, Maeda M, Okamoto R, Nakamura M, Higashida Y. Indices for the evaluation of glandular dose heterogeneity in full-field digital mammography. JOURNAL OF RADIOLOGICAL PROTECTION : OFFICIAL JOURNAL OF THE SOCIETY FOR RADIOLOGICAL PROTECTION 2020; 40:1429-1443. [PMID: 33120368 DOI: 10.1088/1361-6498/abc604] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/22/2020] [Accepted: 10/29/2020] [Indexed: 06/11/2023]
Abstract
This study aims to evaluate the indices of glandular dose heterogeneity in full-field digital mammography. The distributions of GD in a breast phantom with a skin layer of 4 mm were determined using the Monte Carlo method with simulated x-ray fluence spectra. First, the GD to air kerma (GD/Kair) volume histogram was obtained from the GD distributions, which were indicated by the glandular volume (%) as a function of GD/Kair. The GD indices, namely, the maximum glandular dose (GD2%) and glandular volume percentage (%) receiving at least the mean glandular dose (MGD) (VMGD) were calculated from the GD/Kairvolume histogram. Next, the scatter plots of GD2%/MGD andVMGDwere drawn as functions of the normalised mean glandular dose (DgN). Finally, (GD2%)iand (VMGD)iwere obtained from the relationship between the GD indices and DgN for 596 clinical irradiation cases based on individual irradiation conditions. The values of GD2%/MGD were more affected by breast thickness than glandularity and tube voltage, and they decreased according to the power law of DgN for all the target/filter combinations. The values ofVMGDwere proportional to DgN and decreased with increase in the compressed breast thickness. The values of (MGD)iand (GD2%)ifor 596 clinical irradiation cases were estimated to range from 0.6-3.0 mGy to 1.1-7.0 mGy, respectively, and (VMGD)iwas in the range 32%-48%. (GD2%)iand (VMGD)iare mainly affected by breast thickness. These indices are useful for the evaluation of glandular dose heterogeneity in mammography.
Collapse
Affiliation(s)
- Sae Shinohara
- Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan
- Canon Medical Systems Corporation, 1385 Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan
| | - Fujio Araki
- Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan
| | - Megumi Maeda
- Department of Radiology, Sasebo City General Hospital, 9-3 Hirase, Sasebo, Nagasaki 857-8511, Japan
| | - Rumi Okamoto
- Department of Radiology, Social Medical Corporation Hakuaikai Sagara Hospital, 3-31, Matsubara, Kagoshima 892-0833, Japan
| | - Mai Nakamura
- Department of Radiological Technology, Faculty of Fukuoka Medical Technology, Teikyo University, 6-22 Misaki, Ohmuta, Fukuoka 836-8505, Japan
| | - Yoshiharu Higashida
- Department of Radiological Technology, Faculty of Fukuoka Medical Technology, Teikyo University, 6-22 Misaki, Ohmuta, Fukuoka 836-8505, Japan
| |
Collapse
|
6
|
Lee C, Han M, Baek J. Human observer performance on in-plane digital breast tomosynthesis images: Effects of reconstruction filters and data acquisition angles on signal detection. PLoS One 2020; 15:e0229915. [PMID: 32163472 PMCID: PMC7067468 DOI: 10.1371/journal.pone.0229915] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/15/2019] [Accepted: 02/17/2020] [Indexed: 11/29/2022] Open
Abstract
For digital breast tomosynthesis (DBT) systems, we investigate the effects of the reconstruction filters for different data acquisition angles on signal detection. We simulated a breast phantom with a 30% volume glandular fraction (VGF) of breast anatomy using the power law spectrum and modeled the breast mass as a spherical object with a 1 mm diameter. Projection data were acquired using two different data acquisition angles and numbers of projection view pairs, and in-plane breast images were reconstructed using the Feldkamp-Davis-Kress (FDK) algorithm with three different reconstruction filter schemes. To measure the ability to detect a signal, we conducted the human observer study with a binary detection task and compared the signal detectability of human to that of channelized Hotelling observer (CHO) with Laguerre-Gauss (LG) channels and dense difference-of-Gaussian (D-DOG) channels. We also measured the contrast-to-noise ratio (CNR), signal power spectrum (SPS), and β values of the anatomical noise power spectrum (NPS) to show the association between human observer performance and these traditional metrics. Our results show that using a slice thickness (ST) filter degraded the signal detection performance of human observers at the same data acquisition angle. This could be predicted by D-DOG CHO with internal noise, but the correlation between the traditional metrics and signal detectability was not observed in this work.
Collapse
Affiliation(s)
- Changwoo Lee
- Center for Medical Convergence Metrology, Korea Research Institute of Standards and Science (KRISS), Daejeon, South Korea
| | - Minah Han
- School of Integrated Technology and Yonsei Institute of Convergence Technology, Yonsei University, Incheon, South Korea
| | - Jongduk Baek
- School of Integrated Technology and Yonsei Institute of Convergence Technology, Yonsei University, Incheon, South Korea
- * E-mail:
| |
Collapse
|
7
|
Chijoke WO, Adeniji-Sofoluwe AT, Jibiri NN. Evaluation of mean glandular dose and assessment of the risk of radiation induced carcinogenesis in women following screening mammography in a low resource setting. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES 2019. [DOI: 10.1016/j.jrras.2017.07.002] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 10/19/2022]
Affiliation(s)
| | - Adenike T. Adeniji-Sofoluwe
- University of Ibadan, College of Medicine, Department of Radiology, University College Hospital, Ibadan, Nigeria
| | - Norbert N. Jibiri
- University of Ibadan, Faculty of Sciences, Department of Physics, Ibadan, Nigeria
| |
Collapse
|
8
|
McLean D, Ryan L, Robertson L. Analysis of patient breast dose from a mammographic biopsy unit. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2019; 42:193-199. [PMID: 30707352 DOI: 10.1007/s13246-019-00729-6] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/12/2018] [Accepted: 01/21/2019] [Indexed: 11/25/2022]
Abstract
Further assessment of suspicious lesions found during asymptomatic breast cancer screening is critical and involves mammographic follow up with biopsy. The X-ray procedure is complex and variable in nature and until now there is little information on the radiation dose to the breast or associated risks. A survey of radiation doses from a Siemens MammoTest prone biopsy with the support of a Sectra L30 AIR mammographic unit for workup and post clip images has been completed. Procedure details and outcomes, including radiographic and patient related variables have been collected and analysed using standard dosimetric formulation. The partial irradiation of the breast in biopsy and magnification views was considered. The average mean glandular breast dose was 5.13 mGy, comprising of 3.52 mGy from the biopsy procedure and 1.61 mGy from the workup and post clip images, with an average of 8.4 biopsy images and 5.8 workup and post clip images. The risk from these dose levels are dependent on the age of the woman, however are not considered high for a symptomatic X-ray procedure.
Collapse
Affiliation(s)
- Donald McLean
- Medical Physics and Radiation Engineering, The Canberra Hospital, PO Box 11, Woden, ACT, 2606, Australia.
| | - Lewis Ryan
- Medical Physics and Radiation Engineering, The Canberra Hospital, PO Box 11, Woden, ACT, 2606, Australia
| | | |
Collapse
|
9
|
Depuydt J, Viaene T, Blondeel P, Roche N, Van den Broecke R, Thierens H, Vral A. DNA double strand breaks induced by low dose mammography X-rays in breast tissue: A pilot study. Oncol Lett 2018; 16:3394-3400. [PMID: 30127940 DOI: 10.3892/ol.2018.9024] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/07/2017] [Accepted: 04/26/2018] [Indexed: 12/18/2022] Open
Abstract
Breast tissue is very sensitive to ionizing radiation due to the presence of reproductive hormones, including estrogen. In the present pilot study, the efficiency of mammography X-rays to induce DNA double strand breaks (DSB) in mammary epithelial cells was investigated. For this, freshly resected healthy breast tissue was irradiated with 30 kV mammography X-rays in the dose range 0-500 mGy (2, 4, 10, 20, 40, 100 and 500 mGy). Breast specimens were also irradiated with identical doses of 60Co γ-rays as a radiation quality standard. With the γH2AX-foci assay, the number of DNA DSB induced by radiation were quantified in the mammary epithelial cells present in breast tissue. Results indicated that foci induced by 30 kV X-rays and γ-rays followed a biphasic linear dose-response. For 30 kV X-rays, the slope in the low dose region (0-20 mGy) was 8.71 times steeper compared with the slope in the higher dose region (20-500 mGy). Furthermore, compared with γ-rays, 30 kV X-rays were also more effective in inducing γH2AX-foci. This resulted in a relative biological effectiveness (RBE) value of 1.82 in the low dose range. In the higher dose range, an RBE close to 1 was obtained. In conclusion, the results indicated the existence of a low dose hypersensitive response for DSB induction in the dose range representative for mammography screening, which is probably caused by the bystander effect. This could affect the radiation risk calculations for women participating in mammography screening.
