1
|
Bayat F, Elsayed Eldib M, Kavanagh B, Miften M, Altunbas C. Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling. Phys Med Biol 2022; 67:10.1088/1361-6560/ac8268. [PMID: 35853441 PMCID: PMC9378529 DOI: 10.1088/1361-6560/ac8268] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/12/2022] [Accepted: 07/19/2022] [Indexed: 11/12/2022]
Abstract
Objective. The concept of using kilovoltage (kV) and megavoltage (MV) beams concurrently has potential applications in cone beam computed tomography (CBCT) guided radiation therapy, such as single breath hold scans, metal artifact reduction, and simultaneous imaging during MV treatment delivery. However, MV cross-scatter generated during MV beam delivery degrades CBCT image quality. To address this, a 2D antiscatter grid and a cross-scatter correction method were investigated in the context of high dose MV treatment delivery.Approach. A 3D printed, tungsten 2D antiscatter grid prototype was utilized in kV CBCT scans to reduce MV cross-scatter fluence during concurrent MV beam delivery. Remaining cross-scatter in projections was corrected by using the 2D grid itself as a cross-scatter intensity sampling device, referred to as grid-based scatter sampling (GSS). To test this approach, kV CBCT acquisitions were performed while delivering 6 and 10 MV beams, mimicking high dose rate treatment delivery scenarios. kV and MV beam deliveries were not synchronized to eliminate MV beam delivery interruption. MV cross-scatter suppression performance of the proposed approach was evaluated in projections and CBCT images of phantoms.Main results. 2D grid reduced the intensity of MV cross-scatter in kV projections by a factor of 3 on the average, when compared to conventional antiscatter grid. Remaining cross scatter as measured by the GSS method was within 7% of measured reference intensity values, and subsequently corrected. CBCT image quality was improved substantially during concurrent kV-MV beam delivery. Median Hounsfield Unit (HU) inaccuracy was up to 182 HU without our methods, and it was reduced to a median 6.5 HU with our 2D grid and scatter correction approach. Our methods provided a factor of 2-6 improvement in contrast-to-noise ratio.Significance. This investigation demonstrates the utility of 2D antiscatter grids and grid-based scatter sampling in suppressing MV cross-scatter. Our approach successfully minimized the effects of MV cross-scatter in concurrent kV CBCT imaging and high dose MV treatment delivery scenarios. Hence, robust MV cross-scatter suppression is potentially feasible without MV beam delivery interruption or compromising kV image acquisition rate.
Collapse
Affiliation(s)
- Farhang Bayat
- Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, USA
| | - Mohamed Elsayed Eldib
- Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, USA
| | - Brian Kavanagh
- Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, USA
| | - Moyed Miften
- Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, USA
| | - Cem Altunbas
- Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, CO 80045, USA
| |
Collapse
|
2
|
Bayat F, Eldib ME, Altunbas C. Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging. PROCEEDINGS OF SPIE--THE INTERNATIONAL SOCIETY FOR OPTICAL ENGINEERING 2022; 12031:120311K. [PMID: 35465130 PMCID: PMC9028100 DOI: 10.1117/12.2611202] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/14/2023]
Abstract
Simultaneous use of kilovoltage (kV) and megavoltage (MV) beams has numerous potential applications in cone beam computed tomography (CBCT)-guided radiotherapy, such as fast MV+kV CBCT for single breath-hold scan, tumor localization with kV CBCT imaging during MV therapy delivery, and metal artifact suppression. However, the introduction of MV beams results in a large MV-cross scatter fluence incident on the kV Flat Panel Detector (FPD), and thus, deteriorating the low contrast visualization and Hounsfield Unit (HU) accuracy. In this work, we introduced a novel and robust method for reducing the effects of MV cross scatter. First, we implemented a 2D antiscatter grid atop the detector which rejects a large section of MV cross scatter. This hardware-based approach, while effective, allows a fraction of MV cross scatter to be transmitted to the FPD, resulting in artifacts and degraded HU accuracy in CBCT images. We thus introduced a data correction step, which aimed to estimate and correct the remaining MV cross scatter. This approach, referred to as Grid-Based Scatter Sampling, utilized 2D antiscatter grid itself to measure and correct remaining MV cross scatter in projections. We investigated the performance of the proposed approach in experiments by simultaneously acquiring kV CBCT and delivering MV beams with a clinical linac. The results show that the proposed method can substantially reduce HU inaccuracy and increase contrast-to-noise ratio (CNR). Our method does not require synchronization of kV and MV beam pulses, reduction of kV frame acquisition rate, or MV dose rate, and therefore, it is more practical to implement in radiation therapy clinical setting.
Collapse
|
3
|
Li KW, Fujiwara D, Haga A, Liu H, Geng LS. Physical density estimations of single- and dual-energy CT using material-based forward projection algorithm: a simulation study. Br J Radiol 2021; 94:20201236. [PMID: 34541866 DOI: 10.1259/bjr.20201236] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
OBJECTIVES This study aims to evaluate the accuracy of physical density prediction in single-energy CT (SECT) and dual-energy CT (DECT) by adapting a fully simulation-based method using a material-based forward projection algorithm (MBFPA). METHODS We used biological tissues referenced in ICRU Report 44 and tissue substitutes to prepare three different types of phantoms for calibrating the Hounsfield unit (HU)-to-density curves. Sinograms were first virtually generated by the MBFPA with four representative energy spectra (i.e. 80 kVp, 100 kVp, 120 kVp, and 6 MVp) and then reconstructed to form realistic CT images by adding statistical noise. The HU-to-density curves in each spectrum and their pairwise combinations were derived from the CT images. The accuracy of these curves was validated using the ICRP110 human phantoms. RESULTS The relative mean square errors (RMSEs) of the physical density by the HU-to-density curves calibrated with kV SECT nearly presented no phantom size dependence. The kV-kV DECT calibrated curves were also comparable with those from the kV SECT. The phantom size effect became notable when the MV X-ray beams were employed for both SECT and DECT due to beam-hardening effects. The RMSEs were decreased using the biological tissue phantom. CONCLUSION Simulation-based density prediction can be useful in the theoretical analysis of SECT and DECT calibrations. The results of this study indicated that the accuracy of SECT calibration is comparable with that of DECT using biological tissues. The size and shape of the calibration phantom could affect the accuracy, especially for MV CT calibrations. ADVANCES IN KNOWLEDGE The present study is based on a full simulation environment, which accommodates various situations such as SECT, kV-kV DECT, and even kV-MV DECT. In this paper, we presented the advances pertaining to the accuracy of the physical density prediction when applied to SECT and DECT in the MV X-ray energy range. To the best of our knowledge, this study is the first to validate the physical density estimation both in SECT and DECT using human-type phantoms.
