Wauthier L, Di Chiaro L, Favresse J. Sigma Metrics in Laboratory Medicine: A Call for Harmonization.
Clin Chim Acta 2022;
532:13-20. [PMID:
35594921 DOI:
10.1016/j.cca.2022.05.012]
[Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/07/2022] [Revised: 04/27/2022] [Accepted: 05/13/2022] [Indexed: 11/25/2022]
Abstract
BACKGROUND AND AIM
Sigma metrics are applied in clinical laboratories to assess the quality of analytical processes. A parameter associated to a Sigma >6 is considered "world class" whereas a Sigma <3 is "poor" or "unacceptable". The aim of this retrospective study was to quantify the impact of different approaches for Sigma metrics calculation.
MATERIAL AND METHODS
Two IQC levels of 20 different parameters were evaluated for a 12-month period. Sigma metrics were calculated using the formula: (allowable total error (TEa) (%) - bias (%))/(coefficient of variation (CV) (%)). Method precision was calculated monthly or annually. The bias was obtained from peer comparison program (PCP) or external quality assessment program (EQAP), and 9 different TEa sources were included.
RESULTS
There was a substantial monthly variation of Sigma metrics for all combinations, with a median variation of 32% (IQR, 25.6-41.3%). Variation across multiple analyzers and IQC levels were also observed. Furthermore, TEa source had the highest impact on Sigma calculation with proportions of Sigma >6 ranging from 17.5% to 84.4%. The nature of bias was less decisive.
CONCLUSION
In absence of a clear consensus, we recommend that laboratories calculate Sigma metrics on a sufficiently long period of time (>6 months) and carefully evaluate the choice of TEa source.