Collapse
Affiliation(s)
- Julie Depuydt
- Department of Basic Medical Sciences, University of Ghent, 9000 Ghent, Belgium
| | - Tanguy Viaene
- Department of Basic Medical Sciences, University of Ghent, 9000 Ghent, Belgium
| | | | - Nathalie Roche
- Plastic Surgery, Ghent University Hospital, 9000 Ghent, Belgium
| | | | - Hubert Thierens
- Department of Basic Medical Sciences, University of Ghent, 9000 Ghent, Belgium
| | - Anne Vral
- Department of Basic Medical Sciences, University of Ghent, 9000 Ghent, Belgium
| |
Collapse
|
10
|
Acho SN, Boonzaier WPE, Nel IF. EXPOSURE PARAMETERS OF MAMMOGRAMS WITH AND WITHOUT MASS LESIONS FROM A SOUTH AFRICAN BREAST CARE CENTRE. RADIATION PROTECTION DOSIMETRY 2017; 177:348-355. [PMID: 28444247 DOI: 10.1093/rpd/ncx053] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 02/09/2017] [Accepted: 04/04/2017] [Indexed: 06/07/2023]
Abstract
In South African breast care centres, full-field digital mammography units provide breast imaging services to symptomatic and asymptomatic women simultaneously. This study evaluated the technical exposure parameters of 800 mammograms of which 100 mammograms had obvious mass lesions in the fibroglandular tissue. The average breast compression force of mammograms with mass lesions in the fibroglandular tissue was 18.4% less than the average breast compression force of mammograms without mass lesions. The average mean glandular dose (MGD), tube potential (kVp) and compressed breast thickness (CBT) values were 2.14 mGy, 30.5 kVp and 63.9 mm, respectively, for mammograms with mass lesions, and 1.45 mGy, 29.6 kVp and 56.9 mm, respectively, for mammograms without mass lesions. Overall, the average MGD and mean CBT of mammograms with mass lesion were significantly higher compared to those without mass lesions (p < 0.05), although there was no significant difference in their tube potentials (p > 0.05).
Collapse
Affiliation(s)
- Sussan N Acho
- Department of Medical Physics, University of the Free State, Bloemfontein 9300, South Africa
| | - Willem P E Boonzaier
- Department of Medical Physics, University of the Free State, Bloemfontein 9300, South Africa
| | - Ina F Nel
- Department of Clinical Imaging Services, University of the Free State, Bloemfontein 9300, South Africa
| |
Collapse
|
11
|
Heggie JCP, Barnes P, Cartwright L, Diffey J, Tse J, Herley J, McLean ID, Thomson FJ, Grewal RK, Collins LT. Position paper: recommendations for a digital mammography quality assurance program V4.0. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2017; 40:491-543. [PMID: 28914430 DOI: 10.1007/s13246-017-0583-x] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/04/2017] [Indexed: 12/21/2022]
Abstract
In 2001 the ACPSEM published a position paper on quality assurance in screen film mammography which was subsequently adopted as a basis for the quality assurance programs of both the Royal Australian and New Zealand College of Radiologists (RANZCR) and of BreastScreen Australia. Since then the clinical implementation of digital mammography has been realised and it has become evident that existing screen-film protocols were not appropriate to assure the required image quality needed for reliable diagnosis or to address the new dose implications resulting from digital technology. In addition, the advantages and responsibilities inherent in teleradiology are most critical in mammography and also need to be addressed. The current document is the result of a review of current overseas practice and local experience in these areas. At this time the technology of digital imaging is undergoing significant development and there is still a lack of full international consensus about some of the detailed quality control (QC) tests that should be included in quality assurance (QA) programs. This document describes the current status in digital mammography QA and recommends test procedures that may be suitable in the Australasian environment. For completeness, this document also includes a review of the QA programs required for the various types of digital biopsy units used in mammography. In the future, international harmonisation of digital quality assurance in mammography and changes in the technology may require a review of this document. Version 2.0 represented the first of these updates and key changes related to image quality evaluation, ghost image evaluation and interpretation of signal to noise ratio measurements. In Version 3.0 some significant changes, made in light of further experience gained in testing digital mammography equipment were introduced. In Version 4.0, further changes have been made, most notably digital breast tomosynthesis (DBT) testing and QC have been addressed. Some additional testing for conventional projection imaging has been added in order that sites may have the capability to undertake dose surveys to confirm compliance with diagnostic reference levels (DRLs) that may be established at the National or State level. A key recommendation is that dosimetry calculations are now to be undertaken using the methodology of Dance et al. Some minor changes to existing facility QC tests have been made to ensure the suggested procedures align with those most recently adopted by the Royal Australian and New Zealand College of Radiologists and BreastScreen Australia. Future updates of this document may be provided as deemed necessary in electronic format on the ACPSEM's website ( https://www.acpsem.org.au/whatacpsemdoes/standards-position-papers and see also http://www.ranzcr.edu.au/quality-a-safety/radiology/practice-quality-activities/mqap ).
Collapse
Affiliation(s)
| | - P Barnes
- I-Med Radiology Network, Head office Melbourne, Melbourne, Australia
| | - L Cartwright
- Medical Physics Department, Westmead Hospital, Westmead, Australia
| | - J Diffey
- Hunter New England Imaging, John Hunter Hospital, New Lambton Heights, Australia
| | - J Tse
- Medical Physics and Radiation Engineering, Canberra Hospital, Canberra, Australia
| | - J Herley
- Radiation Protection Services Pty Ltd, Brisbane, Australia
| | - I D McLean
- Medical Physics and Radiation Engineering, Canberra Hospital, Canberra, Australia
| | - F J Thomson
- Radiological Physics Consultants Ltd, Warrington, New Zealand
| | - R K Grewal
- Medical Physics Department, Westmead Hospital, Westmead, Australia
| | - L T Collins
- Medical Physics Department, Westmead Hospital, Westmead, Australia
| |
Collapse
|
12
|
Castillo-García M, Chevalier M, Garayoa J, Rodriguez-Ruiz A, García-Pinto D, Valverde J. Automated Breast Density Computation in Digital Mammography and Digital Breast Tomosynthesis: Influence on Mean Glandular Dose and BIRADS Density Categorization. Acad Radiol 2017; 24:802-810. [PMID: 28214227 DOI: 10.1016/j.acra.2017.01.011] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/13/2016] [Revised: 12/16/2016] [Accepted: 01/08/2017] [Indexed: 12/20/2022]
Abstract
RATIONALE AND OBJECTIVES The study aimed to compare the breast density estimates from two algorithms on full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) and to analyze the clinical implications. MATERIALS AND METHODS We selected 561 FFDM and DBT examinations from patients without breast pathologies. Two versions of a commercial software (Quantra 2D and Quantra 3D) calculated the volumetric breast density automatically in FFDM and DBT, respectively. Other parameters such as area breast density and total breast volume were evaluated. We compared the results from both algorithms using the Mann-Whitney U non-parametric test and the Spearman's rank coefficient for data correlation analysis. Mean glandular dose (MGD) was calculated following the methodology proposed by Dance et al. RESULTS Measurements with both algorithms are well correlated (r ≥ 0.77). However, there are statistically significant differences between the medians (P < 0.05) of most parameters. The volumetric and area breast density median values from FFDM are, respectively, 8% and 77% higher than DBT estimations. Both algorithms classify 35% and 55% of breasts into BIRADS (Breast Imaging-Reporting and Data System) b and c categories, respectively. There are no significant differences between the MGD calculated using the breast density from each algorithm. DBT delivers higher MGD than FFDM, with a lower difference (5%) for breasts in the BIRADS d category. MGD is, on average, 6% higher than values obtained with the breast glandularity proposed by Dance et al. CONCLUSIONS Breast density measurements from both algorithms lead to equivalent BIRADS classification and MGD values, hence showing no difference in clinical outcomes. The median MGD values of FFDM and DBT examinations are similar for dense breasts (BIRADS d category).
Collapse
|
13
|
Geeraert N, Klausz R, Muller S, Bloch I, Bosmans H. Evaluation of exposure in mammography: limitations of average glandular dose and proposal of a new quantity. RADIATION PROTECTION DOSIMETRY 2015; 165:342-345. [PMID: 25897143 DOI: 10.1093/rpd/ncv069] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/04/2023]
Abstract
The radiation risk in mammography is traditionally evaluated using the average glandular dose. This quantity for the average breast has proven to be useful for population statistics and to compare exposure techniques and systems. However it is not indicating the individual radiation risk based on the individual glandular amount and distribution. Simulations of exposures were performed for six appropriate virtual phantoms with varying glandular amount and distribution. The individualised average glandular dose (iAGD), i.e. the individual glandular absorbed energy divided by the mass of the gland, and the glandular imparted energy (GIE), i.e. the glandular absorbed energy, were computed. Both quantities were evaluated for their capability to take into account the glandular amount and distribution. As expected, the results have demonstrated that iAGD reflects only the distribution, while GIE reflects both the glandular amount and distribution. Therefore GIE is a good candidate for individual radiation risk assessment.
Collapse
Affiliation(s)
- N Geeraert
- Departement of Radiology, Katholieke Universiteit Leuven, Herestraat 49, Leuven 3000, Belgium Digital Guidance Solutions, GE Healthcare, 283 rue de la Minière, Buc 78530, France Institut Mines-Télécom, Télécom ParisTech, CNRS LTCI, 46 rue Barrault, Paris 75013, France
| | - R Klausz
- Digital Guidance Solutions, GE Healthcare, 283 rue de la Minière, Buc 78530, France
| | - S Muller
- Digital Guidance Solutions, GE Healthcare, 283 rue de la Minière, Buc 78530, France
| | - I Bloch
- Institut Mines-Télécom, Télécom ParisTech, CNRS LTCI, 46 rue Barrault, Paris 75013, France
| | - H Bosmans
- Departement of Radiology, Katholieke Universiteit Leuven, Herestraat 49, Leuven 3000, Belgium
| |
Collapse
|
14
|
Kelaranta A, Toroi P, Timonen M, Komssi S, Kortesniemi M. Conformance of mean glandular dose from phantom and patient data in mammography. RADIATION PROTECTION DOSIMETRY 2015; 164:342-353. [PMID: 25114321 DOI: 10.1093/rpd/ncu261] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/14/2014] [Accepted: 07/18/2014] [Indexed: 06/03/2023]
Abstract
In mammography dosimetry, phantoms are often used to represent breast tissue. The conformance of phantom- and patient-based mean glandular dose (MGD) estimates was evaluated mainly from the aspect of diagnostic reference levels. Patient and phantom exposure data were collected for eight diagnostic and three screening mammography devices. More extensive assessments were performed for two devices. The average breast thickness was close to the nationally used reference of 50 mm in diagnostic (50 mm, SD = 13 mm, n = 5342) and screening (47 mm, SD = 13 mm, n = 395) examinations. The average MGD for all breasts differed by 2% from the MGD determined for breasts in the limited compressed thickness range of 40-60 mm. The difference between phantom- and patient-based MGD estimations was up to 30%. Therefore, phantom measurements cannot replace patient dose data in MGD determination.
