Pestrea C, Gherghina A, Ortan F, Cismaru G, Radu R. Left bundle branch area pacing - restoring the natural order: A Case Report.
Medicine (Baltimore) 2020;
99:e21602. [PMID:
32769916 PMCID:
PMC7593055 DOI:
10.1097/md.0000000000021602]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/30/2019] [Revised: 05/29/2020] [Accepted: 07/08/2020] [Indexed: 11/25/2022] Open
Abstract
INTRODUCTION
Recent studies have shown that His-bundle pacing could be an alternative in patients requiring cardiac resynchronization therapy as it is comparable or better in terms of amelioration of ventricular activation, narrowing of the QRS complex, or clinical outcomes. However, in case of high threshold at the level of His-bundle or inability to correct conduction through a diseased His-Purkinje system other option should be searched like left bundle pacing.
PATIENT CONCERNS
A 77-year-old man presented to the Emergency Department for dizziness and dizziness and lightheadedness due to an intermittent 2:1 atrioventricular block with a QRS complex morphology of a major left branch block.
DIAGNOSIS
Given the documented symptomatic 2:1 AV block, according to the European Guideliness the patient was considered to have a class 1 indication of permanent double chamber cardiostimulation.
INTERVENTIONS
A lead delivery system with a C315 His catheter and a Select Secure 3830 69 cm pacing lead were placed at the His bundle area with important narrowing of the QRS complex but with an unacceptable high threshold. The delivery system was moved towards the apex 1,5 cm and the lead screwed deep into the septum until capture of the left bundle branch was achieved with complete normalization of the conduction troubles.
OUTCOMES
At 3 month follow-up the patient was asymptomatic and the pacing and sensing thresholds remained at same values as during implantation: 0.75/0.4 ms and 14 mV respectively.
CONCLUSION
Left bundle-pacing represents the next step of His-Purkinje system pacing to overcome all difficulties related to His-bundle pacing.