Orbegozo J, Pérez A, Basterra I, Noguera H, González L, Lauzirika G, Piñero DP. Clinical outcomes of a toric continuous range of vision presbyopia-correcting intraocular lens.
J Cataract Refract Surg 2024;
50:724-732. [PMID:
38530027 DOI:
10.1097/j.jcrs.0000000000001428]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/03/2023] [Accepted: 02/12/2024] [Indexed: 03/27/2024]
Abstract
PURPOSE
To evaluate the clinical outcomes of the toric version of a presbyopia-correcting intraocular lens (IOL) based on the combination of a diffractive-based extended depth-of-focus (EDOF) pattern and a diffractive multifocal platform.
SETTING
Miranza COI Bilbao, Bilbao, Spain.
DESIGN
Prospective case series.
METHODS
35 patients (51 to 84 years) with corneal astigmatism ranging from 0.75 to 2.19 diopters (D) undergoing bilateral cataract surgery with implantation of the Synergy Toric II IOL were evaluated during a 3-month follow-up. Visual acuity, refraction, defocus curve, and patient-reported outcomes with the Catquest-9SF questionnaire were analyzed. A vectorial analysis was used to analyze the accuracy of astigmatic correction.
RESULTS
Mean 3-month monocular postoperative uncorrected distance, intermediate (80 cm) and near (40 cm) visual acuities were 0.06 ± 0.11 logMAR, 0.13 ± 0.12 logMAR, and 0.13 ± 0.09 logMAR, respectively. Mean monocular distance-corrected intermediate (80 cm) and near visual acuity (40 cm) were 0.11 ± 0.12 logMAR and 0.10 ± 0.10 logMAR, respectively. Mean binocular defocus curve showed visual acuities of 0.10 logMAR or better for defocus levels from +0.50 to -2.50 D. Residual cylinder was within ±0.50 D in 97.0% of eyes. The surgically induced astigmatism prediction error ranged between -0.49 D and 0.50 D, with a mean value of 0.04 ± 0.16 D. Mean absolute IOL rotation was 3.79 ± 2.94 degrees. Significant improvements were found in all Rasch-calibrated scores obtained with Catquest-9SF ( P < .001).
CONCLUSIONS
The implantation of the toric presbyopia-correcting IOL evaluated provides an efficacious astigmatic correction while providing a fully restoration of the visual function across different distances.
Collapse