Collapse
Affiliation(s)
- A Kelaranta
- STUK-Radiation and Nuclear Safety Authority, Laippatie 4, PO Box 14, Helsinki FI-00881, Finland Helsinki Medical Imaging Center, University Hospital of Helsinki, PO Box 340, Helsinki FI-00290, Finland Department of Physics, University of Helsinki, PO Box 64, Helsinki FI-00014, Finland
| | - P Toroi
- STUK-Radiation and Nuclear Safety Authority, Laippatie 4, PO Box 14, Helsinki FI-00881, Finland
| | - M Timonen
- Helsinki Medical Imaging Center, University Hospital of Helsinki, PO Box 340, Helsinki FI-00290, Finland
| | - S Komssi
- Suomen Terveystalo Ltd., Jaakonkatu 3B, Helsinki FI-00100, Finland
| | - M Kortesniemi
- Helsinki Medical Imaging Center, University Hospital of Helsinki, PO Box 340, Helsinki FI-00290, Finland
| |
Collapse
|
15
|
Leon SM, Brateman LF, Wagner LK. Characterization of scatter in digital mammography from use of Monte Carlo simulations and comparison to physical measurements. Med Phys 2014; 41:111914. [PMID: 25370647 DOI: 10.1118/1.4894808] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE Monte Carlo simulations were performed with the goal of verifying previously published physical measurements characterizing scatter as a function of apparent thickness. A secondary goal was to provide a way of determining what effect tissue glandularity might have on the scatter characteristics of breast tissue. The overall reason for characterizing mammography scatter in this research is the application of these data to an image processing-based scatter-correction program. METHODS mcnpx was used to simulate scatter from an infinitesimal pencil beam using typical mammography geometries and techniques. The spreading of the pencil beam was characterized by two parameters: mean radial extent (MRE) and scatter fraction (SF). The SF and MRE were found as functions of target, filter, tube potential, phantom thickness, and the presence or absence of a grid. The SF was determined by separating scatter and primary by the angle of incidence on the detector, then finding the ratio of the measured scatter to the total number of detected events. The accuracy of the MRE was determined by placing ring-shaped tallies around the impulse and fitting those data to the point-spread function (PSF) equation using the value for MRE derived from the physical measurements. The goodness-of-fit was determined for each data set as a means of assessing the accuracy of the physical MRE data. The effect of breast glandularity on the SF, MRE, and apparent tissue thickness was also considered for a limited number of techniques. RESULTS The agreement between the physical measurements and the results of the Monte Carlo simulations was assessed. With a grid, the SFs ranged from 0.065 to 0.089, with absolute differences between the measured and simulated SFs averaging 0.02. Without a grid, the range was 0.28-0.51, with absolute differences averaging -0.01. The goodness-of-fit values comparing the Monte Carlo data to the PSF from the physical measurements ranged from 0.96 to 1.00 with a grid and 0.65 to 0.86 without a grid. Analysis of the data suggested that the nongrid data could be better described by a biexponential function than the single exponential used here. The simulations assessing the effect of breast composition on SF and MRE showed only a slight impact on these quantities. When compared to a mix of 50% glandular/50% adipose tissue, the impact of substituting adipose or glandular breast compositions on the apparent thickness of the tissue was about 5%. CONCLUSIONS The findings show agreement between the physical measurements published previously and the Monte Carlo simulations presented here; the resulting data can therefore be used more confidently for an application such as image processing-based scatter correction. The findings also suggest that breast composition does not have a major impact on the scatter characteristics of breast tissue. Application of the scatter data to the development of a scatter-correction software program can be simplified by ignoring the variations in density among breast tissues.
Collapse
Affiliation(s)
- Stephanie M Leon
- Department of Diagnostic and Interventional Imaging, University of Texas Medical School at Houston, Houston, Texas 77030
| | - Libby F Brateman
- Department of Radiology, University of Florida, Gainesville, Florida 32610
| | - Louis K Wagner
- Department of Diagnostic and Interventional Imaging, University of Texas Medical School at Houston, Houston, Texas 77030
| |
Collapse
|
16
|
Mittone A, Bravin A, Coan P. Radiation dose in breast CT imaging with monochromatic x-rays: simulation study of the influence of energy, composition and thickness. Phys Med Biol 2014; 59:2199-217. [DOI: 10.1088/0031-9155/59/9/2199] [Citation(s) in RCA: 24] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/23/2022]
|
17
|
Vedantham S, Shi L, Karellas A, O'Connell AM, Conover DL. Personalized estimates of radiation dose from dedicated breast CT in a diagnostic population and comparison with diagnostic mammography. Phys Med Biol 2013; 58:7921-36. [PMID: 24165162 DOI: 10.1088/0031-9155/58/22/7921] [Citation(s) in RCA: 37] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
This study retrospectively analyzed the mean glandular dose (MGD) to 133 breasts from 132 subjects, all women, who participated in a clinical trial evaluating dedicated breast CT in a diagnostic population. The clinical trial was conducted in adherence to a protocol approved by institutional review boards and the study participants provided written informed consent. Individual estimates of MGD to each breast from dedicated breast CT was obtained by combining x-ray beam characteristics with estimates of breast dimensions and fibroglandular fraction from volumetric breast CT images, and using normalized glandular dose coefficients. For each study participant and for the breast corresponding to that imaged with breast CT, an estimate of the MGD from diagnostic mammography (including supplemental views) was obtained from the DICOM image headers for comparison. This estimate uses normalized glandular dose coefficients corresponding to a breast with 50% fibroglandular weight fraction. The median fibroglandular weight fraction for the study cohort determined from volumetric breast CT images was 15%. Hence, the MGD from diagnostic mammography was corrected to be representative of the study cohort. Individualized estimates of MGD from breast CT ranged from 5.7 to 27.8 mGy. Corresponding to the breasts imaged with breast CT, the MGD from diagnostic mammography ranged from 2.6 to 31.6 mGy. The mean (± inter-breast SD) and the median MGD (mGy) from dedicated breast CT exam were 13.9 ± 4.6 and 12.6, respectively. For the corresponding breasts, the mean (± inter-breast SD) and the median MGD (mGy) from diagnostic mammography were 12.4 ± 6.3 and 11.1, respectively. Statistical analysis indicated that at the 0.05 level, the distributions of MGD from dedicated breast CT and diagnostic mammography were significantly different (Wilcoxon signed ranks test, p = 0.007). While the interquartile range and the range (maximum-minimum) of MGD from dedicated breast CT was lower than diagnostic mammography, the median MGD from dedicated breast CT was approximately 13.5% higher than that from diagnostic mammography. The MGD for breast CT is based on a 1.45 mm skin layer and that for diagnostic mammography is based on a 4 mm skin layer; thus, favoring a lower estimate for MGD from diagnostic mammography. The median MGD from dedicated breast CT corresponds to the median MGD from four to five diagnostic mammography views. In comparison, for the same 133 breasts, the mean and the median number of views per breast during diagnostic mammography were 4.53 and 4, respectively. Paired analysis showed that there was approximately equal likelihood of receiving lower MGD from either breast CT or diagnostic mammography. Future work will investigate methods to reduce and optimize radiation dose from dedicated breast CT.
Collapse
Affiliation(s)
- Srinivasan Vedantham
- Department of Radiology, University of Massachusetts Medical School, Worcester, MA 01655, USA
| | | | | | | | | |
Collapse
|
18
|
Borg M, Badr I, Royle GJ. Mammography equipment performance, image quality and mean glandular dose in Malta. RADIATION PROTECTION DOSIMETRY 2013; 156:168-183. [PMID: 23525916 DOI: 10.1093/rpd/nct051] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/02/2023]
Abstract
In this first Maltese national mammography survey, the effectiveness of direct digital (DR) mammography in breast cancer screening has been confirmed. Patient data were made available from three clinics out of the participating nine. A dose survey of mean glandular dose (MGD) calculated for 759 patients examined in the state-owned mammography facilities was performed. An MGD national diagnostic reference level was set at 1.87 mGy for patients with breast compression thicknesses (BCT) between 5.0 and 7.0 cm. This range was selected since patient data were retrieved from three clinics only and the results showed that other international BCT reference levels may be unsuitable for the Maltese population. In fact, the overall average BCT was 5.75 ± 1.4 cm. The survey results have shown that the technical standard of mammographic equipment in the Malta National Breast Screening Programme is on a par with other countries, including its Western European counterparts.
Collapse
Affiliation(s)
- M Borg
- Medical Imaging Department, Mater Dei Hospital, Tal-Qroqq, Msida MSD2090, Malta.
| | | | | |
Collapse
|
19
|
Depuydt J, Baert A, Vandersickel V, Thierens H, Vral A. Relative biological effectiveness of mammography X-rays at the level of DNA and chromosomes in lymphocytes. Int J Radiat Biol 2013; 89:532-8. [PMID: 23484479 DOI: 10.3109/09553002.2013.782447] [Citation(s) in RCA: 22] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022]
Abstract
PURPOSE In many countries, breast cancer screening programs based on periodic mammography exist, giving a large group of women regularly a small dose of ionizing radiation. In order to assess the benefit/risk ratio of those programs the relative biological effectiveness (RBE) of mammography X-rays needs to be determined. MATERIALS AND METHODS Blood of five healthy donors was irradiated in vitro with 30 kV X-rays and (60)Co γ-rays with doses between 5 and 2000 mGy. The phosphorylated histone subtype H2A isoform X-foci (γH2AX-foci) technique was used to quantify the number of DNA double-strand breaks (DSB) after irradiation. Chromosomal damage resulting from non- or misrepaired DNA DSB was quantified with the micronucleus (MN)-assay and the sensitivity was improved by counting only centromere negative micronuclei (MNCM-). RESULTS The threshold detection dose obtained with the γH2AX-foci test was 10 mGy for mammography X-rays compared to 50 mGy for γ-rays. With the MN-assay respectively MN-centromere-assay threshold detection doses of 100, respectively, 50 mGy were obtained for mammography X-rays compared to 200 respectively 100 mGy for γ-rays. An RBE of 1.4 was obtained with the γH2AX-foci assay. With the MN-assays low-dose RBE values between 3 and 4 were determined. CONCLUSION Our results indicate that exposure to mammography X-rays resulted in a modest increase in the induction of DSB compared to γ-rays. However, due to the higher linear energy transfer (LET) of mammography X-rays more clustered DNA damage is produced that is more difficult to repair and results in a more pronounced increase in micronucleus formation.
