Nusrat H, Karim-Picco S, Pang G, Paudel M, Sarfehnia A. Maximum RBE change in
192Ir,
125I, and
169Yb brachytherapy and the corresponding effect on treatment planning.
Biomed Phys Eng Express 2020;
6:015021. [PMID:
33438609 DOI:
10.1088/2057-1976/ab638e]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
PURPOSE
The purpose of this study was to examine RBE variation as a function of distance from the radioactive source, and the potential impact of this variation on a realistic prostate brachytherapy treatment plan.
METHODS
Three brachytherapy sources (125I, 192Ir, and 169Yb) were modelled in Geant4 Monte Carlo code, and the resulting electron energy spectrum in water in 3D space around these sources was scored (voxel size of 2 mm3). With this energy spectrum, microdosimetric techniques were used to calculate the maximum RBE, RBEM, as a function of distance from the source. RBEM of 125I relative to 192Ir was calculated in order to validate simulations against literature; all other RBEM calculations were done by normalizing electron fluence at various distances to the source position. In order to examine the impact of RBEM variation in treatment planning, a realistic 192Ir prostate plan was re-evaluated in terms of RBE instead of absorbed dose.
RESULTS
The RBEM of 125I, 192Ir, and 169Yb at 8 cm away from the source was 0.994 (+/-0.002), 1.030 (+/-0.003), and 1.066 (+/-0.008), respectively. RBEM in the HDR prostate treatment plan exhibited several hot (+3.6% in RBEM) spots.
CONCLUSIONS
The large increase RBEM observed in 169Yb has not yet been described in the literature. Despite the presence of radiobiological hotspots in the HDR treatment, these variations are likely nominal and clinically insignificant.