Collapse
Affiliation(s)
- Kai-Wen Li
- Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, School of Medicine and Engineering, Beihang University, Key Laboratory of Big Data-Based Precision Medicine (Beihang University), Ministry of Industry and Information Technology, Beijing, China.,School of Physics, Beihang University, Beijing, China
| | - Daiyu Fujiwara
- Graduate School of Biomedical Sciences, Tokushima University, Tokushima, Japan
| | - Akihiro Haga
- Graduate School of Biomedical Sciences, Tokushima University, Tokushima, Japan
| | - Huisheng Liu
- Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, School of Medicine and Engineering, Beihang University, Key Laboratory of Big Data-Based Precision Medicine (Beihang University), Ministry of Industry and Information Technology, Beijing, China
| | - Li-Sheng Geng
- Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, School of Medicine and Engineering, Beihang University, Key Laboratory of Big Data-Based Precision Medicine (Beihang University), Ministry of Industry and Information Technology, Beijing, China.,School of Physics, Beihang University, Beijing, China.,Beijing Key Laboratory of Advanced Nuclear Materials and Physics, Beihang University, Beijing, China.,School of Physics and Microelectronics, Zhengzhou University, Zhengzhou, China
| |
Collapse
|
4
|
kV-kV and kV-MV DECT based estimation of proton stopping power ratio - a simulation study. Phys Med 2021; 89:182-192. [PMID: 34390901 DOI: 10.1016/j.ejmp.2021.07.038] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/22/2021] [Revised: 07/30/2021] [Accepted: 07/31/2021] [Indexed: 12/23/2022] Open
Abstract
PURPOSE This study aims to estimate the proton stopping power ratio (SPR) by using 80-120 kV and 120 kV-6 MV dual-energy CT (DECT) in a fully simulation-based approach for proton therapy dose calculations. METHODS Based on a virtual CT system, a two-step approach is applied to obtain the reference attenuation coefficient for image reconstruction. The effective atomic number (EAN) and electron density ratio (EDR) are estimated from two CT scans. The SPR is estimated using a calibration based on known materials to obtain a piecewise linear relationship between the EAN and the logarithm of the mean excitation energy, lnIm. The calibration phantoms are constructed from reference tissue materials taken from ICRU Report 44. Our approach is evaluated through using the ICRP110 human phantom. The respective influences of noise and beam hardening effects are studied. RESULTS With the beam hardening correction applied, the results of 120 kV-6 MV DECT are comparable to those of 80-120 kV DECT in predicting the EAN, but the former demonstrated a clear improvement in predicting the EDR and SPR. The 120 kV-6 MV DECT is able to predict the SPR within an accuracy of 10% for lung tissue and 5% for pelvis tissue, thereby outperforming the 80-120 kV DECT. CONCLUSIONS The 120 kV-6 MV DECT is less sensitive to noise but more susceptible to beam hardening effects. By applying beam hardening correction, the 120 kV-6 MV DECT can predict the SPR more accurately than the 80-120 kV DECT. To utilize our DECT approach most effectively, high-quality reconstructed images are required.
Collapse
|
5
|
Abbasian P, McCowan PM, Rickey DW, Van Uytven E, McCurdy BMC. Modeling the temporal–spatial nature of the readout of an electronic portal imaging device (EPID). Med Phys 2020; 47:5301-5311. [DOI: 10.1002/mp.14440] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/29/2019] [Revised: 07/13/2020] [Accepted: 07/19/2020] [Indexed: 11/07/2022] Open
Affiliation(s)
- Parandoush Abbasian
- Department of Physics and Astronomy University of Manitoba Winnipeg ManitobaR3T 2N2 Canada
| | - Peter M. McCowan
- Department of Physics and Astronomy University of Manitoba Winnipeg ManitobaR3T 2N2 Canada
- Medical Physics Department CancerCare Manitoba 675 McDermot Avenue Winnipeg ManitobaR3E 0V9 Canada
| | - Daniel W. Rickey
- Department of Physics and Astronomy University of Manitoba Winnipeg ManitobaR3T 2N2 Canada
- Medical Physics Department CancerCare Manitoba 675 McDermot Avenue Winnipeg ManitobaR3E 0V9 Canada
- Department of Radiology University of Manitoba 820 Sherbrook Street Winnipeg ManitobaR3A 1R9 Canada
| | - Eric Van Uytven
- Medical Physics Department CancerCare Manitoba 675 McDermot Avenue Winnipeg ManitobaR3E 0V9 Canada
- Department of Radiology University of Manitoba 820 Sherbrook Street Winnipeg ManitobaR3A 1R9 Canada
| | - Boyd M. C. McCurdy
- Department of Physics and Astronomy University of Manitoba Winnipeg ManitobaR3T 2N2 Canada
- Medical Physics Department CancerCare Manitoba 675 McDermot Avenue Winnipeg ManitobaR3E 0V9 Canada
- Department of Radiology University of Manitoba 820 Sherbrook Street Winnipeg ManitobaR3A 1R9 Canada
| |
Collapse
|
6
|
Sajja S, Lee Y, Eriksson M, Nordström H, Sahgal A, Hashemi M, Mainprize JG, Ruschin M. Technical Principles of Dual-Energy Cone Beam Computed Tomography and Clinical Applications for Radiation Therapy. Adv Radiat Oncol 2020; 5:1-16. [PMID: 32051885 PMCID: PMC7004939 DOI: 10.1016/j.adro.2019.07.013] [Citation(s) in RCA: 16] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/25/2019] [Revised: 05/21/2019] [Accepted: 07/20/2019] [Indexed: 02/06/2023] Open
Abstract
PURPOSE Medical imaging is an indispensable tool in radiotherapy for dose planning, image guidance and treatment monitoring. Cone beam CT (CBCT) is a low dose imaging technique with high spatial resolution capability as a direct by-product of using flat-panel detectors. However, certain issues such as x-ray scatter, beam hardening and other artifacts limit its utility to the verification of patient positioning using image-guided radiotherapy. METHODS AND MATERIALS Dual-energy (DE)-CBCT has recently demonstrated promise as an improved tool for tumor visualization in benchtop applications. It has the potential to improve soft-tissue contrast and reduce artifacts caused by beam hardening and metal. In this review, the practical aspects of developing a DE-CBCT based clinical and technical workflow are presented based on existing DE-CBCT literature and concepts adapted from the well-established library of work in DE-CT. Furthermore, the potential applications of DE-CBCT on its future role in radiotherapy are discussed. RESULTS AND CONCLUSIONS Based on current literature and an investigation of future applications, there is a clear potential for DE-CBCT technologies to be incorporated into radiotherapy. The applications of DE-CBCT include (but are not limited to): adaptive radiotherapy, brachytherapy, proton therapy, radiomics and theranostics.
