Lu MY, Ting CY, Jao JC. Effective dose and radiation risk under 640-slice abdominal computed tomography examination without contrast medium injection.
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY 2022;
30:657-666. [PMID:
35367977 DOI:
10.3233/xst-211101]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/14/2023]
Abstract
BACKGROUND
Medical imaging plays a crucial role in modern medicine. In order to provide fast and accurate medical diagnosis, computed tomography (CT) is a commonly used tool in radiological examinations, and 640-slice CT is the most advanced CT imaging modality.
OBJECTIVE
To evaluate the radiation dose and the risk under 640-slice abdominal CT examination.
METHODS
Examinations were performed using a 640-slice CT scanner on an Alderson-Rando anthropomorphic phantom. The used scanning acquisition parameters were the same as those used on abdominal examination without contrast medium injection in clinical practice. To measure the absorbed doses, optically stimulated luminescence dosimeters (OSLDs) were put into liver, stomach, bladder, gonads, colon, small intestine, bone marrow, and skin.
RESULTS
According to the 1990 Recommendations of the International Commission on Radiological Protection (ICRP Publication 60), the calculated effective doses received from this examination were 0.90 mSv in males and 0.89 mSv in females. According to the 2007 Recommendations of the International Commission on Radiological Protection (ICRP Publication 103), the calculated effective dose received from this examination was 0.83 mSv in both sexes.
CONCLUSIONS
Radiation doses obtained from the abdominal 640-slice CT examination are lower than the yearly cumulative doses received from natural radiation, revealing there is no deterministic effect and radiation risk is relatively low; therefore, this CT examination is considered safe.
Collapse