Collapse
Affiliation(s)
- Julie Depuydt
- Department of Basic Medical Sciences, University of Ghent, Belgium
| | | | | | | | | |
Collapse
|
20
|
Myronakis ME, Zvelebil M, Darambara DG. Normalized mean glandular dose computation from mammography using GATE: a validation study. Phys Med Biol 2013; 58:2247-65. [PMID: 23475310 DOI: 10.1088/0031-9155/58/7/2247] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
Mean glandular dose (MGD) is the figure of merit to assess breast dose after a mammographic acquisition. The use of normalized MGD obtained from Monte Carlo computations with measured incident air kerma determines the MGD delivered to patients. The Geant4 Application for Tomographic Emission (GATE) toolkit is a modern Monte Carlo application specifically designed for medical imaging systems modelling. Although there is an increasing number of publications using GATE worldwide for a wide range of medical imaging and therapeutic applications, there is currently no means to obtain normalized MGD. In this work, the GATE toolkit is extended, through the development of two new modules, to provide normalized MGD information for compressed breast phantoms based on simple geometries. The normalized MGD values were validated against published work and provided results at half value layers lower than 0.3 and greater than 0.6 mmAl. In addition, the skin thickness and composition were considered. Normalized MGD was computed after substitution of the adipose layer surrounding the standard breast phantom with skin tissue and the relative difference is reported. Spectrum generation was facilitated by further development of previously published work by other authors. Validation of the new GATE extension showed good agreement with published data and can be used to assess breast dose from mammographic as well as more complex x-ray imaging techniques. Changing skin thickness and composition revealed substantial changes in normalized MGD specifically for compressed breast thickness different than 5 cm and a possible revision of the structure of the standard breast model may be necessary.
Collapse
Affiliation(s)
- Marios E Myronakis
- Division of Radiotherapy and Imaging, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, London, SW3 6JJ, UK.
| | | | | |
Collapse
|
21
|
O'Leary D, Rainford L. A comparison of mean glandular dose diagnostic reference levels within the all-digital Irish National Breast Screening Programme and the Irish Symptomatic Breast Services. RADIATION PROTECTION DOSIMETRY 2013; 153:300-308. [PMID: 22740646 DOI: 10.1093/rpd/ncs112] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/01/2023]
Abstract
Data on image quality, compression and radiation dose were collected from symptomatic breast units within the Republic of Ireland. Quantitative and qualitative data were analysed using SPSS. Recommendations of mean glandular dose (MGD) diagnostic reference levels were made at various levels for film-screen and full field digital mammography units to match levels published worldwide. MGDs received by symptomatic breast patients within Ireland are higher than those received in the all-digital Irish Breast Screening service; 55-65 mm breast: 1.75 mGy (screening) vs. 2.4 mGy (symptomatic) at the 95th percentile; various reasons are proposed for the differences. MGDs achieved in the screening service may be lower because of the exacting requirements for radiographer training, characteristics of the patients and equipment quality assurance levels. More precise imaging guidelines, standards and training of symptomatic radiographers performing mammography are suggested to remediate MGDs delivered to the breasts of Irish women attending the symptomatic breast services.
Collapse
Affiliation(s)
- Desiree O'Leary
- School of Medicine and Medical Science, Diagnostic Imaging, Health Sciences Centre, University College Dublin, Belfield, Dublin 4, Ireland.
| | | |
Collapse
|
22
|
Chen B, Wang Y, Sun X, Guo W, Zhao M, Cui G, Hu L, Li P, Ren Y, Feng J, Yu J. Analysis of patient dose in full field digital mammography. Eur J Radiol 2012; 81:868-72. [DOI: 10.1016/j.ejrad.2011.02.027] [Citation(s) in RCA: 23] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/08/2011] [Accepted: 02/10/2011] [Indexed: 11/16/2022]
|
23
|
Thierry-Chef I, Simon SL, Weinstock RM, Kwon D, Linet MS. Reconstruction of absorbed doses to fibroglandular tissue of the breast of women undergoing mammography (1960 to the present). Radiat Res 2012; 177:92-108. [PMID: 21988547 PMCID: PMC3876279 DOI: 10.1667/rr2241.1] [Citation(s) in RCA: 18] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/03/2022]
Abstract
The assessment of potential benefits versus harms from mammographic examinations as described in the controversial breast cancer screening recommendations of the U.S. Preventive Task Force included limited consideration of absorbed dose to the fibroglandular tissue of the breast (glandular tissue dose), the tissue at risk for breast cancer. Epidemiological studies on cancer risks associated with diagnostic radiological examinations often lack accurate information on glandular tissue dose, and there is a clear need for better estimates of these doses. Our objective was to develop a quantitative summary of glandular tissue doses from mammography by considering sources of variation over time in key parameters, including imaging protocols, X-ray target materials, voltage, filtration, incident air kerma, compressed breast thickness, and breast composition. We estimated the minimum, maximum and mean values for glandular tissue dose for populations of exposed women within 5-year periods from 1960 to the present, with the minimum to maximum range likely including 90% to 95% of the entirety of the dose range from mammography in North America and Europe. Glandular tissue dose from a single view in mammography is presently about 2 mGy, about one-sixth the dose in the 1960s. The ratio of our estimates of maximum to minimum glandular tissue doses for average-size breasts was about 100 in the 1960s compared to a ratio of about 5 in recent years. Findings from our analysis provide quantitative information on glandular tissue doses from mammographic examinations that can be used in epidemiological studies of breast cancer.
Collapse
|
24
|
Samei E, Saunders RS. Dual-energy contrast-enhanced breast tomosynthesis: optimization of beam quality for dose and image quality. Phys Med Biol 2011; 56:6359-78. [PMID: 21908902 DOI: 10.1088/0031-9155/56/19/013] [Citation(s) in RCA: 24] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
Dual-energy contrast-enhanced breast tomosynthesis is a promising technique to obtain three-dimensional functional information from the breast with high resolution and speed. To optimize this new method, this study searched for the beam quality that maximized image quality in terms of mass detection performance. A digital tomosynthesis system was modeled using a fast ray-tracing algorithm, which created simulated projection images by tracking photons through a voxelized anatomical breast phantom containing iodinated lesions. The single-energy images were combined into dual-energy images through a weighted log subtraction process. The weighting factor was optimized to minimize anatomical noise, while the dose distribution was chosen to minimize quantum noise. The dual-energy images were analyzed for the signal difference to noise ratio (SdNR) of iodinated masses. The fast ray-tracing explored 523 776 dual-energy combinations to identify which yields optimum mass SdNR. The ray-tracing results were verified using a Monte Carlo model for a breast tomosynthesis system with a selenium-based flat-panel detector. The projection images from our voxelized breast phantom were obtained at a constant total glandular dose. The projections were combined using weighted log subtraction and reconstructed using commercial reconstruction software. The lesion SdNR was measured in the central reconstructed slice. The SdNR performance varied markedly across the kVp and filtration space. Ray-tracing results indicated that the mass SdNR was maximized with a high-energy tungsten beam at 49 kVp with 92.5 µm of copper filtration and a low-energy tungsten beam at 49 kVp with 95 µm of tin filtration. This result was consistent with Monte Carlo findings. This mammographic technique led to a mass SdNR of 0.92 ± 0.03 in the projections and 3.68 ± 0.19 in the reconstructed slices. These values were markedly higher than those for non-optimized techniques. Our findings indicate that dual-energy breast tomosynthesis can be performed optimally at 49 kVp with alternative copper and tin filters, with reconstruction following weighted subtraction. The optimum technique provides best visibility of iodine against structured breast background in dual-energy contrast-enhanced breast tomosynthesis.
Collapse
Affiliation(s)
- Ehsan Samei
- Carl E Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, Durham, NC 27705, USA.
| | | |
Collapse
|
25
|
Dong SL, Chu TC, Lin YC, Lan GY, Yeh YH, Chen S, Chuanga KS. Determination of equivalent breast phantoms for different age groups of Taiwanese women: an experimental approach. Med Phys 2011; 38:4094-100. [PMID: 21859009 DOI: 10.1118/1.3591989] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE Polymethylmethacrylate (PMMA) slab is one of the mostly used phantoms for studying breast dosimetry in mammography. The purpose of this study was to evaluate the equivalence between exposure factors acquired from PMMA slabs and patient cases of different age groups of Taiwanese women in mammography. METHODS This study included 3910 craniocaudal screen/film mammograms on Taiwanese women acquired on one mammographic unit. The tube loading, compressed breast thickness (CBT), compression force, tube voltage, and target/filter combination for each mammogram were collected for all patients. The glandularity and the equivalent thickness of PMMA were determined for each breast using the exposure factors of the breast in combination with experimental measurements from breast-tissue-equivalent attenuation slabs. Equivalent thicknesses of PMMA to the breasts of Taiwanese women were then estimated. RESULTS The average +/- standard deviation CBT and breast glandularity in this study were 4.2 +/- 1.0 cm and 54% +/- 23%, respectively. The average equivalent PMMA thickness was 4.0 +/- 0.7 cm. PMMA slabs producing equivalent exposure factors as in the breasts of Taiwanese women were determined for the age groups 30-49 yr and 50-69 yr. For the 4-cm PMMA slab, the CBT and glandularity values of the equivalent breast were 4.1 cm and 65%, respectively, for the age group 30-49 yr and 4.4 cm and 44%, respectively, for the age group 50-69 yr. CONCLUSIONS The average thickness of PMMA slabs producing the same exposure factors as observed in a large group of Taiwanese women is less than that reported for American women. The results from this study can provide useful information for determining a suitable thickness of PMMA for mammographic dose survey in Taiwan. The equivalence of PMMA slabs and the breasts of Taiwanese women is provided to allow average glandular dose assessment in clinical practice.