Collapse
Affiliation(s)
- Shailaja Sajja
- Sunnybrook Research Institute, Toronto, Ontario, Canada
- QIPCM Imaging Core Lab, Techna Institute, Toronto, Ontario, Canada
| | - Young Lee
- Department of Radiation Oncology, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada
| | | | | | - Arjun Sahgal
- Department of Radiation Oncology, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada
| | | | | | - Mark Ruschin
- Department of Radiation Oncology, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada
| |
Collapse
|
7
|
Lindsay C, Bazalova‐Carter M, Wang A, Shedlock D, Wu M, Newson M, Xing L, Ansbacher W, Fahrig R, Star‐Lack J. Investigation of combined
kV
/
MV CBCT
imaging with a high‐
DQE MV
detector. Med Phys 2018; 46:563-575. [DOI: 10.1002/mp.13291] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/09/2018] [Revised: 11/01/2018] [Accepted: 11/02/2018] [Indexed: 01/23/2023] Open
Affiliation(s)
- C. Lindsay
- Department of Physics and Astronomy University of Victoria 3800 Finnerty Rd Victoria BC V8P 5C2 Canada
| | - M. Bazalova‐Carter
- Department of Physics and Astronomy University of Victoria 3800 Finnerty Rd Victoria BC V8P 5C2 Canada
| | - A. Wang
- Varian Medical Systems 3120 Hansen Way Palo Alto CA 94304 USA
| | - D. Shedlock
- Varian Medical Systems 3120 Hansen Way Palo Alto CA 94304 USA
| | - M. Wu
- Department of Radiology Stanford University 1201 Welch Rd Stanford CA 94305‐5105 USA
| | - M. Newson
- Department of Physics and Astronomy University of Victoria 3800 Finnerty Rd Victoria BC V8P 5C2 Canada
| | - L. Xing
- Department of Radiation Oncology Stanford University 875 Blake Wilbur Dr Stanford CA 94305‐5847 USA
| | - W. Ansbacher
- Department of Medical Physics BC Cancer Agency ‐ Vancouver Island Centre Victoria BC Canada
| | - R. Fahrig
- Department of Radiology Stanford University 1201 Welch Rd Stanford CA 94305‐5105 USA
| | - J. Star‐Lack
- Varian Medical Systems 3120 Hansen Way Palo Alto CA 94304 USA
| |
Collapse
|
8
|
Automated ultrafast kilovoltage-megavoltage cone-beam CT for image guided radiotherapy of lung cancer: System description and real-time results. Z Med Phys 2018; 28:110-120. [PMID: 29429610 DOI: 10.1016/j.zemedi.2018.01.002] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/08/2017] [Revised: 11/21/2017] [Accepted: 01/15/2018] [Indexed: 12/25/2022]
Abstract
PURPOSE To establish a fully automated kV-MV CBCT imaging method on a clinical linear accelerator that allows image acquisition of thoracic targets for patient positioning within one breath-hold (∼15s) under realistic clinical conditions. METHODS AND MATERIALS Our previously developed FPGA-based hardware unit which allows synchronized kV-MV CBCT projection acquisition is connected to a clinical linear accelerator system via a multi-pin switch; i.e. either kV-MV imaging or conventional clinical mode can be selected. An application program was developed to control the relevant linac parameters automatically and to manage the MV detector readout as well as the gantry angle capture for each MV projection. The kV projections are acquired with the conventional CBCT system. GPU-accelerated filtered backprojection is performed separately for both data sets. After appropriate grayscale normalization both modalities are combined and the final kV-MV volume is re-imported in the CBCT system to enable image matching. To demonstrate adequate geometrical accuracy of the novel imaging system the Penta-Guide phantom QA procedure is performed. Furthermore, a human plastinate and different tumor shapes in a thorax phantom are scanned. Diameters of the known tumor shapes are measured in the kV-MV reconstruction. RESULTS An automated kV-MV CBCT workflow was successfully established in a clinical environment. The overall procedure, from starting the data acquisition until the reconstructed volume is available for registration, requires ∼90s including 17s acquisition time for 100° rotation. It is very simple and allows target positioning in the same way as for conventional CBCT. Registration accuracy of the QA phantom is within ±1mm. The average deviation from the known tumor dimensions measured in the thorax phantom was 0.7mm which corresponds to an improvement of 36% compared to our previous kV-MV imaging system. CONCLUSIONS Due to automation the kV-MV CBCT workflow is speeded up by a factor of >10 compared to the manual approach. Thus, the system allows a simple, fast and reliable imaging procedure and fulfills all requirements to be successfully introduced into the clinical workflow now, enabling single-breath-hold volume imaging.
