Kerr NC, Somes G, Enzenauer RW. The effect of developmentally-at-risk status on the reliability of the iScreen® photorefractive device in young children.
ACTA ACUST UNITED AC 2012;
61:117-23. [PMID:
21856879 DOI:
10.3368/aoj.61.1.117]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/03/2022]
Abstract
INTRODUCTION AND PURPOSE
We evaluated the impact of developmentally-at-risk status on the results of photorefractive screening with iScreen®.
PATIENTS AND METHODS
We sequentially recruited 169 children (aged 2-5 years) to participate in a blinded, prospective study of a photoscreening device. The principle investigator examined the children after photoscreening. Using established standards for amblyogenic factors, the principle investigator and photoscreen interpreters separately made a determination of "normal" or "needs referral" for each child. Sensitivity, specificity, positive predictive value, and negative predictive value for the photoscreening device were determined.
RESULTS
Overall, sensitivity was 85% with a positive predictive value of 98%, and specificity was 87% with a negative predictive value of 47%. Developmentally-at-risk status in 34 children (three refused imaging) did not reduce sensitivity (89%) or specificity (100%) when compared with 130 children (two refused imaging) who did not have developmentally-at-risk factors (sensitivity = 84% and specificity = 80%).
CONCLUSION
The efficacy of photorefractive screening in young children with developmentally-at-risk status is comparable to results found in normal children.