Collapse
Affiliation(s)
- Shang-Lung Dong
- Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, 101 Section 2, Kuang-Fu Road, Hsinchu 30013, Taiwan
| | | | | | | | | | | | | |
Collapse
|
26
|
Development of an Adjustable Model Breast for Mammographic Dosimetry Assessment in Taiwanese Women. AJR Am J Roentgenol 2011; 196:W476-81. [DOI: 10.2214/ajr.09.3700] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
|
27
|
Dong SL, Su JL, Yeh YH, Chu TC, Lin YC, Chuang KS. Development of an imaging-planning program for screen/film and computed radiography mammography for breasts with short chest wall to nipple distance. Br J Radiol 2011; 84:350-7. [PMID: 21123310 PMCID: PMC3473466 DOI: 10.1259/bjr/97507379] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/09/2009] [Revised: 04/09/2010] [Accepted: 04/29/2010] [Indexed: 11/05/2022] Open
Abstract
OBJECTIVE Imaging breasts with a short chest wall to nipple distance (CWND) using a traditional mammographic X-ray unit is a technical challenge for mammographers. The purpose of this study is the development of an imaging-planning program to assist in determination of imaging parameters of screen/film (SF) and computed radiography (CR) mammography for short CWND breasts. METHODS A traditional mammographic X-ray unit (Mammomat 3000, Siemens, Munich, Germany) was employed. The imaging-planning program was developed by combining the compressed breast thickness correction, the equivalent polymethylmethacrylate thickness assessment for breasts and the tube loading (mAs) measurement. Both phantom exposures and a total of 597 exposures were used for examining the imaging-planning program. RESULTS Results of the phantom study show that the tube loading rapidly decreased with the CWND when the automatic exposure control (AEC) detector was not fully covered by the phantom. For patient exposures with the AEC fully covered by breast tissue, the average fractional tube loadings, defined as the ratio of the predicted mAs using the imaging-planning program and mAs of the mammogram, were 1.10 and 1.07 for SF and CR mammograms, respectively. The predicted mAs values were comparable to the mAs values, as determined by the AEC. CONCLUSION By applying the imaging-planning program in clinical practice, the experiential dependence of the mammographer for determination of the imaging parameters for short CWND breasts is minimised.
Collapse
Affiliation(s)
- S L Dong
- Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, Taiwan, Republic of China
| | | | | | | | | | | |
Collapse
|
28
|
Maekawa M. Average glandular dose on the basis of quality control. Nihon Hoshasen Gijutsu Gakkai Zasshi 2010; 66:404-408. [PMID: 20625229 DOI: 10.6009/jjrt.66.404] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 05/29/2023]
|
29
|
Dooley WC, Vargas HI, Fenn AJ, Tomaselli MB, Harness JK. Focused microwave thermotherapy for preoperative treatment of invasive breast cancer: a review of clinical studies. Ann Surg Oncol 2010; 17:1076-93. [PMID: 20033319 DOI: 10.1245/s10434-009-0872-z] [Citation(s) in RCA: 34] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/07/2009] [Indexed: 02/05/2023]
Abstract
BACKGROUND Preoperative focused microwave thermotherapy (FMT) is a promising method for targeted treatment of breast cancer cells. Results of four multi-institutional clinical studies of preoperative FMT for treating invasive carcinomas in the intact breast are reviewed. METHODS Externally applied wide-field adaptive phased-array FMT has been investigated both as a preoperative heat-alone ablation treatment and as a combination treatment with preoperative anthracycline-based chemotherapy for breast tumors ranging in ultrasound-measured size from 0.8 to 7.8 cm. RESULTS In phase I, eight of ten (80%) patients receiving a single low dose of FMT prior to receiving mastectomy had a partial tumor response quantified by either ultrasound measurements of tumor volume reduction or by pathologic cell kill. In phase II, the FMT thermal dose was increased to establish a threshold dose to induce 100% pathologic tumor cell kill for invasive carcinomas prior to breast-conserving surgery (BCS). In a randomized study for patients with early-stage invasive breast cancer, of those patients receiving preoperative FMT at ablative temperatures, 0 of 34 (0%) patients had positive tumor margins, whereas positive margins occurred in 4 of 41 (9.8%) of patients receiving BCS alone (P = 0.13). In a randomized study for patients with large tumors, based on ultrasound measurements the median tumor volume reduction was 88.4% (n = 14) for patients receiving FMT and neoadjuvant chemotherapy, compared with 58.8% (n = 10) reduction in the neoadjuvant chemotherapy-alone arm (P = 0.048). CONCLUSIONS Wide-field adaptive phased-array FMT can be safely administered in a preoperative setting, and data from randomized studies suggest both a reduction in positive tumor margins as a heat-alone treatment for early-stage breast cancer and a reduction in tumor volume when used in combination with anthracycline-based chemotherapy for patients with large breast cancer tumors. Larger randomized studies are required to verify these conclusions.
Collapse
Affiliation(s)
- William C Dooley
- Health Sciences Center, The University of Oklahoma, Oklahoma City, OK, USA.
| | | | | | | | | |
Collapse
|
30
|
Bliznakova K, Kolitsi Z, Speller RD, Horrocks JA, Tromba G, Pallikarakis N. Evaluation of digital breast tomosynthesis reconstruction algorithms using synchrotron radiation in standard geometry. Med Phys 2010; 37:1893-903. [DOI: 10.1118/1.3371693] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
31
|
Yaffe MJ, Boone JM, Packard N, Alonzo-Proulx O, Huang SY, Peressotti CL, Al-Mayah A, Brock K. The myth of the 50-50 breast. Med Phys 2010; 36:5437-43. [PMID: 20095256 DOI: 10.1118/1.3250863] [Citation(s) in RCA: 191] [Impact Index Per Article: 13.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE For dosimetry and for work in optimization of x-ray imaging of the breast, it is commonly assumed that the breast is composed of 50% fibroglandular tissue and 50% fat. The purpose of this study was to assess whether this assumption was realistic. METHODS First, data obtained from an experimental breast CT scanner were used to validate an algorithm that measures breast density from digitized film mammograms. Density results obtained from a total of 2831 women, including 191 women receiving CT and from mammograms of 2640 women from three other groups, were then used to estimate breast compositions. RESULTS Mean compositions, expressed as percent fibroglandular tissue (including the skin), varied from 13.7% to 25.6% among the groups with an overall mean of 19.3%. The mean compressed breast thickness for the mammograms was 5.9 cm (sigma = 1.6 cm). 80% of the women in our study had volumetric breast density less than 27% and 95% were below 45%. CONCLUSIONS Based on the results obtained from the four groups of women in our study, the "50-50" breast is not a representative model of the breast composition.
Collapse
Affiliation(s)
- M J Yaffe
- Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario M4N 3M5, Canada.
| | | | | | | | | | | | | | | |
Collapse
|
32
|
Saunders RS, Samei E, Lo JY, Baker JA. Can Compression Be Reduced for Breast Tomosynthesis? Monte Carlo Study on Mass and Microcalcification Conspicuity in Tomosynthesis. Radiology 2009; 251:673-82. [PMID: 19474373 DOI: 10.1148/radiol.2521081278] [Citation(s) in RCA: 38] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Affiliation(s)
- Robert S Saunders
- Department of Radiology, Duke Advanced Imaging Laboratories, Duke University Medical Center, 2424 Erwin Rd, Suite 302, Durham, NC 27710, USA.
| | | | | | | |
Collapse
|
33
|
Leonard BE, Leonard VF. Mammogram and diagnostic X-rays--evidence of protective Bystander, Adaptive Response (AR) radio-protection and AR retention at high dose levels. Int J Radiat Biol 2009; 84:885-99. [PMID: 19016137 DOI: 10.1080/09553000802460115] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/21/2022]
Abstract
PURPOSE The recently published dose response data by Dr Redpath's research group for low energy (30 kVp) mammography X-rays, displaying Adaptive Response (AR) radio-protective behavior, is significant for millions of American women that undergo annual breast cancer screening. We here, using the recently developed Microdose Model that encompasses the Bystander Effect (BE) and AR behavior, examine the data for BE, AR and high radiation domination by the priming radiations high dose Direct Damage. RESULTS The dose response is divided into three regions, Bystander Effect Region, Adaptive Response Region and Direct Damage Region (with possible retention of the AR protection). The Bystander Effect Region is below the microdose Specific Energy deposition for single photon induced charged particle traversals through the cell nucleus (the microdose Specific Energy Deposition per Traversal value = < z1 > = 0.638 cGy per Hit). Strong evidence is shown that a protective BE of about 50% occurs at a very low dose of 0.054 cGy, the BE is depleted reverting the response back to nearly the zero dose control value at 0.27 cGy, a 42% AR protection then is developed at 1.08 cGy and then the Direct Damage increasingly begins to dominate in the range from 5.4-21.6 cGy. Using the precise Method of Maximum Likelihood Estimator (MLE), the high dose Direct Damage Region is examined. We show that to the dose of 21.6 cGy the AR protection is retained in spite of the significant Direct Damage. We apply the same MLE analysis to the Redpath data for 137Cs gammas and find that the AR protection is completely dissipated at high Direct Damage inducing doses of 100 cGy. CONCLUSIONS The model shows that a protective BE of about 50% occurs at a low factor of 12 below single tracks traversals where less than 10% of the cell nuclei have been hit. Poisson distributed single tracks activates the 42% AR protection. The AR protection is retained at high dose but one needs to understand why 137Cs does not. Other Redpath group AR data sets for 137Cs, 232 MeV protons, and brachytherapy 125I photons did not reveal BE since the lowest data points were above the < z1 > for the radiations, but diagnostic X-rays do.