Collapse
|
9
|
Cai B, Goddu SM, Yaddanapudi S, Caruthers D, Wen J, Noel C, Mutic S, Sun B. Normalize the response of EPID in pursuit of linear accelerator dosimetry standardization. J Appl Clin Med Phys 2017; 19:73-85. [PMID: 29125224 PMCID: PMC5768011 DOI: 10.1002/acm2.12222] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/12/2017] [Revised: 08/06/2017] [Accepted: 09/28/2017] [Indexed: 12/18/2022] Open
Abstract
Normalize the response of electronic portal imaging device (EPID) is the first step toward an EPID‐based standardization of Linear Accelerator (linac) dosimetry quality assurance. In this study, we described an approach to generate two‐dimensional (2D) pixel sensitivity maps (PSM) for EPIDs response normalization utilizing an alternative beam and dark‐field (ABDF) image acquisition technique and large overlapping field irradiations. The automated image acquisition was performed by XML‐controlled machine operation and the PSM was generated based on a recursive calculation algorithm for Varian linacs equipped with aS1000 and aS1200 imager panels. Cross‐comparisons of normalized beam profiles and 1.5%/1.5 mm 1D Gamma analysis was adopted to quantify the improvement of beam profile matching before and after PSM corrections. PSMs were derived for both photon (6, 10, 15 MV) and electron (6, 20 MeV) beams via proposed method. The PSM‐corrected images reproduced a horn‐shaped profile for photon beams and a relative uniform profiles for electrons. For dosimetrically matched linacs equipped with aS1000 panels, PSM‐corrected images showed increased 1D‐Gamma passing rates for all energies, with an average 10.5% improvement for crossline and 37% for inline beam profiles. Similar improvements in the phantom study were observed with a maximum improvement of 32% for 15 MV and 22% for 20 MeV. The PSM value showed no significant change for all energies over a 3‐month period. In conclusion, the proposed approach correct EPID response for both aS1000 and aS1200 panels. This strategy enables the possibility to standardize linac dosimetry QA and to benchmark linac performance utilizing EPID as the common detector.
Collapse
Affiliation(s)
- Bin Cai
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - S Murty Goddu
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Sridhar Yaddanapudi
- Department of Radiation Oncology, University of Iowa Hospitals and Clinics, Iowa City, IA, USA
| | - Douglas Caruthers
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Jie Wen
- Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO, USA
| | | | - Sasa Mutic
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| | - Baozhou Sun
- Department of Radiation Oncology, Washington University, St. Louis, MO, USA
| |
Collapse
|
10
|
Arns A, Blessing M, Fleckenstein J, Stsepankou D, Boda-Heggemann J, Hesser J, Lohr F, Wenz F, Wertz H. Phantom-based evaluation of dose exposure of ultrafast combined kV-MV-CBCT towards clinical implementation for IGRT of lung cancer. PLoS One 2017; 12:e0187710. [PMID: 29125857 PMCID: PMC5681289 DOI: 10.1371/journal.pone.0187710] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/15/2017] [Accepted: 10/24/2017] [Indexed: 11/19/2022] Open
Abstract
PURPOSE Combined ultrafast 90°+90° kV-MV-CBCT within single breath-hold of 15s has high clinical potential for accelerating imaging for lung cancer patients treated with deep inspiration breath-hold (DIBH). For clinical feasibility of kV-MV-CBCT, dose exposure has to be small compared to prescribed dose. In this study, kV-MV dose output is evaluated and compared to clinically-established kV-CBCT. METHODS Accurate dose calibration was performed for kV and MV energy; beam quality was determined. For direct comparison of MV and kV dose output, relative biological effectiveness (RBE) was considered. CT dose index (CTDI) was determined and measurements in various representative locations of an inhomogeneous thorax phantom were performed to simulate the patient situation. RESULTS A measured dose of 20.5mGE (Gray-equivalent) in the target region was comparable to kV-CBCT (31.2mGy for widely-used, and 9.1mGy for latest available preset), whereas kV-MV spared healthy tissue and reduced dose to 6.6mGE (30%) due to asymmetric dose distribution. The measured weighted CTDI of 12mGE for kV-MV lay in between both clinical presets. CONCLUSIONS Dosimetric properties were in agreement with established imaging techniques, whereas exposure to healthy tissue was reduced. By reducing the imaging time to a single breath-hold of 15s, ultrafast combined kV-MV CBCT shortens patient time at the treatment couch and thus improves patient comfort. It is therefore usable for imaging of hypofractionated lung DIBH patients.
