Nugent M, Galbraith JG, Fitzgerald AP, Gul R, Healy NO. Outcomes of a patient recall following early failure of hip hemiarthroplasty.
Ir J Med Sci 2013;
183:521-4. [PMID:
24234942 DOI:
10.1007/s11845-013-1038-3]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/09/2013] [Accepted: 10/30/2013] [Indexed: 11/26/2022]
Abstract
BACKGROUND
Unexpected early loosening was noted in some patients who had bipolar hemiarthroplasties using aspecific combination of head and stem.
AIM
A review of all patients who had received this implant combination was performed to establish whether there was a higher than expected failure rate and to identify those patients who had evidence of early loosening requiring further intervention or monitoring.
METHODS
Theatre records were reviewed to identify those who had undergone bipolar hemiarthroplasty using these products. All surviving patients were contacted and offered an appointment at which they underwent clinical and radiological review. Following review, revision rates were compared to published Australian joint registry data.
RESULTS
Of 247 eligible for recall, 139 attended for clinical and radiological review. The cumulative revision rate was 6.8 % at 4 years, with a mean time to revision of 26 months; however, there was a significantly higher revision rate of 12.1 % in those aged under 75 years at the time of surgery (p = 0.01). This is significantly higher than rates quoted for bipolar hemiarthroplasties in Australian joint registry data.
CONCLUSION
Overall, higher than expected revision rates due to early loosening were seen for this product combination,especially in patients aged\75 years at the time of the initial surgery.