Feissel M, Aho LS, Georgiev S, Tapponnier R, Badie J, Bruyère R, Quenot JP. Pulse Wave Transit Time Measurements of Cardiac Output in Septic Shock Patients: A Comparison of the Estimated Continuous Cardiac Output System with Transthoracic Echocardiography.
PLoS One 2015;
10:e0130489. [PMID:
26126112 PMCID:
PMC4488420 DOI:
10.1371/journal.pone.0130489]
[Citation(s) in RCA: 10] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/19/2015] [Accepted: 05/19/2015] [Indexed: 12/29/2022] Open
Abstract
Background
We determined reliability of cardiac output (CO) measured by pulse wave transit time cardiac output system (esCCO system; COesCCO) vs transthoracic echocardiography (COTTE) in mechanically ventilated patients in the early phase of septic shock. A secondary objective was to assess ability of esCCO to detect change in CO after fluid infusion.
Methods
Mechanically ventilated patients admitted to the ICU, aged >18 years, in sinus rhythm, in the early phase of septic shock were prospectively included. We performed fluid infusion of 500ml of crystalloid solution over 20 minutes and recorded CO by EsCCO and TTE immediately before (T0) and 5 minutes after (T1) fluid administration. Patients were divided into 2 groups (responders and non-responders) according to a threshold of 15% increase in COTTE in response to volume expansion.
Results
In total, 25 patients were included, average 64±15 years, 15 (60%) were men. Average SAPSII and SOFA scores were 55±21.3 and 13±2, respectively. ICU mortality was 36%. Mean cardiac output at T0 was 5.8±1.35 L/min by esCCO and 5.27±1.17 L/min by COTTE. At T1, respective values were 6.63 ± 1.57 L/min for esCCO and 6.10±1.29 L/min for COTTE. Overall, 12 patients were classified as responders, 13 as non-responders by the reference method. A threshold of 11% increase in COesCCO was found to discriminate responders from non-responders with a sensitivity of 83% (95% CI, 0.52-0.98) and a specificity of 77% (95% CI, 0.46-0.95).
Conclusion
We show strong correlation esCCO and echocardiography for measuring CO, and change in CO after fluid infusion in ICU patients.