Collapse
Affiliation(s)
- Anna Arns
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Manuel Blessing
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Jens Fleckenstein
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Dzmitry Stsepankou
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Judit Boda-Heggemann
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Juergen Hesser
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Frank Lohr
- Struttura Complessa di Radioterapia, Dipartimento di Oncologia, Az. Ospedaliero-Universitaria di Modena, Modena, Italy
| | - Frederik Wenz
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| | - Hansjoerg Wertz
- Department of Radiation Oncology, Universitaetsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
| |
Collapse
|
11
|
Hazelaar C, Dahele M, Scheib S, Slotman BJ, Verbakel WF. Verifying tumor position during stereotactic body radiation therapy delivery using (limited-arc) cone beam computed tomography imaging. Radiother Oncol 2017; 123:355-362. [DOI: 10.1016/j.radonc.2017.04.022] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/28/2016] [Revised: 04/26/2017] [Accepted: 04/29/2017] [Indexed: 11/16/2022]
|
12
|
Arns A, Blessing M, Fleckenstein J, Stsepankou D, Boda-Heggemann J, Simeonova-Chergou A, Hesser J, Lohr F, Wenz F, Wertz H. Towards clinical implementation of ultrafast combined kV-MV CBCT for IGRT of lung cancer : Evaluation of registration accuracy based on phantom study. Strahlenther Onkol 2016; 192:312-21. [PMID: 26864049 DOI: 10.1007/s00066-016-0947-2] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/08/2015] [Accepted: 01/14/2016] [Indexed: 12/13/2022]
Abstract
PURPOSE Combined kV-MV cone-beam CT (CBCT) is a promising approach to accelerate imaging for patients with lung tumors treated with deep inspiration breath-hold. During a single breath-hold (15 s), a 3D kV-MV CBCT can be acquired, thus minimizing motion artifacts and increasing patient comfort. Prior to clinical implementation, positioning accuracy was evaluated and compared to clinically established imaging techniques. METHODS AND MATERIALS An inhomogeneous thorax phantom with four tumor-mimicking inlays was imaged in 10 predefined positions and registered to a planning CT. Novel kV-MV CBCT imaging (90° arc) was compared to clinically established kV-chest CBCT (360°) as well as nonclinical kV-CBCT and low-dose MV-CBCT (each 180°). Manual registration, automatic registration provided by the manufacturer and an additional in-house developed manufacturer-independent framework based on the MATLAB registration toolkit were applied. RESULTS Systematic setup error was reduced to 0.05 mm by high-precision phantom positioning with optical tracking. Stochastic mean displacement errors were 0.5 ± 0.3 mm in right-left, 0.4 ± 0.4 mm in anteroposterior and 0.0 ± 0.4 mm in craniocaudal directions for kV-MV CBCT with manual registration (maximum errors of no more than 1.4 mm). Clinical kV-chest CBCT resulted in mean errors of 0.2 mm (other modalities: 0.4-0.8 mm). Similar results were achieved with both automatic registration methods. CONCLUSION The comparison study of repositioning accuracy between novel kV-MV CBCT and clinically established volume imaging demonstrated that registration accuracy is maintained below 1 mm. Since imaging time is reduced to one breath-hold, kV-MV CBCT is ideal for image guidance, e.g., in lung stereotactic ablative radiotherapy.
Collapse
Affiliation(s)
- Anna Arns
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany.
| | - Manuel Blessing
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Jens Fleckenstein
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Dzmitry Stsepankou
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Judit Boda-Heggemann
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Anna Simeonova-Chergou
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Jürgen Hesser
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Frank Lohr
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Frederik Wenz
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| | - Hansjörg Wertz
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Haus 4, Ebene 0, Theodor-Kutzer-Ufer 1-3, 68167, Mannheim, Germany
| |
Collapse
|
13
|
Deep Inspiration Breath Hold-Based Radiation Therapy: A Clinical Review. Int J Radiat Oncol Biol Phys 2015; 94:478-92. [PMID: 26867877 DOI: 10.1016/j.ijrobp.2015.11.049] [Citation(s) in RCA: 154] [Impact Index Per Article: 17.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/03/2015] [Revised: 11/03/2015] [Accepted: 11/29/2015] [Indexed: 01/06/2023]
Abstract
Several recent developments in linear accelerator-based radiation therapy (RT) such as fast multileaf collimators, accelerated intensity modulation paradigms like volumeric modulated arc therapy and flattening filter-free (FFF) high-dose-rate therapy have dramatically shortened the duration of treatment fractions. Deliverable photon dose distributions have approached physical complexity limits as a consequence of precise dose calculation algorithms and online 3-dimensional image guided patient positioning (image guided RT). Simultaneously, beam quality and treatment speed have continuously been improved in particle beam therapy, especially for scanned particle beams. Applying complex treatment plans with steep dose gradients requires strategies to mitigate and compensate for motion effects in general, particularly breathing motion. Intrafractional breathing-related motion results in uncertainties in dose delivery and thus in target coverage. As a consequence, generous margins have been used, which, in turn, increases exposure to organs at risk. Particle therapy, particularly with scanned beams, poses additional problems such as interplay effects and range uncertainties. Among advanced strategies to compensate breathing motion such as beam gating and tracking, deep inspiration breath hold (DIBH) gating is particularly advantageous in several respects, not only for hypofractionated, high single-dose stereotactic body RT of lung, liver, and upper abdominal lesions but also for normofractionated treatment of thoracic tumors such as lung cancer, mediastinal lymphomas, and breast cancer. This review provides an in-depth discussion of the rationale and technical implementation of DIBH gating for hypofractionated and normofractionated RT of intrathoracic and upper abdominal tumors in photon and proton RT.
Collapse
|
14
|
Simeonova AO, Fleckenstein K, Wertz H, Frauenfeld A, Boda-Heggemann J, Lohr F, Wenz F. Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy? Transl Lung Cancer Res 2015; 1:45-53. [PMID: 25806154 DOI: 10.3978/j.issn.2218-6751.10.01] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/17/2011] [Accepted: 10/24/2011] [Indexed: 12/29/2022]
Abstract
In early stage non-small cell lung cancer (NSCLC) definitive radiation therapy is an appropriate alternative to surgery. Recent studies show, that in such patients hypofractionation schedules (for example 3 times 18 Gy or 5 times 12 Gy), can be safely applied, without causing severe toxicities and achieving high local control rates of up to 90% and more. In the last couple of years a lot of knowledge about the cancer biology, technical aspects, clinical outcomes and toxicities has been accumulated from different clinical trials. The purpose of this review is to summarize recent outcomes and developments in stereotactic radiation therapy for patients with early stage NSCLC.