Collapse
Affiliation(s)
- Bobby E Leonard
- International Academy of Hi-Tech Services, Inc., Severna Park, Maryland 21146, USA.
| | | |
Collapse
|
34
|
Saunders RS, Samei E. The effect of breast compression on mass conspicuity in digital mammography. Med Phys 2008; 35:4464-73. [PMID: 18975694 DOI: 10.1118/1.2977600] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
This study analyzed how the inherent quality of diagnostic information in digital mammography could be affected by breast compression. A digital mammography system was modeled using a Monte Carlo algorithm based on the Penelope program, which has been successfully used to model several medical imaging systems. First, the Monte Carlo program was validated against previous measurements and simulations. Once validated, the Monte Carlo software modeled a digital mammography system by tracking photons through a voxelized software breast phantom, containing anatomical structures and breast masses, and following photons until they were absorbed by a selenium-based flat-panel detector. Simulations were performed for two compression conditions (standard compression and 12.5% reduced compression) and three photon flux conditions (constant flux, constant detector signal, and constant glandular dose). The results showed that reduced compression led to higher scatter fractions, as expected. For the constant photon flux condition, decreased compression also reduced glandular dose. For constant glandular dose, the SdNR for a 4 cm breast was 0.60 +/- 0.11 and 0.62 +/- 0.11 under standard and reduced compressions, respectively. For the 6 cm case with constant glandular dose, the SdNR was 0.50 +/- 0.11 and 0.49 +/- 0.10 under standard and reduced compressions, respectively. The results suggest that if a particular imaging system can handle an approximately 10% increase in total tube output and 10% decrease in detector signal, breast compression can be reduced by about 12% in terms of breast thickness with little impact on image quality or dose.
Collapse
Affiliation(s)
- Robert S Saunders
- Duke Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, Durham, North Carolina 27705, USA.
| | | |
Collapse
|
35
|
Bor D, Tukel S, Olgar T, Toklu T, Aydin E, Akyol O. Investigation of mean glandular dose versus compressed breast thickness relationship for mammography. RADIATION PROTECTION DOSIMETRY 2008; 129:160-164. [PMID: 18420560 DOI: 10.1093/rpd/ncn140] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/26/2023]
Abstract
The relationship between the mean glandular dose (MGD) and the compressed breast thickness (CBT) is commonly used for the presentation of mammographic dose survey results and could also be useful for the assessment of individual breast doses retrospectively in case of lack of necessary dosimetric instrumentation. The high data scattering from the best fit reduces the reliability of this technique. The aim of this study was to investigate the accuracy of this relationship using the data collected from a patient survey and phantom experiment. Patients were divided into three different groups according to their breast glandularities, which were predicted from the inspection of previous mammograms. X-ray beam qualities that will be used in patient examinations were determined according to breast thickness and predicted glandularities. The MGD versus CBT relationship for all the examined patients resulted in a poor correlation (R2 = 0.28). This relationship was separately obtained for each glandularity group and also for sub-groups of specific beam qualities. The best correlation (R(2) = 0.73) was obtained for the fatty breast group and Mo/Mo combination. A low correlation (R2 = 0.34) was observed in the mid-glandularity group due to inclusion of a wide range of glandularities in this group. In the case of the dense breast group, although the glandularity range was narrow, there were e still high data scattering (R2 = 0.25). This was probably due to the use of Mo/Rh and Mo/Mo combinations. This is validated by obtaining the MGD-CBT relationship specific to Mo/Mo combination (R2 = 0.61).
Collapse
Affiliation(s)
- Doğan Bor
- Department of Medical Physics, Institute of Nuclear Science, Ankara University, 06 100 Tandogan, Ankara, Turkey.
| | | | | | | | | | | |
Collapse
|
36
|
McLean ID, Heggie JCP, Herley J, Thomson FJ, Grewal RK. Interim recommendations for a digital mammography quality assurance program. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2007; 30:65-100. [PMID: 17682397 DOI: 10.1007/bf03178412] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/27/2022]
Abstract
In 2001 the ACPSEM published a position paper on quality assurance in screen film mammography which was subsequently adopted as a basis for the quality assurance programs of both the Royal Australian and New Zealand College of Radiologists (RANZCR) and of BreastScreen Australia. Since then the clinical implementation of digital mammography has been realised and it has become evident that existing screen-film protocols were not appropriate to assure the required image quality needed for reliable diagnosis or to address the new dose implications resulting from digital technology. In addition, the advantages and responsibilities inherent in teleradiology are most critical in mammography and also need to be addressed. The current document is the result of a review of current overseas practice and local experience in these areas. At this time the technology of digital imaging is undergoing significant development and there is still a lack of full international consensus about some of the detailed Quality Control tests that should be included in quality assurance (QA) programs. This document describes the current status in digital mammography QA and recommends test procedures that may be suitable in the Australasian environment. For completeness, this document also includes a review of the QA programs required for the various types of digital biopsy units used in mammography. In the future, international harmonisation of digital quality assurance in mammography and changes in the technology may require a review of this document. Accordingly, updates of this document will be provided as deemed necessary in electronic format on the ACPSEM's website (see http://www.acpsem.org.au/au/subgroup/radiology/RadiologySG_index.html).
Collapse
Affiliation(s)
- I D McLean
- Medical Physics Department, Westmead Hospital, Wentworthville, Australia.
| | | | | | | | | |
Collapse
|
37
|
Nickoloff EL, Alderson PO. A comparative study of thoracic radiation doses from 64-slice cardiac CT. Br J Radiol 2007; 80:537-44. [PMID: 17704315 DOI: 10.1259/bjr/34603706] [Citation(s) in RCA: 28] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
The goal of this study was to measure radiation doses for 64-slice cardiac CT angiography studies and to study the dose-savings features of these CT scanners. This was done using various phantoms. These radiation doses were compared with those from typical helical body CT scans, fluoroscopy cardiac catheterization studies and mammography examinations. Radiation measurements were made with a CT ionization detector and a solid state dosimeter. A GE 64-slice Lightspeed VCT and a Siemens Somatom Sensation 64 CT were used to scan a standard 32 cm acrylic phantom and an anthropomorphic phantom. Data were collected in axial and various gated cardiac helical modes. Organ doses and the effective doses were calculated from the measurements. In gated CT cardiac mode with the 32 cm acrylic phantom, the measured radiation doses per study were generally three to seven times greater than those from typical body helical CT examinations; the range depended upon selectable scan parameters. With the anatomical phantom, the surface doses in the anteroposterior (AP) plane were typically 20-60% higher than those measured using the 32 cm phantom. The lateral surface doses were -4% to +15%. These results can be attributed to the shorter AP dimension and the air in the lungs. The CT skin entrance radiation doses were 80-90% less than diagnostic cardiac catheterization studies, and organ doses were similar. Because 64-slice cardiac gated CT uses pitches equal to 0.20-0.27 and high mAs values, the patient radiation doses are appreciably higher than in routine body CT examinations. The female breast, which could receive a radiation dose 10-30 times that received from mammography screening, is an organ of particular concern.
Collapse
Affiliation(s)
- E L Nickoloff
- Department of Radiology, Columbia University and New York Presbyterian Hospital, 177 Fort Washington Avenue, Milstein Bldg Room 3-265B, New York, NY 10032-3784, USA.
| | | |
Collapse
|
38
|
Smathers RL, Boone JM, Lee LJ, Berns EA, Miller RA, Wright AM. Radiation Dose Reduction for Augmentation Mammography. AJR Am J Roentgenol 2007; 188:1414-21. [PMID: 17449790 DOI: 10.2214/ajr.06.0998] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Abstract
OBJECTIVE Patients who undergo cosmetic augmentation have larger and denser breasts and receive higher radiation doses during mammography than women without implants. In this study we evaluated the dose increase and techniques for dose reduction. SUBJECTS AND METHODS Mean glandular dose to the breast during screening mammography was measured for 206 women who had undergone breast augmentation. For 13 of these women, mean glandular dose from preoperative mammography also was measured. Effective tube current, peak kilovoltage, and breast thickness were measured, and mean glandular dose was calculated for 1,632 images. Two screen-film combinations and three target-filter combinations were studied. RESULTS For four-view augmentation mammography with a molybdenum-molybdenum (Mo-Mo) target-filter combination, mean glandular dose was reduced 35%, from 10.7 to 7.0 mGy, by changing the screen-film combination from 100 to 190 speed. For four-view augmentation mammography, mean glandular dose was reduced 24% by changing the target-filter combination from Mo-Mo to rhodium-rhodium (Rh-Rh) for full views of breasts containing implants. For four-view augmentation mammography, mean glandular dose was reduced 50% by changing the screen-film combination from 100 to 190 speed and changing the target-filter combination from Mo-Mo to Rh-Rh for implant-full views. CONCLUSION Mean glandular dose per breast from four-view augmentation mammography with the 100-speed screen-film and Mo-Mo target-filter combinations averaged 10.7 mGy, which is 3.1 times higher than the 3.4 mGy for conventional two-view mammography of breasts without implants. In 40 years of screening, this number represents a more than tripled lifetime attributable risk of radiation-induced breast cancer--an unacceptable level. Use of faster screen-film combinations, use of Rh-Rh target-filter combinations, and acquisition of three rather than four views are dose-reduction methods that together result in a 66% dose reduction, from 10.7 to 3.6 mGy. Mean glandular dose should be kept less than 7.0 mGy per breast for screening mammography of patients with breast implants.
