Tsukui H, Abla A, Teuteberg JJ, McNamara DM, Mathier MA, Cadaret LM, Kormos RL. Cerebrovascular accidents in patients with a ventricular assist device.
J Thorac Cardiovasc Surg 2007;
134:114-23. [PMID:
17599496 DOI:
10.1016/j.jtcvs.2007.02.044]
[Citation(s) in RCA: 92] [Impact Index Per Article: 5.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/19/2006] [Revised: 12/02/2006] [Accepted: 02/14/2007] [Indexed: 11/29/2022]
Abstract
OBJECTIVE
A cerebrovascular accident is a devastating adverse event in a patient with a ventricular assist device. The goal was to clarify the risk factors for cerebrovascular accident.
METHODS
Prospectively collected data, including medical history, ventricular assist device type, white blood cell count, thrombelastogram, and infection, were reviewed retrospectively in 124 patients.
RESULTS
Thirty-one patients (25%) had 48 cerebrovascular accidents. The mean ventricular assist device support period was 228 and 89 days in patients with and without cerebrovascular accidents, respectively (P < .0001). Sixty-six percent of cerebrovascular accidents occurred within 4 months after implantation. Actuarial freedom from cerebrovascular accident at 6 months was 75%, 64%, 63%, and 33% with the HeartMate device (Thoratec Corp, Pleasanton, Calif), Thoratec biventricular ventricular assist device (Thoratec Corp), Thoratec left ventricular assist device (Thoratec), and Novacor device (WorldHeart, Oakland, Calif), respectively. Twenty cerebrovascular accidents (42%) occurred in patients with infections. The mean white blood cell count at the cerebrovascular accident was greater than the normal range in patients with infection (12,900/mm3) and without infection (9500/mm3). The mean maximum amplitude of the thrombelastogram in the presence of infection (63.6 mm) was higher than that in the absence of infection (60.7 mm) (P = .0309).
CONCLUSIONS
The risk of cerebrovascular accident increases with a longer ventricular assist device support period. Infection may activate platelet function and predispose the patient to a cerebrovascular accident. An elevation of the white blood cell count may also exacerbate the risk of cerebrovascular accident even in patients without infection. Selection of device type, prevention of infection, and meticulous control of anticoagulation are key to preventing cerebrovascular accident.
Collapse