Collapse
Affiliation(s)
- Anna O Simeonova
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Katharina Fleckenstein
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Hansjörg Wertz
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Anian Frauenfeld
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Judit Boda-Heggemann
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Frank Lohr
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| | - Frederik Wenz
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany
| |
Collapse
|
15
|
Li H, Liu B, Yin FF. Generation of virtual monochromatic CBCT from dual kV∕MV beam projections. Med Phys 2014; 40:121910. [PMID: 24320521 DOI: 10.1118/1.4824324] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/05/2023] Open
Abstract
PURPOSE To develop a novel on-board imaging technique which allows generation of virtual monochromatic (VM) cone-beam CT (CBCT) with a selected energy from combined kilovoltage (kV)∕megavoltage (MV) beam projections. METHODS With the current orthogonal kV∕MV imaging hardware equipped in modern linear accelerators, both MV projections (from gantry angle of 0°-100°) and kV projections (90°-200°) were acquired as gantry rotated a total of 110°. A selected range of overlap projections between 90° to 100° were then decomposed into two material projections using experimentally determined parameters from orthogonally stacked aluminum and acrylic step-wedges. Given attenuation coefficients of aluminum and acrylic at a predetermined energy, one set of VM projections could be synthesized from two corresponding sets of decomposed projections. Two linear functions were generated using projection information at overlap angles to convert kV and MV projections at nonoverlap angles to approximate VM projections for CBCT reconstruction. The contrast-to-noise ratios (CNRs) were calculated for different inserts in VM CBCTs of a CatPhan phantom with various selected energies and compared with those in kV and MV CBCTs. The effect of overlap projection number on CNR was evaluated. Additionally, the effect of beam orientation was studied by scanning the CatPhan sandwiched with two 5 cm solid-water phantoms on both lateral sides and an electronic density phantom with two metal bolt inserts. RESULTS Proper selection of VM energy [30 and 40 keV for low-density polyethylene (LDPE), polymethylpentene, 2 MeV for Delrin] provided comparable or even better CNR results as compared with kV or MV CBCT. An increased number of overlap kV and MV projection demonstrated only marginal improvements of CNR for different inserts (with the exception of LDPE) and therefore one projection overlap was found to be sufficient for the CatPhan study. It was also evident that the optimal CBCT image quality was achieved when MV beams penetrated through the heavy attenuation direction of the object. CONCLUSIONS A novel technique was developed to generate VM CBCTs from kV∕MV projections. This technique has the potential to improve CNR at selected VM energies and to suppress artifacts at appropriate beam orientations.
Collapse
Affiliation(s)
- Hao Li
- Medical Physics Graduate Program, Duke University, Durham, North Carolina 27710 and Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina 27710
| | | | | |
Collapse
|
16
|
Srinivasan K, Mohammadi M, Shepherd J. Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review. Pol J Radiol 2014; 79:181-93. [PMID: 25006356 PMCID: PMC4085117 DOI: 10.12659/pjr.890745] [Citation(s) in RCA: 37] [Impact Index Per Article: 3.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/25/2014] [Accepted: 03/27/2014] [Indexed: 11/23/2022] Open
Abstract
The use of Cone-beam Computed Tomography (CBCT) in radiotherapy is increasing due to the widespread implementation of kilovoltage systems on the currently available linear accelerators. Cone beam CT acts as an effective Image-Guided Radiotherapy (IGRT) tool for the verification of patient position. It also opens up the possibility of real-time re-optimization of treatment plans for Adaptive Radiotherapy (ART). This paper reviews the most prominent applications of CBCT (linac-mounted) in radiation therapy, focusing on CBCT-based planning and dose calculation studies. This is followed by a concise review of the main issues associated with CBCT, such as imaging artifacts, dose and image quality. It explores how medical physicists and oncologists can best apply CBCT for therapeutic applications.
Collapse
Affiliation(s)
- Kavitha Srinivasan
- School of Chemistry and Physics, University of Adelaide, Adelaide, Australia
| | - Mohammad Mohammadi
- School of Chemistry and Physics, University of Adelaide, Adelaide, Australia ; Department of Medical Physics, Royal Adelaide Hospital, Adelaide, Australia
| | - Justin Shepherd
- Department of Medical Physics, Royal Adelaide Hospital, Adelaide, Australia
| |
Collapse
|
17
|
McCowan PM, Rickey DW, Rowshanfarzad P, Greer PB, Ansbacher W, McCurdy BM. An investigation of gantry angle data accuracy for cine-mode EPID images acquired during arc IMRT. J Appl Clin Med Phys 2014; 15:4507. [PMID: 24423849 PMCID: PMC5711235 DOI: 10.1120/jacmp.v15i1.4507] [Citation(s) in RCA: 20] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/29/2013] [Revised: 09/28/2013] [Accepted: 08/29/2013] [Indexed: 12/03/2022] Open
Abstract
EPID images acquired in cine mode during arc therapy have inaccurate gantry angles recorded in their image headers. In this work, methods were developed to assess the accuracy of the gantry potentiometer for linear accelerators. As well, assessments of the accuracy of other, more accessible, sources of gantry angle information (i.e., treatment log files, analysis of EPID image headers) were investigated. The methods used in this study are generally applicable to any linear accelerator unit, and have been demonstrated here with Clinac/Trilogy systems. Gantry angle data were simultaneously acquired using three methods: i) a direct gantry potentiometer measurement, ii) an incremental rotary encoder, and iii) a custom‐made radiographic gantry‐angle phantom which produced unique wire intersections as a function of gantry angle. All methods were compared to gantry angle data from the EPID image header and the linac MLC DynaLog file. The encoder and gantry‐angle phantom were used to validate the accuracy of the linac's potentiometer. The EPID image header gantry angles and the DynaLog file gantry angles were compared to the potentiometer. The encoder and gantry‐angle phantom mean angle differences with the potentiometer were 0.13∘±0.14∘ and 0.10∘±0.30∘, respectively. The EPID image header angles analyzed in this study were within ±1∘ of the potentiometer angles only 35% of the time. In some cases, EPID image header gantry angles disagreed by as much as 3° with the potentiometer. A time delay in frame acquisition was determined using the continuous acquisition mode of the EPID. After correcting for this time delay, 75% of the header angles, on average, were within ±1∘ of the true gantry angle, compared to an average of only 35% without the correction. Applying a boxcar smoothing filter to the corrected gantry angles further improved the accuracy of the header‐derived gantry angles to within ±1∘ for almost all images (99.4%). An angle accuracy of 0.11∘±0.04∘ was determined using a point‐by‐point comparison of the gantry angle data in the MLC DynaLog file and the potentiometer data. These simple correction methods can be easily applied to individual treatment EPID images in order to more accurately define the gantry angle. PACS numbers: 87.53.Kn, 87.55.T‐, 87.56.bd, 87.59.‐e
Collapse
|
18
|
Boda-Heggemann J, Mai S, Fleckenstein J, Siebenlist K, Simeonova A, Ehmann M, Steil V, Wenz F, Lohr F, Stieler F. Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Radiother Oncol 2013; 109:505-9. [PMID: 24128805 DOI: 10.1016/j.radonc.2013.09.014] [Citation(s) in RCA: 38] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/07/2013] [Revised: 09/13/2013] [Accepted: 09/15/2013] [Indexed: 11/18/2022]
Abstract
Hypofractionated image-guided stereotactic ablative radiotherapy (igSABR) is effective in small lung/liver lesions. Computer-assisted breath-hold reduces intrafraction motion but, as every gating/triggering strategy, reduces the duty cycle, resulting in long fraction times if combined with intensity-modulated radiotherapy (IMRT). 10 MV flattening-filter-free IMRT reduces daily fraction duration to <10 min for single doses of 5-20 Gy.