Collapse
Affiliation(s)
- Ralph L Smathers
- Mammography Specialists Medical Group, Inc., 14651 S Bascom Ave., Suite 210, Los Gatos, CA 95032, USA.
| | | | | | | | | | | |
Collapse
|
39
|
Koutalonis M, Delis H, Spyrou G, Costaridou L, Tzanakos G, Panayiotakis G. Monte Carlo generated conversion factors for the estimation of average glandular dose in contact and magnification mammography. Phys Med Biol 2006; 51:5539-48. [PMID: 17047268 DOI: 10.1088/0031-9155/51/21/010] [Citation(s) in RCA: 16] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
Magnification mammography is a special technique used in the cases where breast complaints are noted by a woman or when an abnormality is found in a screening mammogram. The carcinogenic risk in mammography is related to the dose deposited in the glandular tissue of the breast rather than the adipose, and average glandular dose (AGD) is the quantity taken into consideration during a mammographic examination. Direct measurement of the AGD is not feasible during clinical practice and thus, the incident air KERMA on the breast surface is used to estimate the glandular dose, with the help of proper conversion factors. Additional conversion factors adapted for magnification and tube voltage are calculated, using Monte Carlo simulation. The effect of magnification degree, tube voltage, various anode/filter material combinations and glandularity on AGD is also studied, considering partial breast irradiation. Results demonstrate that the estimation of AGD utilizing conversion factors depends on these parameters, while the omission of correction factors for magnification and tube voltage can lead to significant underestimation or overestimation of AGD. AGD was found to increase with filter material's k-absorption edge, anode material's k-emission edge, tube voltage and magnification. Decrease of the glandularity of the breast leads to higher AGD due to the increased penetrating ability of the photon beam in thick breasts with low glandularity.
Collapse
Affiliation(s)
- M Koutalonis
- Department of Medical Physics, School of Medicine, University of Patras, 265 00 Patras, Greece
| | | | | | | | | | | |
Collapse
|
40
|
Gennaro G, di Maggio C. Dose comparison between screen/film and full-field digital mammography. Eur Radiol 2006; 16:2559-66. [PMID: 16733678 DOI: 10.1007/s00330-006-0314-2] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/31/2006] [Revised: 04/05/2006] [Accepted: 04/24/2006] [Indexed: 11/26/2022]
Abstract
The study purpose was the comparison between doses delivered by a full-field digital mammography system and a screen/film mammography unit, both using the same type of X-ray tube. Exposure parameters and breast thickness were collected for 300 screen/film (GE Senographe DMR) and 296 digital mammograms (GE Senographe 2000D). The entrance surface air kerma (ESAK) was calculated from anode/filter combination, kV(p) and mAs values and breast thickness, by simulating spectra through a program based on a catalogue of experimental X-ray spectra. The average glandular dose (AGD) was also computed. Results showed an overall reduction of average glandular dose by 27% of digital over screen/film mammography. The dose saving was about 15% for thin and thick breasts, while it was between 30% and 40% for intermediate thicknesses. Full-field digital mammography dose reduction is allowed by wider dynamic range and higher efficiency of digital detector, which can be exposed at higher energy spectra than screen/film mammography, and by the separation between acquisition and displaying processes.
Collapse
Affiliation(s)
- Gisella Gennaro
- Department of Oncological and Surgical Sciences, Padua University, via Gattamelata 64, 35128, Padua, Italy.
| | | |
Collapse
|
41
|
Zoetelief J, Veldkamp WJH, Thijssen MAO, Jansen JTM. Glandularity and mean glandular dose determined for individual women at four regional breast cancer screening units in the Netherlands. Phys Med Biol 2006; 51:1807-17. [PMID: 16552106 DOI: 10.1088/0031-9155/51/7/012] [Citation(s) in RCA: 15] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The nationwide breast cancer screening programme using mammography has been in full operation in The Netherlands since 1997. There is concern that the mean glandular doses due to mammography might be differing between different regions of the country due to differences in glandularity and compressed breast thickness. To investigate regional differences, glandularity, compressed breast thickness and mean glandular dose were determined for individual breasts during screening at mammography units at four locations in The Netherlands. Differences in glandularity were observed, which could be related qualitatively to differences in age of the participants at the different locations. Mean glandular dose depends on compressed breast thickness, glandularity and technical conditions of screening. The lowest average value of the mean glandular dose was found for the unit in Amsterdam. This is most likely due to the use of the Mo/Rh anode/filter combination at this unit, in addition to the Mo/Mo combination. At the other three units, almost exclusively the Mo/Mo anode/filter combination was used. Differences in mean glandular dose averaged per unit could be related mainly to differences in tube-current exposure-time product values. Consequently, it is concluded that differences in mean glandular dose at different units are marginal.
Collapse
Affiliation(s)
- J Zoetelief
- Department of Radiation, Radionuclides and Reactors, Faculty of Applied Sciences, Delft University of Technology, Mekelweg 15, 2629 JB Delft, The Netherlands.
| | | | | | | |
Collapse
|
42
|
Ng KH, Jamal N, DeWerd L. Global quality control perspective for the physical and technical aspects of screen-film mammography--image quality and radiation dose. RADIATION PROTECTION DOSIMETRY 2006; 121:445-51. [PMID: 16709704 DOI: 10.1093/rpd/ncl051] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/09/2023]
Abstract
The systematic monitoring of image quality and radiation dose is an ultimate solution to ensuring the continuously high quality of mammography examination. At present several protocols exist around the world, and different test objects are used for quality control (QC) of the physical and technical aspects of screen-film mammography. This situation may lead to differences in radiation image quality and dose reported. This article reviews the global QC perspective for the physical and technical aspects of screen-film mammography with regard to image quality and radiation dose. It points out issues that must be resolved in terms of radiation dose and that also affect the comparison.
Collapse
Affiliation(s)
- Kwan-Hoong Ng
- Department of Radiology, University of Malaya, 50603 Kuala Lumpur, Malaysia.
| | | | | |
Collapse
|
43
|
Jansen JTM, Veldkamp WJH, Thijssen MAO, van Woudenberg S, Zoetelief J. Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography. Phys Med Biol 2005; 50:5953-67. [PMID: 16333166 DOI: 10.1088/0031-9155/50/24/013] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The nationwide breast cancer screening programme using mammography has been in full operation in the Netherlands since 1997. Quality control of the screening programme has been assigned to the National Expert and Training Centre for Breast Cancer Screening. Limits are set to the mean glandular dose and the centre monitors these for all facilities engaged in the screening programme. This procedure is restricted to the determination of the entrance dose on a 5 cm thick polymethylmethacrylate (PMMA) phantom. The mean glandular dose for a compressed breast is estimated from these data. Individual breasts may deviate largely from this 5 cm PMMA breast model. Not only may the compressed breast size vary from 2 to 10 cm, but breast composition varies also. The mean glandular dose is dependent on the fraction of glandular tissue (glandularity) of the breast. To estimate the risk related to individual mammograms requires the development of a method for determination of the glandularity of individual breasts. A method has been developed to derive the glandularity using the attenuation of mammography x-rays in the breast. The method was applied to a series of mammograms at a screening unit. The results, i.e., a glandularity of 93% within the range of 0 to 1, were comparable with data in the literature. The glandularity as a function of compressed breast thickness is similar to results from other investigators using differing methods.
Collapse
Affiliation(s)
- J T M Jansen
- Department of Radiation, Radionuclides and Reactors, Faculty of Applied Sciences, Delft University of Technology, Mekelweg 15, 2629 JB Delft, The Netherlands.
| | | | | | | | | |
Collapse
|
44
|
Delis H, Spyrou G, Panayiotakis G, Tzanakos G. DOSIS: a Monte Carlo simulation program for dose related studies in mammography. Eur J Radiol 2005; 54:371-6. [PMID: 15899338 DOI: 10.1016/j.ejrad.2004.07.014] [Citation(s) in RCA: 11] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/03/2004] [Revised: 07/27/2004] [Accepted: 07/29/2004] [Indexed: 11/18/2022]
Abstract
Dosimetric studies in mammography are addressed by means of a Monte Carlo simulation program. The core of this program (DOSIS: dosimetry simulation studies) is a simulation model developed using FORTRAN 90, enriched with a graphical user interface developed in MS Visual Basic. User defined mammographic technique parameters affecting breast dose are imported to the simulation model and the produced results are provided by means of both absolute (surface dose, exposure at detector plane) and relative quantities (percentage depth dose, isodose curves). The program functionality has been demonstrated in the evaluation of various mammographic examination techniques. Specifically, the influence of tube voltage and filtration on the surface dose and the exposure at detector plane has been studied utilizing a water phantom. Increase of tube voltage from 25 to 30 kVp for a Mo/Mo system resulted in a 42% decrease of the surface dose for a thick breast (6 cm), without changing the exposure at the detector plane. Use of 1.02 mm Al filter for a W anode system operating at 30 kVp resulted in a 19.1% decrease of the surface dose delivered to a 5 cm water equivalent breast. Overall, W/Al systems appear to have improved dosimetric performance, resulting up to a 65% decrease of surface dose compared to Mo/Mo systems, for identical exposures at the detector plane and breast thicknesses.