Collapse
Affiliation(s)
- Judit Boda-Heggemann
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Germany.
| | | | | | | | | | | | | | | | | | | |
Collapse
|
19
|
Li H, Giles W, Bowsher J, Yin FF. A dual cone-beam CT system for image guided radiotherapy: initial performance characterization. Med Phys 2013; 40:021912. [PMID: 23387760 DOI: 10.1118/1.4788654] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/03/2023] Open
Abstract
PURPOSE The purpose of this study is to evaluate the performance of a recently developed benchtop dual cone-beam computed tomography (CBCT) system with two orthogonally placed tube∕detector sets. METHODS The benchtop dual CBCT system consists of two orthogonally placed 40 × 30 cm flat-panel detectors and two conventional x-ray tubes with two individual high-voltage generators sharing the same rotational axis. The x-ray source to detector distance is 150 cm and x-ray source to rotational axis distance is 100 cm for both subsystems. The objects are scanned through 200° of rotation. The dual CBCT system utilized 110° of projection data from one detector and 90° from the other while the two individual single CBCTs utilized 200° data from each detector. The system performance was characterized in terms of uniformity, contrast, spatial resolution, noise power spectrum, and CT number linearity. The uniformities, within the axial slice and along the longitudinal direction, and noise power spectrum were assessed by scanning a water bucket; the contrast and CT number linearity were measured using the Catphan phantom; and the spatial resolution was evaluated using a tungsten wire phantom. A skull phantom and a ham were also scanned to provide qualitative evaluation of high- and low-contrast resolution. Each measurement was compared between dual and single CBCT systems. RESULTS Compared to single CBCT, the dual CBCT presented: (1) a decrease in uniformity by 1.9% in axial view and 1.1% in the longitudinal view, as averaged for four energies (80, 100, 125, and 150 kVp); (2) comparable or slightly better contrast (0∼25 HU) for low-contrast objects and comparable contrast for high-contrast objects; (3) comparable spatial resolution; (4) comparable CT number linearity with R(2) ≥ 0.99 for all four tested energies; (5) lower noise power spectrum in magnitude. Dual CBCT images of the skull phantom and the ham demonstrated both high-contrast resolution and good soft-tissue contrast. CONCLUSIONS The performance of a benchtop dual CBCT imaging system has been characterized and is comparable to that of a single CBCT.
Collapse
Affiliation(s)
- Hao Li
- Medical Physics Graduate Program, Duke University, Durham, NC 27710, USA.
| | | | | | | |
Collapse
|
20
|
Liu B, Adamson J, Rodrigues A, Zhou F, Yin FF, Wu Q. A novel technique for VMAT QA with EPID in cine mode on a Varian TrueBeam linac. Phys Med Biol 2013; 58:6683-700. [DOI: 10.1088/0031-9155/58/19/6683] [Citation(s) in RCA: 31] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
21
|
Woodruff HC, Greer PB. 3D Dose reconstruction: Banding artefacts in cine mode EPID images during VMAT delivery. ACTA ACUST UNITED AC 2013. [DOI: 10.1088/1742-6596/444/1/012042] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
|
22
|
Abou-Elenein HS, Attalla EM, Ammar H, Eldesoky I, Farouk M, Zaghloul MS. Megavoltage cone beam computed tomography: Commissioning and evaluation of patient dose. J Med Phys 2012; 36:205-12. [PMID: 22228929 PMCID: PMC3249731 DOI: 10.4103/0971-6203.89969] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/17/2011] [Revised: 07/15/2011] [Accepted: 07/26/2011] [Indexed: 11/04/2022] Open
Abstract
The improvement in conformal radiotherapy techniques enables us to achieve steep dose gradients around the target which allows the delivery of higher doses to a tumor volume while maintaining the sparing of surrounding normal tissue. One of the reasons for this improvement was the implementation of intensity-modulated radio therapy (IMRT) by using linear accelerators fitted with multi-leaf collimator (MLC), Tomo therapy and Rapid arc. In this situation, verification of patient set-up and evaluation of internal organ motion just prior to radiation delivery become important. To this end, several volumetric image-guided techniques have been developed for patient localization, such as Siemens OPTIVUE/MVCB and MVision megavoltage cone beam CT (MV-CBCT) system. Quality assurance for MV-CBCT is important to insure that the performance of the Electronic portal image device (EPID) and MV-CBCT is suitable for the required treatment accuracy. In this work, the commissioning and clinical implementation of the OPTIVUE/MVCB system was presented. The geometry and gain calibration procedures for the system were described. The image quality characteristics of the OPTIVUE/MVCB system were measured and assessed qualitatively and quantitatively, including the image noise and uniformity, low-contrast resolution, and spatial resolution. The image reconstruction and registration software were evaluated. Dose at isocenter from CBCT and the EPID were evaluated using ionization chamber and thermo-luminescent dosimeters; then compared with that calculated by the treatment planning system (TPS- XiO 4.4). The results showed that there are no offsets greater than 1 mm in the flat panel alignment in the lateral and longitudinal direction over 18 months of the study. The image quality tests showed that the image noise and uniformity were within the acceptable range, and that a 2 cm large object with 1% electron density contrast can be detected with the OPTIVUE/MVCB system with 5 monitor units (MU) protocol. The registration software was accurate within 2 mm in the anterior-posterior, left-right, and superior-inferior directions. The additional dose to the patient from MV-CBCT study set with 5 MU at the isocenter of the treatment plan was 5 cGy. For Electronic portal image device (EPID) verification using two orthogonal images with 2 MU per image the additional dose to the patient was 3.8 cGy. These measured dose values were matched with that calculated by the TPS-XiO, where the calculated doses were 5.2 cGy and 3.9 cGy for MVCT and EPID respectively.
Collapse
|
23
|
Abstract
Radiotherapy technology has improved rapidly over the past two decades. New imaging modalities, such as positron emission (computed) tomography (PET, PET-CT) and high-resolution morphological and functional magnetic resonance imaging (MRI) have been introduced into the treatment planning process. Image-guided radiation therapy (IGRT) with 3D soft tissue depiction directly imaging target and normal structures, is currently replacing patient positioning based on patient surface markers, frame-based intracranial and extracranial stereotactic treatment and partially also 2D field verification methods. On-line 3D soft tissue-based position correction unlocked the full potential of new delivery techniques, such as intensity-modulated radiotherapy, by safely delivering highly conformal dose distributions that facilitate dose escalation and hypofractionation. These strategies have already resulted in better clinical outcomes, e.g. in prostate and lung cancer and are expected to further improve radiotherapy results.
Collapse
|
24
|
Wenz F, Belka C, Reiser M, Schönberg SO. [Image-guided radiation therapy. Paradigm change in radiation therapy]. Radiologe 2012; 52:207-12. [PMID: 22274657 DOI: 10.1007/s00117-011-2191-1] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/24/2022]
Abstract
CLINICAL ISSUE The introduction of image-guided radiotherapy (IGRT) has changed the workflow in radiation oncology more dramatically than any other innovation in the last decades. STANDARD TREATMENT Imaging for treatment planning before the initiation of the radiotherapy series does not take alterations in patient anatomy and organ movement into account. TREATMENT INNOVATIONS The principle of IGRT is the temporal and spatial connection of imaging in the treatment position immediately before radiation treatment. DIAGNOSTIC ASSESSMENT The actual position and the target position are compared using cone-beam computed tomography (CT) or stereotactic ultrasound. PERFORMANCE The IGRT procedure allows a reduction of the safety margins and dose to normal tissue without an increase in risk of local recurrence. ACHIEVEMENTS In the future the linear treatment chain in radiation oncology will be developed based on the closed-loop feedback principle. PRACTICAL RECOMMENDATIONS The IGRT procedure is increasingly being used especially for high precision radiotherapy, e.g. for prostate or brain tumors.
Collapse
Affiliation(s)
- F Wenz
- Klinik für Strahlentherapie und Radioonkologie, Universitätsmedizin Mannheim der Universität Heidelberg, Theodor-Kutzer-Ufer 1-3, 68167 Mannheim.
| | | | | | | |
Collapse
|
25
|
Stippel G, van Rooijen DC, Crezee J, Bel A. Automatic delineation of body contours on cone-beam CT images using a delineation booster. Phys Med Biol 2012; 57:N225-36. [DOI: 10.1088/0031-9155/57/13/n225] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
26
|
Roberts DA, Hansen VN, Thompson MG, Poludniowski G, Niven A, Seco J, Evans PM. Kilovoltage energy imaging with a radiotherapy linac with a continuously variable energy range. Med Phys 2012; 39:1218-26. [DOI: 10.1118/1.3681011] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
27
|
van Herk M, Ploeger L, Sonke JJ. A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiation. Radiother Oncol 2011; 100:365-9. [PMID: 21924785 DOI: 10.1016/j.radonc.2011.08.019] [Citation(s) in RCA: 32] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/24/2011] [Revised: 08/14/2011] [Accepted: 08/22/2011] [Indexed: 10/17/2022]
Abstract
BACKGROUND AND PURPOSE Acquisition of cone-beam CT (CBCT) concurrent with VMAT results in scatter of the megavoltage (MV) beam onto the kilovoltage (kV) detector deteriorating CBCT image quality. The aim of this paper is to develop a method to estimate and correct for MV scatter reaching the kV panel. MATERIAL AND METHODS The correction method is based on an alternating pulse sequence such that the kV source is enabled in only every other frame. MV scatter, noise and artifacts in the kV-unexposed frames are separated based on their spatial and temporal frequency characteristics and subsequently subtracted from the exposed frames. The method was tested on a phantom and one lung cancer patient scanned on an Elekta Synergy system simultaneous with a 9 Gy VMAT delivery. RESULTS Without correction, acquisition during VMAT leads to significant cupping and loss of bone contrast. The novel method removes the signal due to MV scatter. Compared to acquisition without MV beam, the reconstruction quality is identical except for some unavoidable noise due to the scattered MV radiation in the exposed frames. Correction is performed in-line with acquisition without introducing delays. CONCLUSIONS A novel, efficient and effective method was developed to correct for MV scatter in CBCT scans acquired concurrent with rotational radiotherapy.
Collapse
Affiliation(s)
- Marcel van Herk
- Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
| | | | | |
Collapse
|
28
|
Boda-Heggemann J, Lohr F, Wenz F, Flentje M, Guckenberger M. kV Cone-Beam CT-Based IGRT. Strahlenther Onkol 2011; 187:284-91. [PMID: 21533757 DOI: 10.1007/s00066-011-2236-4] [Citation(s) in RCA: 145] [Impact Index Per Article: 11.2] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/29/2010] [Accepted: 02/21/2011] [Indexed: 12/25/2022]
Affiliation(s)
- Judit Boda-Heggemann
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany.
| | | | | | | | | |
Collapse
|