Collapse
Affiliation(s)
- H Delis
- Department of Medical Physics, School of Medicine, University of Patras, 26500 Patras, Greece
| | | | | | | |
Collapse
|
45
|
Assiamah M, Nam TL, Keddy RJ. Comparison of mammography radiation dose values obtained from direct incident air kerma measurements with values from measured X-ray spectral data. Appl Radiat Isot 2005; 62:551-60. [PMID: 15701409 DOI: 10.1016/j.apradiso.2004.09.010] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/10/2004] [Accepted: 09/16/2004] [Indexed: 11/26/2022]
Abstract
The application of X-rays and ionising radiations for diagnostic radiology requires that the procedure is justified and optimised and that the exposure to the patient is kept as low as possible, without compromising image information. X-ray mammography is considered to be the most sensitive technique currently available for early detection of breast cancer. The magnitude of the absorbed radiation dose to the breast from mammography X-ray beams forms an important part of the quality control of the mammographic examination since it gives an indication of the performance of the mammographic imaging system as well as an estimated risk to the patient. In this work mean glandular dose (MGD) values were obtained at various tube potentials and tube loadings (TL) using direct measurements of the incident air kerma (ESAK) at the surface of a standard breast phantom and also from spectral measurements acquired with a solid-state detector. Comparisons of the MGD values thus derived are presented and the relationship between MGD, phantom thickness, image quality and tube operating parameters is discussed.
Collapse
Affiliation(s)
- M Assiamah
- Health Physics Service/Schonland Research Institute for Nuclear Sciences, University of the Witwatersrand, Private Bag 3, Wits 2050, Johannesburg, South Africa
| | | | | |
Collapse
|
46
|
Smans K, Bosmans H, Xiao M, Carton AK, Marchal G. Towards a proposition of a diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in Belgium. RADIATION PROTECTION DOSIMETRY 2005; 117:321-6. [PMID: 16464839 DOI: 10.1093/rpd/nci760] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/06/2023]
Abstract
A diagnostic reference level (DRL) is a dose level for a typical X-ray examination of a group of patients with standard body sizes and for broadly defined types of equipment. These levels are expected not to be exceeded for standard procedures when good and normal practice regarding diagnostic and technical performance is applied. In this paper, we have calculated DRLs for screening mammography in Belgium. The 95th percentile of the mean average glandular dose is 2.46 mGy. The DRL based on polymethyl methacrylate (PMMA) measurements was 2.08 mGy. Correlation coefficient (R) between doses from patient studies and phantom studies was 0.90, with an average underestimation of the phantom measurements of 15% for systems that use only Mo/Mo anode/filter. For the centres that use other anode/filters, there is not enough scientific evidence that a single phantom measurement of a standard PMMA block is representative for the patient dose.
Collapse
Affiliation(s)
- K Smans
- UZ Leuven, Herestraat 49, 3000 Leuven, Belgium.
| | | | | | | | | |
Collapse
|
47
|
Chevalier M, Morán P, Ten JI, Fernández Soto JM, Cepeda T, Vañó E. Patient dose in digital mammography. Med Phys 2004; 31:2471-9. [PMID: 15487727 DOI: 10.1118/1.1784591] [Citation(s) in RCA: 46] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
In the present investigation, we analyze the dose of 5034 patients (20,137 images) who underwent mammographic examinations with a full-field digital mammography system. Also, we evaluate the system calibration by analyzing the exposure factors as a function of breast thickness. The information relevant to this study has been extracted from the image DICOM header and stored in a database during a 3-year period (March 2001-October 2003). Patient data included age, breast thickness, kVp, mAs, target/filter combination, and nominal dose values. Entrance surface air kerma (ESAK) without backscatter was calculated from the tube output as measured for each voltage used under clinical conditions and from the tube loading (mAs) included in the DICOM header. Mean values for the patient age and compressed breast thickness were 56 years (SD: 11) and 52 mm (SD: 13), respectively. The majority of the images was acquired using the STD (for standard) automatic mode (98%). The most frequent target/filter combination automatically selected for breast smaller than 35 mm was Mo/Mo (75%); for intermediate thicknesses between 35 and 65 mm, the combinations were Mo/Rh (54%) and Rh/Rh (38.5%); Rh/Rh was the combination selected for 91% of the cases for breasts thicker than 65 mm. A wide kVp range was observed for each target/filter combination. The most frequent values were 28 kVp for Mo/Mo, 29 kVp for Mo/Rh, and 29 and 30 kV for Rh/Rh. Exposure times ranged from 0.2 to 4.2 s with a mean value of 1.1 s. Average glandular doses (AGD) per exposure were calculated by multiplying the ESAK values by the conversion factors tabulated by Dance for women in the age groups 50 to 64 and 40 to 49. This approach is based on the dependence of breast glandularity on breast thickness and age. The total mean average glandular dose (AGD(T)) was calculated by summing the values associated with the pre-exposure and with the main exposure. Mean AGD(T) per exposure was 1.88 mGy (CI 0.01) and the mean AGD(T) per examination was 3.8 mGy, with 4 images per examination on average. The mean dose for cranio-caudal view (CC) images was 1.8 mGy, which is lower than that for medio-lateral oblique (MLO) view because the thickness for CC images was on average 10% lower than that for MLO images. Mean AGD(T) for the oldest group of women (1.90) was 3% higher than the AGD(T) for the younger group (1.85) due to the larger compressed breast thickness of women in the older group (10% on average). Differences between the corresponding AGD(T) values of each age group were lowest for breast thicknesses in the range 40-60 mm, being slightly higher for the women in the older group.
Collapse
Affiliation(s)
- Margarita Chevalier
- Department of Radiology, Facultad de Medicina, Universidad Complutense de Madrid, 28040 Madrid, Spain.
| | | | | | | | | | | |
Collapse
|
48
|
Abstract
While phantoms are used for quality control assessment of the mammography unit, in vivo dose measurements are necessary to account for the variation in size and composition of the female breast. The use of thermoluminescent dosimeters (TLDs) in mammography has been limited due to TLD visibility. The aim of this current investigation was to access the suitability of a paper-thin LiF:Mg,Cu,P TLD (GR-200F) for in vivo dosimetric mammography measurements. The visibility of GR-200F has been directly compared with LiF:Mg,Cu,P TLDs (GR-200A) using a number of commercially available phantoms. The phantoms of thickness 2-5 cm were imaged over the range of tube potentials (24-28 kVp) used clinically. Both types of TLD were placed on the surface of the phantoms allowing assessment of visibility, entrance surface dose (ESD) and field homogeneity. In vivo assessment of ESD and visibility was also carried out on a volunteer undergoing a routine mammography examination. The positions of the GR-200F TLDs were not identified either on the image of the Leeds TOR(MAM) phantom or the patient mammograms. The average ESD for the Leeds phantom was 8.8 mGy, while the patient ESD was 13 mGy. It is now possible to perform in vivo measurements with the potential of increasing the accuracy of the doses measured for women that do not conform to a standard breast thickness or density.
Collapse
Affiliation(s)
- H M Warren-Forward
- University of Newcastle, School of Health Sciences (Medical Radiation Science), Callaghan NSW 2308, Australia
| | | |
Collapse
|
49
|
Jamal N, Ng KH, McLean D, Looi LM, Moosa F. Mammographic breast glandularity in Malaysian women: data derived from radiography. AJR Am J Roentgenol 2004; 182:713-7. [PMID: 14975974 DOI: 10.2214/ajr.182.3.1820713] [Citation(s) in RCA: 20] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Abstract
OBJECTIVE This study was undertaken to estimate mammographic breast glandularity in Malaysian women from radiographic data. MATERIALS AND METHODS A mammography X-ray unit was used to expose different thicknesses of phantom material of varying glandular and adipose composition at 27 kV. A least squares method was then used to fit the combined data of phantom glandularity, thickness, and milliampere-seconds. The subsequent fitted equation was then applied to calculate breast glandularity for 705 women who underwent diagnostic mammography, who were drawn equally from the three major ethnic groups of Malaysia: Malay, Chinese, and Indian. The difference in breast glandularity among ethnic groups was tested for significance using the nonparametric Kruskal-Wallis test. RESULTS The fitted equation gave an absolute error of less than or equal to +/- 8% when applied to the data from phantom exposure. The average breast glandularity of the study sample was 48.9% +/- 18.7%. Breast glandularity was found to decrease with breast thickness and age. CONCLUSION No significant difference was seen in breast glandularity among the ethnic groups (p > 0.05, Kruskal-Wallis test).
Collapse
Affiliation(s)
- Noriah Jamal
- Department of Radiology, University of Malaya, Kuala Lumpur 59100, Malaysia.
| | | | | | | | | |
Collapse
|
50
|
Pani S, Longo R, Dreossi D, Montanari F, Olivo A, Arfelli F, Bergamaschi A, Poropat P, Rigon L, Zanconati F, Dalla Palma L, Castelli E. Breast tomography with synchrotron radiation: preliminary results. Phys Med Biol 2004; 49:1739-54. [PMID: 15152928 DOI: 10.1088/0031-9155/49/9/011] [Citation(s) in RCA: 34] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
A system for in vivo breast imaging with monochromatic x-rays has been designed and built at the synchrotron radiation facility Elettra in Trieste (Italy) and will be operational in 2004. The system design involves the possibility of performing both planar mammography and breast tomography. In the present work, the first results obtained with a test set-up for breast tomography are shown and discussed. Tomographic images of in vitro breasts were acquired using monochromatic x-ray beams in the energy range 20-28 keV and a linear array silicon pixel detector. Tomograms were reconstructed using standard filtered backprojection algorithms; the effect of different filters was evaluated. The attenuation coefficients of fibroglandular and adipose tissue were measured, and a quantitative comparison of images acquired at different energies was performed by calculating the differential signal-to-noise ratio of fibroglandular details in adipose tissue. All images required a dose comparable to the dose delivered in clinical, conventional mammography and showed a high resolution of the breast structures without the overlapping effects that limit the visibility of the structures in 2D mammography. A quantitative evaluation of the images proves that the image quality at a given dose increases in the considered energy range and for the considered breast sizes.
Collapse
Affiliation(s)
- Silvia Pani
- Department of Physics, University of Trieste, INFN, Trieste Section, via Valerio 2, I 34126 Trieste, Italy.
| | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|